Refrigeradores de Laboratorio
Refrigeradores de Laboratorio
Refrigeradores de Laboratorio
Refrigerators
TSX Series
CAUTION All internal adjustments and maintenance must be performed by qualified service personnel.
Material in this manual is for informational purposes only. The contents and the product it describes are
subject to change without notice. Thermo Fisher Scientific makes no representations or warranties with
respect to this manual. In no event shall Thermo be held liable for any damages, direct or incidental,
arising from or related to the use of this manual.
For your future reference and when contacting the factory, please have the following information readily
available. It can be found on the dataplate attached to your unit.
Model Number:
Serial Number:
Date Purchased:
Table of Contents
Models .......................................................................................1
Safety Precautions....................................................................2
Unpacking.................................................................................5
Packing List ..............................................................................6
General Recommendations .....................................................7
Temperature Monitoring ......................................................7
Intended Use ........................................................................7
Operating Standards ...............................................................8
Unit Specifications............................................................... 9
Installation..............................................................................10
Location .............................................................................10
Wiring ................................................................................14
Convenience Outlet (Chromatography Refrigerators) .......15
Shelves ...............................................................................18
Drawers ..............................................................................18
Baskets ...............................................................................23
Temperature Sensor Bottles ............................................... 23
Door Operation ..................................................................23
Remote Alarm (Optional) ..................................................25
Final Checks....................................................................... 26
Startup ....................................................................................27
Initial Startup ......................................................................27
Product Loading and Unloading Guidelines...................... 28
Operation................................................................................29
Control Panel...................................................................... 29
Display Functions .............................................................. 31
Programming Functions .....................................................32
Service Parameters .............................................................32
Temperature Settings .........................................................35
Alarms................................................................................ 35
Chart Recorder (Optional) ...................................................36
Set Up and Operation......................................................... 36
Changing Chart Paper ........................................................37
Power Supply .....................................................................37
Calibration and Adjustments.............................................. 38
Temperature Transmitter (Optional) ..................................39
Powered Temperature Transmitter..................................... 39
Unpowered Temperature Transmitter ................................39
Table of Contents
Maintenance ...........................................................................40
Cleaning the Cabinet Interior .............................................40
Cleaning the Condenser Filter ............................................40
Cleaning the Condenser ..................................................... 40
Automatic Defrost.............................................................. 41
Gasket Maintenance ...........................................................41
Alarm Battery Maintenance ...............................................42
Preparation for Storage ......................................................42
Cleaning the Unit (Chromatography Refrigerators) .......... 42
Replacing Sensor (Chromatography Refrigerators) ...........42
Sensor Maintenance (Chromatography Refrigerators) ...... 42
Troubleshooting .....................................................................43
End of Life Care.....................................................................45
Warranty ................................................................................46
Models
1 Models The table below shows the units covered in this operation and installation
manual by model number. Refer to Section 6.1 for details on the voltage
specifications for the various models.
Energy Star
Unit Model (*)
Model (*)
Thermo Scientific - TSX1205G* V/Y/Z A/D
2 Safety Precautions In this manual, the following symbols and conventions are used:
This symbol when used alone indicates important operating instructions which
reduce the risk of injury or poor performance of the unit.
This symbol indicates surfaces which may become hot during use and may cause
a burn if touched with unprotected body parts.
Before installing, using or maintaining this product, please be sure to read the
manual and product warning labels carefully. Failure to follow these instructions
may cause the product to malfunction, which could result in injury or damage.
This symbol indicates possible pinch points which may cause personal injury.
This symbol indicates a need to use gloves during the indicated procedures. If
performing decontamination procedures, use chemically resistant gloves.
Use this product only in the way described in the product literature and in this
manual. Before using it, verify that this product is suitable for its intended use. If
the equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
Do not modify system components, especially the controller. Use OEM exact
replacement equipment or parts. Before use, confirm that the product has not
been altered in any way.
WARNING: Disconnect the unit from all power sources before cleaning,
troubleshooting, or performing other maintenance on the product or its
controls.
EMC
EMC Registration is done on this equipment for business use only. It may cause
interference when the product would be used in home.
사용자 안내문 이 기기는 업무용 환경에서 사용할 목적으로 적합성평가
를 받은 기기로서 가정용 환경에서 사용하는 경우 전파간섭의 우려가 있
습니다 .
This equipment has been tested and found to comply with the limits for a
Class A digital device. Class A covers devices for usage in all establishments other
than domestic and that are not directly connected to a low voltage power supply
network, which supplies domestic environment.
This equipment has been tested and found to comply with the limits for a
Class A digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference when the
equipment is operated in a commercial environment. This equipment generates,
uses, and can radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely to
cause harmful interference in which case the user will be required to correct the
interference at his own expense.
