G3335-90173 TOF Q-TOF Concepts
G3335-90173 TOF Q-TOF Concepts
G3335-90173 TOF Q-TOF Concepts
Concepts Guide
The Big Picture
Notices
© Agilent Technologies, Inc. 2014 Warranty Safety Notices
No part of this manual may be reproduced The material contained in this docu-
in any form or by any means (including ment is provided “as is,” and is sub-
electronic storage and retrieval or transla- ject to being changed, without notice, CAUTION
tion into a foreign language) without prior in future editions. Further, to the max-
agreement and written consent from Agi- imum extent permitted by applicable A CAUTION notice denotes a haz-
lent Technologies, Inc. as governed by law, Agilent disclaims all warranties, ard. It calls attention to an operat-
United States and international copyright either express or implied, with regard
laws. ing procedure, practice, or the like
to this manual and any information
contained herein, including but not that, if not correctly performed or
Manual Part Number limited to the implied warranties of adhered to, could result in damage
G3335-90173 merchantability and fitness for a par- to the product or loss of important
ticular purpose. Agilent shall not be data. Do not proceed beyond a
Edition liable for errors or for incidental or
consequential damages in connection CAUTION notice until the indi-
Revision A, December 2014 with the furnishing, use, or perfor- cated conditions are fully under-
Agilent Technologies, Inc. mance of this document or of any stood and met.
5301 Stevens Creek Blvd. information contained herein. Should
Santa Clara, CA 95051 Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
WAR NING
Microsoft®, Windows® and Excel® are
registered trademark of Microsoft Corpora- document that conflict with these
terms, the warranty terms in the sep- A WARNING notice denotes a
tion in the US and/or other countries.
arate agreement shall control. hazard. It calls attention to an
Software Revision operating procedure, practice, or
Technology Licenses the like that, if not correctly per-
This guide applies to the Agilent
MassHunter Workstation Software – Data The hardware and/or software described in formed or adhered to, could result
Acquisition program for TOF and Q-TOF ver- this document are furnished under a in personal injury or death. Do not
sion B.06.01 or higher until superseded. license and may be used or copied only in
accordance with the terms of such license. proceed beyond a WARNING
notice until the indicated condi-
Restricted Rights Legend tions are fully understood and
U.S. Government Restricted Rights. Soft- met.
ware and technical data rights granted to
the federal government include only those
rights customarily provided to end user cus-
tomers. Agilent provides this customary
commercial license in Software and techni-
cal data pursuant to FAR 12.211 (Technical
Data) and 12.212 (Computer Software) and,
for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or Com-
puter Software Documentation).
In This Guide...
The Concepts Guide presents “The Big Picture” behind the
Agilent TOF and Q-TOF LC/MS system to help you analyze
samples on your Agilent time-of-flight or quadrupole
time-of-flight mass spectrometer system. This guide helps you
understand how the hardware and software work together.
1 Overview
Learn how the Agilent 6200 Series TOF and 6500 Series Q-TOF
LC/MS system helps you do your job and how the hardware and
software work.
2 Instrument Preparation
Learn the concepts you need to prepare the instrument for
sample acquisition.
4 Data Acquisition
Learn concepts to help you enter information to run individual
samples or a worklist of samples, and to help you acquire data
and monitor runs.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 3
4 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Content
1 Overview
How does the TOF and Q-TOF system help you do your job? 10
Help for applications 11
Help for data acquisition 11
Help for data analysis 13
How do different ion sources work? 16
Electrospray ionization (ESI) and Dual ESI 17
Dual Agilent Jet Stream Electrospray Ionization (Dual AJS
ESI) 21
Atmospheric pressure chemical ionization (APCI) 22
Atmospheric pressure photoionization (APPI) 24
Multimode ionization (MMI) 25
HPLC-Chip 27
How does the Agilent TOF and Q-TOF mass spectrometer
work? 28
Innovative Enhancements in the Agilent 6560 Ion Mobility
Q-TOF 33
Innovative Enhancements in the Agilent 6550 iFunnel
Q-TOF 35
Innovative Enhancements in the Agilent 6545 Q-TOF 37
Innovative Enhancements in the 6540 and 6538 Q-TOF 38
Innovative Enhancements in the 6530 Q-TOF 39
Agilent Jet Stream Thermal Gradient Technology 40
Front-end ion optics 42
2 Instrument Preparation
LC preparation 50
LC module setup 50
Column equilibration and conditioning 53
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 5
TOF and Q-TOF preparation – calibration and tuning 55
TOF mass calibration 56
Tuning choices 58
Tune reports 68
Storage and retrieval of tune results and Instrument Mode 69
Tune Set Point Modifications for Medium and Large
Proteins 71
Real-time displays 72
Instrument Status Window 72
Real-time parameter values (Actuals) 73
Real-time Chromatogram Plot and Spectral Plot windows 75
System logbook 77
6 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
4 Data Acquisition
Interactive single sample setup 110
Sample information 110
Data File information 111
Some of the Additional Information parameters 111
Worklist setup 112
Worklist menus 113
Sample entry 114
Script entry 116
Entry of additional sample information (show, add
columns) 117
Worklist import 118
Report setup 122
Run setup 123
Estimate of worklist file size 124
Data acquisition for samples and worklists 128
What you can monitor during a run 128
What you can do during a run 129
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 7
8 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System
1
Overview
How does the TOF and Q-TOF system help you do your job? 10
Help for applications 11
Help for data acquisition 11
Help for data analysis 13
How do different ion sources work? 16
Electrospray ionization (ESI) and Dual ESI 17
Dual Agilent Jet Stream Electrospray Ionization (Dual AJS ESI) 21
Atmospheric pressure chemical ionization (APCI) 22
Atmospheric pressure photoionization (APPI) 24
Multimode ionization (MMI) 25
HPLC-Chip 27
How does the Agilent TOF and Q-TOF mass spectrometer work? 28
Innovative Enhancements in the 6540 and 6538 Q-TOF 38
Innovative Enhancements in the 6530 Q-TOF 39
Agilent Jet Stream Thermal Gradient Technology 40
Front-end ion optics 42
9
1 Overview
How does the TOF and Q-TOF system help you do your job?
How does the TOF and Q-TOF system help you do your job?
