Datex Ohmeda Modulus II W 7810 Ventilator Maintenance Procedures
Datex Ohmeda Modulus II W 7810 Ventilator Maintenance Procedures
Datex Ohmeda Modulus II W 7810 Ventilator Maintenance Procedures
I. GENERAL INFORMATION
WARNING: This procedure contains limited warnings and cautions. Refer to the product
Service Manual for a full listing of precautions.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
3 Elbow
4 U-Cup Seal
5 O-Ring
6 Spoppet
7 U-Cup Seal
8 O-Ring
no longer required
10 Retaining Ring
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
2 Elbow Fitting
3 Retaining Ring
4 Sleeve
1 Legris Plug
5 Retainer
6 Diaphragm
7 O-Ring
no longer required
8 Connector
9 O-Ring
11 Check Valve
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
1. Ensure that the System Master Switch is in the STANDBY position. Remove all pipeline
connections.
2. Open all cylinder valves. Verify that there is at least 1000 psi (745 psi for N2O) or higher on
the cylinder pressure gauges.
3. Close all cylinder valves. Verify that the pressure does not visibly drop in 1 minute.
4. Loosen the "T" handles and pull the cylinders back to verify that the pin index and strainer
nipples are intact.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
1. Connect all pipeline supplies and cylinder-only gas supplies (for example, Heliox).
a. Open all cylinder valves to ensure that the pneumatic circuits are pressurized.
b. Close cylinder valves for gases which also have pipeline supplies.
2. Turn the System Master Switch to the ON position.
3. Set all flowmeters for 3 L/min.
4. Verify that the cylinder gauges for pipeline supplied gasses have not dropped to 0 psi. Such
a drop would indicate that the flowmeters are being supplied by the cylinders.
5. Remove oxygen supply.
6. Ensure that the following events occur in the following sequence:
a. Audio alarm activates for at least 7 seconds and the FAIL LED is illuminated.
b. Flowmeter floats for all gases except O2 fall to zero.
c. O2 flowmeter float falls to zero.
d. Verify "Low Supply Pres!" is displayed, the LED flashes on the ventilator and the audio
alarm sounds.
7. Turn all flowmeters to their minimum settings.
8. Re-establish O2 supply.
9. Verify that the audio alarm silences and the NORMAL LED is illuminated.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
NOTE: The old style relief valve may allow a negative pressure indication up to -2
cmH2O. If you find an old style relief valve, replace it with the new style valve
assembly.
F. GMS Inspection
NOTE: For units manufactured prior to 1986, verify that the Inhalation Port pressure
sensing nipple, a raised head gasket, a drain valve for adjustable height bag
arms, and check valve disk retainers with stand offs are installed.
2. Periodic Inspection
a. Gauge
1) Remove the gauge assembly and inspect the O-ring; replace as necessary.
Insert the gauge assembly into an approved test device (see tools appendix).
2) Verify the accuracy of the gauge at:
0 ± 1 cmH20 (0 ± 0.7 mmHg)
40 ± 1 cmH20 (29.4 ± 0.7 mmHg)
-5 ± 3 cmH20 (-3.7 ± 2.2 mmHg)
3) Ensure that the patient pressure sensing line is attached to the test device.
Verify that the ventilator's sub-atmospheric pressure alarm activates at -10 ± 5
cmH20 (- 7.4 ± 3.7 mmHg).
4) Return the gauge assembly to the absorber.
5) Ensure that the pressure sensing line is attached to the absorber.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
b. Check the integrity of the sensor port cap and O-ring. Replace as necessary.
c. Check the condition of the Inhalation and Exhalation domes, O-rings, check valves and
check disc retainers. Replace as necessary.
d. Looking through the inhalation check valve seat and O2 sensor port, inspect the GMS
manifold for contamination. If contamination is detected, remove the GMS manifold and
clean.
e. Check the condition of the corrugated return hose. Replace as necessary.
f. Check the condition of the canister gaskets, seals and upper gaskets. Replace as
necessary.
g. Inspect the canister dish; clean as necessary.
h. For elevating arms, check the Bag Arm Tube Drain for cleanliness.
i. Check for correct firm operation of the Locking Lever.
j. Lubricate the GMS Mounting Pin, as necessary. Install a mounting pin gasket if one is
not present.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
NOTE: For passive systems, cap the evacuation output port then proceed.
