Bio Reactor 2016
Bio Reactor 2016
Bio Reactor 2016
CHEG 4137W
Bioreactor
Objective:
The laboratory has acquired a bioreactor for the purpose of growing cell cultures. In the initial attempt
to grow E. coli, an M9 growth media was used, which used the minimal amount of salt and nutrients to
encourage bacterial growth. After a few successful attempts with the M9, groups became unable to
grow E. coli in the media, prompting a switch to LB media, which is much richer in nutrients.
Experimental conditions for the baseline fermentation using M9 media is as follows:
Your goal is to compare the kinetic parameters for bacterial growth using the LB media to those of the
M9 media. You will measure the bacterial optical density (OD) over time and construct a bacterial
growth curve. You will also use a numerical method to estimate the substrate concentration in your
reactor as time progresses. Due to the lengthy nature of a single reactor run, you will likely only
achieve one complete reactor run during the duration of your lab. As such, it is important to
communicate with other groups performing this experiment to prevent tests of repeated conditions.
After collection of your own data, you will be given a sample of previously collected data for both the
M9 media and the LB media.
Theory:
In this experiment, you will be determining some of the kinetic parameters for the growth of E. coli.
Some relevant equations include:
Where X = biomass concentration, S = substrate concentration, YX/S = maintenance coefficient (~0.5 for
this experiment), = growth parameter, Ks = Monod constant, and MAX = the maximum growth rate
parameter. You will be measuring the biomass optically using cell density in a UV-vis spectrometer at
600 nm. Determining other parameters will be the focus of this experiment. For relevant information,
see the references at the end of this document.
Safety Precautions:
E. coli is a biological agent, so the following safety precautions MUST be followed when working with
it. Failure to observe these precautions will result in expulsion from the lab.
1. Gloves, goggles, and lab coats must be worn at all times when working with the bioreactor.
2. If you wish to walk away from the reactor bench, first remove and dispose of your gloves, then
wash your hands with antibacterial soap. DO NOT wear your gloves around the lab or when
working in the communal computer space.
3. Always throw out used gloves in a separate trash bag; do not throw out gloves directly in the
trash!!!
4. Place only contaminated glassware in the wash basin. Anything that enters the wash basin
should immediately be considered contaminated.
5. Before and after working with the bacteria, spray down the countertop with a 1% bleach
solution, then with DI water, then with a 70% ethanol solution.
6. When discarding samples, pour some 70% ethanol solution into a spare plastic bottle. Empty all
samples into the bottle, spraying the cuvettes, centrifuge tubes, etc. out with ethanol. This
waste solution may be poured down the drain at the completion of your experiment. The bottle
must be washed with soap and water after.
7. ALWAYS wash your hands with antibacterial soap after removing your gloves.
Available Variables: Temperature, Glucose Concentration (for M9 media only), Mixing Speed
Due to the lengthy set up of this experiment, the following schedule is recommended:
Lab Session 1
Clean and autoclave the reactor
Clean and autoclave all glassware and pipet tips
Analysis:
Your analysis must include:
1. A determination of the Monod constant, Ks, and the max growth parameter, max for your
selected conditions. How do the Monod constant and maximum growth parameter vary based
on the media you used? What does this observation mean in terms of what is physically
occurring in the bioreactor and in the flask?
Report:
Describe the design of your experiments and the results obtained, including an error analysis. Provide
thoughtful and quantitative discussion of results. Be sure to include your bacterial growth curve (you
may want to consider a semilog plot for this) and glucose concentration curve. Report all appropriate
biokinetic parameters. Express any discrepancies between observed and predicted results in terms of
quantified experimental uncertainties or limitations of the correlations or computational software used.
Pro Tips:
1. The bioreactor is a very expensive piece of equipment. Take the following precautions to avoid
damaging or breaking it:
When moving the reactor vessel away from the bench, always transport it on a cart.
When autoclaving the reactor, always tighten the lid on loosely so the glass has room to
expand as it heats.
When reassembling the bioreactor, the impeller motor should gently slide onto the
impeller. Do NOT try to force it. You may need to readjust the position of the impeller
to get a good fit.
2. Never tightly cap anything you autoclave. Either apply the caps loosely or cover open holes
with aluminum foil.
3. Only glass or metal materials are autoclavable. The only plastic materials that can be
autoclaved are the 100 mL graduated cylinders, the Tygon tubes on the bioreactor, and the pipet
tips. Do not put any other plastic laboratory equipment in the autoclave.
4. You may autoclave glassware many days in advance. It will remain sterile as long as the
aluminum foil cover is not disturbed.
5. Use the thermally reactive autoclave tape on items to ensure the autoclave is functioning as
intended.
6. When leaving the lab with anything that may be biologically active, travel in pairs. One partner
will carry anything with bacteria, and the other partner will open doors for the first to minimize
contamination. Use the basin to carry flasks that contain bacteria.
7. It is wise to prepare an extra inoculation flask. In the event one of your bacteria samples does
not grow, you will have a back-up.
8. When performing the inoculation, make sure the outside of your flasks and the rubber mat in
the incubator/shaker are clean and dry. Do not exceed 300 RPM.
