Model DT450G Dust Monitor: User's Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 41

User’s

Manual Model DT450G


Dust Monitor

IM 11K01C01-01E

IM 11K01C01-01E
5th Edition
<Introduction> i

Introduction
The DT450G Dust Monitor is a process instrument designed to detect solid particles ingas and
integrates a control unit and a sensor probe into a compact enclosure. The DT450G can be
used in a wide variety of applications, including continuous monitoringof the dust content of the
exhaust gas.
This manual describes the procedures for installing, operating, inspecting and maintainingthe
DT450G Dust Monitor. Before using the instruments, thoroughly read themanual to ensure
optimum performance.
This manual consists of six chapters. For your reference, the following table shows theoutlines
of each chapter and indicates which chapters should be read for installation,operation or
maintenance work.
Table of Contents
Relates to
Chapter Outline
Installation Operation Maintenance
1. Overview Standard specifications, Model and Suffix codes
B C B
(or part number), and dimensions.
2. Installation Installation procedures, etc. A C
3. Piping/Wiring Piping and wiring procedures. A C
4. Operation Operation and display in detail. A C
5. Inspection and Inspection/Maintenance
B A
Maintenance
6. Troubleshooting Failures, possible causes, and remedies C A
CMPL (parts list) User replaceable parts list C B
CMPL: Customer Maintenance Parts List

A: Read and completely understand before operating the equipment.


B: Read before operating the equipment, and refer to it whenever necessary.
C: Recommended to read at least once.

Media No.IM 11K01C01-01E (YK) 5th Edition :July 2018 (YK) IM 11K01C01-01E
All Rights Reserved Copyright © 2005, Yokogawa Electric Corporation
ii <Introduction>

 For the safe use of this equipment

CAUTION
Connect the power supply cord only after confirming that the supply voltage matches the rating of
this equipment. In addition, confirm that the power is switched off when connecting power supply.
Some process gas is dangerous to people. When removing this equipment from the process line
for maintenance or other reasons, protect yourself from potential poisoning by using a protective
mask or ventilating the area well.

l About This Manual


• This manual should be passed on to the end user.
• The contents of this manual are subject to change without prior notice.
• The contents of this manual shall not be reproduced or copied, in part or in whole, without
permission.
• This manual explains the functions contained in this product, but does not warrant that those
will suit the particular purpose of the user.
• Every effort has been made to ensure accuracy in the preparation of this manual. However,
should any errors or omissions come to the attention of the user, please contact the nearest
Yokogawa Electric representative or sales office.
• This manual does not cover the special specifications. This manual may not be changed
on any change of specification, construction and parts when the change does not affect the
functions or performance of the product.
• If the product is used in a manner not specified in this manual, the safety of this product may
be impaired.

l Safety and Modification Precautions


• Follow the safety precautions in this manual when using the product to ensure protection
and safety of personnel, product and system containing the product.
• The following safety symbols and wordings are used on the product as well as in this
manual.

l The following safety symbols are used on the product as well as in this
manual.

WARNING
This symbol indicates that an operator must follow the instructions laid out in this manual
in order to avoid the risks, for the human body, of injury, electric shock, or fatalities. The
manual describes what special care the operator must take to avoid such risks.

CAUTION
This symbol indicates that the operator must refer to the instructions in this manual in order
to prevent the instrument (hardware) or software from being damaged, or a system failure
from occurring.

CAUTION
This symbol gives information essential for understanding the operations and functions.

IM 11K01C01-01E
<Introduction> iii
NOTE
This symbol indicates information that complements the present topic.

This symbol indicates Protective Ground Terminal.

This symbol indicates Function Ground Terminal. Do not use this terminal as the protective
ground terminal.

l Special descriptions in this manual


This manual indicates operation keys, displays and drawings on the product as follows:
• Operation keys, displays on the panel
(Ex. message display → "BASE")
(Ex. data display → "102" lit, "102" flashing)

IM 11K01C01-01E
iv <Introduction>

 NOTICE
l Specification check
When the instrument arrives, carefully unpack and check that the instrument has not been
damaged during transportation. Please make sure that the product received meet the purchase
order specifications and all the accessories are present. Specifications can be checked by the
model codes indicated on the nameplate. For details of the model codes, refer to Chapter 1,
“Overview.”

l Details on operation parameters


The DT450G is preset with defaults prior factory shipment. Before measurement, verify that
these factory default settings meet the operating conditions and if necessary, reconfigure
parameters. For checking the parameter settings, refer to Chapter 4.
It is recommended that the reconfigured parameter settings be recorded.

l How to dispose the batteries:


This is an explanation about the new EU Battery Directive (DIRECTIVE 2006/66/EC). This
directive is only valid in the EU.
Batteries are included in this product. Batteries incorporated into this product cannot be removed
by yourself. Dispose them together with this product.
When you dispose this product in the EU, contact your local Yokogawa Europe B.V.office. Do not
dispose them as domestic household waste.

Battery type: silver oxide battery

Notice:
The symbol (see above) means they shall be sorted out and collected as ordained in ANNEX II in
DIRECTIVE 2006/66/EC

IM 11K01C01-01E
<Introduction> v

 After-Sales Warranty
 Do not modify the product.
 During the warranty period, for repair under warranty carry or send the product to the local
sales representative or service office. Yokogawa will replace or repair any damaged
parts and return the product to you.
 Before returning a product for repair under warranty, provide us with the model name
and serial number and a description of the problem. Any diagrams or data explaining the
problem would also be appreciated.
 If we replace the product with a new one, we won’t provide you with a repair report.
 Yokogawa warrants the product for the period stated in the pre-purchase quotation.
Yokogawa shall conduct defined warranty service based on its standard. When the
customer site is located outside of the service area, a fee for dispatching the maintenance
engineer will be charged to the customer.
 In the following cases, customer will be charged repair fee regardless of warranty period.
• Failure of components which are out of scope of warranty stated in instruction manual.
• Failure caused by usage of software, hardware or auxiliary equipment, which Yokogawa
Electric did not supply.
• Failure due to improper or insufficient maintenance by user.
• Failure due to modification, misuse or outside-of-specifications operation which Yokogawa
does not authorize.
• Failure due to power supply (voltage, frequency) being outside specifications or abnormal.
• Failure caused by any usage out of scope of recommended usage.
• Any damage from fire, earthquake, storms and floods, lightning, disturbances, riots,
warfare, radiation and other natural changes.
 Yokogawa does not warrant conformance with the specific application at the user site.
Yokogawa will not bear direct/indirect responsibility for damage due to a specific application.
 Yokogawa Electric will not bear responsibility when the user configures the product into
systems or resells the product.
 Maintenance service and supplying repair parts will be covered for five years after the
production ends. For repair for this product, please contact the nearest sales office
described in this instruction manual.