3 Unpacking At the time of delivery, be sure to inspect the unit packaging for damage before
signing for the shipment. If packaging damage is present, request immediate
product inspection and file a claim with the carrier.
Note Packaging damage does not denote that unit damage exists.
If concealed damage is found (damage that is not apparent until the item has
been unpacked), stop further unpacking and save all packing for carrier's
inspection. Make a written request for inspection to delivering carrier. This must
be done within 15 days after delivery. Then file a claim with the carrier.
Do not return goods to the manufacturer without written authorization.
• USB Drive with this and other user manuals, including translated
versions
• Certificate of Conformance
• Warranty Card
If you have ordered a unit with shelves, the bag will also contain:
If the unit was ordered with a chart recorder, the bag will also contain:
• Calibration information
• Anti-Tip Bracket Kit (See Section 7.1.2, Table 3 for applicable models)
• Thermal Bottle
5 General This section includes some general recommendations for your unit.
Recommendations
5.1 Temperature
Monitoring
IMPORTANT NOTE We recommend the use of a redundant and
independent temperature monitoring system so that the refrigerator can be
monitored continuously for performance commensurate with the value of
product stored.
5.2 Intended Use The 5°C refrigerators described in this manual are high performance units for
professional use. These products are intended for use as cold storage in research
use and as a general purpose laboratory refrigerator, storing samples or inventory
at operating temperatures between 3°C and 7°C.
Expected users of this equipment include but are not limited to personnel from
the following areas: Professional and clinical laboratories, Pharma and Biotech
facilities, Academic, Industrial, and Government facilities or those trained in
laboratory protocols put in place at your facility. The units are not for use by the
general public.
It is not considered a medical device and has therefore not been registered with a
medical device regulatory body (e.g. FDA): that is, it has not been evaluated for
the storage of samples for diagnostic use or for samples to be re-introduced into
the body.
This unit is not intended for use in classified hazardous locations, nor to be used
for the storage of flammable or corrosive inventory.
6 Operating The refrigerators described in this manual are classified for use as stationary
Standards equipment in a Pollution Degree 2 and Overvoltage Category II environment.
These units are designed to operate under the following environmental
conditions:
• Indoor use
• Altitude up to 2000 m (6512 ft.)
• Maximum relative humidity 60% for temperatures from 15 to 32°C
(59 to 90°F).
• Main supply voltage fluctuations not to exceed 10% of the nominal
voltage.
• The refrigerator must not be connected to a GFCI (Ground Fault
Circuit Interrupter) protected outlet as it may be subject to nuisance
tripping.
6.1 Unit Specifications The last character in the model number listed on the refrigerator identifies the
electrical specifications for your unit. The dataplate is located on the upper left
side of the unit.
†
Weights provided are approximate weights for Laboratory Refrigerators with shelves. For TSX1205P* Pharmacy
Refrigerators, add 22 Kg (49 lbs). For TSX2305P*, TSX3005P*, and TSX5005P* Pharmacy Refrigerators,
add 37 Kg (82 lbs) for each set of 6 drawers. Add approximately 6 kg (13 lbs) for TSXxx05C* Chromatography
Refrigerators.
#
TSXxx05C* refrigerators have an additional 4.0 A rated current for the internal receptacle.
7 Installation
WARNING: Do not exceed the electrical rating printed on the dataplate
located on the upper left side of the unit.
CAUTION: Do not move the unit using the drain pan on the back. This
could cause damage to the equipment.
7.1 Location Install the unit on a level area free from vibration with a minimum of 6 inches of
space on the sides and rear and 12 inches at the top. Do not position the
equipment in direct sunlight or near heating diffusers, radiators, or other sources
of heat.
WARNING: Do not move the unit while loaded. Unit shall always be
moved when empty and by pushing slowly at handle level or lower on the
unit. Be especially careful on any uneven surfaces.
7.1.1 Installation Instructions The unit must be level both front to back and side to side when installed. If the
(Models A/D from Table 1) unit is out of level, you may need to shim the corners or casters with thin sheets
of metal. Be sure to set the brakes for units equipped with casters.
7.1.2 Installation Instructions The unit must be level both front to back and side to side when installed. If the
(Models V/Y/Z from Table 1) unit is out of level, you may need to shim the corners or casters with thin sheets
of metal. Be sure to set the brakes for units equipped with casters.
This refrigerator has been designed to meet all recognized industry tip standards
for all normal conditions when anti-tip bracket is installed and properly engaged.
Anti-tip Bracket Installation instructions are provided for wood and concrete
floors. Any other type of construction may require special installation techniques
as deemed necessary to provide adequate fastening of the anti-tip bracket to the
floor. For installation on floors other than wood and concrete, please contact
technical support.
The use of this bracket does not prevent the tipping of the refrigerator when not
properly installed.
