hp6632b Userguide
hp6632b Userguide
hp6632b Userguide
Agilent Technologies certifies that this product met its published specifications at time of shipment from the factory.
Agilent Technologies further certifies that its calibration measurements are traceable to the United States National
Bureau of Standards, to the extent allowed by the Bureau’s calibration facility, and to the calibration facilities of other
International Standards Organization members.
WARRANTY
This Agilent Technologies hardware product is warranted against defects in material and workmanship for a period
of one year from date of delivery. Agilent software and firmware products, which are designated by Agilent for use
with a hardware product and when properly installed on that hardware product, are warranted not to fail to execute
their programming instructions due to defects in material and workmanship for a period of 90 days from date of
delivery. During the warranty period Agilent Technologies will, at its option, either repair or replace products which
prove to be defective. Agilent does not warrant that the operation for the software firmware, or hardware shall be
uninterrupted or error free.
For warranty service, with the exception of warranty options, this product must be returned to a service facility
designated by Agilent. Customer shall prepay shipping charges by (and shall pay all duty and taxes) for products
returned to Agilent for warranty service. Except for products returned to Customer from another country, Agilent shall
pay for return of products to Customer.
Warranty services outside the country of initial purchase are included in Agilent’s product price, only if Customer
pays Agilent international prices (defined as destination local currency price, or U.S. or Geneva Export price).
If Agilent is unable, within a reasonable time to repair or replace any product to condition as warranted, the
Customer shall be entitled to a refund of the purchase price upon return of the product to Agilent.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the
Customer, Customer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the
environmental specifications for the product, or improper site preparation and maintenance. NO OTHER
WARRANTY IS EXPRESSED OR IMPLIED. AGILENT SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES
OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE THE CUSTOMER’S SOLE AND EXCLUSIVE REMEDIES. AGILENT
SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL
DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
ASSISTANCE
The above statements apply only to the standard product warranty. Warranty options, extended support contacts,
product maintenance agreements and customer assistance agreements are also available. Contact your nearest
Agilent Technologies Sales and Service office for further information on Agilent’s full line of Support Programs.
2
Safety Summary
The following general safety precautions must be observed during all phases of operation of this instrument.
Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety
standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability
for the customer’s failure to comply with these requirements.
GENERAL
This product is a Safety Class 1 instrument (provided with a protective earth terminal). The protective features of
this product may be impaired if it is used in a manner not specified in the operation instructions.
Any LEDs used in this product are Class 1 LEDs as per IEC 825-1.
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme à la norme NMB-001 du Canada.
ENVIRONMENTAL CONDITIONS
This instrument is intended for indoor use in an installation category II, pollution degree 2 environment. It is
designed to operate at a maximum relative humidity of 95% and at altitudes of up to 2000 meters. Refer to the
specifications tables for the ac mains voltage requirements and ambient operating temperature range.
Verify that the product is set to match the available line voltage, the correct fuse is installed, and all safety
precautions are taken. Note the instrument’s external markings described under "Safety Symbols".
To minimize shock hazard, the instrument chassis and cover must be connected to an electrical ground. The
instrument must be connected to the ac power mains through a grounded power cable, with the ground wire firmly
connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding)
conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in
personal injury.
ATTENTION: Un circuit de terre continu est essentiel en vue du fonctionnement sécuritaire de l’appareil.
Ne jamais mettre l'appareil en marche lorsque le conducteur de mise … la terre est d‚branch‚.
FUSES
Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be
used. Do not use repaired fuses or short-circuited fuseholders. To do so could cause a shock or fire hazard.
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be
made only by qualified service personnel.
Instruments that appear damaged or defective should be made inoperative and secured against unintended
operation until they can be repaired by qualified service personnel.
3
SAFETY SYMBOLS
Direct current
Alternating current
Terminal is at earth potential. Used for measurement and control circuits designed to be operated
with one terminal at earth potential.
Terminal for Neutral conductor on permanently installed equipment
On (supply)
Off (supply)
Standby (supply). Units with this symbol are not completely disconnected from ac mains when
this switch is off. To completely disconnect the unit from ac mains, either disconnect the power
cord or have a qualified electrician install an external switch.
In position of a bi-stable push control
The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like,
WARNING which, if not correctly performed or adhered to, could result in personal injury. Do not proceed
beyond a WARNING sign until the indicated conditions are fully understood and met.
The CAUTION sign denotes a hazard. It calls attention to an operating procedure, or the like,
Caution which, if not correctly performed or adhered to, could result in damage to or destruction of part
or all of the product. Do not proceed beyond a CAUTION sign until the indicated conditions are
fully understood and met.
4
DECLARATION OF CONFORMITY
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Complies with the essential requirements of the Low Voltage Directive 73/23/EEC and the EMC
Directive 89/336/EEC (including 93/68/EEC) and carries the CE Marking accordingly.
January 1, 2004
Date Bill Darcy/ Regulations Manager
For further information, please contact your local Agilent Technologies sales office, agent or distributor, or
Agilent Technologies Deutschland GmbH, Herrenberger Straβe 130, D71034 Böblingen, Germany
Revision: B.00.00 Issue Date: Created on 11/24/2003 3:14 Document No. Eagle.11.24.doc
PM
5
Acoustic Noise Information
Herstellerbescheinigung
Manufacturer’s Declaration
This statement is provided to comply with the requirements of the German Sound Emission Directive,
from 18 January 1991.
* Sound Pressure Lp <70 dB(A)
* At Operator Position
* Normal Operation
* According to EN 27779 (Type Test).
Printing History
The edition and current revision of this manual are indicated below. Reprints of this manual containing
minor corrections and updates may have the same printing date. Revised editions are identified by a new
printing date. A revised edition incorporates all new or corrected material since the previous printing
date.
Changes to the manual occurring between revisions are covered by change sheets shipped with the
manual. In some cases, the manual change applies only to specific instruments. Instructions provided on
the change sheet will indicate if a particular change applies only to certain instruments.
This document contains proprietary information protected by copyright. All rights are reserved. No part
of this document may be photocopied, reproduced, or translated into another language without the prior
consent of Agilent Technologies. The information contained in this document is subject to change
without notice.
6
Table of Contents
Warranty Information 2
Safety Summary 3
Acoustic Noise Information 6
Printing History 6
Table of Contents 7
QUICK REFERENCE 9
Agilent 66332A Dynamic Measurement DC Source and
Agilent 6611C-6614C System DC Power Supply 9
The Front Panel - At a Glance 10
Front Panel Number Entry 11
Front Panel Annunciators 12
Immediate Action Keys 12
Front Panel Menus - At a Glance 13
SCPI Programming Commands - At a Glance 14
The Rear Panel - At a Glance 15
GENERAL INFORMATION 17
Document Orientation 17
Safety Considerations 18
Options and Accessories 18
Description 19
Capabilities 19
Front Panel Controls 19
Remote Programming 19
Output Characteristic 20
INSTALLATION 21
Inspection 21
Damage 21
Packaging Material 21
Items Supplied 21
Cleaning 21
Location 22
Bench Operation 22
Rack Mounting 22
Input Connections 23
Connect the Power Cord 23
Output Connections 23
Wire Considerations 23
Current Ratings 23
Voltage Drops 24
Multiple Load Connections 24
Remote Sense Connections 25
Sense Leads 25
Stability 26
OVP Considerations 26
Fast/Normal Operation 27
Inductive Loading 27
Capacitive Loading 27
INH/FLT Connections 28
Controller Connections 30
GPIB Interface 30
RS-232 Interface 30
7
TURN-ON CHECKOUT 31
Introduction 31
Using the Keypad 31
Checkout Procedure 31
In Case of Trouble 34
Error Messages 34
Line Fuse 34
FRONT PANEL OPERATION 35
Introduction 35
Front Panel Description 35
System Keys 37
Function Keys 38
Immediate Action Keys 38
Scrolling Keys 38
Metering Keys 39
Output Control Keys 40
Entry Keys 41
Examples of Front Panel Programming 42
1 - Setting the Output Voltage and Current 42
2 - Querying and Clearing Output Protection 43
3 - Making Front Panel Measurements 43
4 - Programming the Digital Output Port 45
5 - Programming the Output Relay (option 760 only) 46
6 - Setting the GPIB Address and RS-232 Parameters 46
7 - Saving and Recalling Operating States 47
SPECIFICATIONS 49
Specifications 49
Supplemental Characteristics 49
VERIFICATION AND CALIBRATION 53
Introduction 53
Equipment Required 53
Test Setup 53
Performing the Verification Tests 54
Turn-On Checkout 54
Voltage Programming and Measurement Accuracy 55
Current Programming and Measurement Accuracy 55
Performing the Calibration Procedure 59
Front Panel Calibration Menu 59
Front Panel Calibration 59
Calibration Error Messages 62
Changing the Calibration Password 62
Calibration Over the GPIB 63
ERROR MESSAGES 65
LINE VOLTAGE CONVERSION 69
Open the Unit 69
Configure the Power Transformer 69
Install the Correct Line Fuse 69
Close the Unit 70
INDEX 71
8
1
Quick Reference
The Agilent 6631B/6632B/6633B/6634B are 100 Watt, high performance dc power supplies with output
current measurement capability in the microampere range. They are well suited for testing portable
battery-powered products.