IM 11K01C01-01E
Blank Page
<CONTENTS> i

Model DT450G
Dust Monitor
IM 11K01C01-01E 5th Edition

CONTENTS
Introduction...............................................................................................................i
 For the safe use of this equipment..................................................................... ii
 NOTICE.................................................................................................................. iv
 After-Sales Warranty............................................................................................ v
1. Overview..................................................................................................... 1-1
1.1 Product Overview.............................................................................................. 1-1
1.2 Specifications..................................................................................................... 1-2
1.3 Model and Suffix Codes.................................................................................... 1-4
1.4 Dimensions......................................................................................................... 1-5
2. Instruction.................................................................................................. 2-1
2.1 Selection of Installation Location.................................................................... 2-1
2.2 Machining of Probe Inlet................................................................................... 2-4
2.3 Installing the DT450G........................................................................................ 2-6
3. Piping and Wiring...................................................................................... 3-1
3.1 Connecting the Pipe for Air Purge................................................................... 3-2
3.2 Connecting the Wires for the Power Supply................................................... 3-3
3.3 Connection of Wires for the Output Signals................................................... 3-4
3.4 Grounding the Duct........................................................................................... 3-5
4. Operation.................................................................................................... 4-1
4.1 Components and Their Functions................................................................... 4-1
4.2 Preparation for Operation................................................................................. 4-2
4.2.1 Inspection of the State of Installation.................................................. 4-2
4.2.2 Supply of Purge Air............................................................................. 4-2
4.2.3 Supply of Power.................................................................................. 4-2
4.2.4 Setting of Parameters......................................................................... 4-3
4.3 Steady-state Operation..................................................................................... 4-6
4.3.1 Examining the Operating Condition.................................................... 4-6
4.3.2 Shutdown............................................................................................ 4-6
4.3.3 Restart................................................................................................. 4-6
5. Inspection and Maintenance.................................................................... 5-1
5.1 Periodic Maintenance........................................................................................ 5-2
5.1.1 Cleaning the Sensor Probe................................................................. 5-2
5.1.2 Correcting (Resetting) the Measurement Range............................... 5-2

IM 11K01C01-01E
ii <CONTENTS>

5.2 Replacing Parts.................................................................................................. 5-2


6. Troubleshooting........................................................................................ 6-1
Customer Maintenance Parts List ........................................CMPL 11K01C01-01E
Revision Information................................................................................................i

IM 11K01C01-01E
<1.Overview> 1-1

1. Overview
This chapter discusses the following topics.
· Product overview
· Specifications
· Model and suffix code
· Dimensions

1.1 Product Overview


Powder is used in many industries including ceramics, cement, chemicals, pharmaceutical and
food. In their processes, powder is collected in order to reduce the loss incurred in the process.
Dust emissions from industrial plants are limited by the air pollution control law, local regulations
or other guidelines. Each industrial facility is required to be maintained so that its dust emissions
do not exceed the allowable level.
Air pollution control systems such as bag filters and electrostatic precipitators are used to
separate particulate or dust from the gas stream before it is exhausted into the atmosphere. To
maintain the dust concentration of the emission gas below the specified level through effective
operation of the control system, it is essential to continuously monitor the dust concentration of
the exhaust gas.
Based on installation and application experiences with the existing DT400G model, the
microprocessor-based DT450G Dust Monitor features improved operationality and reliability.
Features include one-touch automatic range setup in response to process conditions and
automatic drift compensation. Also, the DT450G can measure at a process temperature of up to
250ºC, allowing it to cover a wide range of applications.

l Measuring Principle
When two solid particles come into contact with each other, charge transfer takes place between
them. This charge transfer is known as the principle of triboelectrification or contact electrification
and the DT450G utilizes this principle. More specifically, charge transfer occurs when particles in
a gas stream collide with (or pass nearby) the sensor probe, producing an electric current. The
DT450G amplifies this current to convert it into its output current. The electric current produced
by frictional electricity is subject to the effects of the following factors:
• Amount of the particles' electrostatic charge and a change in the polarity
• Particle size
• Velocity of particles
• Components of particles
The behaviors of these factors vary widely depending on the object of measurement and their
effects are in no way uniform. In the case of fixed-position measurement in a particular plant
however, measurement conditions, such as the components of particles, are almost constant
and therefore the effects are usually marginal. Dust concentration is also constant in the case of
gases emitted from a bag filter, cyclone or electrostatic dust collector.
A current-mode input signal obtained at the detector block is transferred to the converter block,
where the signal is converted to a 4-20 mA DC analog output signal. The converter block outputs
an alarm contact signal at a desired level, in addition to the analog output signal.

IM 11K01C01-01E
1-2 <1. Overview>

1.2 Specifications
Measurement objects: Solid particles in gas
Particle size : 0.3 µm or greater
Measurement range: 0.1 mg/m3 to 1 kg/m3
Measurement principle: Inductive Electrification
Process conditions :
Temperature : 250ºC or less
Pressure : 200 kPa or less
Gas flow rate : 4 m/s or greater (approx. 30 m/s max)
Humidity : 40%RH or less
Input surge voltage : 100 V or less (if 500 V or greater, converter may be damaged.)
Note : When installing downstream of ESP, see Item 2.1.
Measurement range:
Range setup : One-touch automatic setup or manual setup (Measurement
range does not directly represent dust concentration.To know the
absolute value of dust concentration, the relationship between
the output signal and the dust concentration should be obtained
by laboratory analysis or relevant way.)
Note : With one-touch automatic setup function, measurement range is set so that the dust concentra tion of the process
corresponds to a setpoint of setting level. In manual setup, the measurement range is set by gain (x1, x10, x100,
x1000, x10000).
Setup range level : Variable (10 to 50%, 10% step)
Analog output signal : 4-20 mA DC, insulated, load resistance; 500 V
Contact output signal (high/low alarm):
Contact capacity : 5 A, 240 V AC or 24 V DC
Setting range : 1 to 99% (1% step) of measurement span
Contact type: NO (closed when alarm contact is activated) or NC, 2 outputs
Note : Relay is energized when alarm contact is activated.
Damping time constant : 10 to 300 seconds, 10 second step
Relay output time constant : 10 to 180 seconds, 10 second step
Ambient conditions:
Temperature : -20 to 45ºC
Humidity : 95%RH or less (non-condensing)
Vibration : 5 m/s2 (0.5 G) or less
Construction:
Converter : Integrated with detector, protection rating IP 64
Sensor probe : Direct insertion
Materials:
Sensor probe : Stains steel (316 SS)
Insulation of sensor: PEEK
Case : Aluminum
Flange : Stainless steel (304 SS (JIS) or ASTM grade 304)
Finish : Polyester resin coated
Colors:
Case : Silver gray (Munsell 3.2PB 7.4/1.2)
Cover : Mint green (Munsell 5.6PG 3.3/2.9)
Power supply:
Rating : 100 to 120 VAC, 200 to 240 VAC
Operating voltage range : 90 to 132 VAC, 180 to 264 VAC
Power consumption : 8 VA
Wiring connection : 2 ports
• DIN with PG11 cable gland (applicable to cable with 5 to 10 mm OD) for power wiring
• DIN with PG11 cable gland (applicable to cable with 5 to 10 mm OD) for output wiring
Note: Use one cable for analog output and contact signals.

