Service Manual
This is a medical product of class I according to appendix IX of the Medical Product Directive (MDD)
93/42/EEC of 14 June 1993.
The manufacturer declares that the device is compliant with the fundamental requirements according to
Appendix I of MDD, and that a conformity assessment procedure is performed as required for Class I products
according to Appendix VII of MDD. This is documented by the CE symbol.
Reprinting, copying or translating this document in whole or in part is forbidden without the express written
permission of TRUMPF Medizin Systeme GmbH.
All rights under the copyright laws are expressly reserved by TRUMPF Medizin Systeme GmbH.
Within the framework of the legal requirements, the manufacturer is only responsible for the safety
characteristics of this device if the maintenance, repair and modifications performed on this device are carried
out by the manufacturer, or by a person authorised by the manufacturer and in accordance with the
manufacturer’s instructions.
Contents
1 Important Information
1.1 Notes for Service Personnel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Explanation of Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.3 General Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Safety Instructions for Working in the Operating Area . . . . . . . . . . . . . . . . . . . . . . . 8
1.5 Instructions for Protecting the Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2 Repair Instructions
2.1 Electrical Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1.1 Replacing the Accumulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1.2 Adjusting the column in case of breakdown of the electronic unit . . . . . . . . . . . . . . 10
2.1.3 Replacing the Electronic Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.1.4 Checking the fuses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.1.5 Replacing cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.2 Replacing sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.2.1 Position of the sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.2.2 Sensor Tilting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.2.3 Sensor Trendelenburg . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.3 Hydraulic system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.3.1 Replace cylinder, aggregate and control unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.3.2 Replacing hydraulic fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.3.3 Refill oil in the hydraulic aggregate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
2.3.4 Replace pressure switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
2.4 Lifting cylinder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2.5 Servicing the table top . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
2.5.1 Replacing the bowden cable - longitudinal shift . . . . . . . . . . . . . . . . . . . . . . . . . . 22
2.5.2 Replacing the gas spring in the backplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
2.5.3 Replacing the drive of the coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.5.4 Adjusting the coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
2.6 Servicing leg supports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
2.6.1 Replacing the gas springs (leg support in four parts) . . . . . . . . . . . . . . . . . . . . . . . 26
2.6.2 Replacing the bowden cable (leg support in four parts) . . . . . . . . . . . . . . . . . . . . . 27
2.6.3 Replacing the lever for releasing the leg support. . . . . . . . . . . . . . . . . . . . . . . . . . 27
2.7 Replacing the snap-in shaft in head section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3 Test program
3.1 Launching the test program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.2 Launching the various test program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.3 Description of the test programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5 Maintenance
5.1 Needs of maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.2 Maintenance by TRUMPF Medizin Systeme GmbH based on a contract . . . . . . . . . 32
5.3 Service contract . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.4 Self maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.5 Service instruction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.5.1 General Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.5.2 Checking the hydraulic system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.5.3 Checking the electronic unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
8 Connection diagram
Connection diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
1 Important Information
Please follow this manual exactly when performing repairs and keep to all the scheduled
repair steps.
We would be grateful to receive any corrections to this manual and also any tips for
performing repairs more efficiently than described here.
Please contact our Service Department in Saalfeld.
This repair manual is intended for trained service and maintenance personnel.
Refer to the user manual for information about operating the device.
Please contact our Service Department if a problem occurs which cannot be remedied with
the assistance of this repair manual.
Repairs which are not covered by this service manual may only be performed after
consultation with TRUMPF Medizin Systeme GmbH or be performed by
TRUMPF Medizin Systeme GmbH in Saalfeld.
After all repairs or settings, a functional test or test must be performed in accordance with
the instructions in this service manual.
The device may only be handed over to the customer after being tested and in a fully
functional condition.
The handover must be made in writing and must be confirmed by the customer.
The device must be demonstrated to the customer at the handover.
Important information in this user manual is marked with symbols and keywords.
Keywords such as WARNING, CAUTION and ATTENTION indicate the level of danger
involved. The symbols emphasise the message visually. Additional symbols can indicate
injury hazards or danger to life and limb.
WARNING!
Danger to life and limb (for example electric shock)!
Risk of serious injuries and damage to property.
CAUTION!
Risk of injury and damage to property.
ATTENTION!
Risk of damage to property.
NOTE!
Useful additional information and tips.
TEST!
Performance of functional testing, measurements and tests.
ENVIRONMENT!
Notes on disposal in a way which does not harm the environment!
Safety Instructions:
• Observe the hygiene requirements of the medical facilities in respect of clothing,
cleaning regulations and disinfection procedures!
