Thermo-Shaker: User Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 9

1

Thermo-Shaker
Cat.# MS-100

User Manual
ver: 1.1 en

Gingko Biotech

9924 Mesa Rim Road,San Diego CA 92121 www.gingkobiotech.com [email protected]


2

Contents
1 Introduction . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . .3

1.1 Features

1.2 Specifications

2 Safety Precautions and Application Limitation. . . . . . . . . . . . 4

3 Cleaning and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . 4

4 Unpacking and Installation. . . . . . . . . . . . . . . . . . . . . . . . . . 5

5 Control Keys and Display Panel . . . . . . . . . . . . . . . . . . . . . . 6

6 Operation Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

6.1 Turning on device

6.2 Setting temperature, mixing speed and time

6.3 Routine run

6.4 Incubation-only mode

6.5 Mixing-only mode

6.6 Brief mix

7 Change Sample Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8

8 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9

9924 Mesa Rim Road,San Diego CA 92121 www.gingkobiotech.com [email protected]


3

1. Introduction
•Thermo-Shaker MS-100 is designed for incubating and mixing aqueous solutions in sealed
micro test tubes. It combines incubator and shaker functions in one. The incubation temperature
can be precisely maintained between 5 °C above room temperature and 99 °C, and the shaker
speed between 200 and 1500 revolutions per minute (rpm). The value of temperature and speed
are independently set and controlled.

1.1 Features
•Independent control of time, temperature and shaking speed. Two-segment programmable
parameter settings.
•Back lit LCD display.
•Simultaneous display of time, temperature and shaking speed.
•Over heating protection for safety & reliability
•Temperature can be calibrated to meet user's needs.
•Audio signal upon program completion.
•Various optional interchangeable blocks to accommodate different tubes & wells for your
experimental needs. Blocks are easy to clean and sterilize. Customized blocks are available
upon request.
•Four types of standard sample blocks are available: A-Block, 0.2 ml x 96 wells; B-Block, 0.5 ml x
54 wells; C-Block,1.5 ml x35 wells; D-Block: 2.0 ml x 35 wells.

1.2 Specifications
Performance
All specifications are valid at ambient temperatures from 5 °C to 30 °C.
Temperature control range: from 5 °C above room temperature to 99 °C
Temperature accuracy: ± 0.5 °C.
Heating rate: ~ 5.5 °C/min.
Mixing frequency: 200 to 1,500 rpm.
Mixing stroke: 2 mm.
Timer range: 1 min to 99h 59 min.
Physical characteristics
Power supply: 100 - 120VAC, 50/60 Hz, 150W
Fuse: 250V 2.5A Ф5×20
Size (Width x Depth x Height): 30.0 x 21.2 x 18.0 cm
Weight: 7.0 kg
Operation ambient conditions
Relative humidity: < 70 %.
Temperature: 5°C to 30 °C.

9924 Mesa Rim Road,San Diego CA 92121 www.gingkobiotech.com [email protected]


4

2. Safety Precaution and Application Limitation


Thermo-Shaker is for research purpose and should be handled only by experienced laboratory
personnel.
Thermo-Shaker must NOT be operated in a flammable environment and must NOT be used to
mix/incubate explosive or highly reactive substances.
The operation ambient temperature is between 4 °C to 35 °C. The technical data is valid at
ambient temperature between 15 °C and 35 °C.
When mixing and incubating toxic or pathogenic samples, safety regulations must be followed. In
connection with this, please note that the plastic sample block lids is NOT a protective device; it
only provides thermal insulation .
After transferring the device from a cold room to a warmer room, water may condense on the
device. Do not use the device for approximately one hour to allow the condensation to dry.
Thermo-Shaker is an electrical device. Do not spill large quantities of liquid onto the device.

3. Cleaning and Maintenance


Disconnect the unit from the power source before cleaning. The unit should be cleaned with a
soft, damp cloth or paper towel. Do not use harsh cleaner, organic solvent or aerosol spray. Do
not submerge the unit to water. Do not use the unit near water sources.

9924 Mesa Rim Road,San Diego CA 92121 www.gingkobiotech.com [email protected]


5

4. Unpacking and Installation


•Upon receiving your Thermo-Shaker, examine the package for any damage. In the event of shipping damage,
a claim must be filed with the transportation carrier within 10 days.
•Remove the packing material and accessories from the box. Lift the unit out of the box. Do not lift Thermo-
Shaker by its sample block lid!
•The standard package includes: one Thermo-Shaker unit with one sample block pre-installed; one power cord
and one User Manual. If you have ordered additional sample blocks, make sure that they are included in the
special order package. If any item is missing, contact your vendor or Gingko Biotechnologyʼs Service
Department immediately. Save the original packaging until the unit has been shown to work properly.
•Place the device onto a flat, stable bench surface. To ensure sufficient ventilation, keep at least 10 cm of
clearance space around Thermo-Shaker.
•Check the label on the back panel to verify that the unit is rated your local power voltage. Plug the device to a
wall outlet (the electrical power cord is included in the delivery package). Turn on the switch located next to the
power socket on the rear panel. It is advisable to carry out a test run at maximum speed to check the stability of
the device and to ensure that it will move on the bench top during mixing.