IM 11K01C01-01E
<1.Overview> 1-3
Air purge:
Connection : Rc1/4
Air source : Clean, dry air equivalent to instrument air at process pressure
+50 kPa
Air consumption : approx. 50 Nl/min
Mounting : Socket (dedicated) or flange
Weight : Approx. 2.3 kg (excluding flange)
Dimensions : 172 (W) x 173 (H) x 672 (D) mm

IM 11K01C01-01E
1-4 <1. Overview>

1.3 Model and Suffix Codes


Option
Model Suffix Code Description
Code
DT450G ------------------ -------- Dust monitor (heat resistant up to 250ºC
Analog output -5 -------- 4-20 mA DC analog output + fine adjustment for
measuring range
Power supply -2 -------- 200-240 V AC, 50/60 Hz
-5 -------- 100-120 V AC, 50/60 Hz
Probe length -05 -------- 460 mm
Sensor material -1 -------- 316 SS
Air purge -A -------- With air purge
Instruction Manual -J -------- Japanese
-E -------- English
Mounting -W -------- Socket mounting (welding socket)
-A -------- Flange mounting (ANSI Class 150 2 1/2 RF)
-K -------- Flange mounting (JIS 5K 65 FF)
-M -------- Flange mounting (JIS 10K 80 FF)
-P -------- Flange mounting (JIS 10K 100 FF)
-T -------- Flange mounting (JPI Class 150 2 1/2 RF)
— -A -------- Always -A
Option /TC Probe: Teflon coated
Note: Teflon coated probe (option code “/TC”) is recommended to prevent the insufficient insulation by dust and moisture deposited on
sensor probe.
Teflon coated probe is designed to operate at a temperature not exceeding approximately 150ºC.

l Spare Parts
Item Part Number Qty
O-ring B1026ER 1
Flange assembly:
JIS 5K 65 FF K9479DA 1
JIS 10K 80 FF K9479DB 1
JIS 10K 100 FF K9479DC 1
JPI Class 150 2 1/2 RF K9479DE 1
ANSI Class 150 2 1/2 RF K9479DD 1
Socket K9479CA 1
Bolt L9800TC 2

IM 11K01C01-01E
<1.Overview> 1-5

1.4 Dimensions
l Welding Socket Mounted Model
93 579 Unit: mm
172 472
277
250

173
Max.20

Ø10
Ø13
67
Teflon coated
127 30
Power Cable Inlet (Option : /TC)
Output Signal Cable Inlet Rc1/4
(Ø5 to Ø10) (Air Purge Piping Connection) Welding socket
(Ø5 to Ø10) Welding socket type.ai

Duct wall
Weld all around the socket;
be careful not to deform
the socket.
59.5
2-Ø9 Socket
A

(Cutout for 0 Ø38.2 cutout


90° 15 Ø34 Ø37.7-0.1
tubing)

10 30.5
A-A Section Locate the cutout for
A tubing at the bottom
Max. 15
40
(air purge tubing)
Welding Socket (accessory) of
Stainless Steel (316L SS). Welding of Socket Welding Socket.ai

Dimensions of DT450G Dust Monitor (Flange Mounted Model)

IM 11K01C01-01E
1-6 <1. Overview>

l Flange Mounted Model


Unit: mm
93 579 30
172 t 460
(Insertion Length)
265
250

146.5
173

ØD
Ø10
Max.20

Ø13
127 30 67 Teflon coated
Rc1/4 (Option : /TC)
Power Cable inlet 4-Øh
(Ø5 to Ø10) (Air Purge Piping Connection) JIS 5K 65 FF (Example)
Output Signal Cable Inlet
(Ø5 to Ø10)

Flange Dimensions Unit: mm


Class D C g h t
JIS 5K 65 FF 155 130 4 15 14
JIS 10K 80 FF 185 150 8 19 18
ØD
JIS 10K 100 FF 210 175 8 19 18
ANSI Class 150 2 1/2 RF 177.8 139.7 4 19.1 22.4
JPI Class 150 2 1/2 RF 178 139.7 4 19 22.5
Flange type.ai
ØC
g-Øh

JIS 10K 80 FF (Example)

Dimensions of DT450G Dust Monitor (Welding Socket Mounted Model)

IM 11K01C01-01E
<2. Instruction> 2-1

2. Instruction
The DT450G is designed for indoor installation or installation in an outdoor location free from
splashes of rainwater and direct sunlight.
This chapter explains how to install the DT450G in the following order of topics.
• Selection of installation location
• Machining of inlet for sensor probe
• Installation of DT450G

2.1 Selection of Installation Location


The DT450G is designed for mounting on a duct with a width of 650 to 4000 mm (when the
standard probe length is 460 mm) or other alternative locations. Install the DT450G in a location
where optimal measurement can be carried out and inspection and maintenance is easy.
When selecting an installation location, make sure it also satisfies the measuring and ambient
environmental conditions described in Section 1.2. In addition, the installation location must meet
the following requirements.

n Availability of Typical Dust Concentration


The most suitable location along a duct is where there is no flowmeter or valve that may disturb
the gas flow and where the duct is straight and the length is adequate in comparison with the
inside width of the duct. (Ideally, the straight portion should be no less than eight times the length
of the width. Locate the probe inlet in a position 3/8 the length of the straight portion from the
downstream edge and 5/8 the length from the upstream edge.)
Note: Insert the sensor probe into the inlet so that it is perpendicular to the stream of gas.
The DT450G can be positioned in such a manner that its sensor probe is level or facing downward at an angle up to 90º.

Downstream

d ( ≥ 650 mm)

3
8
L ≥ 8d is recommended

5
8

Upstream

Duct
Dust monitor
Install at the center of the duct and
perpendicular to the gas flow. F0701.ai

Figure 2.1 Installation of Dust Monitor on Straight Part of Duct

n No Noise Effects
Make sure the output signal will not be affected by noise arising from plant equipment or other
structures. In an electrostatic dust collector is used, locate the DT450G at least 20 m away from
the dust collector. When monitoring such dust collectors as a bug filter, locate the DT450G in the
upstream of the blower.