NOTE: For a passive system, reattach evacuation output tubing. Step f need not be
performed.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
New style PEEP valves started with serial number AAXS01000. If you find an old style valve, call
the Anesthesia Complaint Coordinator in Madison, at 1-800-521-0086, for disposition.
1. Pre-Operative Checkout
a. Check that the scale and knob labels are legible. If not, replace PEEP valve assembly.
b. Check that the instruction label is legible.
2. Inspection
a. Remove the PEEP Valve from the absorber.
b. Disassemble the lower section of the PEEP Valve and remove the blue check disk.
c. Verify that the pin moves side to side and freely in and out. Any restriction of movement
indicates that cleaning is required.
d. Reassemble the valve and reinstall it on the absorber.
3. Testing
With the patient circuit attached, the ventilator turned "OFF", and the absorber switched to
the ventilator mode, perform the following:
a. Remove the gas supply hose from the common gas outlet of the machine.
b. Attach one end of a 22 mm corrugated hose to the common gas outlet, attach the other
end to the patient outlet of the "Y" connector in the patient circuit.
c. Turn the PEEP Valve fully clockwise to the maximum PEEP setting.
d. Press the flush button while observing the absorber pressure gauge. Typically, the
indicated pressure will be approximately 40 cmH2O. If the pressure exceeds 60 cmH2O,
the valve must be serviced or replaced.
NOTE: If the pressure reading is not in the recommended range, the spring may
need replacing.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
C. Breathing System Pressure Leak Test (Absorber, bellows assembly, manifold, and drive
gas hose)
NOTE: Before performing this test, the bellows must be at the bottom of the housing.
NOTE: If this test is performed with a patient circuit with an occluded 'Y' piece, the
allowable leak rate is 300 ml/min.
5. Remove the corrugated hose, the test plug from the drive gas hose, and occlusions on any
gas sampling connections. Reconnect the drive gas hose.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
NOTE: This test begins now and is concluded later in the procedure.
1. Connect a 12 ft double lumen hose to the SENSE and INFLATE fittings of the Modulus II
Patient Interface Panel. Inspect all cuff and hose connections prior to performing this test.
Repair or replace any worn parts.
2. Attach an NIBP cuff and wrap the cuff around a solid cylinder.
3. Depress the hold button and turn monitor ON. Release the hold button after all 0's appear.
4. Verify the monitor beeps during the test.
B. O2 Monitor Calibration
1. Calibration/Alarms Testing
a. Note the high and low O2 alarm settings.
b. Connect the O2 sensor assembly with test device to the Common Gas Outlet.
c. Set the high O2 Alarm to 95%.
d. Introduce the O2 sensor to 100% concentration of oxygen and allow display to stabilize.
e. Calibrate the monitor to 100% and verify that the High O2 alarm activates.
f. Remove the O2 sensor from 100% Oxygen and expose it room air. Verify that the High
O2 alarm cancels.
g. Set the Low O2 alarm to 25%. Verify that the Low O2 alarm activates. Verify the monitor
stabilizes at 21% ± 3% within 3 minutes.
h. Press alarm silence and verify that the audible alarm is silenced.
i. Set the Low O2 alarm to less than the displayed value. Verify that the Low O2 alarm
cancels.
j. Set the high and low alarms to the original settings.
C. Proportioning System
WARNING: Use the hospital evacuation system with the service gas-evacuator tee. If
hospital evacuation system is not available, perform service in a
well-ventilated area.
1. Every 12 months, perform flowmeter verification on O2 only at the following flows: 500 ml/min,
1L/min, and 5 L/min.
2. Use the N2O Flow Control Valve and perform the tests in Table A progressively from low to
high flows. Do not overshoot any setting. Allow the System O2 monitor to stabilize between
readings.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
TABLE A
TABLE B
Set the O2
Flow Cont. The N2O flow must then read:
Valve to
(L/Min)
Min N2O Max N2O Oxygen
(L/Min) (L/Min) Monitor
6. Decrease N2O to minimum, the O2 float should not decrease at this time.
7. For optional gas(es) use the appropriate table shown below.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
TABLE C
Oxygen He Oxygen
N2 Monitor Range
CO2
TABLE D
TABLE E
8. Set all gas flows to minimum. Verify that the O2 flowmeter indicates 200 ml ± 25 ml (50 ml ±
10 ml for low flow systems).