Label Description
1 Bracket
2 Bolts
3 Anchors
4 Instructions and Installation Template
Wood Construction
a. Drill 15/64" (6 mm) pilot holes in locations marked in step 1.
Label Description
1 Back of Unit
2 1/2" Bolt Location
3 Detailed View
4 Lock Nut
5 1/2" Bolt
6 Floor
7 1/2" Clearance
4. Refrigerator Positioning
Check to see if the anti-tip bracket is installed properly by shining light under
cabinet and confirming bolt in cabinet is secured by bracket on floor.
7.2 Wiring Wiring diagrams are attached on the back of the cabinet.
CAUTION: For personal safety and trouble-free operation, this unit must
be properly grounded while in use. Failure to ground the equipment may
cause personal injury or damage to the equipment. Always conform to the
National Electrical Code and local codes. Do not connect the unit to
overloaded power lines.
CAUTION: Do not position the unit in a way that impedes access to the
disconnecting device or circuit breaker in the back of the unit.
CAUTION: Never cut the grounding prong from the service cord plug. If
the prong is removed, the warranty is invalidated.
7.3 Convenience Outlet Chromatography Refrigerators are equipped with a convenience outlet located
(Chromatography on the back wall. Power to the outlet is protected by a circuit breaker. If there is a
fault, the circuit can be reset by toggling the Chromatography Power Inlet Switch
Refrigerators) (Item D in Figure 4). The total load of the equipment using the convenience
outlet shall not exceed 4 Amps.
• RED (on the top) - When the red LED is ON, it indicates there is no
power to the AC outlet inside the unit.
• GREEN (on the bottom) - When the green LED is ON, it indicates
there is power to the AC outlet inside the unit.
7.3.2 NEMA 5-15 GFCI The NEMA 5-15 GFCI Receptacle has a standard self-test feature with color
Receptacle (optional) indicators for status:
• Solid Red - Indicates the breaker has opened the circuit. If this occurs,
unplug the equipment, check the circuit, and reset the circuit breaker.
• Flashing Red - Indicates “End of Life” and that the receptacle needs to
be replaced.
Red (Solid/Flashing)
Green
7.3.3 Power Inlets The chromatography refrigerator has two power inlets at the back of the unit:
1. Main Power Inlet (A)- This is used to connect the AC mains power
cord to the chromatography refrigerator.
7.3.4 Switches The chromatography refrigerator has three switches at the back of the unit:
1. Mains Power Inlet Switch (B)- This is used to turn ON or OFF the
AC supply to the refrigerator.
Refer to the label adjacent to the switches at the back of the refrigerator for
identification.
7.4 Shelves Single door laboratory refrigerators come standard with 4 full shelves and double
door laboratory refrigerators come standard with 8 full shelves.
Maximum shelf capacity is 45 kg (100 lbs) for full shelves and 22.5 kg (50 lbs)
for half shelves.
For safety in shipping, the shelves are packaged and secured inside the cabinet.
Insert the shelf support hangers (included with the manual inside the unit) into
the built-in shelf supports (located on the inside walls of the cabinet interior) at
the desired locations. Position the shelves on the flat supports (refer to Figure 5).
7.5 Drawers TSX1205 refrigerators have the option to be ordered with 5 drawers, TSX2305
and TSX3005 refrigerators with 6 drawers, and TSX5005 refrigerators with 12
drawers.
7.5.1 Removing the Drawers To remove the drawers, complete the following steps (refer to Figure 6):
1. Pull the drawer towards you until the slides are fully extended.
2. Lift the back of the drawer to disengage the mounting tabs from the
slots on the slides.
Note The drawers fit snugly between the slides. Push the back to the drawer
from underneath to remove the drawer.
3. Raise the back of the drawer almost to a vertical position and disengage
the front mounting clips from the sides.
Label Description
1 Tabs
2 Drawer slide
3 Slot
7.5.2 Reinstalling the Drawers To reinstall the drawers, complete the following steps (refer to Figure 6):
1. Pull both the slides of the drawer towards you until the slides are fully
extended.
2. Position the drawer between the slides and with the back facing end
raised at 45 degree angle, insert the mounting clips at the front into the
slots on front of the slides.
3. Push the back of the drawer down between the slides and insert the
basket tabs into the slots at the back.
Note The drawers fit snugly between the slides. Push on the back of the drawer
from the inside to insert the drawer tabs completely into the slots. Make sure
both the drawer tabs are aligned with the slots on the slides before pushing the
drawer down between the slides.
7.5.3 Changing Drawer Drawer slides have a small wire safety clip at the front pilaster which prevent the
Position slides from falling when the drawer is removed. To change the position of the
drawer slides, complete the following steps:
1. Locate the safety clip.
2. Slip a small screwdriver under the bottom of the wire clip and pry the
clip towards the inside of the refrigerator.