IM 11K01C01E
2-2 <2. Instruction>

n Minimal Vibration
For reasons of sensor structure, avoid installing the DT450G in a location exposed to mechanical
vibration at a level greater than 5 m/s2.
Note: It is virtually impossible for vibration to affect the measured value.

l No Splashes of Rainwater, Direct Sunlight or Abnormal Temperature Rises due


to Heat Radiation from Equipment
The DT450G may be installed outdoors. However, avoid installing the monitor in a location
exposed to direct sunlight or heat radiated from equipment, in order to prevent the temperature of
the main unit’s casing from rising above 45ºC.
Note: The DT450G can be operated within an ambient temperature range of –20ºC to 45ºC.

If the part of the duct assigned for the probe inlet is covered with insulating material that is likely
to contribute to a temperature rise in the casing, remove the insulating material around the probe
inlet in an area approximately 20 cm wide across the inlet.
Insulating material

DT450G

Insulating
materia

Probe inlet

Example of unacceptable mounting


The DT450G is affected by heat
radiation from the insulating material.

400 mm
(Area of insulating DT450G
material to be removed)

Example of acceptable mounting


The temperature rise is minimal
F0202.ai because of air ventilation.

Figure 2.2 Measures to Take When the Temperature of the DT450G’s Casing Is Likely to Rise due
to Insulating Material

IM 11K01C01-01E
<2. Instruction> 2-3

n Ease of Inspection and Maintenance


The DT450G dust monitor must be detached for cleaning the sensor probe. For safe operation
of these maintenance works, a sufficient maintenance space as shown below is required (See
Figure 2.3).

≥1m Duct

0.5 m

0.5 m

≥1m
Maintenance
space

≥1m
F0703.ai

Figure 2.3 Maintenance Space at Installation Location of DT450G Dust Monitor

IM 11K01C01E
2-4 <2. Instruction>

2.2 Machining of Probe Inlet


The DT450G is designed for mounting on a socket or flange attached to an inlet on a duct
or other alternative facility. Equip the inlet with a socket or flange that complies with the
specifications of the DT450G. At this point, make sure the DT450G’s casing and the duct are in
electrical continuity.

l When Mounting to Socket


Use the socket that comes with the DT450G for installation. Weld this socket paying attention to
the following precautions.
• The standard insertion length of the sensor probe is 460 mm.
• For welding the socket, cut out a 38.0 mm diameter hole on the duct or pipe. (The socket’s
outer diameter is 37.7 mm.)
• Weld all around the socket (316L SS). Make sure there are no spots not welded. Be careful
not to deform the socket while welding. Be careful about the orientation of the socket. When
installing the dust monitor on a vertical plane, place the cutout for air tubing at the bottom.

Duct wall Unit: mm


(socket welding plane)

Weld all around


(Be careful not to deform the socket.)

Socket (accessory)

ر38.0 ±0.1
hole

Duct Cutout for Socket


Place the cutout for air purge tubing at the bottom F070202.ai

Figure 2.4 Connection Port for Socket Mounting Model

IM 11K01C01-01E
<2. Instruction> 2-5
l When Mounting to Socket
Prepare the DT450G with a flange that meets the specified standard. Prepare the tie-in flange
at the connection port of the duct or pipe. Note that the standard sensor probe insertion length is
460 mm on the flange specification.

Flange Duct

Inner dia. 38 mm

F070201.ai

Figure 2.5 Connection Port for Flange Mounted Model

A drop of water on the insulated part of the sensor probe will prevent correct measurement.
If the process gas is cooled down around the sensor connection port when measuring high-
temperature and high-humidity gas, be sure to purge the air in order to prevent condensation
at the neck of the sensor connection flange. If there is a possibility of the ambient temperature
decreasing to 0ºC or less, provide thermal insulation at the neck of the connection flange.
(Note)
Mounting
screw
DT450G

Note: The DT450G body can be detached


by removing the mounting screw.

Cable Provide thermal


insulation here.

Duct
F0702012.ai
Tubing for air purge

Figure 2.6 Thermal Installation for Preventing Condensation

IM 11K01C01E
2-6 <2. Instruction>

2.3 Installing the DT450G


Securely fix the DT450G onto a socket or flange on the probe inlet. When fixing the DT450G onto
a socket, use the supplied two M8-size bolts. When fixing the DT450G onto a flange, use bolts
having a size that conforms to the flange specifications.
Be careful not to damage the O-ring when mounting the DT450G onto a socket.

CAUTION
If the gas to be measured is flowing constantly through a duct where the DT450G is mounted, air-
purge the monitor as instructed in subsection 4.2.2. If you leave the monitor exposed to the gas
for a prolonged period without air purging the monitor, the resistance of the insulated part of the
sensor probe will decrease. This may result in a failure of the monitor to start smoothly.

IM 11K01C01-01E
<3. Piping and Wiring> 3-1

3. Piping and Wiring


Before operating the DT450G, the pipe for air purge and wires for output signals must be
connected.
This chapter explains how to connect the pipe and wires to DT450G in the following order of
topics.
• Connecting the pipe for air purge
• Connecting the wires for the power supply
• Connecting the wires for the output signals
• Grounding of a duct or other alternative facilities where the DT450G is mounted

IM 11K01C01-01E
3-2 <3. Piping and Wiring>

3.1 Connecting the Pipe for Air Purge


As a rule, the insulated part of the sensor probe should be air-purged before the DT450G is put
into use. Air purge is mandatory if any of the following cases apply:
• The concentration of dust in a gas is extremely high and the resistance of the insulated part
of the sensor probe is likely to decrease due to dust accumulation.
• The gas temperature at the probe inlet is likely to drop, resulting in condensation.
Note that clean air, such as air specified for instruments, must be used for air purge. (If air is not
available, use gas such as nitrogen instead.)
Note: For reasons of the measurement principle, any dust accumulating on the insulated part of the sensor probe can cause
malfunction. The purpose of air purge is to prevent such dust accumulation. Care must be taken since an air purge with air
containing moisture or oil mist causes the same phenomenon as dust accumulation.

Air purge is normally carried out by continuously supplying air to the DT450G with a pressure
of approximately 50 kPa higher than that of the gas being measured. If the effect of air purge
is inadequate, due to the presence of moisture in the gas being measured, the DT450G may
require air purge at a higher pressure. For air purge, use an air source with a sufficiently large
capacity. (The normal rate of air consumption is approx. 50 Nl/min under the pressure of 50 kPa).
When connecting the pipe to the DT450G for air purge, note the following.
• The DT450G may be removed during inspection or maintenance. When connecting the pipe
to the DT450G, use a one touch connector and/or a pipe made of flexible material to ensure
easy maintenance.
• Install a stop valve or pressure reducing valve in a position as close as possible to the
DT450G.
• Connect a pipe with an inside diameter of 6 to 8 mm to the DT450G. Note that the pipe
connection of the DT450G is Rc1/4 threaded.