Additionally, the combination of bench-top and system features in these dc sources provide versatile
solutions for your design and test requirements.
9
1 – Quick Reference
1 2 3 4
Input
Local Meter
7 8 9
Error OV
+
Prot Address Voltage
CV CC Unr Dis OCP Cal Shift Rmt Addr Err SRQ
4 5 6
_
+ 240 VDC
On Cal Max
5 6 7 8
10
Quick Reference – 1
Use the arrow keys and knob to change voltage or current settings
NOTE The output must be ON to see the displayed values change in Meter mode.
Use the Function keys and knob to change the displayed settings
Use the Arrow keys to edit individual digits in the displayed setting
Increments the flashing digit Moves the flashing digit to the right
Decrements the flashing digit Moves the flashing digit to the left
Use the Function keys and Entry keys to enter a new value
NOTE If you make a mistake, use the Backspace key to delete the number, or press the Meter
key to return to meter mode.
11
1 – Quick Reference
12
Quick Reference – 1
13
1 – Quick Reference
14
Quick Reference – 1
4 3 2 1
5 6 7
15
2
General Information
Document Orientation
This manual describes the operation of the Agilent Model 66332A Dynamic Measurement DC Source
and the Agilent Model 6631B/6632B/6633B/6634B System DC Power Supply. Unless otherwise noted,
both units will be referred to by the description "dc source" throughout this manual. The following
documents are shipped with your dc source:
♦ a User’s Guide (this document), contains installation, checkout, and front panel information
♦ a Programming Guide, contains detailed GPIB programming information
♦ an Agilent VXIplug&play instrument driver, for Windows 95 and Windows NT 4.0
The following Getting Started Map will help you find the information you need to complete the specific
task that you want to accomplish. Refer to the table of contents or index of each guide for a complete list
of the information contained within.
Getting Started Map
Task Where to find information
Installing the unit User’s Guide
Line voltage connections
Computer connections
Load connections
Checking out the unit User’s Guide
Verifying proper operation
Using the front panel
Calibrating the unit
Using the front panel User’s Guide
Front panel keys
Front panel examples
Using the programming interface User’s Guide
GPIB interface Programming Guide
RS-232 interface
Programming the unit using SCPI (and Compatibility) Programming Guide
commands
SCPI commands
SCPI programming examples
Compatibility language
Installing the VXIplug&play instrument driver Programming Guide
17
2 – General Information
Safety Considerations
This dc source is a Safety Class 1 instrument, which means it has a protective earth terminal. That
terminal must be connected to earth ground through a power source equipped with a ground receptacle.
Refer to the Safety Summary page at the beginning of this guide for general safety information. Before
installation or operation, check the dc source and review this guide for safety warnings and instructions.
Safety warnings for specific procedures are located at appropriate places in the guide.
18
General Information – 2
Description
Both the Agilent 66332A Dynamic Measurement DC Source and the Agilent
6631B/6632B/6633B/6634B System DC Power Supply combine two instruments in one unit. It includes
a dc source, which produces dc output with programmable voltage and current amplitude, and a highly
accurate voltage and current meter, with the capability to measure very low-level currents. Additionally,
the Agilent 66332A Dynamic Measurement DC Source has the ability to measure and characterize output
voltage and current of pulse or ac waveforms.
Capabilities
♦ Output Voltage and Current control with 12-bit programming resolution
♦ Extensive measurement capability:
dc voltage and current.
rms and peak voltage and current (Agilent 66332A only).
16-bit measurement resolution (low range accurate down to 2 microamperes).
Triggered acquisition of digitized current and voltage waveforms (Agilent 66332A only).
♦ Front panel control with 14-character vacuum fluorescent display, keypad, and rotary control for
voltage and current settings.
♦ Built-in GPIB and RS-232 interface programming with SCPI command language.
♦ Non-volatile state storage and recall.
♦ Over-voltage, over-current, over-temperature, and RI/DFI protection features.
♦ Extensive selftest, status reporting, and software calibration.
Remote Programming
NOTE: When shipped, Agilent 6631B/6632B/6633B/6634B units are set to the Compatibility
programming language; Agilent 66332A units are set to the SCPI programming
language.
To change the programming language from Compatibility to SCPI, press the front panel
Address key, use ô to scroll to the LANG command, press Ì to select SCPI, then
press Enter. Refer to the Programming Guide supplied with your dc source for further
information about remote programming.
The dc source may be remotely programmed via the GPIB bus and/or from an RS-232 serial port. GPIB
programming is with SCPI commands (Standard Commands for Programmable Instruments), which make
the dc source programs compatible with those of other GPIB instruments. Compatibility commands are
also included to make the dc source compatible with the Agilent 6632A, 6633A, and 6634A Series dc
19
2 – General Information
power supplies (refer to appendix D in the the Programming Guide). Dc source status registers allow
remote monitoring of a wide variety of dc source operating conditions.
Output Characteristic
The dc source’s output characteristic is shown in the following figure. The output of the dc source may be
adjusted to any value within the boundaries shown.
Output
Voltage
CV operating line
Vmax
VSET
CC operating line
2
- +
Output
-Imax -ISET 0 ISET Imax Current
NOTE: If you attempt to operate the dc source beyond its output ratings, the output of the unit
will become unregulated. This is indicated by the UNR annunciator on the front panel.
The output may also become unregulated if the ac input voltage drops below the
minimum rating specified in Appendix A.
20
3
Installation
Inspection
Damage
When you receive your dc source, inspect it for any obvious damage that may have occurred during
shipment. If there is damage, notify the shipping carrier and the nearest Agilent Sales and Support Office
immediately. The list of Agilent Sales and Support Offices is at the back of this guide. Warranty
information is printed in the front of this guide.
Packaging Material
Until you have checked out the dc source, save the shipping carton and packing materials in case the unit
has to be returned. If you return the dc source for service, attach a tag identifying the model number and
the owner. Also include a brief description of the problem.
Items Supplied
The following user-replaceable items are included with your dc source. Some of these items are installed
in the unit.
Table 3-1. Items Supplied
Item Part Number Description
Power Cord contact nearest Agilent A power cord appropriate for your location.
Sales and Support Office
Digital connector 1252-1488A 4-terminal digital plug that connects to the back of the unit.
Output cover 06624-20007 Safety cover that installs over the output screw terminals.
Terminal block screws N/A 6-32 X 3/8 inch screws.