• Obtain this information from the hospital personnel before starting any work!
• Observe all the rules regarding your conduct, especially in the operating area!
• Do not perform any work in the operating area without first obtaining the appropriate
permission!
• Before performing maintenance or repair work on the operating tables make sure that
they have been disinfected!
• The disinfection of the interior components of the operating table must be arranged in
advance and carried out by the medical facility before repair or maintenance work
commences!
• Should symptoms of illness occur, seek medical attention immediately!
After you have worked in dangerous areas, tell the doctor about the nature of these
activities - even months later!
NOTE!
Ask a doctor about the possible dangers and the relevant protective
measures which you should take in your work as a service technician!
2 Repair Instructions
CAUTION!
Disconnect the power supply!
NOTE!
Check the firm fitting of the plug (thick side of the plug points towards the
accumulator)!
Stud must be stick tightly to the accumulators!
6. Measure charging current; optimal value at 800 mA. The charging current depends on
load condition of the accumulators!
7. Reassemble cover.
TEST!
Carry out Test 3!
J5 J4 J7 J6 J1 J2 J3
Electronic unit
CAUTION!
Danger of connecting to the wrong poles!
Lock security elements when connecting membrane keyboard!
CAUTION!
Disconnect the power supply!
1. Remove cover.
2. Remove locking of the bellows.
3. Bring lifting in uppermost position.
4. Disconnect accumulators.
CAUTION!
Danger of short circuit at the connectors of the accumulators! Before
proceeding in the repairs, cover connectors of the accumulators!
brown
Type of Table JU 1A JU 1B
GE X X
GR X
GY X
SM X
SF
JU 1A
JU 4
JU 1B
Type of Accumulator JU 4
12 Ah X
ATTENTION!
The Infrared-Code of the device is permanently stored in the EPROM!
Remove EPROM from the old PCB and insert in the new one. Mind the
marks!
TEST!
Carry out Test 3!
NOTE!
If Trendelenburg does not correspond to the low position of the head
section, exchange the jumpers J6 and J7 (figure 1)!
CAUTION!
Check the guiding of the cables!
Avoid chafe marks and sharp edges!
Danger of connecting to the wrong poles!
Lock security elements when connecting membrane keyboard!
TEST!
Carry out Test 3!
CAUTION!
Disconnect the power supply!
F6
F8
F3
F1 F4
F2
F7
5. Check fuses, replace if necessary. If so, find cause for defect fuse and repair it.
CAUTION!
Danger of connecting to the wrong poles!
Lock security elements when connecting membrane keyboard!
TEST!
Carry out Test 3!
ATTENTION!
Check guidance of the cable!
Avoid chafe marks and sharp edges!
7. Connect accumulator.
CAUTION!
Danger of connecting to the wrong poles!
Lock security elements when connecting shielding foil!
TEST!
Carry out Test 3!
CAUTION!
Danger of connecting to the wrong poles!
Snap in security elements when connecting membrane keyboard!
TEST!
Carry out Test 3
Trendelenburg
1. Remove cover.
2. Disconnect membrane keyboard.
3. Disconnect accumulator.
4. Open the locking of the bellows.
5. Screw off sensor with locking plate.
6. Replace sensor.
7. Tighten the screws of the locking plate, connect sensor to J4.
8. Connect accumulator.
CAUTION!
Danger of connecting to the wrong poles!
Snap in security elements when connecting membrane keyboard!
TEST!
Carry out Test 3!
NOTE!
These works are carried out by TRUMPF Medizin Systeme GmbH Service.
CAUTION!
Make sure to carry out 3. and 4. carefully!
This avoids lowering of the cylinders and strong leaking of oil when opening
the fittings.
ATTENTION!
When replacing elements of the hydraulic system for tilting and
Trendelenburg, move table to final positions to make sure the tubes are
guides properly!
CAUTION!
Danger of connecting to the wrong poles!
Snap in security elements when connecting membrane keyboard!
TEST!
Carry out Test 3!
TEST!
Carry out Test 3!
NOTE!
Do not use any function of the operating table while the switch is
removed
TEST!
Carry out Test 3!
1
2
1 3
4
5
3
4
6
4
7
CAUTION!
Danger of short circuit at the connectors of the accumulators
NOTE!
All loosened and removed screws must be secured using OMNIFIT 100 M!
ATTENTION
Check torsion play!
Less lifting power when tilted too much!
Column should have no recognizable play!
TEST!
Check lifting function with a weight of 135 kg!
NOTE!
Mind guiding in the spar!