D. Vaporizer Test
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
NOTE: Perform steps A through G (Periodic Inspection) only when Section III, Annnual ventilator
Service, is not performed.
TABLE F
Tidal Rate I:E Ratio Min (mL) Nominal (mL) Max (mL)
Volume
(mL)
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
3. Set the anesthesia machine to 2 L/min and switch the BAG/VENT switch to VENT. Ensure
the bellows are filled.
4. Switch the mechanical ventilation switch to ON and wait 40 seconds.
5. Verify that the low minute volume alarm activates. Press alarm silence.
C. Inspiratory Pause
1. Press inspiratory pause and verify the LED is on.
2. Verify I:E ratio changes and bellows pauses at the end of the inspiratory cycle.
3. Turn inspiratory pause OFF.
1. Reinstall the TVX clip backwards (arrows pointing away from the absorber) on the TVX
cartridge.
2. Ensure the reverse flow alarm activates.
3. Reinstall the TVX clip with the arrows pointing toward the absorber and ensure that the
reverse flow alarm cancels after two ventilation cycles (breaths). Turn the Mechanical
Ventilation switch OFF.
4. Return the Reverse Flow setting in the Set Up page to its original position.
5. Turn the Mechanical Ventilation switch ON.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
NOTE: If the monitor has the old style rotary vane pump, it should be replaced with a
pump upgrade kit. Refer to Appendix B for the stock number.
5. If the present pump has a silver colored motor housing, replace the sample pump after
12 months. If the present pump is labeled 3003-056-VFB, and has a larger, black anodized
"escap" pump motor housing, replace as needed. For both pumps, replace only with the
larger black anodized pump motor assembly.
6. Reinstall the electronics module cover and plug in the power cord.
B. Periodic Inspection.
1. Place the monitor in the service mode by holding down the Waveform button and turning the
power ON.
2. If equipped, check sample inlet filter; replace as necessary.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
C. Flow Verification
1. Connect a sample line to the Sample Inlet.
2. Connect a flowmeter 0-600 mL/min to the exhaust port of the monitor.
3. Occlude the sample line and verify the flow drops to less than 10ml/min.
4. Select Low Flow and verify the flow is between 90 - 170 ml/min.
5. Select High Flow and verify the flow is between 200-400 ml/min.
NOTE: Be sure the barometric pressure used for comparison is the local value
and is not corrected to sea level.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
E. Calibration Procedure
NOTE: The monitor should be on 5 minutes prior to calibration. First "CO2 balance N2" is
used to check CO2 span. Then, CO2 with N2O is used to check N2O span and CO2
compensation. Verify that the cal gas has not expired.
1. Zero Calibration
a. Verify monitor is in Low Flow.
b. With the full length of sample tubing connected to the monitor allow room air to be drawn
into the monitor (Ensure the room air is free of appreciable levels of nitrous oxide and
carbon dioxide).
c. If required, adjust the N2O Zero and the CO2 Zero controls until the upper display
indicates 0% (N2O) and 0.0% (CO2).
2. CO2 Calibration Only
a. Connect the free end of the sample tube to the 5% CO2 balance N2 cal gas can.
Do not block off the open end of the adapter tube. The opening is required to bleed off
excess pressure in the line.
b. Press down continuously on the valve stem of the gas canister. When the readings are
stable, adjust the front panel CO2 Span control until the reading matches the value of the
CO2 calibration gas (5% ±0.1).
NOTE: To ensure the calibration can is providing a sufficient flow of calibration gas to
the monitor, continue to press down on the valve stem for five additional
seconds and verify the readings remain stable. If the readings steadily
decrease, an insufficient volume of calibration gas is indicated (depleted
canister). Using a new can of calibration gas, repeat the Calibration procedure.
F. Operational Tests
1. With a sample line connected to the monitor, power it up and place it in the Ready mode.
Select % for the display.
CAUTION: Do not touch the sample line to your mouth or face. When breathing into
the sample line, hold it two to three inches away.
2. Blow into the sample line and verify a CO2 waveform is displayed. Also verify the End Tidal
Digital Display reads between 3 - 6%.
3. Set the high End tidal and Low End Tidal alarm values to trip these alarms.
4. Verify the audio and visual alarm indicators are activated for each of these alarms.
5. Clear the alarms and set the Apnea Delay switch to 30 seconds.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
6. Do not breath into the sample line. Verify the Apnea flashing "Alarm" indicator and audio
tone are active after 30 seconds.