Line Fuse 2110-0055 F4A for 100/120 Vac operation
2110-0002 F2A for 220/230 Vac operation
Feet 5041-8801 feet for bench mounting
User's Guide 5962-8196 Contains installation, checkout, and front panel information.
Programming Guide 5962-8198 Contains detailed GPIB programming information.
Cleaning
Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to
clean internally.
21
3 - Installation
Location
Figure 3-1 gives the dimensions of your dc source. The dc source must be installed in a location that
allows sufficient space at the sides and back for adequate air circulation (see Bench Operation).
NOTE: This dc source generates magnetic fields that may affect the operation of other
instruments. If your instrument is susceptible to operating magnetic fields, do not locate
it in the immediate vicinity of the dc source. Typically, at three inches from the dc
source, the electromagnetic field is less than 5 gauss. Many CRT’s, such as those used in
computer displays, are susceptible to magnetic fields much lower than 5 gauss. Check
susceptibility before mounting any display near the dc source.
Bench Operation
A fan cools the dc source by drawing air in through the sides and exhausting it out the back. Minimum
clearances for bench operation are 1 inch (25 mm) along the sides. Do not block the fan exhaust at the
rear of the unit.
Rack Mounting
The dc source can be mounted in a standard 19-inch rack panel or cabinet. Table 2-1 documents the
Agilent part numbers for the various rack mounting options that are available for the dc source.
Installation instructions are included with each rack mount option.
NOTE: Support rails or an instrument shelf is required when rack mounting units.
22
Installation - 3
Input Connections
Connect the Power Cord
1. Unscrew the line fuse cap from the rear panel and verify that the fuse rating matches what is
specified on the FUSES label on the rear panel. Reinstall the fuse. (See table 3-1 for part numbers.)
2. Connect the power cord to the IEC 320 connector on the rear of the unit. If the wrong power cord
was shipped with your unit, contact your nearest Agilent Sales and Support Office (refer to the list at
the back of this guide) to obtain the correct cord.
NOTE: The detachable power cord may be used as an emergency disconnecting device. Removing
the power cord from the ac input connector will disconnect ac input power to the unit.
Output Connections
The output terminal block has connections for the + and − output, the + and − sense inputs, and an earth
ground terminal. The terminal block screws are 6-32 X 3/8 inch. Remove the safety cover by inserting a
flat bladed screwdriver in the opening on the left side of the cover and pushing the locking tab to the left.
This will release the cover.
Optional front panel binding posts are available to connect load wires for bench operation. The front
panel binding posts are paralleled with the rear panel + and − connections. Before using the front panel
binding posts, make sure that the output terminals are jumpered for local sensing.
NOTE: Front panel binding posts are provided for convenience. Only the rear panel terminals are
optimized for noise, regulation, and transient response as documented in Appendix A.
Wire Considerations
To minimize the possibility of instability on the output,
♦ keep load leads as short as possible
♦ bundle or twist the leads tightly together to minimize inductance
Current Ratings
Fire Hazard To satisfy safety requirements, load wires must be large enough not to overheat when
carrying the maximum short-circuit current of the dc source. If there is more than one
load, then any pair of load wires must be capable of safely carrying the full-rated current
of the dc source.
The following table lists the characteristics of AWG (American Wire Gage) copper wire.
Table 3-2. Ampacity and Resistance of Stranded Copper Conductors
AWG No. Ampacity (in free air) Resistance (at 20 deg. C)
Ω/m Ω/ft
20 8.33 0.0345 0.01054
18 15.4 0.0217 0.00663
16 19.4 0.0137 0.00417
14 31.2 0.0086 0.00262
12 40 0.0054 0.00165
23
3 - Installation
Voltage Drops
The load wires must also be large enough to avoid excessive voltage drops due to the impedance of the
wires. In general, if the wires are heavy enough to carry the maximum short circuit current without
overheating, excessive voltage drops will not be a problem. The voltage drops across the load wires
should be limited to less than two volts. Refer to Table 3-2 to calculate the voltage drop for some
commonly used AWG copper wire.
If cabling considerations require the use of distribution terminals that are located remotely from the dc
source, connect the dc source’s output terminals to the remote distribution terminals by a pair of twisted
or bundled wires. Connect each load to the distribution terminals separately. Remote voltage sensing is
recommended under these circumstances. Sense either at the remote distribution terminals, or if one load
is more sensitive than the others, sense directly at the critical load.
+S + - -S
twist or bundle
each pair
LOAD 1
LOAD 2
LOAD 3
24
Installation - 3
The output terminal block of the unit is shipped configured for local sensing, with the +S terminal
jumpered to +, and the −S terminal jumpered to −. Remove these jumpers when making your sense wire
connections. The terminal block screws are 6-32 X 3/8 inch.
Sense Leads
The sense leads are part of the dc source’s feedback path and must be kept at a low resistance (less than
several ohms) to maintain optimal performance. Connect the sense leads carefully so that they do not
become open-circuited. If the sense leads are left unconnected or become open during operation, the dc
source will regulate at the output terminals, resulting in a 3% to 5% increase in output over the
programmed value. Shorting the sense leads trips the OVP circuit.
NOTE: It is good engineering practice to twist and shield all signal wires to and from the sense
connectors. Connect the shield at the dc source end only. Do not use the shield as one of
the sensing conductors.
remove
+ 240 VDC MAX TO
sense -
jumpers
+S + - -S
twisted or -
shielded pair
LOAD
25
3 - Installation
The overvoltage protection circuit senses voltage near the output terminals, not at the load. Therefore the
signal sensed by the OVP circuit can be significantly higher than the actual voltage at the load. When
using remote sensing, you must program the OVP trip voltage high enough to compensate for the voltage
drop between the output terminals and the load. Also, if the sum of the programmed voltage and the load-
lead drop exceeds the dc source’s maximum voltage rating, this may also trip the OV protection circuit.
Stability
When the unit is configured for remote sensing, it is possible for the impedance of the load wires and the
capacitance of the load to form a filter, which becomes part of the unit’s feedback loop. This can degrade
the unit’s stability and result in poor transient response performance. In extreme cases it may also cause
oscillations. The wiring guidelines previously discussed under "Wire Considerations" will eliminate most
stability problems associated with load lead inductance. If additional measures are required:
♦ keep the load capacitance as small as possible
♦ use larger diameter load wires to reduce resistance
OVP Considerations
The dc source’s OVP circuit contains a crowbar SCR, which effectively shorts the output of the dc
source whenever the OVP trips. If an external voltage source such as a battery is connected across the
output and the OVP is inadvertently triggered, the SCR will continuously sink a large current from the
battery, possibly damaging the dc source.
To avoid this, program the OVP setting to its maximum value to prevent it from inadvertently tripping.
Additionally, an internal fuse is connected in series with the SCR. This fuse will open to prevent large
currents from damaging the SCR. If this internal fuse has opened, The FS status annunciator will be set.
Refer to the Service Manual for instructions about replacing this fuse.
In addition, the OVP circuit’s SCR crowbar has been designed to discharge capacitances up to a specific
limit. These limits are:
Agilent 6631B 127,000 µF.
Agilent 6632B and 50,000 µF.
66322A
Agilent 6633B 20,000 µF.
Agilent 6634B 10,000 µF.
If your load capacitance approaches this limit, it is recommended that you do not intentionally trip the
OVP and discharge the capacitance through the SCR as part of your normal testing procedure, as this
may lead to long-term failure of some components.
26
Installation - 3
Fast/Normal Operation
A switch on the rear of the unit lets you switch between operating in either Fast mode or Normal mode.
When set to Fast mode, this switch disconnects the output capacitor that is located inside the unit. Fast
mode lets you improve or enhance certain operating characteristics; while at the same time degrading
other operating characteristics.
1. In Fast mode, the programming time for voltage programming is faster than for normal operation,
however, output noise is greater.
2. In Fast mode, the absence of the internal output capacitor results in increased output impedance and
therefore, greater stability when driving inductive loads. Conversely, the addition of external
capacitive loads in Fast mode will reduce the stability of the unit during constant voltage operation.
