Model DT450G Dust Monitor: User's Manual
Model DT450G Dust Monitor: User's Manual
Model DT450G Dust Monitor: User's Manual
IM 11K01C01-01E
IM 11K01C01-01E
5th Edition
<Introduction> i
Introduction
The DT450G Dust Monitor is a process instrument designed to detect solid particles ingas and
integrates a control unit and a sensor probe into a compact enclosure. The DT450G can be
used in a wide variety of applications, including continuous monitoringof the dust content of the
exhaust gas.
This manual describes the procedures for installing, operating, inspecting and maintainingthe
DT450G Dust Monitor. Before using the instruments, thoroughly read themanual to ensure
optimum performance.
This manual consists of six chapters. For your reference, the following table shows theoutlines
of each chapter and indicates which chapters should be read for installation,operation or
maintenance work.
Table of Contents
Relates to
Chapter Outline
Installation Operation Maintenance
1. Overview Standard specifications, Model and Suffix codes
B C B
(or part number), and dimensions.
2. Installation Installation procedures, etc. A C
3. Piping/Wiring Piping and wiring procedures. A C
4. Operation Operation and display in detail. A C
5. Inspection and Inspection/Maintenance
B A
Maintenance
6. Troubleshooting Failures, possible causes, and remedies C A
CMPL (parts list) User replaceable parts list C B
CMPL: Customer Maintenance Parts List
Media No.IM 11K01C01-01E (YK) 5th Edition :July 2018 (YK) IM 11K01C01-01E
All Rights Reserved Copyright © 2005, Yokogawa Electric Corporation
ii <Introduction>
CAUTION
Connect the power supply cord only after confirming that the supply voltage matches the rating of
this equipment. In addition, confirm that the power is switched off when connecting power supply.
Some process gas is dangerous to people. When removing this equipment from the process line
for maintenance or other reasons, protect yourself from potential poisoning by using a protective
mask or ventilating the area well.
l The following safety symbols are used on the product as well as in this
manual.
WARNING
This symbol indicates that an operator must follow the instructions laid out in this manual
in order to avoid the risks, for the human body, of injury, electric shock, or fatalities. The
manual describes what special care the operator must take to avoid such risks.
CAUTION
This symbol indicates that the operator must refer to the instructions in this manual in order
to prevent the instrument (hardware) or software from being damaged, or a system failure
from occurring.
CAUTION
This symbol gives information essential for understanding the operations and functions.
IM 11K01C01-01E
<Introduction> iii
NOTE
This symbol indicates information that complements the present topic.
This symbol indicates Function Ground Terminal. Do not use this terminal as the protective
ground terminal.
IM 11K01C01-01E
iv <Introduction>
NOTICE
l Specification check
When the instrument arrives, carefully unpack and check that the instrument has not been
damaged during transportation. Please make sure that the product received meet the purchase
order specifications and all the accessories are present. Specifications can be checked by the
model codes indicated on the nameplate. For details of the model codes, refer to Chapter 1,
“Overview.”
Notice:
The symbol (see above) means they shall be sorted out and collected as ordained in ANNEX II in
DIRECTIVE 2006/66/EC
IM 11K01C01-01E
<Introduction> v
After-Sales Warranty
Do not modify the product.
During the warranty period, for repair under warranty carry or send the product to the local
sales representative or service office. Yokogawa will replace or repair any damaged
parts and return the product to you.
Before returning a product for repair under warranty, provide us with the model name
and serial number and a description of the problem. Any diagrams or data explaining the
problem would also be appreciated.
If we replace the product with a new one, we won’t provide you with a repair report.
Yokogawa warrants the product for the period stated in the pre-purchase quotation.
Yokogawa shall conduct defined warranty service based on its standard. When the
customer site is located outside of the service area, a fee for dispatching the maintenance
engineer will be charged to the customer.
In the following cases, customer will be charged repair fee regardless of warranty period.
• Failure of components which are out of scope of warranty stated in instruction manual.