7. Breath into the sample line. Verify the Apnea alarm is cancelled.
8. Momentarily block the sample line. Verify the monitor goes into the purge mode.
9. Turn the monitor OFF and return the alarm settings to their original positions.
Table G
Table G shows the barometric pressures (mmHg) which correspond to the pressure transducer's output
voltage (P OUT) as displayed in Service Mode (Test 7);
TABLE G
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
TABLE G
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
A. VISUAL INSPECTION
1. Verify that the probe connection and input and output jacks are in good condition.
2. Replace the battery every 48 months. Record the replacement date on the battery.
NOTE: Later revisions of software refer to SpO2 instead of SaO2. For testing purposes, they
are the same.
1. Turn the Modulus II Plus System Master Switch to the OFF position.
2. Set the oximeter to run on battery power.
3. Enter the service mode by holding down the Alarm Volume button and turning the power ON.
Follow the instruction on the display to perform the diagnostic and calibration tests.
NOTE: Proceed with audio alarm testing only after the initial two minute alarm silence
period is over (Rev P and Later Software). The Alarm Silence Period for Rev M
software is 1 minute after power up.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
b. Depress the TREND 20/60 control. Verify that the displays for 20 minutes and 60
minutes of trend data are available.
c. Turn the oximeter OFF.
X. 2120 NIBP
NOTE: The overpressure switch will interrupt monitor power if excessive pressure is sensed in
the system. Do not adjust the overpressure switch unless the trip point deviates more
than -10 mmHg or +15 mmHg from the nominal specification (270 mmHg).
5) Continue to slowly pressurize the system and note at what pressure the monitor
cycles off to release the pressure (and go into the HOLD mode). The
overpressure trip point should be 260 mmHg or greater but not more than 285
mmHg. As the pressure reaches the expected trip point, do not increase the
pressure by more than 2 mmHg per second.
2. Back-up Timer
a. The monitor should be in the calibrate mode and about 50 mmHg pressure applied to the
inflate port.
b. Start a stop watch and depress the Alarm Silence button to start the Back-up Timer.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
B. Periodic Inspection
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
NOTE: Perform electrical Safety Tests every 6 months, or if the covers to a machine(s) is
removed.
NOTE: If tests are conducted with power from an isolated electrical power system, conduct
the tests as detailed below and record "ISOLATED POWER SYSTEM" on the MSR
detail block. Do not use an electrical jumper plug to defeat the isolation of the
power system.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
A. Disconnect all test devices and reconnect pipeline hoses, monitors and accessories that were
previously removed.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Appendix A
Required Tools
Item # Description Part Number Equipment
6 Electrical Analyzer BET 300AD / Dale 600 0175-2305-000 All but GMS
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Appendix B
Parts
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
7810 Ventilator
1. 12-month maintenance kit** 1500-8036-000 12 Months
2. Bellows, Adult ABA 1500-3378-000 As Required
3. Bellows, Pediatric Non ABA 0229-1018-700 As Required
4. Diaphragm and Seat Assy, ABA 1500-3377-000 As Required
5. Disk/ Ring /Bumper Assy, ABA 1500-3381-000 As Required
6. O-ring, large, pop-off valve Non ABA 1500-3267-000 As Required
7. O-ring, small, pop-off valve Non ABA 0210-0593-300 As Required
8. O-ring, drive gas nipple 0210-0593-300 As Required
9. Retaining ring, Ped Bellows Non ABA 0229-0036-300 As Required