3. In Normal mode, the internal output capacitor helps control peak voltage excursions away from the
the nominal value for sudden changes in load current. In Fast mode, larger peak voltage excursions
will show up at the output of the unit during sudden load current changes.
Capacitive Loading
In Normal mode, the dc source will be stable for many load capacitances, however, large load
capacitances may cause ringing in the dc source’s transient response. If this occurs, the problem may be
solved by increasing or decreasing the total load capacitance.
In Fast mode, the dc source can maintain stability only for small capacitive loads. These limits are:
Agilent 6631B 2.2 µF.
Agilent 6632B and 66322A 1.0 µF.
Agilent 6633B 0.22 µF.
Agilent 6634B 0.10 µF.
Inductive Loading
Fast mode is recommended for driving inductive loads, although you will typically experience stable
conditions in Normal mode. Inductive loads present no loop-stability problems in constant voltage
operation, but may cause problems in constant current operation if the load impedance is outside the
boundaries mentioned below.
The ratio of the load inductance (L) to the sum of the Equivalent Series Resistance (ESR) for the load
(R1) and dc source (Rs), determines whether the source can drive a load without becoming unstable. To
help determine this ratio, the value of Rs is listed as:
Agilent 6631B 0.1 ohm.
Agilent 6632B and 0.2 ohm.
66322A
Agilent 6633B 1.0 ohm.
Agilent 6634B 2.0 ohm.
If the ratio L/(R1 + Rs) is less than 0.005, the dc source can reliably drive the load.
27
3 - Installation
INH/FLT Connections
This rear panel connector, has a fault output port and an inhibit input port. The fault (FLT) output, also
referred to as the DFI (discrete fault indicator) signal in the front panel and SCPI commands, is an open
collector circuit that pulls the positive output low with respect to the negative (chassis referenced)
common. The high impedance inhibit (INH) input, also referred to as the RI (remote inhibit) signal in the
front panel and SCPI commands, is used to shut down the power supply output whenever the INH + is
pulled low with respect to INH (chassis referenced) common.
The connector accepts wires sizes from AWG 22 to AWG 12. Disconnect the mating plug to make your
wire connections.
NOTE: It is good engineering practice to twist and shield all signal wires to and from the digital
connectors. If shielded wire is used, connect only one end of the shield to chassis ground
to prevent ground loops.
Figure 3-4 shows how you can connect the FLT/INH circuits of the dc source.
In example A, the INH input connects to a switch that shorts the Inhibit pin (+) to common whenever it
is necessary to disable output of the unit. This activates the remote inhibit (RI) circuit, which turns off
the dc output. The front panel Prot annunciator comes on and the RI bit is set in the Questionable Status
Event register. To re-enable the unit, first open the connection between pins INH + and common and
then clear the protection circuit. This can be done either from the front panel or over the GPIB/RS-232.
In example B, the FLT output of one unit is connected to the INH input of another unit. A fault
condition in one of the units will disable all of them without intervention either by the controller or
external circuitry. The controller can be made aware of the fault via a service request (SRQ) generated by
the Questionable Status summary bit. Note that the FLT output can also be used to drive an external relay
circuit or signal other devices whenever a user-definable fault occurs.
28
Installation - 3
FLT INH
....
NOTE: Connectors + - +
are removable
1 2 3 4
FLT Output
FLT INH
+ - +
1 2 3 4
1 2 3 4
INH Input INH Common
Switch
(Normally INH Input
Open)
FLT Output
A) INH Example with One Unit B) FLT Example with Multiple Units
29
3 - Installation
Controller Connections
The dc source connects to a controller either through an GPIB or an RS-232 connector.
GPIB Interface
Each dc source has its own GPIB bus address, which can be set using the front panel Address key as
described in chapter 5. GPIB address data is stored in non-volatile memory. The dc source is shipped
with its GPIB address set to 5.
Dc sources may be connected to the GPIB interface in series configuration, star configuration, or a
combination of the two, provided the following rules are observed:
♦ The total number of devices including the controller is no more than 15.
♦ The total length of all cables used is no more than 2 meters times the number of devices connected
together, up to a maximum of 20 meters. (Refer to table 2-2 for a list of GPIB cables available from
Agilent Technologies.)
♦ Do not stack more than three connector blocks together on any GPIB connector.
♦ Make sure all connectors are fully seated and the lock screws are firmly finger-tightened.
RS-232 Interface
The dc source has an RS-232 programming interface, which is activated by commands located in the
front panel Address menu. All SCPI and COMPatibility commands are available through RS-232
programming. When the RS-232 interface is selected, the GPIB interface is disabled.
The RS-232 connector is a DB-9, male connector. Adapters are available to connect the dc source to any
computer or terminal with a properly configured DB-25 connector (see Table 2-2).
1 2 3 4 5
6 7 8 9
30
4
Turn-On Checkout
Introduction
Successful tests in this chapter provide a high degree of confidence that the dc source is operating
properly. For verification tests, see appendix B. Complete performance tests are given in the Service
Guide.
NOTE: This chapter provides a preliminary introduction to the dc source front panel. See chapter
5 for more details.
31
4 - Turn-On Checkout
Checkout Procedure
The test in this section checks for output voltage and current on the dc source.
NOTE: To perform the checkout procedure, you will need a wire for shorting the output
terminals together.
The following procedure assumes that the unit turns on in the factory-default state. If you need more
information about the factory default state, refer to the *RST command in chapter 4 of the Programming
Guide. Note that the values shown in the Display column may not exactly match the values that appear
on the front panel of your unit.
If you have not already done so, connect the power cord to the unit and plug it in.
Table 4-1. Checkout Programming Values
Model Voltage OV Prot Current Note:
6631B 8 8.8 10 This checkout procedure is written for models
6632B/66332A 20 22 5 6632B and 66332A. If you have another
6633B 50 55 2 model, enter the correct values from the table
6634B 100 110 1 Where the procedure calls for an <input>.
NOTE: Press the Meter key to exit a menu at any time and return to meter mode. If the Err
annunciator on the display is on, press the Shift key followed by the Error key to see the
error number. See table 4-2 at the end of this chapter.
2. Check that the dc source You should be able to hear the fan and feel the
fan is on air coming from the back of the unit.
3. Press Voltage, VOLT 0.000 Programs the output to the selected voltage. After
<2, 0>, Enter VOLT <20> the value is entered, the display returns to Meter
mode. Because the output has not been enabled,
the meter still indicates 0 volts.
4. Press Output On/Off <20.003V> Turns the output on. The Dis annunciator should
0.0006A be off and CV should be on.
5. Press Shift, OV VOLT:PROT Display shows the overvoltage protection trip
<22.00> voltage for your unit.
32
Turn-On Checkout - 4
33
4 - Turn-On Checkout
In Case of Trouble
Error Messages
Dc source failure may occur during power-on selftest or during operation. In either case, the display may
show an error message that indicates the reason for the failure.
Selftest Errors
Pressing the Shift, Error keys will show the error number. Selftest error messages appear as: ERROR
<n> where "n" is a number listed in the following table. If this occurs, turn the power off and then back
on to see if the error persists. If the error message persists, the dc source requires service.
Table 4-2. Power-On Selftest Errors
Error No. Failed Test
Error 0 No error
Error 1 Non-volatile RAM RD0 section checksum failed
Error 2 Non-volatile RAM CONFIG section checksum failed
Error 3 Non-volatile RAM CAL section checksum failed
Error 4 Non-volatile RAM STATE section checksum failed
Error 5 Non-volatile RST section checksum failed
Error 10 RAM selftest
Error 11 to 14 VDAC/IDAC selftest 1 to 4
Error 15 OVDAC selftest
Error 80 Digital I/O selftest error
Line Fuse
If the dc source appears "dead" with a blank display and the fan not running, check your power source to
be certain line voltage is being supplied to the dc source. If the power source is normal, the dc source
fuse may be defective.
