Allied Autovent 2000 3000 Manual
Allied Autovent 2000 3000 Manual
Allied Autovent 2000 3000 Manual
1720 SUBLETTEAVE,
ST.LOUIS, WO, 63110
L90900S-224
TEL. (314)771-2400, FAX; (314)771-3465 REV. A
TABLE O F CONTENTS Section P a g e Njimber
INTRODUCTION
GENERAL DESCRIPTION ^
CONTROL MODULE ^
Source G a s Inlet
Patient Valve Outlets
Breaths Per Minute (BPM)
Tidal Volume (Vf)
Inspiratory Time {Til
Patient Valve Supply Tubing and Oxygen Line
Non-Rebreathing Vglve and Corrugated Hose
U N P A C K I N G A N D INSPECTION O F THE 8
AUTOVENT 2 0 0 0 / 3 0 0 0
OPERATING INSTRUCTIONS 8
Instructions for Use of the AutoVent 2 0 0 0 / 3 0 0 0
v/ith Oxygen Cylinders
Regulator Attachment
Control Module Attachment
Instructions for Use of the AutoVent 2 0 0 0 / 3 0 0 0
with a W a l l Outlet Employing a Quick Connection Adapter
Instructions for use of the AutoVent 2 0 0 0 / 3 0 0 0
with an A i r / O x y g e n Blender
Ventilator Check out
PATIENT USE 12
W A R R A N T Y INFORMATION 18
THE AUTOVENT
2000/3000
USE OF THE Design This operaUng manual for the Life Support Products
OPERATING MANUAL AutoVent 2 0 0 0 / 3 0 0 0 Automatic Ventilators has been
designed for ease of use in the paramedical and field
transport setting.
PERFORMANCE
CHARACTERISTICS The AutoVent 2 0 0 0 / 3 0 0 0 are time cycled, constant
A N D FEATURES ^ l " ^ ^ ' 9°^ powered ventilators. This feature allows the
ventilators' automatic rate to be set by the operator from 8
to 15 breaths per minute (BPM), in the AutoVent 2 0 0 0
version, and 8 to 2 0 BPM in the AutoVent 3 0 0 0 version.
Should the patient require additional breaths, these can be
obtained on demand by making an Inspiratory effort on the
patient valve.
PATIENT VALVE The patient valve delivers compatible with masks and Figure 2 indicates the
ASSEMBLY both controlled and other airway devices. individual components
spontaneous breaths to the The patient valve allows which make up the Patient
patient. It consists of a spontaneous breothing Valve Assembly.
demand valve, a visual upon demand if the patient
Indicator, pressure limit makes an insplrotory effort **NOTE: Since fhe patient
alarm, and exhalation o f - 2 c m . H2O. (The BPM vaive and control module
valve. It has a standard control may be turned to are a matched sef wifh
15mm. inside the " 0 " position if desired identical serial numbers,
diameter/22mm. outside for this purpose.) do not separate. If used
diameter adopter, which is wifh other units, setting
may not be accurate. **
VISUAL INDICATOR
^ S O U R C E OAS INLETS
PRESSURE LIMIT A U R M
TOP VIEW
- OUTLET ADAPTER
(EXHAUTION VALVE INSIDE)
Visual Indicator The visual indicator is located on the top of the Patient VISUAL INDICATOR
PRESSURE LIMIT A L A R M
Pressure Limit A l a r m An audible pressure limif alarm is located in the Patient
Module Valve Assembly. This alarm sounds v/henever the patient
airv/ay pressure approaches the designed pressure limit.
The Pressure Limit Alarm v/ill p'phfmue to sound during the
inspiratory phase until either the airway pressure decreases
or the ventilator cycles off to begin the expiratory phase.
W A R N I N G : If t h e i h a x i m u m pressure limit is
reached/ the pre-set tidal v o l u m e m a y not be
delivered to the patient. Inspiratory time will
remain constant, h o w e v e r , a n d a n inspiratory
hold will be maintained with no additional
volume being delivered until the venfilator cycles
to the expiratory phase. This waming also
appears under Tidal Volume in the Performance
Characterisfics,
Source G a s Inlets Located on the side of the Patient Volve Assembly the
inlets connect the Patient Valve with the Control Module.
The top inlet: (1) Supplies the actuator assembly and is a
nippled connector. The bottom inlet; (2) Supplies source
gas to the patient and is a diameter index safety system
(DISS) oxygen connector.
FIGURE 5 .
5
FIGURE 6.