You can set up an Agilent 6200 Series Time-of-Flight LC/MS
(TOF) system and the Agilent 6500 Series Quadrupole
Time-of-Flight LC/MS (Q-TOF) system in several configurations:
ESI – Electrospray Ionization • For normal flow LC/MS with a binary pump, quaternary
APCI – Atmospheric Pressure pump, well-plate sampler (or autosampler or HTC/HTS
Chemical Ionization autosampler) and ESI or Dual ESI with Agilent Jet Stream
APPI - Atmospheric Pressure Thermal Gradient Technology - Agilent 6530 Quadrupole
Photo Ionization Time-of-Flight, Agilent 6540 Quadrupole Time-of-Flight,
HPLC-Chip/MS – High Agilent 6545 Quadrupole Time-of-Flight, Agilent 6550
Performance Liquid iFunnel Quadrupole Time-of-Flight, Agilent 6560 Ion Mobility
Chromatography on a Chip Quadrupole Time-of-Flight, and Agilent 6230 Time-of-Flight.
MALDI – Matrix-Assisted Laser • For normal flow LC/MS with a binary pump, quaternary
Desorption Ionization pump, well-plate sampler (or autosampler or HTC/HTS
MMI - Multimode Ionization autosampler) and ESI, Dual ESI, APCI, APPI, or MMI ion
sources.
• For microflow LC/MS with a capillary pump, micro well-plate
sampler and ESI, Dual ESI, APCI or MMI ion sources.
• For nanoflow LC/MS with a nanopump, micro well-plate
sampler and nano-ESI source or HPLC-Chip/MS interface
(used in place of standard or dual nanospray source) to
increase reliability and boost performance with narrow peak
dispersion and lower dead volumes.
• Agilent 6200 Series TOF or 6500 Series Q-TOF LC/MS system
with an AP-MALDI or PDF-MALDI source.
10 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Help for applications
Paired with Agilent Infinity and Infinity II Series LCs, the 6500
Series Q-TOF LC/MS delivers fast, sensitive, reproducible
analyses of small and large molecules.
• Reproducible mass accuracy
• Low picogram and femtomole limits of detection
• Fast MS/MS operation (for the Q-TOF)
To help you use the Agilent 6200 Series TOF and the 6500 Series
Q-TOF LC/MS systems for these applications, the software lets
you perform the following tasks in a single window:
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 11
1 Overview
Help for data acquisition
12 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Help for data analysis
Acquire data
• Enter sample information and pre- or post-analysis programs
(scripts) and run single samples interactively
A worklist is a list of a sequence of • Enter and automatically run both individual samples and
samples that you enter and run sequences of samples in a worklist
automatically with the Data
Acquisition program.
• Set up pre- and post-analysis scripts to run between samples
in a worklist.
• Set up and run a worklist.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 13
1 Overview
Help for data analysis
14 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Help for data analysis
For the Report Designer Add-in, please refer to the online Help
or the Reporting Training DVD. The Report Designer Add-in
allows you to customize the templates that are used when you
print a report. The Report Designer is used with the
MassHunter Qualitative Analysis program and with the
MassHunter Quantitative Analysis program.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 15
1 Overview
How do different ion sources work?
16 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Electrospray ionization (ESI) and Dual ESI
solvent
spray
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 17
1 Overview
Electrospray ionization (ESI) and Dual ESI
Formation of ions
Ion formation in API-electrospray occurs through more than
one mechanism. If the chemistry of analyte, solvents, and
buffers is correct, ions are generated in solution before
nebulization. This results in high analyte ion concentration and
good API-electrospray sensitivity.
Nebulization
Nebulization (aerosol generation) takes the sample solution
through these steps:
a Sample solution enters the spray chamber through a
grounded needle called a nebulizer.
b For high-flow electrospray, nebulizing gas enters the spray
chamber concentrically through a tube that surrounds the
needle.
c The combination of strong shear forces generated by the
nebulizing gas and the strong voltage (2–6 kV) in the spray
18 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Electrospray ionization (ESI) and Dual ESI
Coulomb repulsion – repulsion When the force of the Coulomb repulsion equals that of the
between charged species of the surface tension of the droplet, the droplet explodes, producing
same sign smaller charged droplets that are subject to further
evaporation. This process repeats itself, and droplets with a
high density of surface-charges are formed. When charge
density reaches approximately 108 V/cm3, ion evaporation
occurs (direct ejection of bare ions from the droplet surface).
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 19
1 Overview
Electrospray ionization (ESI) and Dual ESI
Multiple charging
Electrospray is especially useful for analyzing large
biomolecules such as proteins, peptides, and oligonucleotides,
but can also analyze smaller molecules like drugs and
environmental contaminants. Large molecules often acquire
more than one charge. Because of this multiple charging, you
can use electrospray to analyze molecules as large as 150,000 u
even though the mass range (or more accurately mass-to-charge
range) for a typical quadrupole LC/MS instrument is around
3000 m/z. For example:
100,000 u / 10 z = 1,000 m/z
The optional Agilent MassHunter When a large molecule acquires many charges, a mathematical
BioConfirm Software performs the process called deconvolution is used to determine the actual
calculations to accomplish molecular weight of the analyte.
deconvolution.
20 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Dual Agilent Jet Stream Electrospray Ionization (Dual AJS ESI)
Figure 3 Dual AJS ESI source plumbed for high flow application
Both the Dual AJS ESI and the Dual ESI source support two
nebulizers for different applications: the standard nebulizer
(G1958-60098) and the capillary (or microflow) LC/MS
nebulizer (G1946-60260). The main differences are:
• The machined tip of the capillary LC/MS nebulizer has a
smaller exit orifice and internal taper.
• The internal diameter (ID) of the internal needle for the
capillary LC/MS nebulizer is 50 μm versus 120 μm for the
“standard” nebulizer needle.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 21
1 Overview
Atmospheric pressure chemical ionization (APCI)
vaporizer
(heater)
drying gas
corona capillary
discharge
needle
APCI requires that the analyte be in the gas phase for ionization
to occur. To vaporize the solvent and analyte, the APCI source is
typically operated at vaporizer temperatures of 400 to 500°C.
22 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Atmospheric pressure chemical ionization (APCI)
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 23
1 Overview
Atmospheric pressure photoionization (APPI)
APPI and APCI are similar, with APPI substituting a lamp for
the corona needle for ionization. APPI often also uses an
additional solvent or mobile phase modifier, called a “dopant”,
to assist with the photoionization process.
vaporizer
(heater)
drying gas
capillary
UV lamp
24 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Multimode ionization (MMI)
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 25
1 Overview
Multimode ionization (MMI)
Unlike the APCI and APPI sources where the temperature of the
vaporizer is monitored, in the multimode source the actual
vapor temperature is monitored. As a result, the vaporizer is
typically set to between 200 and 250°C.
