Chlorine: Performance

Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

Chlorine Cl 2 No.

8LL

n=1
ppm

8LL
0.1
0.2

0.4

0.6

0.8

1.0
detecting layer (white)

Performance The minimum scale value (0.05ppm) is not printed on the tube, but only the scale line is printed.
Measuring range 0.025 to 0.05 ppm (0.05)to 1.0 ppm 1.0 to 2.0ppm
Number of pump strokes 2(200ml) 1(100 ml) 1/2(50 ml)
Correction factor 0.5 1 2
Sampling time 4 min 2 min 1 min
Detecting limit : 0.005 ppm(2 pump strokes)
Colour change : White → Pale green
Corrections for temperature & humidity : Unnecessary
Relative standard deviation : 10 %(for 0.05 to 0.2 ppm), 5 %(for 0.2 to 1.0 ppm)
Shelf life : 1 year(in the refrigerator)

Reaction principle
Cl2 + ABTS → Reaction product
Possible coexisting substances and their interferences
Substance Concentration Interference Changes colour by itself to
Ammonia No
Carbon monoxide No
No
Carbon dioxide No
Hydrogen chlorine No
Nitrogen dioxide +
Chlorine dioxide + Pale green
Halogen gas(ex. Chlorine) +
Sulphur dioxide No No

Calibration gas generation


Permeation tube method