_ )
, TIDAL V O L U M E
CONTROL MODULE The Control Modules for
SOURCE G A S INLET
the AutoVent 2 0 0 0 / 3 0 0 0
are designed to be
compact, durable ond easy
to use. The units are
constructed to perform in PATIENT V A L V E
OUTLETS
the difficult environments of
paramedic or transport
>
operations. Their features
include an impact-resistant
case with shock absorbing
AUTOVENT 2000
bumpers and easy to read
controls for independently
BREATHS PER MINUTE C O N T R O L
setting Breaths Per Minute
(BPM), Tidal Volume and
Inspiratory Time (AutoVent
3 0 0 0 only). The reverse
side of the case also has
simplified operating
instructions for ease of
operation.
Patient V a l v e Outlets
Located on the right side of the Control Module body
and marked with arrows, the patient valve outlets connect
the Control Module with the Patient Valve Assembly. The
top outlet (1) Supplies source gas to the Patient Valve
Assembly at a constant flow dnd is a diameter index safety
system (DISS) oxygen connector. The boftom outlet: (2}
FIGURE 9 .
Supplies the Patient Valve Actuator and is a unique 7 / 1 6
inch threaded female connector.
Breaths Per Minute This control sets the ventilator rate from 8 to 15 BPM In the
(BPM) AutoVent 2000 version, and from 8 to 20 BPM in the
AutoVent 3000 version. Adjusting the knob clockwise
decreases the breathing rate and adjusting counter clockwise
increases the rate. Source gas is available on demand, even
FIGURE 10.
in the " 0 " position up to 36 LPM depending on Tidal Volume
AUTOVENT 2000 AUTOVENT 3000
setting, from the Patient Valve Assembly to allow the patient to
breathe spontaneously.
Tidal V o l u m e (VO This control adjusts the volume available to the patient
during a breath and is adjustable from 400 ml to 1200 ml in
the AutoVent 2000 version, and from 200 to 1200 mil in the
AutoVent 3000 version. Turning the knob clockwise increases
tidal volume. Turning the knob counter clockwise decreases
tidal volume. Foiiowing a volume adjustment change, the tidal FIGURE 11.
volume stabilizes after one breath and remains constant. AUTOVENT 2000 AUTOVENT 3000
Inspiratoiy Time (Ti) This control knob in the center position of the AutoVent 3000
allows adjustment of the potient's inspiratory time. The two
settings allow selection of Adult and Child inspiratory time
respectively. The inspiratory time for the circled orange child
setting is approximately 1 second and when selected corresponds
to the circled orange settings on the BPM and Tidal Volume
control knobs. The inspiratory time for the white Adult setting is
approximately 2 seconds and when selected corresponds to the FIGURE t i A .
white settings on the BPM and Tidal Volume contro! knobs. The
AutoVent 2000 has a pre-set inspiratory time of approximately 2
seconds.
P.E.E.P. PORT
FIGURE 14.
10
Instructions for Use of * *NOTE: Check ail line and tubing connections for leaks.
the AutoVent If a leak occurs, check that the previous fitting instructions
2 0 0 0 / 3 0 0 0 with have been followed correctly. **
O x y g e n Cylinders
(Continued) CAUTION: Read all CAUTION: A l w a y s CAUTION: A l w a y s
instructions m a k e sure a n verify that the
thoroughly before adequate supply of cylinder valve is in the
opening the cylinder o x y g e n or source gas closed or off position
v a l v e . Connect all is available for patient (fully clockwise) prior
oxygen/source gas use a n d transport. It is tb disconnecting the
pressure lines to the advisable to have a tubing assembly or
LSP AutoVent back-up regulator removing the
2000/3000 and available to facilitate regulator from the
change-over in the o x y g e n cylinder.
Patient V a l v e
Assembly prior to use. event a cylinder
Assure all high transfer needs to be
pressure outlets ore made.
plugged and cylinders
turned off or closed
w h e n not in use.
— JrTTT—
**NOTE: Immediately check all pressure lines and tubing
for leaks. If a leak occurs, check thaf fhe previous fitting
insfructions have been followed correctly. Check thaf all
line and tubing connections have been adequately hand
tightened.**
FIGURE 15.
Insfructions for Use of Instructions for this apphcation ore essentially identical to
the AutoVent those for use wifh either cylinder or wall outlet systems.
2 0 0 0 / 3 0 0 0 with a n
The high-flow blender system should be connected to the
A i r / O x y g e n Blender
source gases as per manufacturers instructions. The
blender then becomes the attachment site for the source
gas supply line to the Control Module.