26 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
HPLC-Chip
HPLC-Chip
Traditional nanospray mass spectrometry has proven its
potential as a cost-effective, sensitive and reproducible
technique for the identification of peptides at femtomole to
atomol levels. However, connecting nano capillaries, columns
and valves frequently is a tedious procedure and requires user
skills and routine. When handled incorrectly, nano flow
connections are prone to leakage which are difficult to detect
and result in poor system performance and extended downtime
of the complete system. Quartz nanospray needles are prone to
blockages and require frequent replacement.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 27
1 Overview
How does the Agilent TOF and Q-TOF mass spectrometer work?
How does the Agilent TOF and Q-TOF mass spectrometer work?
6200 Series TOF The Agilent TOF is an orthogonal acceleration time-of-flight
mass spectrometer (oa-TOF). The acceleration pulse applied to
send the ions down the flight tube is orthogonal to the direction
that ions are entering the mass analyzer. This geometry
minimizes the effect of the entrance velocity on the flight time,
leading to higher resolution.
6500 Series The Agilent 6500 Series Q-TOF LC/MS is a liquid
Q-TOF chromatograph Q-TOF mass spectrometer that performs MS/MS
using a quadrupole, a hexapole (collision cell) and
a time-of-flight unit to produce spectra. The quadrupole selects
precursor ions that are fragmented in the collision cell into
product ions, which are then impelled to the detector, at an
angle perpendicular to the original path.
The Agilent 6500 Series Q-TOF LC/MS supports several
atmospheric pressure ionization (API) sources. A common
atmospheric sampling interface introduces ions from these
various sources into the mass spectrometer vacuum system.
28 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
How does the Agilent TOF and Q-TOF mass spectrometer work?
Figure 9 Schematic of Agilent 6530 Q-TOF LC/MS with major improvements circled
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 29
1 Overview
How does the Agilent TOF and Q-TOF mass spectrometer work?
Figure 10 Schematic of Agilent 6540 UHD Accurate-Mass Q-TOF LC/MS with major improvements circled
30 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
How does the Agilent TOF and Q-TOF mass spectrometer work?
Figure 11 Schematic of Agilent 6545 Q-TOF LC/MS with major improvements circled
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 31
1 Overview
How does the Agilent TOF and Q-TOF mass spectrometer work?
Figure 12 Schematic of Agilent 6550 iFunnel Q-TOF LC/MS with major improvements circled
32 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Innovative Enhancements in the Agilent 6560 Ion Mobility Q-TOF
Figure 13 Schematic of Agilent 6560 Ion Mobility Q-TOF LC/MS with major improvements circled
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 33
1 Overview
Innovative Enhancements in the Agilent 6560 Ion Mobility Q-TOF
Trapping funnel
The trapping funnel accumulates and releases ions into the
drift tube. The continuous ion beam from the electrospray
process has to be converted into a pulsed ion beam prior to ion
mobility separation. The trapping funnel operates by first
storing and then releasing discrete packets of ions into the drift
tube.
Drift tube
The drift cell is approximately 80 cm long and generally
operated at 20 V/cm or less drift field. Ions are separated as
they pass through the drift tube based on their collision cross
section and charge. Ions with larger collision cross sections
undergo a higher number of collisions with drift gas molecules
compared to ions with smaller collision cross section.
Therefore, larger ions travel through the drift tube slower than
the smaller ions. Also, ions with higher charge states experience
a higher electric force, and hence travel at a higher velocity,
compared to ions with lower charge states. The drift tube is
operated under low field limit conditions allowing the
instrument to generate accurate structural information for
compounds. Under the low electric field conditions the mobility
is not dependent on the electric field but rather on the structure
of the molecule and its interaction with the buffer gas. In
addition to separating ions based on their structures,
experiments can be performed to quantify this collision cross
section value.
Rear funnel
Ions exiting the drift tube enter the rear funnel which
efficiently refocuses and transfers ions to the mass analyzer
through a hexapole ion guide.
34 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Innovative Enhancements in the Agilent 6550 iFunnel Q-TOF
iFunnel Technology
The iFunnel Technology encompasses two enhancements to the
Agilent 6550 iFunnel Q-TOF: the Agilent Jet Stream technology,
a hexabore capillary and the Dual Ion Funnel technology.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 35
1 Overview
Innovative Enhancements in the Agilent 6550 iFunnel Q-TOF
36 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Innovative Enhancements in the Agilent 6545 Q-TOF
Slicer design
The slicer design on the 6545 is identical to the 6550. It has 2
different sizes of the opening: a larger one for high sensitivity
applications, and a smaller one for high resolution applications.
The positions can be changed in the Tune Context and are
saved as part of the tune file. In total, 10 positions are usable: 4
positions for the high sensitivity, and 6 positions for high
resolution.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 37
1 Overview
Innovative Enhancements in the 6540 and 6538 Q-TOF
38 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Innovative Enhancements in the 6530 Q-TOF
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 39
1 Overview
Agilent Jet Stream Thermal Gradient Technology
page 40).
40 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Agilent Jet Stream Thermal Gradient Technology
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 41
1 Overview
Front-end ion optics
After the API source forms ions, the Agilent 6200 Series TOF or
6500 Series Q-TOF LC/MS system performs the following
operations, organized according to the stages of the ion path
and the vacuum stages of the TOF or Q-TOF. See Figure 8 on
page 28 for details.
In an Agilent 6220 TOF, the ions exit the first ion guide and pass
into the third stage of the vacuum system. In the third stage of
the vacuum, the ions are passed onto a second octopole
assembly (octopole 2) which then sends the ion on to the beam
shaping assembly.
42 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Front-end ion optics
The following sections are only part of the Q-TOF LC/MS instrument. The
N O TE next section in the TOF instrument is the Beam shaping (Vacuum stage 4
for both 6220 TOF and Q-TOF) on page 45.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 43
1 Overview
Front-end ion optics
The collision cell contains nitrogen, the same gas that is used in
the ion source. The small diameter of the hexapole assembly
assists in capturing fragmented ions.
44 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Front-end ion optics
Beam shaping (Vacuum stage 4 for both 6220 TOF and Q-TOF)
In the Agilent 6538 and 6540 UDH Accurate-Mass Q-TOF, ions
enter the ion beam compression technology. Ion beam
compression provides up to a 10-fold compression and cooling
which helps in creating a much denser and thinner ion beam
that passes through a narrower slit leading into the slicer and
pulser region.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 45
1 Overview
Front-end ion optics
direction in the flight tube, they never return to the ion pulser,
but move to where the ion detector is mounted.