Sample Block

Control/
Display Panel

Power Switch

Power Socket

Fuse holder

9924 Mesa Rim Road,San Diego CA 92121 www.gingkobiotech.com [email protected]


6

5. Control Keys and Display Panel

True Value

Set Value
Increase
temperature /
speed / time
Turn on or off
temperature / Start and stop a run;
speed / time Press and hold to
control terminate current run

Decrease
temperature /
speed / time

Move cursor

9924 Mesa Rim Road,San Diego CA 92121 www.gingkobiotech.com [email protected]


7

6. Operation Instructions
6.1 Turning on device
•Turn on the power switch located on the back. “System-Testing”
should appear for about 6 seconds during a routine self-test of the
System-Testing
device. Then, the parameter set values from previous run will appear,
......
and the device will automatically heat up to the stored temperature set
value.

6.2 Setting temperature, mixing speed and time


PV: 25.5 0 00:00
•Setting control values of a run: After pressing “set” key once, device
SV: 60.0 1000 20:00
enters its programming mode, and a cursor appears. Press “set” key
M
further to move the cursor to the value to be adjusted (temperature /
speed / time). Adjust the value by pressing ▲ and ▼ keys. PV: 25.5 0 00:00
•Setting off control of temperature / speed / time: move cursor to the
SV: OFF 1000 20:00
parameter to be turned off (i.e. uncontrolled) and press “off” key.
•“set” key is not functional while shaker is running.
•Time “OFF” means continuous operation.
•Parameters can not be set off altogether at the same time.

6.3 Routine run


•After setting the values, wait for 4 seconds until the cursor disappears.
•Press “start/stop” key to start the run.
•The temperature increases to the set value first, and then, shaker will start and time display starts to count
down. After completion, device stops automatically; “Program Ends” appears on display panel and audio alert
goes off.
•Press and hold down “start/stop” key for few seconds to manually stop the shaker. The temperature maintains
its set value, and time display becomes “00:00”. Press “start/stop” again to restart the run.
6.4 Incubation-only mode
•Use “set” key to move cursor to under speed set value. Press “off” key to change the value to be ”OFF”.
•Wait about 4 sec till cursor disappears. Press “start/stop” key to start the incubator without shaking.
6.5 Mixing-only mode
•Use “set” key to move cursor to temperature set value. Press “off” key to change the value to ”OFF”.
•Wait till cursor disappears. Press “start/stop” key to start the shaker without temperature control.
6.6 Brief mix
•Use the set key to move the cursor under time set value. Press ▲ or ▼ key to set the value to “00:00”.
•Press the “start/stop” key, and the device starts shaking for 6 sec at the set value of speed.

9924 Mesa Rim Road,San Diego CA 92121 www.gingkobiotech.com [email protected]


8

7. Change Sample Block


Before change the block, unplug the power cord to the unit!
The block may be hot. Protect hands if the unit has been used at elevated temperatures.

Block Removal
a) Open the transparent plastic lid and unscrew the four
bolts as indicate

b) Close the lid, and move the block from the unit.

Block Installation
a) Line up the holes on the block with the holes on the
device.

b) Place the four bolts to mounting holes and tighten


the bolts with a hex wrench.

9924 Mesa Rim Road,San Diego CA 92121 www.gingkobiotech.com [email protected]


9

8. Troubleshooting

Error Cause Solution

No display. No power. Check and restore power.


Blown fuse. Replace fuse.
Broken switch. Replace power switch.
Others. Contact vendor.

Large discrepancy between the broken sensor. Contact vendor.


displayed temperature and the
actual.

Temperature display shows Ambient temperature < 0 ºC. Increase ambient temp.
“OPEN” or “SHORT”, and audio Broken thermal sensor. Contact vendor.
alarm turns on.

Sample block does not heat up. Broken thermal sensor. Contact vendor.

Control key(s) doesʼt respond. Broken membrane switch. Contact vendor.

9924 Mesa Rim Road,San Diego CA 92121 www.gingkobiotech.com [email protected]

You might also like