• Failure caused by usage of software, hardware or auxiliary equipment, which Yokogawa
Electric did not supply.
• Failure due to improper or insufficient maintenance by user.
• Failure due to modification, misuse or outside-of-specifications operation which Yokogawa
does not authorize.
• Failure due to power supply (voltage, frequency) being outside specifications or abnormal.
• Failure caused by any usage out of scope of recommended usage.
• Any damage from fire, earthquake, storms and floods, lightning, disturbances, riots,
warfare, radiation and other natural changes.
Yokogawa does not warrant conformance with the specific application at the user site.
Yokogawa will not bear direct/indirect responsibility for damage due to a specific application.
Yokogawa Electric will not bear responsibility when the user configures the product into
systems or resells the product.
Maintenance service and supplying repair parts will be covered for five years after the
production ends. For repair for this product, please contact the nearest sales office
described in this instruction manual.
IM 11K01C01-01E
Blank Page
<CONTENTS> i
Model DT450G
Dust Monitor
IM 11K01C01-01E 5th Edition
CONTENTS
Introduction...............................................................................................................i
For the safe use of this equipment..................................................................... ii
NOTICE.................................................................................................................. iv
After-Sales Warranty............................................................................................ v
1. Overview..................................................................................................... 1-1
1.1 Product Overview.............................................................................................. 1-1
1.2 Specifications..................................................................................................... 1-2
1.3 Model and Suffix Codes.................................................................................... 1-4
1.4 Dimensions......................................................................................................... 1-5
2. Instruction.................................................................................................. 2-1
2.1 Selection of Installation Location.................................................................... 2-1
2.2 Machining of Probe Inlet................................................................................... 2-4
2.3 Installing the DT450G........................................................................................ 2-6
3. Piping and Wiring...................................................................................... 3-1
3.1 Connecting the Pipe for Air Purge................................................................... 3-2
3.2 Connecting the Wires for the Power Supply................................................... 3-3
3.3 Connection of Wires for the Output Signals................................................... 3-4
3.4 Grounding the Duct........................................................................................... 3-5
4. Operation.................................................................................................... 4-1
4.1 Components and Their Functions................................................................... 4-1
4.2 Preparation for Operation................................................................................. 4-2
4.2.1 Inspection of the State of Installation.................................................. 4-2
4.2.2 Supply of Purge Air............................................................................. 4-2
4.2.3 Supply of Power.................................................................................. 4-2
4.2.4 Setting of Parameters......................................................................... 4-3
4.3 Steady-state Operation..................................................................................... 4-6
4.3.1 Examining the Operating Condition.................................................... 4-6
4.3.2 Shutdown............................................................................................ 4-6
4.3.3 Restart................................................................................................. 4-6
5. Inspection and Maintenance.................................................................... 5-1
5.1 Periodic Maintenance........................................................................................ 5-2
5.1.1 Cleaning the Sensor Probe................................................................. 5-2
5.1.2 Correcting (Resetting) the Measurement Range............................... 5-2
IM 11K01C01-01E
ii <CONTENTS>
IM 11K01C01-01E
<1.Overview> 1-1
1. Overview
This chapter discusses the following topics.
· Product overview
· Specifications
· Model and suffix code
· Dimensions
l Measuring Principle
When two solid particles come into contact with each other, charge transfer takes place between
them. This charge transfer is known as the principle of triboelectrification or contact electrification
and the DT450G utilizes this principle. More specifically, charge transfer occurs when particles in
a gas stream collide with (or pass nearby) the sensor probe, producing an electric current. The
DT450G amplifies this current to convert it into its output current. The electric current produced
by frictional electricity is subject to the effects of the following factors:
• Amount of the particles' electrostatic charge and a change in the polarity
• Particle size
• Velocity of particles
• Components of particles
The behaviors of these factors vary widely depending on the object of measurement and their
effects are in no way uniform. In the case of fixed-position measurement in a particular plant
however, measurement conditions, such as the components of particles, are almost constant
and therefore the effects are usually marginal. Dust concentration is also constant in the case of
gases emitted from a bag filter, cyclone or electrostatic dust collector.