1. Turn off the front panel power switch and unplug the power cord.
2. Remove the fuse from the rear panel.
3. If the fuse is defective, replace it with a fuse of the same type (see "Input Connections" in chapter 3).
4. Turn on the dc source and check the operation.
NOTE: If the dc source has a defective fuse, replace it only once. If it fails again, the dc source
requires service.
34
5
Front panel Operation
Introduction
Here is what you will find in this chapter:
♦ a complete description of the front panel controls
♦ front panel programming examples
NOTE: The dc source must be in set to Local mode to use the front panel controls. Press the
Local key on the front panel to put the unit in local mode.
1 2 3 4
Input
Local Meter
7 8 9
Error OV
+
Prot Address Voltage
CV CC Unr Dis OCP Cal Shift Rmt Addr Err SRQ 4 5 6
_
+ 240 VDC
On Cal Max
5 6 7 8
35
5 - Front Panel Operation
36
Front Panel Operation - 5
System Keys
Refer to the examples later in this chapter for more details on the use of these keys.
SYSTEM
Local
Error
Address
Save
Recall
Shift
This is the blue, unlabeled key, which is also shown as Shift in this guide.
Pressing this key accesses the alternate or shifted function of a key (such as
ERROR ). Release the key after you press it. The Shift annunciator is lit,
indicating that the shifted keys are active.
Local Press to change the dc source’s selected interface from remote operation to local
(front panel) operation. Pressing the key will have no effect if the interface state
is already Local, Local-with-Lockout, or Remote-with-Lockout.
Address Press to access the system address menu. This menu lets you configure the dc
source’s interface. Address Menu entries are stored in non-volatile memory.
Display Command Function
ADDRESS <value> Sets the GPIB Address
INTF <char> Selects an interface (GPIB or RS232)
BAUDRATE<value> Selects baud rate (300, 600, 1200, 2400, 4800, 9600)
PARITY <char> Message parity (NONE, EVEN, ODD, MARK, SPACE)
FLOW <char> Flow control (XON-XOFF, RTS-CTS, DTR-DSR, NONE)
LANG <char>I Selects language (SCPI or COMP)
value = a numeric value
char = a character string parameter
Use p and q to scroll through the command list.
Use Ë and Ì to scroll through the parameter list.
Recall Press to place the dc source into a previously stored state. You can recall up to 4
(0 through 3) previously stored states.
Shift Error Press to display the system error codes stored in the SCPI error queue. This
action also clears the queue. If there is no error in the queue, 0 is displayed.
Shift Save Press to store an existing dc source state in non-volatile memory. The
parameters saved are listed under *SAV in the dc source Programming Guide.
You can save up to 4 states (0 through 3).
37
5 - Front Panel Operation
Function Keys
Refer to the examples later in this chapter for more details on the use of these keys.
FUNCTION
Input
Meter
OV
Voltage
Current Protect
Cal
Output Output
on/off
Scrolling Keys
Scrolling keys let you move through the commands in the presently selected function menu.
ô Press
q to bring up the next command in the list. Press
p to go back
to the previous command in the list. Function menus are circular; you can
return to the starting position by continuously pressing either key. The
following example shows the commands in the Input function menu:
q CURR:RANGE <char>
q CURR:DET <char>
38
Front Panel Operation - 5
Metering Keys
Metering keys control the metering functions of the dc source. When the unit is operating in front panel
meter mode, all front panel measurements are calculated from a total of 2048 readings taken at a 46.8
microsecond sampling rate. Therefore, the total acquisition time for a single front panel measurement is
about 100 milliseconds. Refer to “Making Front Panel Measurements” for more information.
NOTE: You can vary the both the sampling rate and the number of data points in each
measurement when controlling the unit over the GPIB interface. (Refer to chapter 3 in
the Programming Guide).
Meter Press this key to access the meter menu list. Also use this key to exit a menu at
any time and return to meter mode.
Display Measurement
<reading>V <reading>A Measures output dc voltage and current
<reading>V MAX Measures peak output voltage1
<reading>V MIN Measures minimum output voltage1
<reading>V HIGH Measures the high level of a voltage pulse waveform1
<reading>V LOW Measures the low level of a voltage pulse waveform1
<reading>V RMS Measures rms voltage1
<reading>A MAX Measures peak output current1
<reading>A MIN Measures minimum output current1
<reading>A HIGH Measures the high level of a current pulse waveform1
<reading>A LOW Measures the low level of a current pulse waveform1
<reading>A RMS Measures rms current1
Shift Input Press this key to access the following metering functions.
Display Command Function
CURR:RANGE <char> Select current range (AUTO, LOW or HIGH)
CURR:DET <char> Select current measurement bandwidth (ACDC or DC)1
Notes:
1
Available on Agilent Model 66332A in SCPI language mode only.
reading = the returned measurement
value = a numeric value
char = a character string parameter
Use p and q to scroll through the menu commands.
Use Ë and Ì to scroll through the menu parameters.
Use É and Ê to select a digit in a numeric entry field.
39
5 - Front Panel Operation
Shift Cal
This key accesses the calibration menu (Refer to Appendix B to calibrate your
dc source).
Notes:
1
These parameters are stored in non-volatile memory
2
These status summary bits are explained in chapter 3 of the Programming Guide
value = a numeric value
char = a character string parameter
Use p and q to scroll through the menu commands.
Use Ë and Ì to scroll through the menu parameters.
Use É and Ê to select a digit in a numeric entry field.
40
Front Panel Operation - 5
Entry Keys
Refer to the examples later in this chapter for more details on the use of these keys.
ENTRY
7 8 9
4 5 6
Clear Enter
1 2 3 Entry
Back
0 . - space
41
5 - Front Panel Operation
42
Front Panel Operation - 5
When controlling the unit over the GPIB interface, you can vary both the sampling rate and the number
of data points in each measurement. If you are using the Agilent 66332A dc source to measure waveform
data, the GPIB interface also lets you qualify the triggers that initiate the measurements. With this
flexibility, measurement accuracy can be improved for waveforms with frequencies as low as several
Hertz. Refer to chapter 3 in the Programming Guide for more information.
Two current measurement ranges can be selected in the Input menu. A high current range is available for
measuring output currents up to the 30% higher than the maximum rating of the dc source. A low current
range is available for improved resolution when measuring output currents below 20 milliamperes. The
low current measurement range is accurate to 0.1% of the reading ±2.5 microamperes. When the current
Range is set to AUTO, the unit automatically selects the range that provides the best measurement
resolution.
NOTE: If the front panel display indicates OVLD, the output has exceeded the measurement
capability of the instrument. If the front panel display indicates
-- -- -- -- -- -- , an GPIB measurement is in progress.
As previously mentioned, the Agilent 66332A dc source has the capability of measuring output
waveform parameters such as peak, minimum, high level, and low level as illustrated in the following
figure.
43
5 - Front Panel Operation
V or A MAX
46.8 microsecond
sampling rate
V or A HIGH
NOTE: Measurement samples may not
coincide with the actual maximum
or minimum point of the waveform.
V or A LOW
V or A MIN
100 millisecond
acquisition time
Figure 5-5. Front Panel Pulse Measurement Parameters (Agilent 66332A only)
Note: In the Low current measurement range, the current detector is fixed at DC. With the
current detector in dc, accurate current measurements cannot be made on waveforms
with frequency contents over a few kHz.
44
Front Panel Operation - 5
45
5 - Front Panel Operation
46
Front Panel Operation - 5
You can save up to 4 states (from location 0 to location 3) in non-volatile memory and recall them from
the front panel. All programmable settings are saved.
47
A
Specifications
Specifications
Table A-1 lists the specifications of the dc source. Specifications are warranted over the ambient
temperature range of 0 to 55 °C. Unless otherwise noted, specifications apply when measured at the rear
terminals after a 30-minute warm-up period.