10. Seal, ABA 1500-3359-000 As Required
11. Seat, Free Breathing Valve 0207-5590-100 As Required
12. U Cup Seal Non ABA 0210-0784-300 As Required
**NOTE: Any part in the 12 month maintenance kit that is not scheduled to be replaced on at the 12 month
interval, should be discarded.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
GMS Absorber
1. APL Valve Service Kit 0216-6785-870 As Required
1 ea Instr Instal GMS Apl Repl Parts
1 ea Label Set ABS GMS
1 ea Shaft Needle
1 ea Seat Valve APL
2 ea O-Ring 1-1/16OD 15/16ID
1 ea Disk APL 1.03D BL Delrin GMS ABS
2 ea Washer .946ID 1.1960D .016T Brass
2 ea Ring Retainer 15/16Shaft Truarc #5103-93H ST ST
1 ea Grommet APL Valve
1 ea Ring Locking APL Valve
1 ea Spring, Com 12.70D 38.1L
2. Canister Gaskets 0210-1214-100 As Required
3. Canister Seals 0210-1218-300 As Required
4. Disk Valve - PEEP 0210-5297-100 As Required
5. Disk Retainer - PEEP 0207-1642-542 As Required
6. Dome - PEEP 0207-1638-100 As Required
7. Instruction Tab - PEEP 1400-3006-000 As Required
8. Manifold Gasket with Risers 0229-2077-100 As Required
9. Washer Pin Gasket 0202-0094-300 As Required
10. O-Ring - Bag Arm Drain 0210-0483-300 As Required
11. O-Ring - Gauge 0210-0566-300 As Required
12. O-Ring - Lower Dish Drain, New Style 0210-0594-300 As Required
13. O-Ring - Lower Dish Drain, Old Style 0210-0481-300 As Required
14. O-Ring 0210-0574-300 As Required
15. Pivot Bracket 0229-2085-200 As Required
16. Retaining Ring - PEEP 0203-5249-300 As Required
17. Return Tube 1400-3009-000 As Required
18. Screen 0214-7167-535 As Required
19. In/Exhalation Check Valve Service Kit 1400-8011-000 As Required
2 ea O-Ring 2-1/16OD 1-7/8ID
1 ea O-Ring 9/16OD 7/16ID
2 ea Disc Chk Vlv Rvsbl
2 ea Pad Die Cut 2-1/2 X 1-1/4
2 ea Dome Valve Check
2 ea Cup Seal
1 ea O-Ring .437ID .625OD
20. Cup Seal (2) 0229-2007-100 24 Months
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
GMS Absorber
21. Service Kit - Selector Valve GMS 0216-6782-800 As Required
1 ea Bearing .130 DIA X. 250L
1 ea Bearing Nylon .250ID
1 ea Bushing Selector Valve
1 ea O-Ring .500 ID .625 OD .062W EPR 80 Duro
1 ea O-Ring 1/8X1/4X1/16 EPR
1 ea O-Ring .250ID .375OD .062W EPR 78 Duro
1 ea Roller
1 ea Sleeve Pivot Pin
1 ea Instr Instal Kit Selector Valve & Gauge Sensing
1 ea Spring Com
2 ea Ring Truarc .125 Shaft #5133-12H E-Ring
1 ea Adapter Barbed End 1/16ID Tube X 6-32 THD
1 ea Net, HT 6-32, Full MCH, SST
1 ea Pin Pivot
1 ea Retainer Cover
1 ea Shaft Retaining Ring
1 ea Bracket Support
1 ea Shaft Selector Valve
1 ea Bracket Pivot
2 ea Washer .438OD X .128ID X .065T Delrin
1 ea Tube Connector Patient Press 1/16ID Silicone Med
1 ea Washer No 6 Flat SST
Modulus II Plus
1. Battery System 6V 0690-1000-312 48 Months
2. Fan Filter 0219-3849-300 As Required
3. Gasket, Yoke Cylinder Valve 0210-5022-300 As Required
4. Index Pin 0201-0503-500 As Required
5. Inlet Strainer Nipple w/ Filter 0206-2806-725 As Required
6. O-Ring Tec Port Manifold 0210-0736-300 12 Months
7. O-Ring Manifold Machining Port 1102-3015-000 12 Months
8. Yoke Plug 0206-3040-542 As Required
9. Yoke Check Valve Repair Kit 0236-5035-870 As Required
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Appendix C
Ventilator Service Calibration Software Revision 3.XX or 4.XX
1. Turn the Anesthesia System Master switch to OFF or STANDBY. Place the Mechanical Ventilation
switch in the OFF position. Plug the 7800 Ventilator test Fisxture into connector J10 of the control
board. Connect a DMM to the test fixture. Place DIP switch SW101 switch 1 in the OFF position
and switches 2 and 3 in the ON position.
2. Turn the Anesthesia System Master switch to ON.
a. Verify long repeating 1khz beeps.
b. Verify within 25 seconds short 3khz beeps.
c. Press and hold alarm silence to enter service calibration. Verify screen displays “Entering
Service Calibration.”
