Thermo-Shaker: User Manual
Thermo-Shaker: User Manual
Thermo-Shaker: User Manual
Thermo-Shaker
Cat.# MS-100
User Manual
ver: 1.1 en
Gingko Biotech
Contents
1 Introduction . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . .3
1.1 Features
1.2 Specifications
6 Operation Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
8 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
1. Introduction
•Thermo-Shaker MS-100 is designed for incubating and mixing aqueous solutions in sealed
micro test tubes. It combines incubator and shaker functions in one. The incubation temperature
can be precisely maintained between 5 °C above room temperature and 99 °C, and the shaker
speed between 200 and 1500 revolutions per minute (rpm). The value of temperature and speed
are independently set and controlled.
1.1 Features
•Independent control of time, temperature and shaking speed. Two-segment programmable
parameter settings.
•Back lit LCD display.
•Simultaneous display of time, temperature and shaking speed.
•Over heating protection for safety & reliability
•Temperature can be calibrated to meet user's needs.
•Audio signal upon program completion.
•Various optional interchangeable blocks to accommodate different tubes & wells for your
experimental needs. Blocks are easy to clean and sterilize. Customized blocks are available
upon request.
•Four types of standard sample blocks are available: A-Block, 0.2 ml x 96 wells; B-Block, 0.5 ml x
54 wells; C-Block,1.5 ml x35 wells; D-Block: 2.0 ml x 35 wells.
1.2 Specifications
Performance
All specifications are valid at ambient temperatures from 5 °C to 30 °C.
Temperature control range: from 5 °C above room temperature to 99 °C
Temperature accuracy: ± 0.5 °C.
Heating rate: ~ 5.5 °C/min.
Mixing frequency: 200 to 1,500 rpm.
Mixing stroke: 2 mm.
Timer range: 1 min to 99h 59 min.
Physical characteristics
Power supply: 100 - 120VAC, 50/60 Hz, 150W
Fuse: 250V 2.5A Ф5×20
Size (Width x Depth x Height): 30.0 x 21.2 x 18.0 cm
Weight: 7.0 kg
Operation ambient conditions
Relative humidity: < 70 %.
Temperature: 5°C to 30 °C.
Sample Block
Control/
Display Panel
Power Switch
Power Socket
Fuse holder
True Value
Set Value
Increase
temperature /
speed / time
Turn on or off
temperature / Start and stop a run;
speed / time Press and hold to
control terminate current run
Decrease
temperature /
speed / time
Move cursor
6. Operation Instructions
6.1 Turning on device
•Turn on the power switch located on the back. “System-Testing”
should appear for about 6 seconds during a routine self-test of the
System-Testing
device. Then, the parameter set values from previous run will appear,
......
and the device will automatically heat up to the stored temperature set
value.
Block Removal
a) Open the transparent plastic lid and unscrew the four
bolts as indicate
b) Close the lid, and move the block from the unit.
Block Installation
a) Line up the holes on the block with the holes on the
device.
8. Troubleshooting
Temperature display shows Ambient temperature < 0 ºC. Increase ambient temp.
“OPEN” or “SHORT”, and audio Broken thermal sensor. Contact vendor.
alarm turns on.
Sample block does not heat up. Broken thermal sensor. Contact vendor.