3. Lift up the slide at the front. The slide is free to move from the front
pilaster.
6. Determine desired location for the slide and insert the slide into the rear
pilaster at a 45 degree angle towards the center of the cabinet.
7. Once the slide is seated in the rear pilaster, rotate the slide until it is
parallel with the unit wall and insert slide into front pilaster.
CAUTION: Ensure the slide is level. If the slide is not seated in a level
orientation, the drawers/baskets will not function as intended. This could
cause injury or product damage.
8. Replace the safety clip by hooking the clip under the front pilaster slide
bracket. Then rotate the wire safety clip to position the other bend of
the wire safety clip to position the other bend of the clip over the top of
the same bracket.
7.6 Baskets TSX1205P pharmacy refrigerators come standard with 5 wire baskets, TSX2305P
and TSX3005P pharmacy refrigerators with 6 wire baskets, and TSX5005P
pharmacy refrigerators with 12 wire baskets.
Baskets are shipped in a factory installed position that promotes uniform spacing
between the baskets. The basket slides are adjustable to higher and lower
positions in the cabinet. You can position these slides in the vertical slots which
are spaced at one-inch intervals. The baskets must be removed in order to adjust
the position of the slides. The procedure for removing, reinstalling, and changing
position of the baskets can be found in Section 7.5.1, Section 7.5.2 and
Section 7.5.3.
7.7 Temperature Sensor The temperature displayed on the control panel is measured by a probe inserted
Bottles in the sensor bottle inside the cabinet. The sensor probe must be inserted in the
glycol bottle prior to operation. The sensor bottle is positioned on the left wall
towards the top of the unit.
To install the sensor(s) into the bottles, complete the following steps:
1. Remove the bottle from the bracket and verify that each bottle is full of
liquid.
2. Remove the solid cap from the bottle and save for the next time the unit
needs to be moved.
3. Remove the taped lid and temperature sensor(s) from the wall.
4. Insert the temperature probe(s) into the bottle.
5. Secure the cap to the bottle and return the bottle to the bracket.
7.8 Door Operation Laboratory refrigerators have either swinging or sliding doors.
There are port holes in the walls of the cabinet to help facilitate routing of
independent temperature sensors. Be sure to seal the holes after routing the wires
to prevent undesired air exchange.
7.8.1 Swinging Doors The doors on all swinging door units are designed to stay open if opened 90
degrees or more. The door spring tension cannot be adjusted. If the self-closing
door does not work properly, make sure the unit is level.
CAUTION: Keep hands and body parts clear of closing doors. The moving
parts create a potential pinch point.
7.8.2 Sliding Doors The doors on the 45 cu.ft. models are self-closing sliding glass doors. If the
self-closing mechanism is not working properly, check to make sure that the unit
is level.
CAUTION: Keep hands and body parts clear of closing doors. The moving
parts create a potential pinch point.
The sliding glass doors can be locked once they are in the closed position using
the key provided. To lock these doors:
1. Locate the lock in the middle of the cabinet at the bottom of the doors.
There is a hook in the middle of the cabinet at the top of the doors that can be
used to keep the doors open when required.
7.9 Remote Alarm All units have factory-installed local alarm contacts that can be used for remote
(Optional) alarm systems.
18 840 420
16 1330 665
14 2120 1060
12 3370 1685
Remote alarm terminals are located at the rear of the machine compartment. The
three terminals are: COMMON, OPEN ON FAIL (Normally Closed), and
CLOSE ON FAIL (Normally Open).
The contacts will trip in the event of a power outage, high temperature alarm or
low temperature alarm.
7.10 Final Checks Before start up, be sure to complete the following steps:
1. Make sure that the unit is free of all wood or cardboard shipping
materials, both inside and outside.
2. Check the positions of the shelves, drawers and baskets. If you want to
adjust the positions, see instructions in Section 7.4, Section 7.5 and
Section 7.6.
3. Verify that the unit is connected to a dedicated circuit.
8 Startup
8.1 Initial Startup To start up the refrigerator, complete the following steps:
1. Verify that the sensor probe(s) have been installed in the bottle(s) as
described in Section 7.7.
2. Connect the AC mains power cord to the mains power inlet connector.
3. Ensure the double pole circuit breaker switch located next to the power
inlet is in “ON” position (i.e. “I” position).
4. Insert the silver colored key in the switch and turn to the Power On
position. The display will show the actual cabinet temperature and the
compressor and evaporator should start within approximately
10 minutes.
5. Allow the unit to reach operating temperature before loading it with any
product. To stabilize the temperature profile, a 24-hour waiting period
is recommended.
6. If you desire to enable the alarms, turn the three position key switch one
turn further clockwise to the Alarm On position. To avoid nuisance
alarms, wait until the unit has pulled down to the desired operating
temperature (5°C default).