Note: The rate of air consumption is approximate 50 Nl/min.


In some cases, the pressure of air supply may
need to set higher than the specified.

Coupling thread
Pressure of gas under measurement
Piping material Rc1/4
plus approximately 50 kPa
with 6- to 8-mm
inside diameter
Air source

Stop valve Pressure


F0301.ai
control valve

Figure 3.1 Method of Piping for Air Purge

If air purge need not be carried out, attach a blind plug to the pipe connection.

IM 11K01C01-01E
<3. Piping and Wiring> 3-3

3.2 Connecting the Wires for the Power Supply


CAUTION
The DT450G has no power switch. For reasons of safety during inspection and maintenance,
install a double-pole switch in the power line.

CAUTION
Since the DT450G may be removed for inspection or maintenance, allow each cable to have a
sufficient margin of length.

Connect a power supply that conforms to the voltage and frequency in the specifications of the
DT450G. For wiring, use a three-core cable of a 5 to 10 mm outside diameter with the nominal
cross-section of the conductor being at least 1.5 mm2. Ground one core appropriately. Lead the
cable from the power supply through the DT450G’s cable connection fitted with a Pg11 cable
gland into the DT450G’s inner assembly. Then, connect the cable to the AC1 terminal (non-
grounded side) and AC2 terminal (grounded side).
Terminate the end of the grounding wire with an M5 crimp terminal and then connect it to the
ground terminal inside the enclosure. The ground resistance should be 100 Ω or less (JIS Class
D grounding).
Note: The DT450G main unit and the duct on which the monitor is mounted are in electrical continuity. Should the measured value be
adversely affected by the method of two point grounding at the duct (see Section 3.4) and the terminal, remove either of the two
grounding lines.

Power Connector

AC1
AC2

Ground

N L
Power Switch

Power Wiring.ai

AC power supply
Figure 3.2 Wiring diagram of the DT450G to a power supply

IM 11K01C01-01E
3-4 <3. Piping and Wiring>

3.3 Connection of Wires for the Output Signals


This section discusses the wiring necessary to transmit analog output signals and alarm contact
signals to each device.
Wire the analog output and alarm contact signals of the DT450G using a single cable.
Use a cable with an outside diameter of 5 to 10 mm, comprising three shielded twisted pair wires
with the nominal cross-section of the conductor being not more than 0.5 mm2.
If only the analog output signal is used and it is unlikely to be affected by noise (as the signal
is wired at a location distant from the power line), it is not always necessary to use twisted-pair
wires.
Insert the cable through the cable connection fitted with a Pg11 cable gland into the internal
assembly of the DT450G. Connect the core wires of the analog output signal to the mA+ (positive
side) and mA- (negative side) terminals. Also, connect the core wires of the alarm contact signal
to the c of RL1 or RL2 and nc terminals (for an open contact upon alarm output) or to the c and
no terminals (for a closed contact upon alarm output).
Ground the shielded wires of the cable on the receiver side.

CAUTION
The output signal may be mixed with electric noise if the cable swings. If the cable is likely to be
affected by wind, fin the cable.

DT450G
Not used

mA-
Output
mA+
no Power Supply connector
RL2  
c (Relay contact) AC1
nc
AC2
no
c RL1  
(Relay contact)
nc

(*2)
接地

SHIELD

(Not used)
Junction Box
N L
Power Switch

Alarm contact signal

AC power supply

Receiver

Analog Output Signal(*3)

DT450G Wiring.ai
Grounding
(*1) The nomonal cross-sevtion of each conductor should be 1.5 mm2
(*2) The nomonal cross-sevtion of each conductor should be not more than 0.5 mm2
(*3) This terminal is used for the grouding on the receiver side.
Notes: The above wiring diagram shows the one when using no contacts. Connect wires to terminals nc and c
when using nc contacts.

Figure 3.3 Wiring Diagram

IM 11K01C01-01E
<3. Piping and Wiring> 3-5

3.4 Grounding the Duct


CAUTION
When such equipment as an electrostatic dust collector is used, be sure to ground both the duct
and the DT450G. Otherwise, there may be a danger of electrical shock due to high potential
static electricity.

If a duct or other alternative facility where the DT450G is mounted is charged with electricity, the
electricity will serve as noise, possibly affecting the signal under measurement. Ground the duct
where the DT450G is mounted, as well as other nearby structures, in order to prevent the duct
from being charged with electricity.

IM 11K01C01-01E
Blank Page
<4. Operation> 4-1

4. Operation
The ranges and other parameters of the DT450G must be configured before you start operation.
This chapter explains how to operate the DT450G, including preparations, in the following order
of topics.
• Components and their functions
• Preparation for operation
• Regular operation

4.1 Components and Their Functions

Auto Setup Button

Not used Display lamp Green: Operation


Red : Autotuning
LED Display
mA-
Output
mA+
no Power Supply Connector
RL2
c (Relay contact) AC1
nc
AC2
[A]Key [B]Key [C]Key
no
RL1 Parameter Setting Button
c (Relay contact)
nc

SHIELD

Not used Name&Func.ai

Figure 4.1 Components of the DT450G and Their Functions

IM 11K01C01-01E
4-2 <4. Operation>

4.2 Preparation for Operation


Before you start operating the DT450G, carry out the following tasks.
• Inspection of the state of installation
• Supply of purge air
• Supply of power
• Parameter (Setting of measurement range. Setting of alarm points, Setting of damping
action)

4.2.1 Inspection of the State of Installation


Check that all the specifications are completely satisfied with the installation of the DT450G.
Also check that the DT450G wiring and piping have been correctly connected. Especially, make
sure the following are true.
• The DT450G is securely mounted onto the probe inlet and is free from effects of strong
mechanical vibration. In addition, the short tubing of the probe inlet is adequately heat-
insulated and the main unit’s casing is not exposed to direct sunlight that causes an
abnormal temperature rise or subject to heat radiated from other equipment.
• The duct or other alternative facilities where the DT450G is mounted is reliably grounded.
• Wires are correctly connected to the terminals of the DT450G. In addition, the cables are
securely fixed by cable glands so as to prevent any corrosive gas or moisture from entering
the monitor.
• The air-purge piping and each cable connected to the DT450G have a sufficient margin of
length necessary for maintenance.