Because the calculation for the mass of each ion depends on its
flight time in the flight tube, the background gas pressure must
be very low. Any collision of an ion with residual gas slows the
ion on its path to the detector and affects the accuracy of the
mass calculation.
Ion detection Figure 20 shows a schematic of the Agilent 6500 Series Q-TOF
LC/MS detector.
Figure 20 Agilent 6200 Series TOF or 6500 Series Q-TOF detector, with
potentials shown for positive operation
46 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Overview 1
Front-end ion optics
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 47
1 Overview
Front-end ion optics
48 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System
2
Instrument Preparation
LC preparation 50
LC module setup 50
Column equilibration and conditioning 53
TOF and Q-TOF preparation – calibration and tuning 55
TOF mass calibration 56
Tuning choices 58
Tune reports 68
Storage and retrieval of tune results and Instrument Mode 69
Tune Set Point Modifications for Medium and Large Proteins 71
Real-time displays 72
Instrument Status Window 72
Real-time parameter values (Actuals) 73
Real-time Chromatogram Plot and Spectral Plot windows 75
System logbook 77
Learn about the concepts that can help you prepare the
instrument for use.
49
2 Instrument Preparation
LC preparation
LC preparation
To install, configure and start the To prepare the LC for sample runs, you usually do three tasks:
LC modules, see the Installation • Set up the LC modules for operation
Guide.
• Equilibrate or condition the column
• Monitor the plot baseline to assure pump and column
stability (See “Real-time displays” on page 72.)
See the Quick Start Guide and You can also view the system logbook for explanations of errors.
online Help for instructions on how (See “System logbook” on page 77.)
to prepare the LC for a sample run.
LC module setup
You set up the LC modules in the Instrument Status window
through the shortcut menus. The Instrument Status window is
also called the Dashboard. You can get help on any of the
devices by clicking the button in the device pane.
Figure 21 Instrument Status window (or Dashboard) with shortcut menu for Column Comp.
Table 1 shows you the tasks that you may perform to set up the
LC modules and the menu items where you can do the tasks.
You can do most of these tasks using the shortcut menu.
50 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Instrument Preparation 2
LC module setup
Well-Plate Sampler Change volumes for the installed syringe Done in the Device Configuration user
(WPS) or µWPS or interface
h-ALS or h-ALS-SL or
h-ALS-SL+
Select tray type and its position Right-click the device panel and click Assign
Wellplates, Edit Wellplate Types
Reset injector, Move Home, Needle Up or Down, Right-click the device panel and click Reset
set the valve mainpass or the bypass Sampler, Home Arm, Needle Up/Down,
Switch Valve to Bypass/Mainpass, Switch
Thermostat On, Wash Needle
HTC/HTS Update the plate assignments on the computer Right-click the device panel and click Update
Plate Assignment
Reset the injector Right-click the device panel and click Reset
injector
Binary, isocratic, and Turn the pump on, off or place in Standby Right-click the device panel and click Switch
quaternary pumps Pump On/Off
Set the date and time for automatic pump turnon Right-click the device panel and click Control
Set up to monitor solvent levels Right-click the device panel and click Bottle
Fillings
CapPump Do same tasks as binary pump Same menu items as binary pump
Purge the pump Right-click the device panel and click Purge
On/Off
Set mixer and filter volumes Right-click the device panel and click
Configuration
Enable a fast change in solvent composition Right-click the device panel and click Start
Fast Composition Change
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 51
2 Instrument Preparation
LC module setup
Nanopump Do the same tasks as CapPump Same menu items as CapPump except for
Accuracy Calibration (flow sensor)
Thermostat. Column Turn the right or left controller on or off Right-click the device panel and click
Compartment (TCC) Configuration
Set the maximum controller temperatures Right-click the device panel and click Set
Temperature
Enter information for the instrument columns Tools > Analytical Column Setup
Diode Array Det. (DAD) Turn UV or Visible lamp on Right-click the device panel and click UV
Lamp On; Vis Lamp On
Set the date or time for automatic DAD turnon Right-click the device panel and click Control
Calibrate the DAD wavelength Right-click the device panel and click
Calibration
Bring the baseline of the plot to zero Right-click the device panel and click
Balance
Show an intensity plot for the detector Right-click the device panel and click
Intensity plot
Flexible Cube Switch pump on or off Right-click the device panel and click Switch
pump on or Switch pump off.
Display the Control dialog box. Right-click the device panel and click
Control.
Display the Method Setup dialog box Right-click the device panel and click
Method.
Supercritical Fluid Display the Injector Wash dialog box Right-click the device panel and click
Chromatography (SFC) Injector Wash.
Display the Method Setup dialog box Right-click the device panel and click
Method.
Display the SFC Control dialog box Right-click the device panel and click
Control.
52 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Instrument Preparation 2
Column equilibration and conditioning
Fluorescence Detector Ignite the UV lamp Right-click the device panel and click Switch
(FLD) on. To turn off the lamp, click Switch off.
Display the detector’s method setup dialog box Right-click the device panel and click
Method.
Display the detector’s control dialog box Right-click the device panel and click
Control.
Equilibration
You can equilibrate a column in one of three ways:
• Interactively
You change the loaded method set points to the solvent
composition for the run, no volume for the injection, higher
than normal flow rates and no data storage. You can then
immediately apply these set points to the instrument and
interactively stop the run when the column is ready.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 53
2 Instrument Preparation
Column equilibration and conditioning
Conditioning
You can condition a column in one of three ways:
• With one of the first two procedures described in the
Equilibration paragraphs.
You enter pump conditions to bring the column to its initial
condition. You can also condition the column by setting a
post-run time in the method.
• With a script in a worklist
The LC conditioning script, SCP_LCCondition, is part of the
Data Acquisition software. When you enter the script into the
worklist, you specify the method that you will use for the run.
If a TOF or Q-TOF is connected to the LC, you can also enter
a parameter that diverts the LC eluent to waste. With this
script, there is no injection and no data storage.
54 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Instrument Preparation 2
TOF and Q-TOF preparation – calibration and tuning
Agilent Auto Tunes for the Q-TOF include Quick Tune, Initial
Tune, Mass Calibration / Check, Standard Tune, Transmission
Tune, FPS System Tune, System Tune, and Set Detector Gain.
