Asnt RT Question Bank
Asnt RT Question Bank
Asnt RT Question Bank
2. Gauge pressure is :
a. the harmonic pressure within a gage
b. atmosphere pressure plus absolute pressure
c. the difference between absolute and atmosphere pressure
d. a sum of the partial pressures of air at standard temperature and
pressure
18. If the HMS leak detector’s forepump belt breaks while the instrument
is in operation and unattended:
a. No damage will occur since the automatic protection system will be
activated
b. The vacuum system will stay at low vacuum as long as the cold trap
remains filled
c. Backstreaming of the diffusion pump oil vapors will occur
contaminating the system
d. The system will continue to function properly as long as the
diffusion pump is operating
22. When vacuum testing large volumes with a mass spectrometer leak
detector, the response and cleanup times depend on :
a. the volume of the system under test
b. the conductance of the connecting tubes
c. the pumping speed of the mass spectrometer leak detector unit
d. all of the above
23. Operating the mass spectrometer tube at a pressure greater than that
recommended by the manufacturer is undesirable because :
a. too many electrons will be produced
b. the spectrometer filament may burn out prematurely
c. the diffusion pump cannot handle it without wear
d. the pressure relay switches will arc
29. When using the mass spectrometer leak detector probe test
technique, it is :
a. difficult to detect small leaks that are close to large leaks
b. impossible to segregate leaks
c. desirable to have a draft blowing across the test surface
d. simple to accurately size detected leaks
31. A vacuum pump using a vapor jet as the pumping means is called a :
a. mechanical vacuum pump
b. cold trap
c. diffusion pump
d. fission pump
34. Tracer gas ions in a helium mass spectrometer leak detector are
separated from the remainder of the ion beam by the :
a. ion collector
b. magnetic field
c. hot tungsten filament
d. cold trap
35. The evacuation of the test object to the point at which its volume can
be exposed to a MSLD system is called the:
a. Diffusion cycle
b. Vacuum cycle
c. Roughing cycle
d. Pressurizing cycle
36. The gage that is used to monitor the pressure in the high vacuum
section of a mass spectrometer leak detector is called a :
a. cold-cathode discharge gage
b. Bourdon pressure-vacuum gage
c. Pirani gage
d. bulk pressure gage
37. The mass spectrometer vacuum gage that is used to monitor the
pressure during the roughing cycle is called a :
a. cold-cathode discharge gage
b. thermal conductivity gage
c. Bourdon pressure meter
d. bulk pressure gage
38. Which of the following factors would affect the sensitivity of a mass
spectrometer leak detector?
a. use of a remote control system
b. the pumping speed of the mass spectrometer leak detector vacuum
system
c. design of the roughing pump
d. type of throttle valve used
43. Increasing probing speed over the area being tested will result in :
a. an increase in response time
b. an increase in test sensitivity
c. a decrease in test sensitivity
d. a decrease in instrument sensitivity
44. The scale of a mass spectrometer leak detector output meter reads in :
a. SCFM
b. micrometers (µm)
c. torr (µm Hg)
d. none of the above
46. For a mass spectrometer leak detector, what is the function of the
roughing pump?
a. to evacuate the system external to the mass spectrometer leak
detector
b. to pressurize the system external to the mass spectrometer leak
detector
c. to evaluate the internal mass spectrometer leak detector high
vacuum section
d. none of the above
56. When performing a helium mass spectrometer tracer probe test, the
probing should be conducted in a sequence from the:
a. Highest to the lowest areas in the system
b. Farthest to the nearest areas from the instrument
c. Most difficult to the easiest areas to test
d. Lowest to the highest areas in the system
59. The safest effective and economical cryogenic fluid commonly used
in cold traps is liquid:
a. Oxygen
b. Argon
c. Nitrogen
d. Hydrogen
60. In the event of power failure while operating any helium mass
spectrometer, the first thing the operator should do is:
a. Look for some other source of electrical power
b. Disconnect the power cord to the machine so that when the power
comes back “ON”, the electronics will not come “ON” before the
roughing pump
c. Close the diffusion pump isolation (accumulator) valve
d. Vent the system to atmosphere, remove the cold trap and clean it.
61. When starting up a helium mass spectrometer from total shut down,
the high vacuum system diffusion pump takes about
__________________ to begin pumping after it is turned “ON”.
a. 2 minutes
b. 20 minutes
c. 60 minutes
d. 1-1/2 hours
www.sisndt.com / www.ndtsis.com P a g e 10
Southern Inspection Services
www.sisndt.com / www.ndtsis.com P a g e 11
Southern Inspection Services
www.sisndt.com / www.ndtsis.com P a g e 12