A current-mode input signal obtained at the detector block is transferred to the converter block,
where the signal is converted to a 4-20 mA DC analog output signal. The converter block outputs
an alarm contact signal at a desired level, in addition to the analog output signal.
IM 11K01C01-01E
1-2 <1. Overview>
1.2 Specifications
Measurement objects: Solid particles in gas
Particle size : 0.3 µm or greater
Measurement range: 0.1 mg/m3 to 1 kg/m3
Measurement principle: Inductive Electrification
Process conditions :
Temperature : 250ºC or less
Pressure : 200 kPa or less
Gas flow rate : 4 m/s or greater (approx. 30 m/s max)
Humidity : 40%RH or less
Input surge voltage : 100 V or less (if 500 V or greater, converter may be damaged.)
Note : When installing downstream of ESP, see Item 2.1.
Measurement range:
Range setup : One-touch automatic setup or manual setup (Measurement
range does not directly represent dust concentration.To know the
absolute value of dust concentration, the relationship between
the output signal and the dust concentration should be obtained
by laboratory analysis or relevant way.)
Note : With one-touch automatic setup function, measurement range is set so that the dust concentra tion of the process
corresponds to a setpoint of setting level. In manual setup, the measurement range is set by gain (x1, x10, x100,
x1000, x10000).
Setup range level : Variable (10 to 50%, 10% step)
Analog output signal : 4-20 mA DC, insulated, load resistance; 500 V
Contact output signal (high/low alarm):
Contact capacity : 5 A, 240 V AC or 24 V DC
Setting range : 1 to 99% (1% step) of measurement span
Contact type: NO (closed when alarm contact is activated) or NC, 2 outputs
Note : Relay is energized when alarm contact is activated.
Damping time constant : 10 to 300 seconds, 10 second step
Relay output time constant : 10 to 180 seconds, 10 second step
Ambient conditions:
Temperature : -20 to 45ºC
Humidity : 95%RH or less (non-condensing)
Vibration : 5 m/s2 (0.5 G) or less
Construction:
Converter : Integrated with detector, protection rating IP 64
Sensor probe : Direct insertion
Materials:
Sensor probe : Stains steel (316 SS)
Insulation of sensor: PEEK
Case : Aluminum
Flange : Stainless steel (304 SS (JIS) or ASTM grade 304)
Finish : Polyester resin coated
Colors:
Case : Silver gray (Munsell 3.2PB 7.4/1.2)
Cover : Mint green (Munsell 5.6PG 3.3/2.9)
Power supply:
Rating : 100 to 120 VAC, 200 to 240 VAC
Operating voltage range : 90 to 132 VAC, 180 to 264 VAC
Power consumption : 8 VA
Wiring connection : 2 ports
• DIN with PG11 cable gland (applicable to cable with 5 to 10 mm OD) for power wiring
• DIN with PG11 cable gland (applicable to cable with 5 to 10 mm OD) for output wiring
Note: Use one cable for analog output and contact signals.
IM 11K01C01-01E
<1.Overview> 1-3
Air purge:
Connection : Rc1/4
Air source : Clean, dry air equivalent to instrument air at process pressure
+50 kPa
Air consumption : approx. 50 Nl/min
Mounting : Socket (dedicated) or flange
Weight : Approx. 2.3 kg (excluding flange)
Dimensions : 172 (W) x 173 (H) x 672 (D) mm
IM 11K01C01-01E
1-4 <1. Overview>
l Spare Parts
Item Part Number Qty
O-ring B1026ER 1
Flange assembly:
JIS 5K 65 FF K9479DA 1
JIS 10K 80 FF K9479DB 1
JIS 10K 100 FF K9479DC 1
JPI Class 150 2 1/2 RF K9479DE 1
ANSI Class 150 2 1/2 RF K9479DD 1
Socket K9479CA 1
Bolt L9800TC 2
IM 11K01C01-01E
<1.Overview> 1-5
1.4 Dimensions
l Welding Socket Mounted Model
93 579 Unit: mm
172 472
277
250
173
Max.20
Ø10
Ø13
67
Teflon coated
127 30
Power Cable Inlet (Option : /TC)
Output Signal Cable Inlet Rc1/4
(Ø5 to Ø10) (Air Purge Piping Connection) Welding socket
(Ø5 to Ø10) Welding socket type.ai
Duct wall
Weld all around the socket;
be careful not to deform
the socket.