4.2.2 Supply of Purge Air


Using the pressure control valve, adjust the supply pressure of purge air as noted below.
Then, fully open the stop valve.
• Supply pressure = Pressure of gas under measurement + approximately 50 kPa

4.2.3 Supply of Power


Turn on the power switch installed in the power line to feed power to the DT450G.
When the power is turned on, the green LED indicator lamp on the monitor comes on.
It takes approximately 30 minutes for the DT450G to reach steady-state operation.

IM 11K01C01-01E
<4. Operation> 4-3

4.2.4 Setting of Parameters


The dust monitor has a 4-20 mA output and two relay contact alarm outputs. The effect of sudden
change of measured value can be removed by setting a user configurable time constant.
The output current of the 4-20 mA signal can be trimmed.
The relay output (s) change their state when the measured dust concentration exceeds a user
selected alarm trigger level. The relay output (s) are also protected against the effect of sudden
change of input signal, with a user configurable time constant, common for both relays.
All the user selectable parameters have factory default values, which in many cases will suffice.
If your application requires changing parameters, proceed as follows.
How to Change Parameters
Open the cover of the dust monitor. You see a front panel, a display unit and three parameter
buttons (A, B and C).
It is needed to enter PIN code before setting parameters.
1. Press (A) key and select first PIN code digit 0 to 9.
2. Press (B) key and select second PIN code digit 0 to 9.
3. Press (C) key and select third PIN code digit 0 to 9.
Factory default for PIN code is 1.1.1.
Parameters can be changed after the right PIN code is selected and simutaneously press and
hold down (B) and (C) key for one second. (Parameter setting mode)
Press the leftmost (A) key of the three keys below the display, until the leftmost digit indicates
the number of the parameter you wish to change (refer to Table 4-1). The two other display digits
now show the parameter value. Pressing the middle (B) key and rightmost (C) key, you may now
change the parameter value. See instructions below for each individual parameter.
When you have set all the parameters, press the leftmost (A) key again until the display shows
“-SP.” Press one of the other keys (B or C) to save all the parameters.
NOTE: All parameters must be as you wish to have them, when you save them. If you want to make more changes, or decide not to
save the parameter you have changed, press the left most (A) key again.

Table 4.1 Parameter Change Table


Display Left Factory
Parameter Range Key Effect Key B Key Effect Key C
Number Default
Alarm contact (relay 1) 1 1 to 99% Increase number by 10% Increase number by 1% 5%
Alarm contact (relay 2) 2 1 to 99% Increase number by 10% Increase number by 1% 60%
4 mA trim 3 Increase number Decrease number Set to 4 mA
20 mA trim 4 Increase number Decrease number Set to 20
mA
Alarm contact 10 to Increase number by 10 s Decrease number 10 s
5
time constant 180 s
Analog output signal 10 to Increase number by 10 sDecrease number by 10 s
6
time constant 300 s 10 s
Percent level of full 10 to Increase number by 10% Decrease number by 20%
7
range 50% 10 %
8
Not available
9
Fine adjustment for 1 to
0. Increase number Decrease number 20
measuring range 100%
10.*
Not available
Pin code 1.
Set measuring range 1.0* 0 to 5 Increase number Decrease number 0
Save parameters -SP* Save all Save all −
*1.0, 10., in 2 digits -sp in 3 digits

Do not set parameters except “0” for the leftmost digit “8”, “9” and “10.”.

IM 11K01C01-01E
4-4 <4. Operation>

n Setting Alarm Contacts


Press the leftmost (A) key until the leftmost digit shows “1” or “2”. Set the alarm contact of relay
1 or 2 in percent of the output full scale value. Each press of the middle (B) or the rightmost (C)
key only increases the number, which will roll back to 1% after it reaches 99%. The percentage is
displayed by the two rightmost display digits.

n Trimming Analog Output Signals (4 mA and 20 mA)


Press the leftmost (A) key until the leftmost digit shows “3” or “4”. Increase or decrease the
output current. The display shows the nominal mA value with its two rightmost digits. Connect
a calibrated mA meter to the analog output terminal to check the actual current value. Allow 30
minutes warm up time before trimming the mA output.

n Setting Time Constants


Press the leftmost (A) key until the leftmost digit shows “5” or “6”. Change the time constant of the
alarm contact or the analog output signal in increments of 10 seconds.
The alarm contact time constant will change back to 0 s from 180 s, the analog output signal
time constant from 300 s. The two leftmost digits of the value are displayed by the two rightmost
display digits.

n Setting of the Measuring Range


The measuring range can be set manually or by Auto Setup button.

l Manual Setting
In manual setting the measuring range is set by amplifying the signal detected by the sensor in 5
stages.
Press the leftmost (A) key until the leftmost digit shows “1.0”, and press the middle (B) or the
rightmost (C) key until the rightmost digit shows the number of the desired parameter. Table 4-2
shows that the rightmost digit number and its parameter.
Table 4-2 Measuring Range Setting
Display
Parameter Description
Right Number
0 Automatic range setup
1 Maximum range (For high concentration of dust)
2 Gain x 10
3 Gain x 100
4 Gain x 1000
5 Gain x 10000 (For low concentration of dust)
When parameter 1 is selected, the measuring range is maximized. If the dust concentration is
high, select this parameter. The output sensitivity becomes higher as the parameter number
increases from “1” to “5”. Chose the most appropriate parameter for your process.

l Auto Setup
In the process where the dust concentration is a certain level and stable during normal operation,
range setting by Auto Setup button is also available.
Before the Auto Setup, the setting level should be confirmed.
The dust monitor has an Automatic Setup function. You can change the default percent of full
range (20%) in steps of 10% from 10 to 50%. Press the leftmost (A) key until the leftmost digit
shows “7” and increase or decrease the value using the middle (B) or the rightmost (C) key.

IM 11K01C01-01E
<4. Operation> 4-5
The set point of 20% means that the 4-20 mA output at the typical dust concentration will be 4 mA
+ (20 mA - 4 mA)  20% which makes 7.2 mA. If the dust concentration is five times the typical
value or higher, the output will be 20 mA, and if the dust concentration is zero, the output will be 4
mA.
Next, the measuring range should be set. Press the leftmost (A) key until the leftmost digit shows
“1.0”. Press the middle (B) key or the rightmost (C) key to increase or decrease the value until the
rightmost digit shows “0” (see Table 4-1). Last, press the leftmost (A) key until the display shows
“-SP.” Press the middle (B) or the rightmost (C) key to save the setting.
For Automatic Setup, make sure that the process is running with a normal dust concentration.
Then open the cover of the dust monitor. After move the parameters setting mode, initiate
Automatic Setup by pressing the small key button near the indicator lamp on the lower printed
circuit board (see Figure 4.1). Note: The indicator lamp should be green before you press the key.
Note: Initiate Automatic Setup in parameters setting mode.
Collecting data of the dust concentration will take an hour or less. The three digit display counts
down towards zero, and the indicator lamp is red as long as the data are being collected. To avoid
interference from the background electrical noise, you should close the cover of the enclosure
right after having initiated the Automatic Setup. After an hour, the setup is ready and the dust
monitor is in normal operation. The indicator turns back to green when setup is ready.
Note: The setting level is valid only when the automatic range setup (parameter “0” ) is used.