49
A - Specifications
Supplemental Characteristics
Table A-2 lists the supplemental characteristics, which are not warranted but are descriptions of typical
performance determined either by design or type testing.
Table A-2. Supplemental Characteristics
Parameter Agilent 6631B Agilent 6632B Agilent 6633B Agilent 6634B
Agilent 66332A
Input Ratings 100 Vac mains: 100 Vac nominal, 50/60 Hz, 3.5 A, 380 VA max.
(at full load) 120 Vac mains: 120 Vac nominal, 50/60 Hz, 3 A, 380 VA max.
220 Vac mains: 220 Vac nominal, 50/60 Hz, 1.6 A, 380 VA max
230 Vac mains: 230 Vac nominal, 50/60 Hz, 1.5 A, 380 VA max
Output Programming Voltage: 0-8.190 V 0−20.475 V 0−51.188 V 0−102.38 V
Range Current: 0-10.237A 0−5.1188 A 0−2.0475 A 0−1.0238 A
OVP: 0−22 V 0−55 V 0−110 V
0-12V
Average Programming Voltage: 2 mV 5 mV 12.5 mV 25 mV
Resolution Current: 2.63 mA 1.32 mA 0.526 mA 0.263 mA
OVP: 100 mV 250 mV 500 mV
60 mV
OVP Accuracy 2.4% + 100 mV 2.4% + 240 mV 2.4% + 600 mV 2.4% + 1.2 V
Maximum Current 14.3 A 6.66 A 2.43 A 1.21 A
Measurement
Average Current High Range: 436 uA 213 µA 74 µA 37 µA
Measurement Resolution Low Range: 0.6 uA 0.6 µA 0.6 µA 0.6 µA
Sink Current -10 A −5 A −2 A −1 A
1
Sink Current Tracking SCPI mode : 0.4% + 4 mA 0.4% + 2 mA 0.4% + 1 mA 0.4% + 0.5 mA
Compatibility mode: -500 mA −250 mA −100 mA −50 mA
Minimum Current in 40 mA 20 mA 8 mA 4 mA
Compatibility Mode
Programming Accuracy Voltage: 0.01% + 0.15 mV 0.01% +0.25 mV 0.01% + 0.5 mV 0.01% + 1 mV
Temperature Coefficient Current: 0.01% + 60 uA 0.01% + 30 µA 0.01% + 12 µA 0.01% + 6 µA
(change/°C) OVP: 0.01% + 2 mV 0.015% + 4 mV 0.015% + 10 mV 0.015% + 20 mV
Readback Accuracy Voltage: 0.01% + 60 uV 0.01% + 150 µV 0.01% + 500 µV 0.01% + 750 µV
Temperature Coefficient Current (ACDC): 0.05% + 320 uA 0.05% + 160 µA 0.05% + 80 µA 0.05% + 40 µA
(change/°C) Current (DC): 0.02% + 50uA 0.02% + 25 µA 0.02% + 10 µA 0.02% + 5 µA
Current (Low range): 0.01% + 0.3 uA 0.01% + 0.3 µA 0.01% + 0.3 µA 0.01% + 0.3 µA
Drift2 Voltage: 0.01% + 0.25 mV 0.01% + 0.5 mV 0.01% + 1 mV 0.01% + 1 mV
Current: 0.01% + 100uA 0.01% + 50 µA 0.01% + 20 µA 0.01% + 10 µA
Output Voltage Rise/Fall Normal mode: 2 ms
Time Fast mode: 400 µs
(for a change from 10% to 90% or 90% to 10% of the total excursion)
Output Voltage Settling Normal mode: 6 ms
Time Fast mode: 2 ms
(to settle within 1 LSB or 0.025% times the rated voltage of the final value)
1
For 8 volt unit, between 0 and 10 mA, the sink current remains at –10 mA.
For 20 volt unit, between 0 and 5mA, the sink current remains at –5 mA.
For 50 volt unit, between 0 and 2.5 mA, the sink current remains at –2.5 mA.
For 100 volt unit, between 0 and 1.25 mA, the sink current remains at –1.25 mA.
2
Following a 30 minute warm-up, the change in output over 8 hours, under ambient temperature,
constant load, and line operating conditions.
50
Specifications - A
51
A - Specifications
52
B
Verification and Calibration
Introduction
This appendix includes verification and calibration procedures for the Agilent 66332A and Agilent
6631B/6632B/6633B/6634B dc source. Instructions are given for performing the procedures either from
the front panel or from a controller over the GPIB.
The verification procedures do not check all the operating parameters, but verify that the dc source is
performing properly. Performance Tests, which check all the specifications of the dc source, are given in
the applicable dc source Service Manual.
Important Perform the verification procedures before calibrating your dc source. If the dc source
passes the verification procedures, the unit is operating within its calibration limits and
does not need to be recalibrated.
Equipment Required
The equipment listed in the following table, or the equivalent to this equipment, is required for
verification and calibration.
Table B-1. Equipment Required
Equipment Characteristics Recommended Model
Digital Voltmeter Resolution: 10 nV @ 1 V Agilent 3458A
Readout: 8.5 digits
Accuracy: >20 ppm
Current Monitor 1 15 A (0.1 Ω), ±0.04%, TC=5ppm/°C Guildline 9230/15
Load Resistor 400Ω (Agilent 6631B calibration, used for
(3 W min. TC=20ppm/°C) verification of all models) p/n 0811-2878
1.1 kΩ (Agilent 6632B/66332A calibration)
2.5 kΩ (Agilent 6633B calibration)
5 kΩ (Agilent 6634B calibration)
Power Supply 8 V @ 10 A Agilent 6631B
GPIB Controller Full GPIB capabilities HP Series 200/300 or
equivalent
1
The 4- terminal current shunt is used to eliminate output current measurement error caused by voltage drops in the
load leads and connections. It has special current-monitoring terminals inside the load connection terminals. Connect
the voltmeter directly to these current-monitoring terminals.
Test Setup
Figure B-1 shows the setup for the tests. Be certain to use load leads of sufficient wire gauge to carry the
full output current (see chapter 3).
53
B - Verification and Calibration
+ 240 VDC MAX TO
+ 240 VDC MAX TO -
-
+S + - -S
+S + - -S
+ DC +
DC Current Load
- -
A. B.
Load Resistor
400 6631B
1.1K 6632B
-
+ 240 VDC MAX TO
& 66332A -
+ 240 VDC MAX TO
+S + - -S 2.5 K 6633B +S + - -S
5 K 6634B
DC + +
Load DC
resistor Current
Ammeter Voltmeter monitor
- -
+ - + -
External External
C. D.
DC supply DC supply
NOTE: The verification procedure can only be performed using the SCPI language commands.
Use either the front panel Address key to access the LANG command, or use the
SYSTem:LANGuage command to change the programming language to SCPI.
The following procedures assume you understand how to operate the dc source from the front panel as
explained in chapter 5.
When performing the verification tests from an GPIB controller, you may have to consider the relatively
slow settling times and slew rates of the dc source as compared to computer and system voltmeters.
Suitable WAIT statements can be inserted into the test program to give the dc source time to respond to
the test commands.
Perform the following tests for operation verification in the order indicated.
1. Turn-On Checkout
2. Voltage Programming and Measurement Accuracy
3. Current Programming and Measurement Accuracy
54
Verification and Calibration - B
Turn-On Checkout
Perform the Turn-On Checkout as directed in chapter 4.
NOTE: The dc source must pass turn-on selftest before you can proceed with the verification
tests.
55
B - Verification and Calibration
6. Divide the voltage drop across the current monitor by its resistance Readings within high current
to convert the value to amperes. Record this value and the current limits (see tables B-3,4, 5 or 6).
reading on the front panel display.
Current Measurement (Low Range)
Action Normal Result
7. Turn off the dc source and connect it as shown in figure B-1B
with the 400 ohm load resistor. Set the DMM to operate in
current mode.