59.5
2-Ø9 Socket
A
10 30.5
A-A Section Locate the cutout for
A tubing at the bottom
Max. 15
40
(air purge tubing)
Welding Socket (accessory) of
Stainless Steel (316L SS). Welding of Socket Welding Socket.ai
IM 11K01C01-01E
1-6 <1. Overview>
146.5
173
ØD
Ø10
Max.20
Ø13
127 30 67 Teflon coated
Rc1/4 (Option : /TC)
Power Cable inlet 4-Øh
(Ø5 to Ø10) (Air Purge Piping Connection) JIS 5K 65 FF (Example)
Output Signal Cable Inlet
(Ø5 to Ø10)
IM 11K01C01-01E
<2. Instruction> 2-1
2. Instruction
The DT450G is designed for indoor installation or installation in an outdoor location free from
splashes of rainwater and direct sunlight.
This chapter explains how to install the DT450G in the following order of topics.
• Selection of installation location
• Machining of inlet for sensor probe
• Installation of DT450G
Downstream
d ( ≥ 650 mm)
3
8
L ≥ 8d is recommended
5
8
Upstream
Duct
Dust monitor
Install at the center of the duct and
perpendicular to the gas flow. F0701.ai
n No Noise Effects
Make sure the output signal will not be affected by noise arising from plant equipment or other
structures. In an electrostatic dust collector is used, locate the DT450G at least 20 m away from
the dust collector. When monitoring such dust collectors as a bug filter, locate the DT450G in the
upstream of the blower.
IM 11K01C01E
2-2 <2. Instruction>
n Minimal Vibration
For reasons of sensor structure, avoid installing the DT450G in a location exposed to mechanical
vibration at a level greater than 5 m/s2.
Note: It is virtually impossible for vibration to affect the measured value.
If the part of the duct assigned for the probe inlet is covered with insulating material that is likely
to contribute to a temperature rise in the casing, remove the insulating material around the probe
inlet in an area approximately 20 cm wide across the inlet.
Insulating material
DT450G
Insulating
materia
Probe inlet
400 mm
(Area of insulating DT450G
material to be removed)
Figure 2.2 Measures to Take When the Temperature of the DT450G’s Casing Is Likely to Rise due
to Insulating Material
IM 11K01C01-01E
<2. Instruction> 2-3
≥1m Duct
0.5 m
0.5 m
≥1m
Maintenance
space
≥1m
F0703.ai
IM 11K01C01E
2-4 <2. Instruction>
Socket (accessory)
ر38.0 ±0.1
hole
IM 11K01C01-01E
<2. Instruction> 2-5
l When Mounting to Socket
Prepare the DT450G with a flange that meets the specified standard. Prepare the tie-in flange
at the connection port of the duct or pipe. Note that the standard sensor probe insertion length is
460 mm on the flange specification.
Flange Duct
Inner dia. 38 mm
F070201.ai
A drop of water on the insulated part of the sensor probe will prevent correct measurement.
If the process gas is cooled down around the sensor connection port when measuring high-
temperature and high-humidity gas, be sure to purge the air in order to prevent condensation
at the neck of the sensor connection flange. If there is a possibility of the ambient temperature
decreasing to 0ºC or less, provide thermal insulation at the neck of the connection flange.