8. If desired, lock the cabinet door using the gold colored key. Place
duplicate key copies in a safe place.
All controls should now be fully operational, the alarm active (if enabled), and all
visual indicators active.
Note The Chromatography refrigerator is shipped with the Manual reset switch
at the back of the unit (refer Section 7.3.4) as ON. The Manual reset switch
should always be in ON position.
Upon startup, the red LED on the chromatography safety circuit turns ON.
After a power delay of approximately 4 minutes, the red LED on the
chromatography safety circuit turns OFF and the green LED on the chromo
safety circuit turns ON.
Note When the red LED is ON, the outlet will not be powered. The LEDs
indicate the status of the safety circuit only.
Note Operating equipment inside the refrigerator may shift the temperature
profile in the unit. Monitor the refrigerator as needed for any such shift.
8.2 Product Loading and When loading your laboratory refrigerator, take care to observe the following
Unloading Guidelines guidelines:
• For critical applications, be sure that the alarm systems are working and
active before you load any product.
• Ensure clearance between the top of the cargo and the bottom of the
shelf/drawer/basket. Lack of clearance may affect unit performance or
impede operation of drawer/basket. Keep cargo within the bounds of
the shelf/drawer/basket.
• For initial product loading and after removing the drawers or baskets for
cleaning, be sure to load the unit from the bottom drawer or basket up
to ensure the components are seated properly. (Applicable for drawers
and baskets only.)
• Never load the unit above the load limit line. This is important to
ensure that air can circulate properly and evenly distribute the
temperature throughout the interior.
9 Operation
13 6 7 8 9
12
Figure 14. Refrigerator Control Panel
The control panel is located on the top right side of your refrigerator. You can use
the three pushbuttons (#5, #8, and #9 in Figure 14) to change the temperature
display (#1) or to adjust temperature and alarm setpoints, as given in Section 9.3.
The thermometer display (#2) provides a quick visual indicator of current
cabinet temperature and alarm conditions.
5 Bars 5
When cabinet temperature exceeds the warm alarm setpoint, the top bar of the
thermometer flashes. When temperature is lower than the cold alarm setpoint,
the bulb flashes. When you are in programming mode (described in Table 9) the
thermometer shows the setpoint value you are changing.
6. Door ajar - illuminates when the refrigerator door is open longer than
3 minutes approximately (when the alarm is activated and the key
switch is turned to the alarm position).
9. Scan - pushbutton used to change the main display and for various
other functions.
10. Audible alarm - illuminates during warm and cold alarm conditions.
11. Light Switch (glass doors only) - pushbutton switch to toggle the
refrigerator light on and off without opening the door. Light will always
come on when the door is opened.
12. Key Switch - Switch used to turn the power and alarms on and off for
the unit.
13. Alarm On - Setpoint Security - when the key switch is in the alarm on
position, the alarms are on and the setpoints cannot be modified.
14. Power On - when the key switch is in the on position, the unit power is
on with no alarms active.
15. Off - when the key switch is in the off position, the unit is off.
9.3 Programming You can enter the programming mode by pressing the Scan button ( ) and
Functions holding for approximately 5 seconds. The display will then flash “Prg” to indicate
that you have entered the programming mode. Use ( ) and ( ) buttons to
modify the values and press the Scan ( ) button to save the value and go to the
next screen.
Note If the alarms are enabled, all parameters are read only and cannot be edited.
If the unit is left idle for a period of approximately 30 seconds during program
mode, the program mode shall be exited.
If at any point the scan button is pressed and held for a period of approximately
10 seconds, the unit shall enter the service mode.
9.4 Service Parameters You can enter the service mode by pressing the Scan button ( ) and holding
for approximately 10 seconds while in Programming mode (refer to Section 9.3).
The display will then flash “SEr” to indicate that you have entered the service
mode followed by the software checksum values flashing on the screen. The
service icon ( ) will also illuminate. Use up ( ) and down ( ) buttons to
modify the values and press the scan ( ) button to save the value and go to the
next screen.
Note If the alarms are enabled, the unit will not enter Service Mode.
3 Press Scan ( ) CuFt Size of the unit Setting for size of the unit.
9 Press Scan ( ) dEF Int Defrost Interval The time interval between defrost cycles in hours.
12 Press Scan ( ) qUA Quality Mode Normally OFF(00). Used for production line testing.
Control Probe
15 Press Scan ( ) Cnt Prb Displays control probe temperature (Read only).
Temperature
Lower Display Displays Err since this refrigerator does not have a
17 Press Scan ( ) d lo Prb
Probe lower display probe.
Defrost Probe
18 Press Scan ( ) dEF Prb Displays defrost probe temperature (Read only).
Temperature
If the unit is left idle for a period of approximately 30 seconds during service
mode, the unit will exit service mode and return to normal operation.