Some of these tunes only work for tuning the TOF part of the
instrument. All of the TOF auto tune tools perform both
automatic calibration and tuning.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 55
2 Instrument Preparation
TOF mass calibration
56 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Instrument Preparation 2
TOF mass calibration
Figure 22 Tune & Calibration tab of the Tune window for a 6545 Q-TOF
See Chapter 3, “Methods with During sample analysis the system corrects the calibration with
Acquisition Parameters,” starting the introduction of a standard containing reference masses, if
on page 81, to learn more about you enable the correction through the method.
mass correction using reference
standards. The calibration equations used are proprietary.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 57
2 Instrument Preparation
Tuning choices
Tuning choices
You can see the tuning choices available to you on the Tune &
Calibration tab (Figure 24). Notice that you must specify the
part of the instrument to tune for the Q-TOF instrument. Also,
starting with the B.02.01 release, not all of the tuning choices
are available with all sources.
With the ESI, Dual ESI, Dual AJS ESI, Multimode, APPI and
APCI ion sources, you can run Check Tune, Quick Tune, and
TOF Mass Calibration. You can only run Standard Tune, Set
Detector Gain and Initial Tune (TOF) if one of the supported
sources in Table 2 is installed. The Dual ESI source can run all
tuning choices with all instruments.
6530A/B Q-TOF ESI, Dual ESI, AJS ESI, Dual AJS ESI
6540A/B Q-TOF ESI, Dual ESI, AJS ESI, Dual AJS ESI
58 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Instrument Preparation 2
Tuning choices
The Mass Range also affects which tunes are supported. You
can run all tunes if the Mass Range is Standard (3200 m/z). If
the Mass Range is Low (1700 m/z), you cannot run an Initial
Tune or a Set Detector Gain. If the software is version B.05.01
SP1 or later, it is possible and recommended for small molecule
applications to perform an additional Quad Tune in the 1700
m/z range. Doing this leads to a performance increase for
molecules < 300 m/z. If you have an Agilent 6545 Q-TOF and you
select Low (1700 m/z) for the Mass Range, then you have three
options for tuning for this Mass Range plus an option to tune
for Fragile Ions for the lowest Mass Range.
All the automatic tuning choices calibrate the TOF using eight
to ten masses, except for the 1700 mass range, which calibrates
using six masses.
Figure 24 Tune & Calibration tab for a 6545 Q-TOF instrument with Fast
Polarity Switching enabled
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 59
2 Instrument Preparation
Tuning choices
Table 2 on page 58 shows all of the sources that you can use to
run an Initial Auto Tune. On all instrument models, you can
perform an Initial Tune with a Dual ESI source.
60 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Instrument Preparation 2
Tuning choices
Instrument Time for Initial Auto Tune Time for Initial Auto Tune
(TOF) (Quadrupole)
During the Initial Auto Tune process for the TOF, the system
goes through the following steps without your intervention.
a Resets any current tune parameters to original defaults
b Performs a coarse TOF mass axis calibration
c Adjusts all the tunable parameters automatically
d Performs a final TOF mass axis calibration
e Prints a tune report at the end of the Autotune process
During the Initial Auto Tune process for the quadrupole, the
system goes through steps a, c, d and e.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 61
2 Instrument Preparation
Tuning choices
• APCI
Standard Tune
These tools perform many of the same optimization operations
as the Initial Tune tools but use the current settings as starting
values and does a limited set of ion optic ramps.
Table 2 on page 58 shows all of the sources that you can use to
run a Standard Tune. On all instrument models, you can
perform a Standard Tune with a Dual ESI source.
The Standard Tune for the Quadrupole and for the TOF part of
the instrument do subsets of the Initial Auto Tune actions, and
they use the current tune parameters as starting points rather
than using the factory defaults.
62 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Instrument Preparation 2
Tuning choices
• APPI
• APCI
During Quick Tune the system goes through the following steps:
a Automatically opens the calibrant valve
b Adjusts the vertical Q (Vert Q) and bottom slit parameters
in the Beam Shaping Optics (optimizes transmission)
c Adjusts the middle mirror of the compound TOF ion
mirror (optimizes resolution)
d Does a mass axis calibration after the final adjustment
e Prints a tune report
Table 2 on page 58 shows all of the sources that you can use to
run a Set Detector Gain. On all instrument models, you can
perform a Set Detector Gain with a Dual ESI source.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 63
2 Instrument Preparation
Tuning choices
If you select Low (1700 m/z) for the Mass Range, then you have
three additional low mass tune options. You can select a smaller
range to use when tuning. If you select one of the two lower
mass ranges for the auto tune, you can also mark whether or not
there are Fragile Ions.
64 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Instrument Preparation 2
Tuning choices
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 65
2 Instrument Preparation
Tuning choices
Figure 27 SWARM multi dimensional tuning to find the best combination of all parameters
Manual Tune
See the online Help for detailed instructions to help you
manually tune the TOF or Q-TOF instrument. Only perform a
manual tune if the Autotune options produce a result that fails.
66 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Instrument Preparation 2
Tuning choices
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 67
2 Instrument Preparation
Tune reports
Tune reports
At the end of every Check Tune, Quick Tune, Standard Tune,
Transmission Tune, or System Tune, the system generates a
printable Tune report in Excel. The TOF and Quad tune reports
let you know if optimization limits are met. For a TOF tune, the
report also lets you know if the mass calibration is satisfactory
or not. To print previous tune reports, you click the Tune
Report button in the Tune & Calibration tab.
68 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Instrument Preparation 2
Storage and retrieval of tune results and Instrument Mode
Instrument Modes
You select from four different instrument modes.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 69
2 Instrument Preparation
Storage and retrieval of tune results and Instrument Mode
70 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Instrument Preparation 2
Tune Set Point Modifications for Medium and Large Proteins
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 71
2 Instrument Preparation
Real-time displays
Real-time displays
On—Ready Green
72 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Instrument Preparation 2
Real-time parameter values (Actuals)
The parameters and states for each module listed below are
available for display.
See the online Help for
descriptions of each of these Table 4 Actuals available for display
parameters and states.