(Note)
Mounting
screw
DT450G
Duct
F0702012.ai
Tubing for air purge
IM 11K01C01E
2-6 <2. Instruction>
CAUTION
If the gas to be measured is flowing constantly through a duct where the DT450G is mounted, air-
purge the monitor as instructed in subsection 4.2.2. If you leave the monitor exposed to the gas
for a prolonged period without air purging the monitor, the resistance of the insulated part of the
sensor probe will decrease. This may result in a failure of the monitor to start smoothly.
IM 11K01C01-01E
<3. Piping and Wiring> 3-1
IM 11K01C01-01E
3-2 <3. Piping and Wiring>
Air purge is normally carried out by continuously supplying air to the DT450G with a pressure
of approximately 50 kPa higher than that of the gas being measured. If the effect of air purge
is inadequate, due to the presence of moisture in the gas being measured, the DT450G may
require air purge at a higher pressure. For air purge, use an air source with a sufficiently large
capacity. (The normal rate of air consumption is approx. 50 Nl/min under the pressure of 50 kPa).
When connecting the pipe to the DT450G for air purge, note the following.
• The DT450G may be removed during inspection or maintenance. When connecting the pipe
to the DT450G, use a one touch connector and/or a pipe made of flexible material to ensure
easy maintenance.
• Install a stop valve or pressure reducing valve in a position as close as possible to the
DT450G.
• Connect a pipe with an inside diameter of 6 to 8 mm to the DT450G. Note that the pipe
connection of the DT450G is Rc1/4 threaded.
Coupling thread
Pressure of gas under measurement
Piping material Rc1/4
plus approximately 50 kPa
with 6- to 8-mm
inside diameter
Air source
If air purge need not be carried out, attach a blind plug to the pipe connection.
IM 11K01C01-01E
<3. Piping and Wiring> 3-3
CAUTION
Since the DT450G may be removed for inspection or maintenance, allow each cable to have a
sufficient margin of length.
Connect a power supply that conforms to the voltage and frequency in the specifications of the
DT450G. For wiring, use a three-core cable of a 5 to 10 mm outside diameter with the nominal
cross-section of the conductor being at least 1.5 mm2. Ground one core appropriately. Lead the
cable from the power supply through the DT450G’s cable connection fitted with a Pg11 cable
gland into the DT450G’s inner assembly. Then, connect the cable to the AC1 terminal (non-
grounded side) and AC2 terminal (grounded side).
Terminate the end of the grounding wire with an M5 crimp terminal and then connect it to the
ground terminal inside the enclosure. The ground resistance should be 100 Ω or less (JIS Class
D grounding).
Note: The DT450G main unit and the duct on which the monitor is mounted are in electrical continuity. Should the measured value be
adversely affected by the method of two point grounding at the duct (see Section 3.4) and the terminal, remove either of the two
grounding lines.
Power Connector
AC1
AC2
Ground
N L
Power Switch
Power Wiring.ai
AC power supply
Figure 3.2 Wiring diagram of the DT450G to a power supply
IM 11K01C01-01E
3-4 <3. Piping and Wiring>
CAUTION
The output signal may be mixed with electric noise if the cable swings. If the cable is likely to be
affected by wind, fin the cable.
DT450G
Not used
mA-
Output
mA+
no Power Supply connector
RL2
c (Relay contact) AC1
nc
AC2
no
c RL1
(Relay contact)
nc
(*2)
接地
SHIELD
(Not used)
Junction Box
N L
Power Switch
AC power supply
Receiver
DT450G Wiring.ai
Grounding
(*1) The nomonal cross-sevtion of each conductor should be 1.5 mm2
(*2) The nomonal cross-sevtion of each conductor should be not more than 0.5 mm2
(*3) This terminal is used for the grouding on the receiver side.
Notes: The above wiring diagram shows the one when using no contacts. Connect wires to terminals nc and c
when using nc contacts.