Insert new bowden cable using the old one to pull!
TEST!
Check longitudinal shift!
CAUTION!
Make sure not to endanger people and goods when replacing gas springs!
Gas springs are still under pressure when removed from the operating
table! In faulty gas springs, the releasing pin of the piston rod can fly off like
in an explosion!
ATTENTION!
Do not operate the backplate!
NOTE!
Check releasing of the backplate!
Both gas springs must release simultaneously!
NOTE!
Note and remember the positions of the disk springs
ATTENTION!
Adjust coupling!
10x
10x
1 3 2 4 5
NOTE!
There must be no braking when the leg support is fixed!
Check:
• Lever [1] (coupling) locked (vertical position):
At the end of the padding of the leg support 40 kg must be carried safely!
• Lever [1] (coupling) loosened (horizontal position):
There must be no braking effect when moving the leg support!
1. Loosen lock nut using special wrench (abraded open-jawed wrench 17 mm).
2. Remove bolt from the bottom of the gas spring (loosen threaded pin).
3. Screw out gas spring holding it at the bottom.
CAUTION!
Make sure not to endanger people and goods when replacing gas springs!
Gas springs are still under pressure when removed from the operating
table! In faulty gas springs, the releasing pin of the piston rod can fly off like
in an explosion!
4. Mount gas spring in fully extended condition, grease thread, screw locknut in to the
end of the thread.
5. Compare screw-in-distance of gas spring with other leg support.
NOTE!
The angles of the leg supports must be equal in fully extended condition!
6. Tighten locknut.
7. Align drill hole of the gas spring with the one of the transition piece, fix using a bolt,
secure with a threaded pin.
TEST!
Check functionality of the release!
NOTE!
Guide bowden cable in the lever under the deflector to make sure the
handle is freely accessible.
6. Guide spar-end of the bowden cable through the setting screw and hinge in.
7. Screw in setting screw completely.
8. Hinge bowden cable into lever, lock handles with the screws.
9. Screw in gas springs 5 turns, tighten locknut, insert bolt, secure using the threaded pin.
10. Adjust bowden cable using the setting screw until gas spring works properly.
TEST
Check functionality!
2 3 4 1 4
TEST!
Check function of the head section adjustment!
3 Test program
1. Tilting right
2. Lifting up
3. Trendelenburg
4. Lifting down
5. Tilting left
Launching Confirmation
Test Button Confirmation
with button beep
Test 1
function of the 1 x 0.2 s
column activation button
zero button
Test 2
2 x 0.2 s
Sensors
1 x 1 s beep
Test 3
3 x 0.2 s red LED
5 courses
0.5 Hz
any button
Test 4
4 x 0.2 s
long time test
Test 5
5 x 0.2 s activation button
switching off
Part/Material Disposal
Part/Material Disposal
Part/Material Disposal
5 Maintenance
Meas.
Method of Set value,
No. Items value,
examination Display
remarks
Meas.
Method of Set value,
No. Items value,
examination Display
remarks
Meas.
Method of Set value,
No. Items value,
examination Display
remarks
Meas.
Method of Set value,
No. Items value,
examination Display
remark
Meas.
Method of Set value,
No. Items value,
examination Display
remark
For customers with knowledge and experience of TRUMPF Medizin Systeme GmbH
operating tables:
The appliance programme can be tailored to the customer’s specific needs.
Duration: 1 day
Timetable:
6.4 Services/Costs
TRUMPF Medizin Systeme GmbH pays the costs of:
• Training
• Overnight accommodation
• Full board
• Training materials
Travel costs and continued payment of wages are born by the participant.