Module Parameter or State
All modules Run Time, Run State, Ready State, Ready Type, Not
Ready Text Long, Not Ready Text Short, Rawdata State,
Error State
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 73
2 Instrument Preparation
Real-time parameter values (Actuals)
WPS, µWPS, h-ALS, The same as ALS except no Vial and addition of Drawn
h-ALS-SL and Volume, Sample Position and Needle Position
h-ALS-SL+
Capillary Pump Same as binary pump with no Fill A1, A2, B1, B2 and
with the addition of Solvent Ratio C, Solvent Ratio D, Fill
C, Fill D, Primary Flow, EMPV, Purge Status, Purge
Channel, Purge Time and Pump Op Mode
Thermostatted Left Temp, Left Temp Set, Right Temp, Right Temp set,
Column Column Valve, Therm Power
Compartment
74 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Instrument Preparation 2
Real-time Chromatogram Plot and Spectral Plot windows
Signal plots Change range of intensity or time, freeze and zoom the
plot
Pump parameters Change range of intensity or time, freeze and zoom the
plot
TOF or Q-TOF spectra Autoscale the axes, freeze and zoom the plot; toggle
between a line spectrum and a profile spectrum; toggle
between a mass x-axis and a time x-axis, change to a
display of DAD spectra (from line spectra only)
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 75
2 Instrument Preparation
Real-time Chromatogram Plot and Spectral Plot windows
The default display is to plot abundance vs. mass, but you can
change the x-axis to time. To do this, right-click the Spectrum
Plot window and click Show Time from the shortcut menu. Click
Show Mass to return to the Abundance vs. Mass plot.
76 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Instrument Preparation 2
System logbook
System logbook
What you can view in the system logbook
The system logbook does not list The Logbook Viewer displays the dates and times when system
any changes to a method or events take place:
worklist. • Run starts, stops and aborts
• Method loaded
• Tuning and calibration operations
• Unexpected software errors
• Device driver errors, warnings and alerts including leak
detection, vial not found, lamp burned out
• Device powered off or reconfigured
• Start up and shutdown
• Worklist events
• Method log files
When you open the Logbook Viewer, the software loads and
displays the most recent entries in the log file.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 77
2 Instrument Preparation
System logbook
Event Source The module that produced the event (Worklist, Instr Mgr,
App UI, Launcher, DA Mgr)
Most of the tasks that you can do with the logbook help you
view the entries you need to see more easily.
78 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Instrument Preparation 2
System logbook
Change how long entries are saved in Tools > Purge Settings
the logbook
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 79
2 Instrument Preparation
System logbook
80 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System
3
Methods with Acquisition Parameters
Parameter entry 82
General TOF and Q-TOF parameters 85
TOF and Q-TOF acquisition parameters 88
Ion source parameters 86
Setup of TOF and Q-TOF reference mass correction (recalibration) 95
TOF and Q-TOF chromatogram setup 99
Setting parameters to acquire a data file in All Ions MS/MS
mode 102
Method saving, editing and reporting 105
Saving a method with data acquisition parameters 105
Method editing 107
Method reporting 108
See the online Help for You use this chapter to learn about the concepts that can help
instructions on setting up methods you set up methods containing data acquisition parameters.
and parameter descriptions.
81
3 Methods with Acquisition Parameters
Parameter entry
Parameter entry
LC parameter entry
Most of the LC parameter entries are the same as those that you
can change with the Agilent 1200 LC control module and with
other Agilent software products, such as Agilent ChemStation.
82 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Methods with Acquisition Parameters 3
Automatic TOF and Q-TOF parameter changes during a run
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 83
3 Methods with Acquisition Parameters
Acquisition tab
Per time segment set points are labeled with (Seg). Per
experiment set points are labeled with (Expt). All other set
points are per run. These set points can be found in these tabs.
• General Tab
• Source Tab
• Acquisition Tab
• Advanced Parameters Tab (for 6560 only) - These parameters
also can be different for different time segments.
Acquisition Mode
For the Agilent 6560 Ion Mobility Q-TOF, you can toggle between
IM-QTOF and QTOF-Only mode. In QTOF-Only mode, the
trapping funnel does not trap ions, but it simply lets all ions
pass through into the drift tube continuously.
Acquisition tab
You can set up time segments and experiments for changing
these parameters.
84 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Methods with Acquisition Parameters 3
General TOF and Q-TOF parameters
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 85
3 Methods with Acquisition Parameters
Ion source parameters
86 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Methods with Acquisition Parameters 3
Ion source parameters
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 87
3 Methods with Acquisition Parameters
TOF and Q-TOF acquisition parameters
Figure 38 Acquisition tab for the 6550 in the Method Editor window
Figure 39 Acquisition tab for the 6560 in the Method Editor window
88 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Methods with Acquisition Parameters 3
TOF and Q-TOF acquisition parameters
• If you know the compounds and hence the precursor ions you
are looking for, click Targeted MS/MS (Seg) mode.
MS mode
In this mode the Q-TOF instrument behaves solely as a TOF
instrument with no quad isolation applied. You specify the mass
range and the acquisition rate and time to collect spectra.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 89
3 Methods with Acquisition Parameters
TOF and Q-TOF acquisition parameters
Q-TOF
Front End
detector
ion pulser
90 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Methods with Acquisition Parameters 3
TOF and Q-TOF acquisition parameters
For general method development you set the slope and offset for
a curve whose collision energy the system selects to match the
precursor ion, m/z, on the curve.
Targeted List For compounds for which you already know the
precursor ions, you can place the collision energy in the
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 91
3 Methods with Acquisition Parameters
TOF and Q-TOF acquisition parameters
targeted list for the precursor ion and vary its value to optimize
the abundance of the product ion.
92 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Methods with Acquisition Parameters 3
TOF and Q-TOF acquisition parameters
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 93
3 Methods with Acquisition Parameters
TOF and Q-TOF acquisition parameters
94 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Methods with Acquisition Parameters 3
Setup of TOF and Q-TOF reference mass correction (recalibration)
If you have an Agilent 6560 Ion Mobility Q-TOF, you can only
enable reference mass correction if the Acquisition Mode is
QTOF-Only. If the Acquisition Mode is IM-QTOF, then you
instead can use the IM-MS Data File Reprocessing Utility. See
the online Help for the IM-MS Browser program for more
information on this utility.
You can introduce the reference standard into the TOF system
in one of three ways:
• A calibrant delivery system (CDS) that automatically
introduces the standard to a reference sprayer (second
sprayer)— Dual ESI (0.2 to 1.5 mL/min flow rates only), Dual
AJS ESI and Dual nano ESI
• An external dispense pump—other ion sources
• Addition to the sample—other ion sources
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 95
3 Methods with Acquisition Parameters
Setup of TOF and Q-TOF reference mass correction (recalibration)
If you mark the Enable check box, the system uses reference
masses of the mass reference standard for automatic
recalibration of each acquired spectrum.
