Section 9
Oxygen Analyzer
The GENERON HP Nitrogen Generator is provided with a Fuel Cell type oxygen analyzer,
which is an integral part of the vent-and-shutdown control system.
2. Set 3-way valve (20-1) to “SPAN”. This allows the inlet calibration gas to flow through
the analyzer.
3. Adjust (FI 20-1) to a flow of 1 scfh and check that pressure reducing valve (PRV 20-1) is
set to 5 psig. Allow the meter reading to stabilize for several minutes.
Routine Operation of the analyzer includes regular calibrations and cell changes. The Fuel
cell has a nominal life of 0.5 - 1 year. Analyzer calibration should be performed every 4wo
weeks, or whenever abnormal performance is suspected. Check frequently, that the
pressure reducing valve (PRV-20-1) is set to 5 PSIG and the Flow Indicator to 1 SCFH. For
more details about the analyzer, refer to the analyzer instruction manual.