3. Press alarm silence to advance to next test. Adjust contrast per screen instructions.
4. Press alarm silence to advance to next test. Select 5Vdc position on Test device. Verify voltage is
+5.2 Vdc + 0.00, -0.20. Adjust R7 on Universal Power Supply (no adjustment on Original).
5. Press alarm silence to advance to next test. Adjust R105. Select the test fixture +4.5 Vdc position
(J10 Pin 5 and Pin 15). The voltage reading should be 4.5 Vdc +0.005 Vdc. Adjust R105 on control
board as necessary.
6. Press alarm silence to advance to next test. Adjust R101. Select the test fixture O2 (R101) position
(J10 Pin 2 and Pin 15). The voltage reading should be 1.000Vdc +0.000 Vdc -0.005 Vdc. Adjust
R101 on control board as necessary.
7. Press alarm silence to advance to next test. Verify A/D: O2 PASSED. If the PASSED and FAILED
messages alternate, continue to the next step.
8. Press alarm silence to advance to next test. Adjust R102. Select the text fixture Pt Press (R102)
position (J120 Pin 1 and Pin 15). The voltage reading should be 1.000 Vdc+0.000Vdc -0.005 Vdc.
Adjust R102 on control board as necessary.
9. Press alarm silence to advance to next test. Verify A/D: PRES. PASSED.
10. Press alarm silence to advance to next test. Verify CPU PASSED.
11. Press alarm silence to advance to next test. Verify EPROM PASSED.
12. Press alarm silence to advance to next test. Verify EEPROM PASSED.
13. Press alarm silence to advance to next test. Verify RAM PASSED.
14. Press alarm silence to advance to next test. Toggle mechanical ventilation switch ON and OFF.
Verify condition on screen.
15. Press alarm silence to advance to next test. Verify correct operation of inspiratory pause switch.
Verify green light is OFF when switch is OFF. Verify condition on screen.
16. Press alarm silence to advance to next test. Verify DIP switch positioning. 0=off. 1=on. x=unused.
17. Press alarm silence. Adjust tidal volume knob. Verify correct operation throughout entire range,
50-1500ml.
18. Press alarm silence to advance to next test. Adjust rate knob. Verify correct operation throughout
entire range, 2-100B/min. Verify the knob and display correspond. Adjust as necessary.
19. Press alarm silence to advance to next test. Adjust inspiratory flow knob. Verify correct operation
throughout entire range, 10-100 L/min. Verify the knob and display correspond. Adjust as
necessary.
20. Press alarm silence to advance to next test. Adjust HI press. limit knob. Verify correct operation
throughout entire range, 20-100 cmH2O. Verify the knob and display relativelycorrespond. Adjust as
necessary.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
21. Press alarm silence to advance to next test. Advance low Ve limit pushwheels throughout entire
range, 0-9. Verify the pushwheels and display values correspond.
22. Press alarm silence to advance to next test. Advance low O2 limit pushwheels throughout entire
range, 0-9. Verify the pushwheels and display values correspond.
23. Press alarm silence to advance to next test. Advance high O2 limit pushwheels throughout entire
range, 0-9. Verify the pushwheels and display values correspond.
24. Press alarm silence to advance to next test. Check display contrast. Adjust R114 on control board
as necessary.
25. Press alarm silence to advance to next test. Check volume throughout entire range, 1-10. Reset to
original volume.
26. Press alarm silence to advance to next test. Verify EXHALATION VALVE TEST : PASS.
27. Press alarm silence to advance to next test. Verify INLET GAS VALVE TEST : PASS.
28. Press alarm silence to advance to next test.
Patient pressure sensor test:
a. Verify 0 +0.5 cmH2O is on the display. Adjust R102 on the pressure transducer board as
necessary.
b. Using approved device, apply 100cmH2O (73.5 mmHg) to the patient pressure port. Verify
100 +1.0 cmH2O (73.5 + 1 mmHg) is shown on the display. Adjust R101 on the pressure
transducer board as necessary.
c. Recheck zero value. If value is off, readjust until A and B criterion are met.
d. Apply 50 cm H2O (36.7 mmHg) to the patient port. The display must read within + 2 cmH2O
(1.5 mmHg) of the input pressure.
e. Remove the test device from the patient port.
29. Press alarm silence to advance to the next test.
High pressure sensor test:
a. Verify the display reads within 1.5% of the test device pressure gauge reading (kpa= psi X
6.895).
b. If the display reading is not within 1.5% of the gauge reading, adjust R103 on pressure
transducer board as necessary.