You must select the reference masses for each polarity in the
method. Reference masses are not only ion source dependent,
but also polarity dependent. To adjust both the a and t0 terms
requires at least two reference masses, spread across the
application mass range, such that the first reference mass is
<330 m/z and the second is 500 m/z or more above the first.
Using more reference masses improves recalibration results. Up
to 20 masses can be selected.
96 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Methods with Acquisition Parameters 3
Setup of TOF and Q-TOF reference mass correction (recalibration)
If you mark the Use Bottle A check box, the TOF or Q-TOF
controls the valve to introduce the internal reference standard
to the reference sprayer via the CDS.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 97
3 Methods with Acquisition Parameters
Setup of TOF and Q-TOF reference mass correction (recalibration)
The software averages only spectra from the same scan group,
and therefore a spectral average spans a spectral cycle.
A spectral cycle contains one spectrum for each of the
Experiments defined in a given Segment.
98 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Methods with Acquisition Parameters 3
TOF and Q-TOF chromatogram setup
Signals above 200,000 counts for the reference mass ion can
negatively affect the reference mass correction.
You can select the signal to plot (TIC, EIC, BPC, Set point,
Actual), the experiment type, the offset of the baseline for the
plot and the valid range of values in counts. You can also set the
mass range for the EIC.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 99
3 Methods with Acquisition Parameters
Q-TOF Advanced Parameters for Ion Mobility
100 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Methods with Acquisition Parameters 3
Q-TOF Advanced Parameters for Ion Mobility
IM Front Funnel The Trap Funnel RF is the most critical parameter for all
applications, as a too low value leads to defocused ions,
resulting in a loss in sensitivity. A value that is too high can lead
to excessive ion-heating, resulting in fragmentation of labile
molecules, or structural changes of molecules. A minor effect is
observed from the High Pressure Funnel RF; a lower value is
typically used for small and labile compounds.
IM Trap Trap grids low are the values when pulsing the ions out of the
trap. These values are the values when operating the
instrument in Q-TOF mode. Usually the exit grids do not need
any further adjustment from the autotune delivered with the
instrument. The high voltages (gray) are the sums of the grid
delta (top values) and the low values; these are the actual
applied voltages during trapping. Trap Exit Grid 2 Delta needs
to be reduced for small molecules and increased for large
molecules. In some cases, it might be advantageous to decrease
the total DC field in the trap by further reducing the delta
between Trap Exit and Trap Entrance (to as low as 0.5V)
combined with lowering the Trap Entrance Grid Low.
IM Drift Tube The best IM-MS resolution is achieved at maximum field
strength. This is calculated as the effective field ((Drift Tube
Entrance Voltage minus Drift Tube Exit Voltage) divided by
drift tube length). For accurate determination of collision cross
sections, this voltage needs to be changed stepwise. See the
Familiarization Guide for more information.
IM Rear Funnel For IM Rear Funnel parameters, the RF need to be reduced for
labile molecules. Depending on the total decrease, a drop in
IM-MS resolution might occur and will be part of the individual
compound optimization. In general, a delta of 2V between the
Rear Funnel Exit and IMS Hex Entrance should be maintained
for almost all applications. The total value of these voltages is
determined by the Octopole Entrance Lens; it should be another
delta of 1 to 2V between these.
All IM-MS parameters described in the Tune window are
accessible in the Advanced Parameters tab of the Q-TOF tab.
The table contains the Use Method column. If you mark this
column, then the value you specify is used instead of the value
in the tune file. Different samples require different parameters
to be changed from the tune file.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 101
3 Methods with Acquisition Parameters
Setting parameters to acquire a data file in All Ions MS/MS mode
The Agilent 6560 Ion Mobility Q-TOF can only have one fixed
collision energy in IM-QTOF mode.
102 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Methods with Acquisition Parameters 3
Setting parameters to acquire a data file in All Ions MS/MS mode
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 103
3 Methods with Acquisition Parameters
Setting parameters to acquire a data file in All Ions MS/MS mode
When you are setting up the experiments, you have the first
experiment have the low energy value, and then you increase
the Collision Energy or Fragmentor voltage for each
successive experiment. So, the highest Collision Energy or
Fragmentor voltage is in the last experiment.
104 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Methods with Acquisition Parameters 3
Method saving, editing and reporting
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 105
3 Methods with Acquisition Parameters
Saving a method with data acquisition parameters
106 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Methods with Acquisition Parameters 3
Method editing
You can view the name and path of the currently loaded method
in the Properties tab of the Method Editor window (Figure 53).
Method editing
You can edit methods containing data acquisition parameters
using the Method Editor window of the Data Acquisition
software.
Use this location for method development. The set points are
sent to the instrument when the method is loaded and when
you click the Apply button after changing a parameter.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 107
3 Methods with Acquisition Parameters
Method reporting
Method reporting
You can see the parameters in a method in one of three ways:
• You click Method Report from the File > Print menu.
• You review the parameters in the Method Editor.
• You can see method parameters associated with a data file in
the Agilent MassHunter Workstation Qualitative Analysis
software.
108 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System
4
Data Acquisition
Interactive single sample setup 110
Worklist setup 112
Data acquisition for samples and worklists 128
109
4 Data Acquisition
Interactive single sample setup
Sample information
The sample information that the system records with the data
file is the sample name, vial position and other information in
the Sample Run window, such as Sample Type and Injection
Volume.
Injection Volume
You can specify an injection volume for the sample in this box.
You select the value As Method if you want to use the injection
volume specified in the method.
110 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Data Acquisition 4
Data File information
Auto Increment If you want to use the same file name and have
the system automatically change the number at the end of the
file name, you turn auto-increment on. Then, you type a file
name that ends in the number 001, and the system makes sure
that a new file is created every time you re-run that sample.
Other folders
You can save your data to any folder on the system. You must
use the browse button to select a different folder. To open any
data file saved to a different folder than D:\MassHunter\Data,
you must browse to the different folder.
Method Type
Acquisition Only, DA Only, or Both Acquisition and DA.
Run Types
This value is not shown by default. You can add it by clicking the
+ button and adding it using the Add Parameters dialog box.
Standard Start for LC runs
Manual Run for infusion runs with a syringe pump
External Start for AP-MALDI
LC Only Run for when you do not acquire MS data
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 111
4 Data Acquisition
Worklist setup
Worklist setup
Agilent developed worklists for the primary purpose of running
many samples automatically and then reporting on the
compounds found in the samples.