8. Turn on the dc source, access the Input menu, and set the current CURR:RANG LOW
range to LOW.
9. Set the output voltage to 0 V and the current to its full scale value Output current near 0 A.
(see table B-2). Press Output On/Off to enable the output.
10. Record the current reading from the DMM as well as from the Readings within low current
front panel display. The difference between the two readings measurement (see table B-3, 4,
should be within the limits specified in the table. 5 or 6).
11. Set the output voltage to 8 volts. Output current near +20 mA.
12. Record the current reading from the DMM as well as from the Readings within high current
front panel display. The difference between the two readings measurement (see table B-3, 4,
should be within the limits specified in the table. 5 or 6).
Current Sink Measurement
Action Normal Result
13. Turn off the dc source and connect an external supply to the
output of the unit as shown in figure B-1C using a 400 ohm load
resistor. Set the DMM to operate in current mode.
14. Turn on the dc source, access the Input menu, and set the current CURR:RANG LOW
range to LOW.
15. Access the Input menu and set the current sense detector to DC. CURR:DET DC
16. Turn on the external supply and program its output for 8 volts Output current near −20 mA.
and 1 A. Program the dc source to 0 V and 1 A. Press Output
On/Off to enable the output.
17. Record the current reading from the DMM as well as from the Readings within low current
front panel display. Enter the difference between the two readings sink measurement
should be within the limits specified in the table. (see table B-3,4, 5 or 6).
18. Turn off the dc source and connect the current monitor and
external supply to the output of the unit as shown in figure B-1D.
Set the DMM to operate in dc voltage mode.
19. Turn on the external supply and program its voltage to 5 volts Output near −full scale current.
and current for the full scale current rating of the UUT. Program
the dc source to 0 V and full scale current (see table B-2). Press
Output On/Off to enable the output.
20. Divide the voltage drop across the current monitor by its Readings within high current
resistance to convert the value to amperes. Record this value as sink measurement
well as the current reading on the front panel display. The (see table B-3, 4, 5 or 6).
difference between the two readings should be within the limits
specified in the table.
56
Verification and Calibration - B
57
B - Verification and Calibration
58
Verification and Calibration - B
Table B-1 lists the equipment required for calibration. Figure B-1 shows the test setup.
You do not have to do a complete calibration each time. If appropriate, you may calibrate only the
voltage or current and proceed to "Saving the Calibration Constants". However, the voltage or current
calibration sequence must be performed in its entirety. The following parameters may be calibrated:
♦ voltage programming and measurement
♦ overvoltage protection (OVP)
♦ current programming and measurement
♦ negative current programming
♦ low range measurement
♦ ac current measurement
59
B - Verification and Calibration
Action Display
5. Connect the DMM (dc volts mode) directly to the dc source. Do not
connect the load resistor or current shunt.
6. Press Shift Cal, scroll to CAL VOLT, and press Enter. CAL:VOLT
7. Press Shift Cal, scroll to CAL LEV, and press Enter to select the CAL:LEV P1
first calibration point.
8. Press Shift Cal, scroll to CAL DATA, and use the Entry keypad to CAL:DATA 0.00
enter the voltage value displayed on the DMM.
9. Press Shift Cal, scroll to CAL LEV, use Ì to scroll to P2 (the CAL:LEV P2
second calibration point), and press Enter.
10. Press Shift Cal, scroll to CAL DATA, and use the Entry keypad to CAL:DATA 0.00
enter the second voltage value displayed on the DMM.
Action Display
11. Press Shift Cal, scroll to CAL VOLT PROT, and press Enter. CAL:VOLT:PROT
12. Wait for the dc source to compute the OVP calibration constant. The
display returns to Meter mode when the calculation is complete.
60
Verification and Calibration - B
Action Display
13. Connect the appropriate current monitor as shown in figure B-1A.
Connect the DMM (in dc mode) across the current shunt.
14. Press Shift Cal, scroll to CAL CURR, and press Enter. CAL:CURR
15. Press Shift Cal, scroll to CAL LEV, and press Enter to select the CAL:LEV P1
first calibration point.
16. Press Shift Cal and scroll to CAL DATA. Wait for the DMM reading CAL:DATA 0.00
to stabilize. Then read the DMM and compute the first current value
(DMM reading ÷ shunt resistance). Use the Entry keypad to enter the
first current value.
17. Press Shift Cal, scroll to CAL LEV, use ä to scroll to P2 (the CAL:LEV P2
second calibration point), and press Enter.
18. Press Shift Cal and scroll to CAL DATA. Wait for the DMM reading CAL:DATA 0.00
to stabilize. Then read the DMM and compute the second current
value (DMM reading ÷ shunt resistance). Use the Entry keypad to
enter the second current value.
Action Display
19. Connect only an external supply directly to the dc source. Connect the
+ terminal to the + output and the - terminal to the - output. Set the
external supply to 1 V, 1 A. The dc source should be in CC mode.
20. Press Shift Cal, scroll to CAL CURR NEG, and press Enter. CAL:CURR:NEG
21. Wait for the dc source to compute the negative current calibration
constant. The display returns to Meter mode when the calculation is
complete. Disconnect the external supply.
Action Display
22. Press Shift Cal, scroll to CAL CURR MEAS LOW, and press Enter. CAL:CURR:MEAS:LOW
23. Connect the UUT as shown in figure B-1B with the appropriate load
resistor for your model. Set the DMM to operate in current mode.
24. Press Shift Cal, scroll to CAL LEV, and press Enter to select the CAL:LEV P1
first calibration point.
25. Press Shift Cal and scroll to CAL DATA. Wait for the DMM reading CAL:DATA 0.00
to stabilize. Then use the Entry keypad and enter the current reading
displayed on the DMM.
61
B - Verification and Calibration
Action Display
26. Disconnect all loads from the dc source. Make sure the mode switch
on the rear panel is set to Normal.
27. Press Shift Cal, scroll to CAL CURR MEAS AC, and press Enter. CAL:CURR:MEAS AC
28. Wait for the dc source to compute the ac current calibration constant.
The display returns to Meter mode when the calculation is complete.
WARNING: Storing calibration constants overwrites the existing ones in non-volatile memory. If you
are not sure you want to permanently store the new constants, omit this step. The dc
source calibration will then remain unchanged.
Action Display
29. Press Shift Cal, scroll to CAL SAVE, and press Enter. CAL:SAVE
30. Press Shift Cal, select CAL OFF, and press Enter to exit Calibration CAL OFF
mode. *RST and *RCL will also set the calibration state to OFF.
62
Verification and Calibration - B
NOTE: If you want the calibration function to operate without requiring any password, change
the password to 0 (zero).
The SCPI calibration commands are explained in chapter 3 of the dc source Programming Guide.
Calibration error messages that can occur during GPIB calibration are shown in table B-3.
63
C
Error Messages
The following table lists the errors that are associated with SCPI syntax errors and interface problems. It
also lists the device dependent errors. Information inside the brackets is not part of the standard error
message, but is included for clarification.
When errors occur, the Standard Event Status register records them in bit 2, 3, 4, or 5 as described in the
following table:
65
C - Error Messages
66
Error Messages - C
Selftest Errors 0 through 99 (sets Standard Event Status Register bit #3)
0 No error
1 Non-volatile RAM RD0 section checksum failed
2 Non-volatile RAM CONFIG section checksum failed
3 Non-volatile RAM CAL section checksum failed
4 Non-volatile RAM STATE section checksum failed
5 Non-volatile RST section checksum failed
10 RAM selftest
11 VDAC/IDAC selftest 1
12 VDAC/IDAC selftest 2
13 VDAC/IDAC selftest 3
14 VDAC/IDAC selftest 4
15 OVDAC selftest
80 Digital I/O selftest error
Device-Dependent Errors 100 through 32767 (sets Standard Event Status Register bit #3)
213 Ingrd receiver buffer overrun
216 RS-232 receiver framing error
217 RS-232 receiver parity error
218 RS-232 receiver overrun error
220 Front panel uart overrun
221 Front panel uart framing
222 Front panel uart parity
223 Front panel buffer overrun
224 Front panel timeout
401 CAL switch prevents calibration
402 CAL password is incorrect
403 CAL not enabled
404 Computed readback cal constants are incorrect
405 Computed programming cal constants are incorrect
406 Incorrect sequence of calibration commands
407 CV or CC status is incorrect for this command
408 Output mode switch must be in NORMAL position
601 Too many sweep points
602 Command only applies to RS-232 interface
603 CURRent or VOLTage fetch incompatible with last acquisition
604 Measurement overrange
67
D
Line Voltage Conversion
WARNING: Shock Hazard Operating personnel must not remove instrument covers.