Note If an alarm occurs while in the service mode, it will not show until exiting
this mode.
CAUTION: Performing this reset will overwrite any changes that were
made after the receipt of this unit.
9.5 Temperature The factory default temperature setting is 5°C for all laboratory refrigerators. To
Settings change the factory settings, refer to programming instructions in Section 9.3.
9.6 Alarms The alarm system is designed to provide visual and audible warning signals for
both power failure and rise in temperature. The alarm is equipped with a battery
backup.
Default cold and warm alarm values are 2°C and 8°C. These values may be
adjusted, following instructions in Section 9.3.
The alarm system is activated only when the key switch is turned to the Alarm
On position. The audible warning signal sounds when there is a power failure,
temperature alarm condition, or when the door is ajar for more than 3 minutes
approximately.
The Mute function (pressing the ( ) button) allows you to turn off the audio
warning without turning off the visual indicators. The alarm will ring back after
approximately 6 minutes if the alarm is still active.
During a warm alarm, a flashing of the uppermost bar of the thermometer, the
temperature display, and the speaker icon will occur at the same time. If there has
been a warm alarm since the last alarm reset, but the temperature is not currently
in an alarm state, the uppermost bar of the thermometer and the speaker icon
will slowly alternate in flashing.
During a cold alarm, a flashing of the bulb of the thermometer, the temperature
display, and the speaker icon will occur at the same time. If there has been a cold
alarm since the last alarm reset, but the temperature is not currently in an alarm
state, the bulb of the thermometer and the speaker icon will slowly alternate in
flashing.
During a power failure, the power failure icon will illuminate, the thermometer
will display without the bulb, and the temperature will flash approximately every
3 seconds. If there has been a power failure since the last alarm reset, the power
failure icon and the speaker icon will slowly alternate in flashing.
During a low battery condition, the low battery icon will illuminate.
The alarms can be reset by pressing and holding the and simultaneously.
10 Chart Recorder Panel-mounted six inch recorders are available as a factory-installed option.
(Optional) Recorder operation begins when the system is powered on.
10.1 Set Up and To prepare the recorder to function properly, complete the following steps:
Operation 1. Open the recorder door to access the recorder.
2. Connect the nine volt DC battery located at the recorder’s upper right
corner. This battery provides back-up power.
3. Install clean chart paper (refer to Section 10.2 at the next page).
4. Close the recorder door.
Note The recorder may not respond until the system reaches temperatures
within the recorder’s range.
Label Description
1 Pressure Sensitive Chart Buttons
2 Imprinting Stylus
3 Reference Mark
5 Chart
10.2 Changing Chart To change the chart paper, complete the following steps:
Paper 1. Locate the pressure sensitive buttons at the front, upper left of the
recorder panel.
2. Press and hold the change chart button (#3 in Figure 16) for
approximately one second. The pen will move off the scale.
3. Unscrew the center nut, remove the old chart paper, and install new
chart paper. Carefully align the day and time with the reference mark on
the recorder panel (a small groove on the left side of the panel, as shown
in Figure 15).
4. Replace the center nut and hand tighten. Press the change chart button
again (#3) to resume temperature recording.
Note Be careful not to pull up on the chart recorder arm while changing paper.
Pulling on the arm may damage the recorder.
10.3 Power Supply The recorder normally uses AC power when the system is operating. If AC power
fails, the LED indicator on the recorder flashes to alert you to a power failure.
The recorder continues sensing cabinet temperature and the chart continues
turning for approximately 24 hours with back-up power provided by the
nine-volt battery.
The LED indicator glows continuously when main power is functioning and the
battery is charged.
When the battery is low, the LED flashes to indicate that the battery needs to be
changed.
10.4 Calibration and This recorder has been accurately calibrated at the factory and retains calibration
Adjustments even during power interruptions. If required, however, adjustments can be made
as follows:
1. Run the unit continuously at the control setpoint temperature.
Continue steady operation for at least two hours to provide adequate
time for recorder response.
2. Measure cabinet center temperature with a calibrated temperature
monitor.
3. Compare the recorder temperature to the solution temperature. If
necessary, adjust the recorder by pressing the left and right chart buttons
(#1 and #2 from Figure 16) for approximately 5 seconds.
Note The stylus does not begin to move until the button is held for
approximately 5 seconds.
11 Temperature
Transmitter
(Optional)
11.1 Powered A powered temperature transmitter is an option for this unit. This temperature
Temperature transmitter allows for remote temperature monitoring of your system.
Transmitter To connect your remote temperature monitoring system to the powered
temperature transmitter, refer to Figure 17.
Label Description
1 Terminal Strip
2 Rear of Unit
3 Controller and/or Indicator
4 Customer Provided System
Note The building management system load wiring must have a resistance less
than 340 ohms.