• Set the Inspiratory Time control knob to the adult setting ^ClO >
on the AutoVent 3 0 0 0 , (Inspiratory time is pre-set on
the AutoVent 2 0 0 0 ) . Rotate the control knob clockwise
FIGURE 1 6 .
until it is against the end stop on the adult setting.
60 SECONDS
• Count the number of complete ventilator cycles for a
fuil minute., At the checkout setting, there should be 12 I N S P I R A T O R Y TIME
BPM delivered, with a 2 second inspiration and 3 2 SEC.
12 B P M
second expiration per breath. E X P I R A T O R Y TIME
3 SEC.
2. Occlude the outlet of the patient Valve Assembly. An
FIGURE 1 7 .
audible pressure limit alarm should sound after the
ventilator cycles, indicating the designed pressure limit
has been reached.
**NOTE: The pressure limit alarm should sourid
throughout the latter portion of the breath after reaching fhe
pressure limif, and stop when the ventilator cycles to
expiration.**
**NOTE: AutoVents & patient Valves are serialized and
calibrated to work together and sfiould remain
FIGURE 18.
together for the life of the products
3. Test the unit for proper function prior to each patient use. Refer to Maintenance section
for this procedure (page 14).
4. Should the unit fail to operate properly at any time, refer to the Troubleshooting Guide
(page 16.) Disconnect the patient from the ventilator any time the unit does not appear
to be operating properly. Ir unable to determine the cause of problem, contact Ufe
Support Products for service.
5. Clean the unit after each use (refer to the Maintenance section for detailed instructions
on page 14).
PATIENT USE **NOTE: ifthe LSP Automatic Ventilator is to be powered by a cylinder, be sure to turn on
the cylinder valve slowly. **
Check for obstructions in the patient's throat or mouth (vomitus, foreign bodies, broken
dentures, etc.), and remove if present, in occordance with prevailing standards.
Set fhe volume to equal 8 to 10 ml. for every kg. of body weight; e.g., 7 0 kg. patient
equal 7 0 0 ml. volume.
PATIENT USE Set the BPM control knob to the desired setting. Refer to the
(Continued) quick set-up insfructions on the bock of the Control Module for
guidelines.
Set the Tidal Volume control knob to the desired volume. FIGURE 19.
(1) BPM - Usina fhe second hand on your watch, count the
number of breaths delivered to the patient for one full
minute. Ifyou need to increase or decrease fhe rate,
adjust the BPM confrol knob accordingly, checking the
rote again once the adjustment has been mode.
**NOTE: If fhe pressure Increase the volume and attempt to ventilate via
limit alarm sounds during deiivered to the patient, other means if adjustments
the inspiratory phase and until adequate chest do not result in satisfactory
adequate chest movement movement occurs, by" ventilation ofthe patient.
does not occur, an rotating the Tidal Volume For additional information,
increase in airway control knob in a clockwise refer to fhe Troubleshooting
resistance, a blocked direction. Disconnect the Guide. * * (page 16)
airway and/or a stiff lung patient from the ventilator
is indicated.
14
Cleaning a n d 1 . Cleaning and Leave hoses connected so you d o not get water inside.
Disinfecting Disinfecting the Control
Follow established protocol regarding frequency of
Equipment Module
cleaning.
FIGURE 24.
AxiioVent, 16
•>','• —
Cleaning a n d LOCATING DRACKEtS CAUTION: If the flapper valve is twisted or the
Disinfecting Equipment locating bosses are not properly positioned, the
(Continued) Patient V a l v e Assembly will not function
properly. A l w a y s m a k e sure the valve is flat
a n d properly seated.
3. Cleaning ihe Non- Clean after each use. All components of the P / N L496
Rebreathing Valve Non-Rebreathing Valve ore autoclavable. To disassemble,
unscrew the valve inlet from the outlet and remove the
duckbill diaphragm. The individual components can now
be cleaned. If autoclaving systems are not available, you
may also sterilize via cidex or other bacteriocidal solution.
Be sure to rinse and dry all parts thoroughly before re-
assembling. After cleaning, inspect all ports for damage or
breakage. Replace any damaged or broken parts.
STORAGE OF THE Store the unit in a clean, dry area within a temperature
LSP A U T O V E N T range of-40°F to 160°F.
2000/3000 After long periods of storage, the unit should be fully
tested before use in accordance with the checkout
procedures in this manual.