l Fine adjustment for measuring range


Fine adjustment adjusts measuring range from 1-100%. Press the leftmost (A) key until the
leftmost digit shows “0.”, set the value from 1-100% .
• In case of setting 10%, the maximum range is 2 times higher than current range.
• In case of setting 50%, the maximum range is 0.4 times higher than current range.
• In case of setting 100%, the maximum range is 0.2 times higher than current range.
The following is relationship between the setting value and multiplication for measuring range.
20
(Multiplicationnnn for measuring range) =
(Setting value)

• Relay connection and parameter setting mode


(1) When both relays connected to NO
Relay 1 contact : Open Relay 2 contact : Close
(2) When relay 1 connected to NO and relay 2 connected to NC
Relay 1 contact : Open Relay 2 contact : Open
(3) When relay 1 connected to NC and relay 2 connected to NO
Relay 1 contact : Close Relay 2 contact : Close
(4) When both relays connected to NC
Relay 1 contact : Close Relay 2 contact : Open

IM 11K01C01-01E
4-6 <4. Operation>

4.3 Steady-state Operation


Proceed to steady-state operation after making sure there is no problem with the DT450G during
start-up.
Before going into steady-state operation, make sure the cover is securely attached to the
DT450G main unit. Moisture or corrosive gas entering into the DT450G may result in malfunction.
This section discusses the following topics concerning the handling of the DT450G during
steady-state operation.
• Examination of operating condition (carried out only at the beginning of operation)
• Shutdown
• Restart

4.3.1 Examining the Operating Condition


At the beginning of operation, examine the following points to ensure that the DT450G continues
to operate in its optimum condition.

 Effects of External Noise, Etc.


If the DT450G is being affected by noise to such a degree that normal operation becomes
problematic, first make sure the DT450G is installed and wired correctly as instructed by this
manual. Then, take appropriate noise countermeasures.

 State of Dust Accumulation


Accumulation of a large amount of dust on the sensor probe of the DT450G may adversely
affect its normal operation. If the measured value is found to be abnormal, remove the DT450G
and examine the state of dust contamination of the sensor probe. If the cause of the abnormal
measured value is dust accumulation, clean the sensor probe periodically. If you are using the
DT450G in a high-temperature or moist gas atmosphere, also examine the effectiveness of
air purge. If the amount of dust accumulating on the outlet of purge air is found to be gradually
increasing, increase the supply pressure of purge air. If condensation is observed around the
probe inlet, it is advisable that heat insulation measures be taken.

4.3.2 Shutdown
To shut down the DT450G, turn off the power switch installed in the power line. If the DT450G will
not be used for a prolonged period, only turn off the power switch without turning off the air-purge
line.
The air-purge line need not be left turned on if there is no flow of gas to be measured (provided
that the ambient humidity is not too high).

4.3.3 Restart
There is nothing in particular that needs to be done before restarting the DT450G. If the DT450G
has not been in use for a prolonged period however, make sure there are no abnormalities with
the sensor probe before you restart the DT450G. When there are no abnormalities with the
DT450G, it will come into steady-state operation in approximately 30 minutes after power-on.

IM 11K01C01-01E
<5. Inspection and Maintenance> 5-1

5. Inspection and Maintenance


WARNING
If the DT450G is installed in the downstream of an electrostatic dust collector, there is a danger
of serious injury or death from electric shock due to high-potential static electricity. Carry out
maintenance without disconnecting the grounding line that was connected when the DT450G
was installed. If the DT450G is not yet grounded with a grounding line, ground it before you start
maintenance.

This chapter explains inspection and maintenance carried out to repair or replace worn out or
deteriorated parts and thereby maintain the DT450G in its optimal operating condition.

CAUTION
When inspecting and maintaining the DT450G, remove it from where it is installed.
Be careful not to inflict severe impact to the monitor during inspection and maintenance,
otherwise it may be damaged.

IM 11K01C01-01E
5-2 <5. Inspection and Maintenance>

5.1 Periodic Maintenance


If the sensor probe is contaminated with a large amount of dust or significantly worn out by
dust, the output signal level for the same dust concentration may differ from the level when the
DT450G was initially put into operation. To be able to revert to the initial output signal level, the
sensor probe must be cleaned or the measurement range must be reset. Do not disassemble
such components as the printed circuit board from the internal unit of the DT450G; otherwise, the
DT450G may fail to operate correctly.

5.1.1 Cleaning the Sensor Probe


If the sensor probe tends to gather a significant amount of dust, it is advisable that the probe be
cleaned periodically. If dust accumulation is not very serious, its effect on the output signal is in
most cases negligible; therefore, it is not necessary to clean the sensor probe frequently.
Remove dust on the sensor probe with a dry cloth. For correct and normal measurement, keep
the insulation resistance of the purge-air outlet at a high level. To ensure this, completely remove
contaminants from the outlet. When cleaning the sensor probe, also remove dust that has
accumulated on the probe inlet.

5.1.2 Correcting (Resetting) the Measurement Range


The sensor probe may be worn out by dust contained in the gas under measurement.
Although wear in the sensor probe may not directly lead to the inability to measure, the level of
current arising from frictional electricity may change, affecting the output signal. If you need to
know precise dust concentration, correct (or reset) the measurement range according to your
degree of necessity. Correct (reset) the measurement range according to the procedure of
measurement range setting discussed in subsection 4.2.4. If the wear on the sensor probe is
extreme, it must be replaced.

5.2 Replacing Parts


In order to keep the DT450G serviceable over a prolonged period, you must replace parts that
are worn out or have deteriorated as a result of the gas under measurement.

n Replacing the O-rings


The O-ring that needs replacement is the one that prevents leakage of the gas under
measurement from the probe inlet. If the O-ring is found to be damaged in any way, such as
scratches, or cracks or permanent deformity due to deterioration, replace it.
Note: O-rings are also used in the DT450G main unit and on the contact surfaces of the sensor probe, though they are not visible from
outside. Ask Yokogawa Electric Corporation for replacement of these O-rings.