11.2 Unpowered An unpowered temperature transmitter is an option for this unit. This
Temperature temperature transmitter allows for remote temperature monitoring of your
system.
Transmitter
To connect your remote temperature monitoring system to the unpowered
temperature transmitter, refer to the diagram attached on the back of the unit.
12 Maintenance
WARNING: Disconnect equipment from main power before attempting
any maintenance to equipment or its controls unless stated otherwise.
12.1 Cleaning the To clean the cabinet interior, remove the shelves, drawers, or baskets following
Cabinet Interior the instructions in Section 7.4, Section 7.5 and Section 7.6. Use a solution of
water and a mild detergent for cleaning. Rinse the interior storage components
and wipe them dry with a soft cloth.
12.2 Cleaning the Clean the condenser filters every three months. There is one condenser filter
Condenser Filter located in the back cage of the unit that can be accessed without removing the
back cage or disconnecting the power.
1. Remove the filter by pulling upwards through the slot in the back cage.
4. Shake the excess water from the filter and let it dry.
5. Reinstall filter by pushing downwards through the slot in the back cage.
CAUTION: Do not pull the filter downwards from the bottom. The
condenser has sharp surfaces.
CAUTION: Never clean around the condensers with your fingers. Some
surfaces are sharp.
The condenser is located in the top rear of the machine compartment. To clean
the condenser, complete the following steps:
1. Disconnect the power.
5. Reconnect power.
12.4 Automatic Defrost The defrosting process on all models is primarily accomplished by air circulated
during off-cycle periods. This heat-free process ensures that the temperature is
not affected by the defrost cycle. The default defrost cycle runs approximately
once per hour and terminates once a preset evaporator temperature or timer
criteria is reached.
Defrost water is collected in a pan in the rear of the unit and evaporated using
system heat. No maintenance is required.
12.5 Gasket Periodically check the gaskets around the door for punctures or tears. Leaks are
Maintenance indicated by condensation or frost which forms at the point of gasket failure.
Make sure that the cabinet is level (refer to Section 7.1 for leveling information).
Keep the door gaskets clean and frost free by wiping gently with a soft cloth.
2. Insert a strip of paper (a couple of inches wide) between the door gasket
and the cabinet flange and close the door.
3. Slowly pull the paper strip from the outside. You should feel some
resistance.
4. Repeat this test at 4-inch intervals around the door. If the door does not
seal properly, replace the gasket.
12.6 Alarm Battery Have a certified technician replace the alarm battery every twelve months at most
Maintenance or when the alarm is active. The part number for a replacement battery is
322533H01.
12.7 Preparation for If the unit is going to be stored in an off condition, allow the unit to warm up
Storage and dry out with the door open before moving into storage.
12.8 Cleaning the Unit The refrigerator is designed to be cleaned using a mild detergent (such as Lysol®
(Chromatography wipes (non-aerosol) or Formula 409®) and water. Lightly spray the interior
storage components and wipe them dry with a soft cloth or spray the cloth first
Refrigerators) and then wipe interior surfaces. Do not spray directly on the center outlet
column. Use a damp cloth to clean around the outlet.
Use of other cleaners may contain chemicals that will turn off power to the
outlet. If this occurs, continue to clean the refrigerator and leave the doors open
for approximately 5 minutes to allow any vapors to exit the refrigerator. In the
event that the outlet power turns off due to chemicals, the red light will
illuminate and the user will need to manually reset the safety circuit using the
“Switch Manual Reset Chromatography” at the back of the unit before power is
restored. Toggling this switch off then on will restart the safety system and the
green light should illuminate after approximately 4 minutes. If the light
repeatedly changes from red to green or the red light stays illuminated, ensure
there are no chemical vapors present in the refrigerator and reset the system. If
this continues, please contact Customer Service.
12.9 Replacing Sensor The sensor needs replacement if the temperature goes below freezing in the
(Chromatography cabinet. Use system alarms to ensure the temperature inside the unit is always
above 0°C. If exposed to freezing conditions the sensor shall be replaced.
Refrigerators)
If the display shows an error and sounds an alarm (in case of a control, defrost,
bottle or ambient probe failure), the sensor needs to be replaced.
12.10 Sensor The sensor shall be replaced every five years by a trained service provider.
Maintenance
(Chromatography
Refrigerators)
13 Troubleshooting
WARNING: Troubleshooting procedures involve working with high
voltages which can cause injury or death. Troubleshooting should only be
performed by trained personnel.
Check that the double pole circuit breaker located next to the power inlet is in
Unit does not “ON” position (i.e “I” position). Try cycling the switch to OFF position (i.e
operate or “O” position) and then bring to ON position (“I”).
Power supply
Power Failure
Indicator is on Test the voltage and verify that it is correct for your unit (refer to Table 2).