IM 11K01C01-01E
<3. Piping and Wiring> 3-5
If a duct or other alternative facility where the DT450G is mounted is charged with electricity, the
electricity will serve as noise, possibly affecting the signal under measurement. Ground the duct
where the DT450G is mounted, as well as other nearby structures, in order to prevent the duct
from being charged with electricity.
IM 11K01C01-01E
Blank Page
<4. Operation> 4-1
4. Operation
The ranges and other parameters of the DT450G must be configured before you start operation.
This chapter explains how to operate the DT450G, including preparations, in the following order
of topics.
• Components and their functions
• Preparation for operation
• Regular operation
SHIELD
IM 11K01C01-01E
4-2 <4. Operation>
IM 11K01C01-01E
<4. Operation> 4-3
Do not set parameters except “0” for the leftmost digit “8”, “9” and “10.”.
IM 11K01C01-01E
4-4 <4. Operation>
l Manual Setting
In manual setting the measuring range is set by amplifying the signal detected by the sensor in 5
stages.
Press the leftmost (A) key until the leftmost digit shows “1.0”, and press the middle (B) or the
rightmost (C) key until the rightmost digit shows the number of the desired parameter. Table 4-2
shows that the rightmost digit number and its parameter.
Table 4-2 Measuring Range Setting
Display
Parameter Description
Right Number
0 Automatic range setup
1 Maximum range (For high concentration of dust)
2 Gain x 10
3 Gain x 100
4 Gain x 1000
5 Gain x 10000 (For low concentration of dust)
When parameter 1 is selected, the measuring range is maximized. If the dust concentration is
high, select this parameter. The output sensitivity becomes higher as the parameter number
increases from “1” to “5”. Chose the most appropriate parameter for your process.
l Auto Setup
In the process where the dust concentration is a certain level and stable during normal operation,
range setting by Auto Setup button is also available.
Before the Auto Setup, the setting level should be confirmed.
The dust monitor has an Automatic Setup function. You can change the default percent of full
range (20%) in steps of 10% from 10 to 50%. Press the leftmost (A) key until the leftmost digit
shows “7” and increase or decrease the value using the middle (B) or the rightmost (C) key.
IM 11K01C01-01E
<4. Operation> 4-5
The set point of 20% means that the 4-20 mA output at the typical dust concentration will be 4 mA
+ (20 mA - 4 mA) 20% which makes 7.2 mA. If the dust concentration is five times the typical
value or higher, the output will be 20 mA, and if the dust concentration is zero, the output will be 4
mA.
Next, the measuring range should be set. Press the leftmost (A) key until the leftmost digit shows
“1.0”. Press the middle (B) key or the rightmost (C) key to increase or decrease the value until the
rightmost digit shows “0” (see Table 4-1). Last, press the leftmost (A) key until the display shows
“-SP.” Press the middle (B) or the rightmost (C) key to save the setting.
For Automatic Setup, make sure that the process is running with a normal dust concentration.
Then open the cover of the dust monitor. After move the parameters setting mode, initiate
Automatic Setup by pressing the small key button near the indicator lamp on the lower printed
circuit board (see Figure 4.1). Note: The indicator lamp should be green before you press the key.
Note: Initiate Automatic Setup in parameters setting mode.
Collecting data of the dust concentration will take an hour or less. The three digit display counts
down towards zero, and the indicator lamp is red as long as the data are being collected. To avoid
interference from the background electrical noise, you should close the cover of the enclosure
right after having initiated the Automatic Setup. After an hour, the setup is ready and the dust
monitor is in normal operation. The indicator turns back to green when setup is ready.
Note: The setting level is valid only when the automatic range setup (parameter “0” ) is used.
IM 11K01C01-01E
4-6 <4. Operation>
4.3.2 Shutdown
To shut down the DT450G, turn off the power switch installed in the power line. If the DT450G will
not be used for a prolonged period, only turn off the power switch without turning off the air-purge
line.
The air-purge line need not be left turned on if there is no flow of gas to be measured (provided
that the ambient humidity is not too high).