Component replacement and internal adjustments must be made only by qualified
service personnel.
spare jumper
jumper
grey
grey
white/red/grey
white/red/grey
jumper
jumper
69
D - Line Voltage Conversion
List of Fuses
F500 2110-0055 1 Fuse, F4A, 250V (for 100/120 Vac input)
F500 2110-0002 1 Fuse, F2A, 250V (for 220/230 Vac input)
F300 2110-0712 1 Fuse, sub-min, M4A, 125V
F301 2110-0712 1 Fuse, sub-min, M4A, 125V
F302 2110-0697 1 Fuse, sub-min, M15A, 32V (66332A, 6632B, 6633B)
F302 2110-0685 1 Fuse, sub-min, M7A, 125V (6634B)
F303 2110-0697 1 Fuse, sub-min, M15A, 32V (66332A, 6632B, 6633B)
F303 2110-0685 1 Fuse, sub-min, M7A, 125V (6634B)
F304 2110-0699 1 Fuse, sub-min, M5A, 125V
F305 2110-0777 1 Fuse, M8A, 32V (66332A, 6632B, 6633B)
F305 0811-3776 1 Res Fusible 0.5 Ohm 5W (6634B)
F310 2110-0025 1 Fuse, M15A, 32V (6631B)
70
Index
cleaning, 21
clear protection, 43
—— controller connections, 30
-- -- -- -- -- --, 43 controls and indicators, 19
conversion, ac line, 69
crowbar circuit, 26
—+— current measurement range, 43
+/- terminals, 23 CV mode, 20, 42
+S/-S terminals, 23
—D—
—0— damage, 21
0 ... 9, 41 DC current detector, 44
DFI signal, 28
DIGIO, 45
—A— digital connector, 21, 28
digital I/O, 28
ac line conversion, 69
digital output port, 45
accessories, 18
dimensions, 22
ACDC current detector, 44
disabling multiple units, 28
airflow, 22
annunciators
Addr, 36 —E—
Cal, 36
CC, 36 enable output, 42
CV, 36 enter value, 42
entry keys, 41
É
Dis, 36
Err, 36 , 41
OCP, 36 0 ... 9, 41
Prot, 36 Backspace, 41
Rmt, 36 Clear Entry, 41
Shift, 36 Enter Number, 41
SRQ, 36 error messages, 34
Unr, 36 error numbers, 65
AWG ratings, 23
—F—
—C— fast mode, 27
cables, 18 features, 19
calibration, 59 FLT, 45
current programming - high range, 61 FLT output, 28
current programming - low range, 61 front panel, 19, 35
enable, 60 annuncuiators, 36
equipment, 53 connections, 23
error messages, 62 controls and indicators, 35
menu, 59 keys, 36
OVP, 60 FS, 38, 43
password, 63 FS status, 26
peak current, 62 function keys, 38
saving, 62 p, 38
setup, 53 Cal, 40
voltage measurement, 60 Current, 40
voltage programming, 60 immediate action, 38
capacitive loading, 27 Input, 39
capacitor discharge limit, 26 Meter, 39
CC mode, 42 OCP, 38
checkout procedure, 32 Output, 40
71
Index
Output On/Off, 38 OT, 43
OV, 40 output
Prot Clear, 38 characteristic, 20
Protect, 40 connections, 23
Voltage, 40 connector, 21
fuses, 23 control keys, 40
current setting, 42
enable, 42
—G— rating, 20
ground, earth, 18 voltage setting, 42
guide, programming, 17 output relay, 46
guide, user’s, 17 OV, 43
OVLD, 34, 43
OVP
—H— capacitor discharge limit, 26
history, 6 circuit, 26
GPIB, 46
address, 46 —P—
connections, 30
interface, 30 power cord, 21, 23
power receptacle, 18
preliminary checkout, 31
—I— print date, 6
inductive loading, 27 protection
INH, 45 FS, 43
INH input, 28 OC, 43
input OT, 43
connections, 23 OV, 43
power, 18 RI, 43
inspection, 21
—Q—
—L— query protection, 43
line fuse
replacing, 34 —R—
line voltage conversion, 69
load line, 20 rack mount kit, 18
load voltage drops, 24 rack mounting, 22
location, 22 rear panel
connections, 28, 30
recalling operating states, 47
—M— remote programming, 19
making measurements, 43 remote sensing, 25
manuals, 21 stability, 26
measurement ranges, 43 repacking, 21
multiple loads, 24 RI, 43
signal, 28
RIDFI, 45
—N— RS-232, 46
connections, 30
non-volatile memory
interface, 30
clearing, 47
storing, 37, 40
normal mode, 27 —S—
safety class, 18
—O— safety warning, 18
saving operating states, 47
OC, 43
selftest errors, 34
operating features, 19
sense connections, 25
option 760, 46
service guide, 18
options, 18
72
Index
setting voltage/current, 42 —V—
shift annunciator, 31
shift key, 31 verification
shorting switch, 28 current measurement accuracy, 55
specifications, 49 current programming, 55
stability with remote sensing, 26 equipment, 53
supplemental characteristics, 50 setup, 53
system errors, 65 test record, 57
system keys, 37 voltage measurement accuracy, 55
Address, 37 voltage programming, 55
Error, 37 verification tests, 31
Interface, 37
Local, 37
RCL, 37
—W—
Save, 37 warranty, 2
Shift, 37 wire
current ratings, 23
—T— wiring considerations, 23
73
Agilent Sales and Support Office
For more information about Agilent Technologies test and measurement products, applications, services,
and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir
You can also contact one of the following centers and ask for a test and measurement sales
representative.
Japan:
Agilent Technologies Japan Ltd.
Measurement Assistance Center
9-1, Takakura-Cho, Hachioji-Shi,
Tokyo 192-8510, Japan
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840
74
Manual Updates
The following updates have been made to this manual since the November 1997 printing indicated on the
Printing History page.
7/8/99
Additional information about CV mode operation has been added to page 20.
A note has been added to Table A-1 for the Ripple and Noise specifications and a correction has been
made for Model Agilent 6631B.
Steps 22 and 23 on page 61 under Low Range Current Measurement Calibration have been switched.
1/7/00
All references to HP have been changed to Agilent.
All references to HP-IB have been changed to GPIB.
A note about magnetic fields has been added to page 22.
On page 49 the model reference for note 2 has been corrected.
2/21/01
A note has been added to Table A-1 for the Programming Accuracy specification.
Information about Option 760 has been added to Table A-1, note 5
The values for Average Programming Resolution, Current in table A-2 have been adjusted.
5/01/02
Step 19 on page 56 under Negative Current Programming Calibration has been corrected.
10/2/03
The Ripple and Noise specification for Model 6634B has changed in table A-1.
3/9/04
The ac input ratings and fuse information has been updated as per IEC 61010-1 requirements throughout
the manual.
The Declaration of Conformity has been updated on page 5.
5/11/09
A URL has been added to the declarations pages to obtain the latest declaration of conformity.
Corrections have been made to Figure 3-5.
The RF field annotation note has been removed from Table A-1.