30. Press alarm silence to advance to next test. Adjust servo zero test.
Note: Having the Mechanical Ventilation switch ON allows switching between servo zero
and servo gains test.
a. Select the test fixture 60 mVdc position (J10 Pin 4 and Pin 10).
b. Place Mechanical Ventilation switch in the ON position.
c. Adjust R103 on the control board until a voltage of 60 +0.6 mVdc is displayed.
31. Press alarm silence to advance to next test.
Adjust servo gain test:
a. Select the test fixture 330 mVdc position (J10 Pin 4 and Pin 10). Adjust R104 on the control
board until a voltage reading of 330 +3.0 mVdc is displayed.
b. If R104 was adjusted, press the alarm silence and verify adjustments for R103 (60 +0.6
mVdc) and R104 (330 +3.0 mVdc) are still correct.
c. Turn OFF Mechanical Ventilation switch.
32. Press alarm silence to advance to next test. Verify CALIBRATION HIGH PRES REGULATOR is on
screen. This test will be performed later in conjunction with flow delivery test.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
33. Press alarm silence twice to advance to next test. Verify VOL TRANSDUCER INPUT TEST is
displayed on screen.
a. Press alarm silence to advance to next test.
b. Set 8L/min O2 flow on anesthesia machine flowmeter.
c. Connect the TVX clip, with the arrows pointing out, to the common gas outlet of the
anesthesia machine.
d. Verify the ventilator displays 8 +2.0L/min with a forward flow. (→).
e. Reverse the sensor clop and verify the ventilator displays 8 +2.0L/min reverse flow (←).
f. Remove TVX clip from the common gas outlet and set the O2 flow to minimum.
34. Press alarm silence to advance to next test.
Verify SYSTEM TEST A/D FLOW VALUE: PASS
35. Press alarm silence to advance to next test.
Verify HI PRES LIMIT SAFETY CKT TEST is displayed.
a. Press alarm silence to advance to next test.
b. Verify HI PRES LIMIT CKT TEST: FAIL appears on display.
c. Connect an adjustable pressure source (90-135 cmH2O) to the drive gas output connection.
d. Raise the pressure very slowly and check that the pressure drops sharply between 105 and
115 cm H2O (767-86 mmHg).
e. The display should read PASS after the pressure drops.
f. Remove test equipment from patient outlet.
Note: To repeat test, turn ventilator mechanical switch ON and press alarm silence to
return to Step b. To exit test, turn Mechanical Ventilation switch OFF.
Caution: If POSSIBLE FLOW CALIBRATION is displayed, turn DIP switch SW101 switch 4
OFF before continuing.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Note: The accuracy of the measuring device shall not be greater than +3%. The total flow
specification for the ventilator and measuring device is +10%. Respirometer
readings have been adjusted to compensate for respirometer response curve.
e. Verify the 26 psi (180 kPa) regulator has a reading 26 + 1.0 psi (180
+7 kPa) at 10L/min and a reading of 26 +0, -2.0 psi (180 +0, -14
kPa) at 100L/min.
f. Place the mechanical ventilation switch in the OFF position.
41. Press alarm silence to advance to next test.
42. Press alarm silence to advance to next test.
Perform Watchdog Timer Circuit Test:
a. Press alarm silence.
b. Verify a system fail alarm sounds, the yellow LED blinks on the alarm silence button, and the
display goes blank.
43. Turn the Anesthesia System Master Switch to Standby.
44. Return DIP switch SW101 switches 2 and 3 to the OFF position and switch 1 to the ON position.
45. Remove all test fixtures from Ventilator.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.
Maintenance Procedures
Appendix D *
Maintenance Procedure Revision Log
* Note: This document has been under revision control, but a revision history has not
been required to be included with this procedure. From this revision going forward, this
log will be updated to inform the user of changes made to this document.
MP 1014 The information contained in this document is CONFIDENTIAL INFORMATION and IS INTENDED ONLY FOR THE USE OF EMPLOYEES OF
GENERAL ELECTRIC COMPANY, IT’S SUBSI DIARIES, AFFILIATES OR DESIGNEES. Distribution or copying of this document by non-
employees or non-designees is strictly prohibited without the express written consent of General Electric Company or subsidiary thereof.