9. Do not turn on the bioreactor unless water is flowing from the faucet line. Do not try to run the
reactor unless the reactor is on the base and connected to the water lines.
10. When running the bioreactor, be sure to keep all open ports capped with foil unless taking
samples.
11. When preparing diluted samples, always use fresh pipet tips. Be careful not to insert a pipet tip
contaminated with bacteria into your fresh media, as this will force you to discard the media
solution.
References:
1. Shuler, M.L., Kargi, F., Bioprocess Engineering: Basic Concepts. 2nd Ed., Prentice Hall, New York, (2001).
2. Blanch, H. W., Clark. D. S. (1997), Biochemical engineering
3. James E. Bailey, David F. Ollis, Biochemical Engineering Fundamentals, 1986
4. James M. Lee, Biochemical Engineering, Prentice, 1992 (or copy of ebook, 2003).
Bioreactor & Benchtop Shaker Operations Method
Note: For this full-day experiment, you will run the bioreactor and the benchtop shaker simultaneously.
*Steps 3-6 are performed the day before the full-day experiment*
1. If needed, turn on the water faucet line connected to the reactor, then turn on the bioreactor.
2. Using the autoclaved glass funnel, add 2L of LB Media into the bioreactor.
3. Calibrate the DO probe (see Appendix for procedure).
4. Place the impeller motor onto the reactor. It should sit firmly atop the reactor. You may need to
manipulate the impeller manually to get a good fit. You should not try to force the motor onto
the reactor.
5. Set the temperature control setting to “Prime” and press “Enter.” Let the reactor run at this
setting for roughly 30 seconds, then set the temperature control setting to “P-I-D.”
6. Set the desired mixing rate and temperature using the control panel. Make sure that the
“Control” setting for both agitation and temperature is set to “P-I-D.”
7. Fill one 50 mL centrifuge tube with about 25 mL media. Cap and save for later; this will be
used if the samples need to be diluted.
8. Fill the contaminated basin with approximately 3 inches of water and 2 capfuls detergent
9. Pour some ethanol into the e.Coli waste bottle.
10. Pipet 1000 L of fresh media from the centrifuge tube into a clean cuvette. Use this blank to
zero the spectrophotometer at 600 nm.
11. Empty the contents of the cuvette into the bottle of ethanol, then place all contaminated
cuvettes/glassware into the decontamination basin.
12. Begin the e.Coli growth in the bioreactor
a. Once the media in the reactor is at temperature, pour 50 mL of your inoculated bacteria
into the reactor carefully.
b. Immediately extract 1 mL from the reactor using a syringe, place it into a clean cuvette
and measure the absorbance at 600 nm. Empty the contents into the ethanol bottle and
place all contaminated equipment into the decontamination basin.
13. Begin the e.Coli growth in the benchtop shaker
a. Fill each of the three 250-mL flasks with 100 mL of fresh media
b. Pipet 2.5 mL of your inoculated bacteria into each flask
c. Immediately extract 1 mL from each of the flasks in the shaker using the micropipette,
place into a clean cuvette and measure the absorbance at 600 nm. Empty the contents
into the ethanol bottle and place all contaminated equipment into the decontamination
tub.
d. Place the flasks in the shaker to begin the cell growth.
14. Collect optical density readings for the bioreactor and benchtop shaker flasks every 30 minutes.
If your readings on the spectrophotometer begin to get high (above 1.000), you will need to
dilute the next sample you take with fresh media. Be sure to note the composition of each
sample (i.e. how much sample and how much is fresh media).
15. Once your measurements are stabilized for an hour, the reaction has completed. Take your final
samples, then pour about 1 capful detergent into the reactor for at least 10 min of sterilization.
Make sure you turn the temperature setting to “OFF”
16. Discard the contents of your e.Coli waste container in the sink, and place the container in the
disinfection basin for cleaning.
17. Turn the agitation setting to “OFF” after sterilizing with detergent.
Step 8: Clean Up
1. Empty the decontamination tub and refill with soap and water. Leave your glassware to soak.
2. Disconnect the cooling water tubes and impeller motor on the bioreactor.
3. Spray down the countertop, control station surface, and biosafety cabinet with 1% bleach
solution, then distilled water, then ethanol.
4. Empty the contents of the bioreactor into a sink, and scrub all surfaces with soap and water.
This includes the inside of the reactor, the impeller and tubes inserted into the reactor, and all
exterior surfaces.
5. Return the bioreactor to the control base.
6. Clean the rest of your glassware with soap and water. Leave glassware and bottles on the
drying rack to dry. Rinse off the cuvettes with water and dispose of them in the designated
waste bag.
7. Clean the countertop and the biosafety cabinet with bleach, DI water, and ethanol
Appendix
pH Probe Calibration
Note: If autoclaving the probe, then pH probe calibration is done prior to autoclaving
When the system is operated for the first time, or when the electrode has been disconnected from the
voltage source for longer than 5-10 minutes, the electrode must be connected to the operating O2
amplifier for polarization prior to calibration. The electrode is polarized and ready for operation after
six hours of polarization time. Do not calibrate until the electrode is polarized.
Calibration should be performed prior to inoculation. The DO probe should be immersed in the liquid
medium during calibration.