Selected Each line in the worklist table has a check box that allows you
sample to mark that line for processing or not. This can be useful if you
execution need to restart a worklist after some lines have already been
run.
112 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Data Acquisition 4
Worklist menus
Worklist menus
You find all the tasks to create a worklist in the worklist menus.
• Add a single sample one at a time or add multiple single
samples all at once
• Add scripts before or after the worklist or between samples
in a worklist
• Add or show more sample information columns
• Add, insert or delete rows and columns
• Set up to print a worklist report or track a worklist run
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 113
4 Data Acquisition
Sample entry
Sample entry
One-at-a-time entry
You may want to do this to equilibrate the system before
running a worklist.
114 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Data Acquisition 4
Sample entry
Sample You can select the sample positions without having to type in
Position their values from the Sample Position tab on the Add Multiple
Samples dialog box.
Figure 58 Sample Position tab of the Add Multiple Samples dialog box
Sample methods
You can create a .m method containing either acquisition
parameters, data analysis parameters or both. See the Quick
Start Guide or the online Help for instructions for creating the
method.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 115
4 Data Acquisition
Script entry
Script entry
Scripts are special programs, that You can enter scripts to be run at the following times:
execute automatically. Agilent • Before or after samples as part of the method
includes scripts with the Agilent
MassHunter Workstation The sample method can include pre- and post-analysis
Software, and you can write your scripts. (See Chapter 3, Acquisition Methods)
own scripts. • Before or after samples in the worklist or batch (insert or
add scripts, respectively)
• Before or after a worklist and after data acquisition
(Figure 63 on page 123)
116 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Data Acquisition 4
Entry of additional sample information (show, add columns)
You can access these capabilities You can add more columns in one of two ways:
through the worklist menu. • Show or hide columns that contain sample information
already available in the worklist
• Add columns for new sample information
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 117
4 Data Acquisition
Worklist import
Worklist import
You can populate a worklist with sample information from other
files in multiple ways:
• Use the Study Manager program which automatically creates
a worklist when you submit a study.
• Copy individual columns one at a time from an Excel
spreadsheet (or a csv file imported into Excel) and paste (or
fill) them into the TOF or Q-TOF worklist under the correct
header
You do this when you need to transfer information
infrequently or the information is different for each transfer.
• Import a csv file directly
You cannot import a partial list of You do this when you need to use the same parameters in a
the samples within the csv file. worklist frequently.
118 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Data Acquisition 4
Worklist import
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 119
4 Data Acquisition
Worklist import
120 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Data Acquisition 4
Worklist import
DA DataAnalysis Method
[Dynamic Mapping]
(//Added column)
InternalStdA Compound 1
Caffeine Compound
[Data Value
Mapping]
Sample Unknown
QualControl QC
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 121
4 Data Acquisition
Report setup
Report setup
After you run the worklist, you can send a worklist report to
one or more of these locations: Screen, Printer, Excel File or
PDF File. You specify the report destination and the file path in
the Worklist Report Options dialog box.
You can also specify to print all of the columns that are part of
the table or to print only the visible columns, and you can select
a different Worklist Report template. You can modify the
worklist report template if you know how to modify an RDL
report.
122 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Data Acquisition 4
Run setup
Run setup
Before you run a worklist you select parameters for the entire
worklist.
• Start run types and the part of the method to run
• Paths for the acquisition method, data analysis method and
data file
• Whether or not to combine export output when also running
a Qualitative Analysis method
• Scripts to run before or after the worklist
• Free disk threshold
The free disk threshold is the amount of disk space in
gigabytes that must be available before the worklist starts.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 123
4 Data Acquisition
Estimate of worklist file size
Overlapped Injection
You can load a sample into the sample loop during a run that is
still completing. This option allows you to save time that is
required to load the sample before the run.
124 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Data Acquisition 4
Estimate of worklist file size
For an IM-QTOF data file, the typical file size is about 3 times
larger than what the equivalent non-IM LC-MS data file would
have been. A typical data file is very sparse (it has a lot of
zero-abundance points which are not written to disk), but how
sparse depends on the sample and lots of different conditions.
As an example of the size of a very complex data file, a 1 minute
method was run to acquire a raw data file of an injection of an
E-coli digest with collision energy applied; the data file was
around 180 MBytes in size (or around 3 MBytes/sec.) Many data
files are not this large or complex.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 125
4 Data Acquisition
Estimate of worklist file size
Mass range You can reduce data file size by reducing the mass
range of interest. Profile data are stored as abundance for each
flight time. If you restrict the mass range, the number of data
points in the spectrum depends on the mass calibration curve.
The table below shows the number of possible data points for a
given mass range using a typical mass calibration curve. The
Agilent 6540 UHD Accurate-Mass Q-TOF has approximately 25%
more data points.
Note that low masses require
more data points than high masses Table 14 Number of data points per spectrum for given mass ranges
because of the non-linear nature of
Mass range Number of data Number of data Number of data
the time to mass conversion. (Daltons) points (1 GHz) points (2 GHz) points (4 GHz)
126 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Data Acquisition 4
Estimate of worklist file size
Storage of method parameters for each run also affects file size
but to a lesser degree than the file size of the spectrum. They
take up about 10,000 bytes before sample injection starts. For a
10-sample worklist, they take up less than 0.1 MB.
If you have a 2 GHz instrument, the disk space needed for these
samples is between 204 and 345 MB. If you have a 4 GHz
instrument, the disk space needed for these sample is between
408 and 690 MB.
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 127
4 Data Acquisition
Data acquisition for samples and worklists
When you start a worklist run, the first sample row turns dark
blue, indicating that the sample in this row is running and data
is being acquired. When the data acquisition finishes, the first
sample row turns light blue to indicate that Data Analysis is
running.
128 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
Data Acquisition 4
What you can do during a run
Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide 129
4 Data Acquisition
What you can do during a run
130 Agilent 6200 Series TOF and 6500 Series Q-TOF LC/MS System Concepts Guide
www.agilent.com
In This Book
The Concepts Guide
presents “The Big Picture”
behind the Agilent 6200
Series TOF and 6500 Series
Q-TOF LC/MS system to
help you to understand
how to use the TOF and
Q-TOF LC/MS system
components.
This guide includes
concepts for:
• Inner workings of the
TOF and Q-TOF MS
• Instrument Preparation
• Methods with
Acquisition Parameters
• Data Acquisition
Printed in USA
Revision A, December 2014
*G3335-90173*
G3335-90173