4.3.3 Restart
There is nothing in particular that needs to be done before restarting the DT450G. If the DT450G
has not been in use for a prolonged period however, make sure there are no abnormalities with
the sensor probe before you restart the DT450G. When there are no abnormalities with the
DT450G, it will come into steady-state operation in approximately 30 minutes after power-on.
IM 11K01C01-01E
<5. Inspection and Maintenance> 5-1
This chapter explains inspection and maintenance carried out to repair or replace worn out or
deteriorated parts and thereby maintain the DT450G in its optimal operating condition.
CAUTION
When inspecting and maintaining the DT450G, remove it from where it is installed.
Be careful not to inflict severe impact to the monitor during inspection and maintenance,
otherwise it may be damaged.
IM 11K01C01-01E
5-2 <5. Inspection and Maintenance>
IM 11K01C01-01E
<6. Troubleshooting> 6-1
6. Troubleshooting
This chapter explains measures that should be taken when the output signal gives abnormal
readings. Note that not all anomalies are due to failure (disorder or malfunction) in the DT450G.
Some of them may be due to a change in the physical properties types of components, the flow
velocity, etc.) of the gas being measured. If such a change is likely, first check for the following.
Table 6.1 Causes of Failure and Corrective Actions
State of Failure Cause Corrective Action
The output signal • Air purge is not effective. • Check that the gap at the purge-air exit is
overshoots as specified. If the exit is contaminated with
toward the 20 mA dust, increase the purge pressure (remove
side. the dust on the exit).
• Condensation occurs at the probe inlet. • Take appropriate heat insulation measures.
The output signal • The purge-air exit is contaminated with • Clean the exit (see subsection 5.1.1).
gives abnormal dust.
readings. • The sensor probe is in a state of insulation • Use normal, dry purge air.
failure because of inappropriate purge air
(e.g., contains moisture).
• The gas under measurement has entered • Set the pressure relatively higher and
the purge-air piping, causing an insulation continue air purge for a while.
failure in the sensor probe. (If contaminants still remain in the piping,
contact Yokogawa Electric for assistance.)
The output • The flow velocity of the gas under • Change the measurement point to a
signal does not measurement is not constant. position in the straight portion of the duct
stabilize. where the flow of gas is regular.
• The humidity of the gas under • If the change occurs constantly, check if
measurement changes. measuring the gas is practicable at all.
The output • The wiring is inappropriate and therefore • Check the states of grounding and wiring
signal becomes picks up noise. and correct a problem if there is any.
disturbed • There is equipment nearby serving as a • Take measures for protecting the DT450G
momentarily. source of noise. from the source of noise.
There is no • The fuse is blown or there is something • Turn off the power switch temporarily if the
output signal. wrong with the wiring connections. green indicator lamp is unlit.
Then, examine the wiring connections and
check if the fuse has blown.
Consult Yokogawa if the fuse has blown.
• There is something wrong with the electric • Request repair to be carried out by
circuit components. Yokogawa.
IM 11K01C01-01E
Blank Page
Customer Model DT450
Maintenance Dust Monitor
Parts List
3 2
All Rights Reserved, Copyright © 2005, Yokogawa Electric Corporation. CMPL 11K01C01-01E
Subject to change without notice.
2nd Edition : Feb. 2012 (YG)
Yokogawa Electric Corporation
i
Revision Information
Title: Model DT450G Dust Monitor
Manual No.: IM 11K01C01-01E
IM 11M13A01-04E
Blank Page
n If you want have more information about Yokogawa products, you can visit
Yokogawa’s home page at the following website.
Home page: http://www.yokogawa.com/an/
n Written by Environmental & Analytical Products PMK Dept.
IA Div., Product Business Center
Yokogawa Electric Corporation
n Published by Yokogawa Electric Corporation
2-9-32 Nakacho, Musashino-shi, Tokyo 180-8750, JAPAN
n Printed by KOHOKU PUBLISHING & PRINTING INC.