The unit should not be connected to a GFCI (Ground Fault Circuit Interrupter)
protected outlet as it may be subject to nuisance tripping.
Temperature
Make sure that the control is set correctly. Refer to Section 9.3.
Control
Condenser Make sure the condenser and filter are clean. Refer to Section 12.2 and
Temperature clogged Section 12.3.
fluctuates
Make sure the solution bottles for the temperature sensors are full. The
Solution bottle
solution is a 50/50 mixture of glycerin and distilled water.
If the temperature control is set correctly, the condenser is clean, but
Other causes
temperature continues to fluctuate, call an authorized service representative.
12V backup
Low battery Replace the battery. It is located on the top right hand side of the cabinet. Call
battery needs to
icon is lit an authorized service representative.
be replaced.
Incorrect
Perimeter Heater
Duty Cycle Increase the Perimeter Heater Duty Cycle, refer to Section 9.4.
(swinging doors
Condensation only)
around door
frame Gaps exist in Ensure all port holes in the cabinet top, sides, and back are sealed properly to
unit port holes prevent warm airflow into the cabinet. Seal any gaps.
Door seal is Verify nothing is placed through the door seal such as a sensor. Check the door
broken seal following instructions in Section 12.5.
14 End of Life Care Be sure to follow local regulations when disposing of an old unit. Some
suggestions are listed in the following:
1. Remove items and defrost unit. Be sure to clean up any biological safety
hazards.
15 Warranty Domestic Warranty • 2 Years Parts and Labor Plus an Additional 8 Years on
V-Drive
During the first twenty four (24) months from shipment, Thermo Fisher
Scientific Inc, through its authorized Dealer or service organizations, will at its
option and expense repair or replace any part found to be non-conforming in
material or workmanship with the exception of V-drive which is covered for an
additional 8 years (96 months) from the time of the shipment.Thermo Fisher
Scientific Inc reserves the right to use replacement parts, which are used or
reconditioned. Replacement or repaired parts will be warranted for only the
unexpired portion of the original warranty.
This warranty does not apply to damage caused by (i) accident, misuse, fire, flood
or acts of God; (ii) failure to properly install, operate or maintain the products in
accordance with the printed instructions provided, (iii) causes external to the
products such as, but not limited to, power failure or electrical power surges, (iv)
improper storage and handling of the products, (v) use of the products in
combination with equipment or software not supplied by Thermo Fisher; or (vi)
installation, maintenance, repair, service, relocation or alteration of the products
by any person other than Thermo Fisher or its authorized representative. To
obtain proper warranty service, you must contact the nearest authorized service
center or Dealer. Thermo Fisher Scientific, Inc’s own shipping records showing
date of shipment shall be conclusive in establishing the warranty period. At
Thermo Fisher’s option, all non-conforming parts must be returned to Thermo
Fisher postage paid and replacement parts are shipped FOB Thermo Fisher’s
location.
Limitation of Liability
WEEE Konformittät. Dieses Produkt muss die EU Waste Electrical & Electronic Deutschland
Equipment (WEEE) Richtlinie 2012/19/EU erfüllen. Das Produkt ist durch folgendes
Symbol gekennzeichnet. Thermo Fisher Scientific hat Vereinbarungen getroffen mit
Verwertungs-/Entsorgungsanlagen in allen EU-Mitgliederstaaten und dieses Produkt muss
durch diese Firmen widerverwetet oder entsorgt werden. Mehr Informationen über die
Einhaltung dieser Anweisungen durch Thermo Scientific, dieVerwerter und Hinweise die
Ihnen nützlich sein können, die Thermo Fisher Scientific Produkte zu identizfizieren, die
unter diese RoHS. Anweisungfallen, finden Sie unter www.thermofisher.com/WEEERoHS
Conformità WEEE. Questo prodotto deve rispondere alla direttiva dell’ Unione Europea Italia
2012/19/EU in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici (WEEE).
È marcato col seguente simbolo.Thermo Fischer Scientific ha stipulato contratti con una o
diverse società di riciclaggio/smaltimento in ognuno degli Stati Membri Europei. Questo
prodotto verrà smaltito o riciclato tramite queste medesime. Ulteriori informazioni sulla
conformità di Thermo Fisher Scientific con queste Direttive, l’elenco delle ditte di
riciclaggio nel Vostro paese e informazioni sui prodotti Thermo Scientific che possono
essere utili alla rilevazione di sostanze soggette alla Direttiva RoHS sono disponibili sul sito
www.thermofisher.com/WEEERoHS
www.thermofisher.com
327929H01 Rev. E
Contact Us
North America: 866-984-3766
Europe: France +33 2 2803 2180 Germany +49 9184 90 9640 UK/Ireland + 44 870 609 9203
Asia: China +86 21 6865 4588 India 1800 22 8374 Japan +81 45 453 9220