Instruments
DC Ammeter
• The PMMC instrument is an ammeter.
• Pointer deflection is directly proportional to the current
flowing in the coil.
• An ammeter is always connected in series with a circuit in
which current is to be measured.
• A shunt resistance is connected in parallel with the
instrument coil. Because when large currents are to be
measured, it is necessary to bypass a major part of the current
through shunt resistance.
• The resistance of the shunt can be calculated using conventional circuit analysis.
Referring to figure 1,
• Since the shunt resistance is in parallel with the meter movement, the voltage drop
across the shunt and movement must be the same.
Therefore,
Vs= Vm
Is Rs = Im Rm
Rs = Im Rm / Is
But,
Is = I – Im
Hence,
Rs = Im Rm / (I – Im)
In figure 1,
Suppose that the meter resistance is exactly 99 ohms and the shunt resistance is 1 ohm.
The shunt current will be 99 times the meter current. In this situation, if the meter gives
full scale deflection for a coil current of 0.1 mA, the scale should be calibrated to read
100 × 0.1 mA or 10 mA at full scale.
Is = 99Im
I= Im +99Im
3.3 D
• The total ammeter current in the previous
example is 10mA when the moving coil
instrument indicates FSD. Therefore the meter
scale can be calibrated for FSD to indicate 10mA.
When the pointer indicates 0.5FSD and 0.25FSD,
the current levels are 5mA and 2.5mA,
respectively. Thus, the ammeter scale may be
calibrated to linearly represent all currents level
from zero to 10mA.
• So if a shunt having a smaller value is used, the
shunt current and total ammeter current will be
larger than the levels calculated. In fact shunt
resistance values can be determined to convert a
PMMC instrument into an ammeter for
measuring any desired values of current. (e.g. see
the next example)
Example
A PMMC instrument has FSD of 100 uA and coil resistance of 1 K ohm .
Calculate the required shunt resistance value to convert the instrument into
an ammeter with
a. FSD = 100 mA
b . FSD= 1 Ampere
3.4 D
Multirange ammeter using switched
shunts
Example
Design a multirange ammeter with range of 0-1A, 5A and 10A employing
individual shunt in each A D’Arsonval movement with an internal resistance of
500 ohm and full scale deflection of 10 mA is available.
3.3 K
Multirange ammeter using ayrton
shunts or universal shunts
3.5 D
General requirements of a shunt resistance:
• The temperature coefficients of the shunt and instrument should be low and nearly
identical.
Manganin is usually used as a shunt for dc instruments, since it gives a low value of
thermal emf with copper.
DC Voltmeter
• The basic PMMC instrument can be converted into a dc voltmeter by adding a series
resistor known as multiplier, as shown in figure 3.
A basic PMMC instrument movement with a full scale deflection of 50 micro Amperes
and internal resistance of 500 ohms is used as a voltmeter. Determine the value of the
multiplier resistance needed to measure a voltage range of 0 -10 V.
Example 2 4.2b K
Calculate the value of multiplier resistance on the 50 V range of a dc voltmeter that uses a
500 micro Ampere meter movement with an internal resistance of 1 kilo ohms.
Example 3 3.5D
Voltmeter sensitivity
the sensitivity of voltmeter is measured in terms of its resistance. This resistance is the
required minimum resistance which allows to flow FSD current through PMMC.
1 / 500x10-6 A = 2 KΩ / volt
•If the sensitivity is known, the total voltmeter resistance is easily calculated. i.e.
From equation 1 :
4.11 K
Multirange Voltmeter
In multirange voltmeter, the multipliers are connected in a series string, and the range
selector selects the appropriate amount of resistance required in series with the movement.
Example 5 (Dictate)
•In the previous example if you ponder upon the circuits you will realize that the
circuit in figure (b) is less expensive as compared to the circuit in figure (a) because
all of the multiplier resistor in (a) must be special (non-standard) resistors while in
circuit (b) all are regular standard resistor except R1.
3.7 D