EN - TOTAL PROTEIN - BAOSR6x32 - US
EN - TOTAL PROTEIN - BAOSR6x32 - US
EN - TOTAL PROTEIN - BAOSR6x32 - US
OSR6132 4 x 25 mL R1
4 x 25 mL R2
OSR6232 4 x 48 mL R1
4 x 48 mL R2
OSR6632* 4 x 164 mL R1
4 x 164 mL R2
Intended Use
System reagent for the quantitative determination of Total Protein in human serum on Beckman Coulter AU analyzers.
*Total Protein reagent OSR6632 for use on the AU2700/5400 system only.
Summary
Measurements of Total Protein are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as
other metabolic and nutritional disorders.
The total serum protein is the sum of all circulating proteins and is a major component of blood. It is often useful, however, in interpreting the
1
significance of the total protein concentration to have more specific knowledge of individual fractions such as albumins and globulins.
Methodology
2
This Total Protein procedure is based on the modification of Weichselbaum. Cupric ions in an alkaline solution react with proteins and polypeptides
containing at least two peptide bonds to produce a violet colored complex. The absorbance of the complex at 540/660 nm is directly proportional to
the concentration of protein in the sample.
OH ¯
2+
Protein + Cu Blue violet complex
System Information
e e
For AU400/400 /480, AU600/640/640 /680 and AU2700/5400/AU5800 Beckman Coulter Analyzers.
Reagents
Final concentration of reactive ingredients:
Sodium hydroxide 200 mmol/L
Potassium sodium tartrate 32 mmol/L
Copper sulfate 18.8 mmol/L
Potassium iodide 30 mmol/L
Precautions
1. For in vitro diagnostic use.
2. WARNING! CORROSIVE! Do not pipet by mouth. Avoid contact with eyes, skin or clothing. In case of contact, immediately flush affected areas
with plenty of water for 15 minutes. Obtain medical attention immediately for eye contact or ingestion.
Preparation of Reagents
For OSR6132 and OSR6232, the Total Protein Reagents are ready for use. No preparation is required. For OSR6632, insert the pipe supplied into
the 180 mL reagent vial before use on the analyzer. Care must be taken when handling the pipe to avoid contamination. The pipe is for single use
only. Do not remove the large cap.
Indications of Deterioration
Visible signs of microbial growth, turbidity, precipitate, or any change in color in the Total Protein reagent may indicate degradation and warrant
discontinuance of use. R1 should be a clear blue solution and R2 should be a clear, colorless solution.
Interfering Substances
4
Results of studies show that the following substances interfere with this Total Protein procedure.
The criteria for no significant interference is recovery within 10% of the initial value.
Bilirubin: No significant interference up to 40 mg/dL Bilirubin
Hemolysis: No significant interference up to 500 mg/dL Hemolysate
Lipemia: No significant interference up to 1000 mg/dL Intralipid*
BAOSR6x32.02 OSR General Chemistry
2012-01
Total Protein
* Intralipid, manufactured by KabiVitrium Inc., is a 20% IV fat emulsion used to emulate extremely turbid samples.
The information presented is based on results from Beckman Coulter studies and is current at the date of publication. Beckman Coulter Inc. makes
no representation about the completeness or accuracy of results generated by future studies. For further information on interfering substances, refer
5
to Young for a compilation of reported interferences with this test.
Procedure
A complete list of test parameters and operational procedure can be found in the User’s Guide appropriate to the analyzer.
Materials Provided
Total Protein Reagent
Pipe (one per each 180 mL vial)
Calibration
Calibration of this total protein procedure is accomplished by the use of the Chemistry Calibrator (Cat # DR0070), which is traceable to the National
Institutes of Standards and Technology (NIST) Standard Reference Material (SRM) 927a.
Recalibration of this test is required when any of these conditions exist:
1. An observed drift in QC values of > 5%.
2. A change of bottle/Lot number.
3. Major preventative maintenance was performed on the analyzer.
4. A critical part was replaced.
Quality Control
During operation of the Beckman Coulter AU analyzer, at least two levels of an appropriate quality control material should be tested a minimum of
once a day. In addition, controls should be performed after calibration, with each new lot of reagent, and after specific maintenance or
troubleshooting steps described in the appropriate User’s Guide. Quality control testing should be performed in accordance with regulatory
requirements and each laboratory’s standard procedure.
Results
Automatically printed out for each sample in g/dL at 37°C. For SI units (g/L) results must be multiplied by 10.
Dynamic Range
The Total Protein procedure is linear from 3 to 12 g/dL for serum determinations. Samples exceeding the upper limit of linearity should be diluted and
repeated. The sample may be diluted, repeated and multiplied by the dilution factor automatically utilizing the AUTO REPEAT RUN.
Expected Values
6
3 years to adult: 6.0 - 8.3 g/dL
6
Newborns: 4.6 - 7.0 g/dL
Beckman Coulter Determined Reference 6.4 - 8.9 g/dL
7
Range:
Expected values may vary with age, sex, diet and geographical location. Each laboratory should determine its own expected values as dictated by
good laboratory practice.
Sensitivity
e
Typical change in absorbance per minute for 1 g/dL of Total Protein is 23.7 mAbsorbance in the AU400/400 , 51.3 mAbsorbance in the
e
AU600/640/640 , and 62.5 mAbsorbance in the AU2700/5400/680/480 analyzers.
Manufactured by: Beckman Coulter, Inc., 250 S. Kraemer Blvd. Brea, CA 92821, USA