Lifting column
3 9 16
10
4
11, 12
13
18
17
6, 7 14
19 15
20
27
23
28
21
24
25
29
26
30
Running
20
2
11
7 8
6
10
9 17
12
11
13
5 14
15 16
18
4
19
9 brakes
Running gear, directed wheel with
11
10
11
12 15
6 11
6 13
15
16
13 17
11
14
18
3 19
4
20
2 4
6 5
7
8
9
8
2
9
10
9
12 13
4
9
1
3 14
12
4
5
14
13 9
9
6
11
21
22
8
23
20
19
18
19
10
18
9 11
10
12
12
4
5 13
6 17
14
7
15
16
1
8
2
9
3 16
13
12 15
10
11 12
10 14
22
20
23
17
26
28
25
27
24
25 1
23
8
10
24
9 11
22 28 12
10
23
12
18
19
13
18
20 14
15
19 21 16
20
21
27
28
4 17
5
6 22
7
17
1
8
2
9
3 16
13
12 15
11
10 12
10 14
15
29
31 15
15
25
23
33
30 21 24
15
16
20
22
19
17 18
26
7 28
8
27
9
32
6 10
3 14
13
11
4 12
4
5 6
1
1 2
6
4
1
4
2 5
3
1 6
6
5
5
6
7
6
5
9
10
10
9
12
10
13
5
2
6 8
7
3
4
5
2
3
Coupling
3
2
3
1
17
14
16 15 18
13
9
8 12
7
4
11
10
5 6
24
25/14x
24
22 26/16x
21
20
19 23
19
12
2
3 13
4
14
11 16 15
1
17
6 5
7 19 21
10 18
20
8
9
10
8
10
8
21
20
21
22
19
17 19
17
18 23
19
17 20
1 2 3 6 26
5
25
4 9
7
8
10
11
12
24
13
14
15
16
13
23
20
47
30
29
49 24
50
27 42 39
39
40
28 43
41 46
44
27 45
26
28 51
25
31
32
34 33
35
38
37
36
1 2 3 6
5
4 9
7
8
10
11
12
48
19
18
13 17
14
15
16
13
22
21
23
20
47
30
29
49 24
50
27 42 39
39
40
28 43
41 46
44
27 45
26
28 51
25
31
32
34 33
35
38
37
36
1 2 3 6
5
4 9
7
8
10
11
12
48
19
18
13 17
14
15
16
13
2
1
Sensor height
1 1
Sensor height
3 4
5 6
Z3A
Z1A
Z2B
Z1B
Z3B
Z2A
Z1A
Z1B
Z3B
Z2B
1
2
Z3A
Z2A
Z3B
Z3A
Z2B
Z2A
Z1B
Z1A
Gaskets
Banjo bolt
ATTENTION!
Insert two sealing rings per banjo bolt!
5 6
2
M3
M4
ATTENTION!
1 Plug connector of the hydraulic motor in J2, left!
2 Tighten screws of J6 and J7!
3 Clamp screens of the cables under the plates!
5 6
2
M3
M4
Pin assignment J4
Pin assignment J5
8 Connection diagram
white
green
brown
white
green
brown
Cable remote control shield Pin 1
blue
Don´t connect white 12 V, 12 Ah 12 V, 12 Ah Pin 2
head end black
white
green
brown
shield! brown
Pin 3
white
Connection diagram
brown motor
green black
L1
yellow shield
grey N
red
white
brown
green yellow
Connection diagram
yellow green
grey
red grey M1
brown Pin 1
violet white brown M2
black black Pin 2
Pin 1 cable 10 x 0,14 qmm, shielded M3
blue pink
Pin 2 blue Pin 3
M4
pink red Pin 4
Pin 3 violet green M5
shield Pin 5
M6
shield Pin 6
green/yellow yellow M7
Pin 7
M8
Pin 8
grey M9
Pin 9
M 10
Pin 10
3 x 0,75 qmm pink
Pin 11
Jul 1B Pin 12
shield
Pin 13
Pin 14
Jul 1A J 20
Pin 15
pink GND
Pin 25
GND
Pin 24
GND
Pin 23
GND
Pin 22
GND
Pin 21
GND
Pin 20
primary 230 V black GND
Pin 19
blue
GND
Pin 18
GND
Pin 17
brown black GND
Pin 16
blue white
brown
blue
Connection 115 V
primary
brown
Filter
Hydraulic diagram control unit
Chock Choke
Hydraulic diagram control unit
M5 M6 M7 M8 M9 M10
Magnetic valve
Blind plate
M2 M1 M4 M3
Magnetic valve
Hydraulic aggregate
433 mm
3 2 1
NOTE!
Bleed the system:
- Lift back section to highest position.
- Use the function “Back section up/down” several times.
Then the system is automatically bleeded.
TEST!
Check all functions!
7
1
A
5
8 6 6
5
2
34 6 5
5 6
A
9
10
ATTENTION!
For SATURN GR do not use any other cable remote control than
item no. 4 545 605!
5. The position in the spar must be as with the old bowden cable!
Reassemble in reversed order
TEST!
Check after check list!
Load on one side 70 kg.
2
1
1
Figure 13: SATURN GY without pads
1 Toggle link socket
TEST!
Synchronize positions of both leg sections at lower stop (in lowest position)!
Adjust at thread of toggle link socket.
A
5
4 4
6
3 13
A 8
6 13
2
4
4 11
5 12 9
10
7
1
CAUTION!
For SATURN GY do not use any other cable remote control than
item no. 4 545 607!