TROUBLESHOOTING Increased tidal volume or Volume setting to high Check Control Module
increased chest expansion settings and adjust volume
GUIDE
OS required
(Continued)
Pressure limit alarm during Increased airway resistance Evaluate patient and adjust
inspiratory phase ventilators as required
Failure of the pressure limi Alarm outlet is plugged with Remove and clean, or
alarm debris or has malfunctioned replace
18
i4utoVent:
**NO)'E: Some warnings and cautions appear more fhan once throughout the manuai.
They appear in this summary to help direct the user to the proper page and section ofthis
manual.**
Cautions
S U M M A R Y OF Page N u m b e r Cautions
WARNINGS AND
10 Read all instructions f/ioroug/i/y before opening the
CAUTIONS cylinder valve. Connect all oxygen/source gos pressure
(Continued) lines to the LSP AutoVent 2 0 0 0 / 3 . 0 0 0 and Patient Valve
Assenibly prior to use. Assure all high pressure outlets are
capped and cylinders turned off or closed when not in use.
APPENDIX A : Support Equipment For your ordering convenience, the following is a list of
adjunct equipment which may be used in conjunction with
or in support of the LSP AutoVent 2 0 0 0 / 3 0 0 0 .
\
AuioVenilOOO/ZOOO 21
Aspirator L146
G a s Consumption
Driving G a s : 0.4 LPM Maximum
Weight: 16o2./450g.
Material
Cover: Polycarbonate
Outlet; Polysulfone
Weight: 24 o z . / 6 8 0 g .
APPENDIX E: The conversion chart values are calculations which approximate actual performance at
various altitudes, but do not represent guaranteed performance specifications.
AutoVent 2 0 0 0 / 3 0 0 0
Life Altitude Conversion Chart
Support
Products Tidal Volume Settings (ml.)
Altitude 200 300 400 500 600 700 800 900 1000 1200
(ft.)
1000 3280 226 339 452 565 678 79] 904 1017 1130 1356
2000 I 6560 I 254 | 381 I 508 j 635 | 762 | 889 I 1016 I 1143 I 1270 1524
3000 9840 288 432 576 720 864 1008 1152 1296 1440 1728
4000 13120 328 492 656 820 984 1146 1312 1476 1640 1958
5000 16400 374 561 748 935 1122 1309 1496 1683 1870 2244
6000 19680 460 690 920 1 150 1380 1610 1840 2070 2300 2760
i4utoVent20C=0/3000 25
APPENDIX F: The values in this chart are calculations which approximate actual performance at
various settings, but do not represent guaranteed performance specifications.
AutoVent 2000/3000
Oxygen Cylinder Depielion Times*
Breaths per minute
12 14 16 18 20 Cyi.
Tidol
Volumo
1200
66 62 52 45 40 36 32 E
1000 64 58 49 42 37 33 30 Jmb. D
41 38 32 28 24 22 20 D
800
83 72 64 58 52 E
600 78 68 60 54 49 Jmb. D
50 44 39 35 32 D
500
300
NOTES:
26
NOTES:
Limited O n e (1) Y e a r W a r r a n t y
LSP worronls ihis producl to be free from defecls in maleriol ond Disclaimer of Other Warranties
workmonsliip for a period of one (1) year from lhe dole of manufaclure.
This Worfonly is expressly condilioned on connplionce wilh oii Inspeclion THiS W A R R A N T Y IS IN PLACE A N D IN UEU OF ALL OTHER
and prevenlalive mainlenance requirements os set by oppllcoble WARRANTIES OR REPRESENTATIONS, EXPRESSED O R IMPLIED,
governmenl agencies ond os specified by LSP. I N C L U D I N G , B U T N O T LIMITED T O , W A R R A N T I E S O F
MERCHANTABIUTY OR FITNESS FOR A SPECIFIC P U R P O S E , B Y
This Warranty is extended by ISP only to the first purchaser of the OPERATION O F L A W OR OTHERWISE.
k conform. LSP shall repair or replace (at LSP's opiion) this product if II is
detective, molfunclions or foils to conform lo Ihis Worronljr, ond sholl return
exposures to heot, sun, water, ozone, or lo olher deleriorolive elemenls.
K LSP shall In no event be liable for any consequentiol damages, nor for
loss, domages or expenses direclly or indirectly orising from lhe use of Ihis Producfs
product.
1720 Sublelte Avenue • Sl. Louis, M O 63110
TEL: 314/771-2400 • FAX: 314/771-0650
\mi'
^ '<'!ll