IM 11K01C01-01E
<6. Troubleshooting> 6-1

6. Troubleshooting
This chapter explains measures that should be taken when the output signal gives abnormal
readings. Note that not all anomalies are due to failure (disorder or malfunction) in the DT450G.
Some of them may be due to a change in the physical properties types of components, the flow
velocity, etc.) of the gas being measured. If such a change is likely, first check for the following.
Table 6.1 Causes of Failure and Corrective Actions
State of Failure Cause Corrective Action
The output signal • Air purge is not effective. • Check that the gap at the purge-air exit is
overshoots as specified. If the exit is contaminated with
toward the 20 mA dust, increase the purge pressure (remove
side. the dust on the exit).
• Condensation occurs at the probe inlet. • Take appropriate heat insulation measures.
The output signal • The purge-air exit is contaminated with • Clean the exit (see subsection 5.1.1).
gives abnormal dust.
readings. • The sensor probe is in a state of insulation • Use normal, dry purge air.
failure because of inappropriate purge air
(e.g., contains moisture).
• The gas under measurement has entered • Set the pressure relatively higher and
the purge-air piping, causing an insulation continue air purge for a while.
failure in the sensor probe. (If contaminants still remain in the piping,
contact Yokogawa Electric for assistance.)
The output • The flow velocity of the gas under • Change the measurement point to a
signal does not measurement is not constant. position in the straight portion of the duct
stabilize. where the flow of gas is regular.
• The humidity of the gas under • If the change occurs constantly, check if
measurement changes. measuring the gas is practicable at all.
The output • The wiring is inappropriate and therefore • Check the states of grounding and wiring
signal becomes picks up noise. and correct a problem if there is any.
disturbed • There is equipment nearby serving as a • Take measures for protecting the DT450G
momentarily. source of noise. from the source of noise.
There is no • The fuse is blown or there is something • Turn off the power switch temporarily if the
output signal. wrong with the wiring connections. green indicator lamp is unlit.
Then, examine the wiring connections and
check if the fuse has blown.
Consult Yokogawa if the fuse has blown.
• There is something wrong with the electric • Request repair to be carried out by
circuit components. Yokogawa.

Table 6.2 Corrective Action against Errors at Automatic Setup


Display Description Corrective action
-E1 Input signal is too low Try Automatic Setup again
Manually set measurement range to Range 5
-E2 Input signal is too High Try Automatic Setup again
Manually set measurement range to Range 1

IM 11K01C01-01E
Blank Page
Customer Model DT450
Maintenance Dust Monitor
Parts List

3 2

Item Part Number Qty Description


1 B1026ER 1 O-ring
2 Flange assembly:
K9479DA 1 JIS 5K 65 FF
K9479DB 1 JIS 10K 80 FF
K9479DC 1 JIS 10K 100 FF
K9479DE 1 JPI Class 150-2 1/2 RF
K9479DD 1 ANSI Class 150-2 1/2 RF
3 K9479CA 1 Socket
4 L9800TC 2 Bolt

DT450G for CMPL.eps

All Rights Reserved, Copyright © 2005, Yokogawa Electric Corporation. CMPL 11K01C01-01E
Subject to change without notice.
2nd Edition : Feb. 2012 (YG)
Yokogawa Electric Corporation
i

Revision Information
 Title: Model DT450G Dust Monitor
 Manual No.: IM 11K01C01-01E

Edition Date Remark (s)


6th Jul. 2018
Revised and Corrected overall
Modified specification

5th Jun. 2012


Revised and Corrected over all
1.3.2 “Spare Parts”: Deleted the “Note”.

4th Apr. 2012


1.2 “Specification” Seiup level: Deleted of the description;
1.3.1 “Dust Monitor” Table: Deleted of the suffix code (-1, -3);
1.3.2 “Spaer parts”: Deleted of the parts number;
3.3 “Connection of Wires for the Output Signals” Figure 3.3: Change of the description with *
mark;
4.2.4 “Setting of parameters”: Deleted of the description, and the table 4-1 and 4-2,
“Setting of parameters” (2) Auto Setup: Deleted of the description;
CMPL “CMPL 11K01C01-01E”: Revision to 2nd edition.

3rd Jul. 2009


1.2 “Specifications”: Added item of DT450G-5;
1.3.1 “Dust Monitor” table: Added suffix code for Analog output in “-5” ;
3.3 “Conection of Wires for the Output Signals”Figure 3.3: Added item DT450G-5 in (*1);
4.2.4 “Setting of Parameters”: Added item of DT450G-5,
“Setting of Parameters”: Added table 4-3,
“Setting of Parameters”: Added item (3);

2nd Aug. 2007


2. Deleted some descriptions in “No Noise Effects.” ;
2.2 “Machining of Probe Inlet”: Deleted some descriptions in “When Mounting to Socket.” ;
3.2 “Connecting the Wires for the Power Supply”: Deleted some descriptions;
3.3 “Connection of Wires for the Output Signals”: Deleted “Note.” ;
3.4 “Grounding the Duct”: Deleted Figure 3.4 and some descriptions;
4.4 “Setting Time Constants,” “Automatic Range Setup” and “Manual Setting of the Measuring
Range”: Fully revised;
7.9.2 “Checking Calibration Contact Output”: Made corrections in Table 7.11;
8.3 “Output Hold Setting,” “Table 8.4 Analog Output Hold Setting”: Added Note;
8.3.2 “Preference Order of Output Hold Value”: Deleted “or blow-back”;
8.3.3 “Output Hold Setting”: Made corrections and changed descriptions in Table 8.5, added note;
8.3.4 “Default Values”: Changed descriptions in Table 8.6;
8.5.1 “Output Contact”: Made corrections;
8.5.2 “Setting Output Contact”: Table 8.10, Changed descriptions,
WARNING: Deleted second warning;
8.5.3 Changed layout;
8.7.4 “Setting Purging”: Added item;
10.4 “Reset,” “Table Output-related Items in Group C”: Added Note;
10.6 Output-related Items in Group C. Deleted some codes;
Contact-related Items in Group E. Deleted some codes;
12.2.1 “What is an Alarm?”: Table 12.2, Added Alarms 11 and 13;
12.2.2 Alarm 6: Changed descriptions, Alarm 7: Changed descriptions,
Added Alarm11, Alarm 13;
CMPL CMPL11M13A01-04E Changed part numbers.

1st Sep. 2005 Newly published

IM 11M13A01-04E
Blank Page

n If you want have more information about Yokogawa products, you can visit
Yokogawa’s home page at the following website.
Home page: http://www.yokogawa.com/an/
n Written by Environmental & Analytical Products PMK Dept.
IA Div., Product Business Center
Yokogawa Electric Corporation
n Published by Yokogawa Electric Corporation
2-9-32 Nakacho, Musashino-shi, Tokyo 180-8750, JAPAN
n Printed by KOHOKU PUBLISHING & PRINTING INC.

You might also like