App Note DC DC Stability V2
App Note DC DC Stability V2
App Note DC DC Stability V2
Strongly supported by
Note: Basic procedures such as setting-up, adjusting and calibrating the Bode 100 are described
in the Bode 100 user manual. You can download the Bode 100 user manual at
www.omicron-lab.com/bode-100/downloads#3
Note: All measurements in this application note have been performed with the Bode Analyzer
Suite V2.41 SR1. Use this version or a higher version to perform the measurements shown
in this document. You can download the latest version at
www.omicron-lab.com/bode-100/downloads
For the characterization of the loop we measure the open loop gain by using the voltage injection
method. This method is commonly used to analyze the stability of voltage regulators such as switched
mode power supplies.
To ensure that the measured loop gain equals the real loop gain we need to find a point where the
loop is restricted to one single path and where the impedance looking in the direction of the loop is
much bigger than the impedance looking backwards.
The following figure shows the feedback loop of the circuit and indicates the suitable injection point.
The impedance looking backwards equals the output impedance of the converter which is very low (in
the range of several ). The impedance looking in direction of the loop is formed by the
compensator and the voltage divider and is in the range of several .
More details on the selection of the injection point and the theory of the voltage injection method can
be found in the article Loop Gain Measurement which is available for download at:
http://www.omicron-lab.com/bode-100/application-notes-know-how/articles-use-cases.html#3
The disturbance voltage is applied in parallel to the injection resistor using the B-WIT 100 injection
transformer. The transformer is necessary to isolate the output of the Bode 100 from the DC operating
point of the feedback loop. The following figure shows how the Bode 100 is connected to the circuit.
Figure 3: Connecting the Bode 100 to measure the loop response of the regulator
The inputs of the Bode 100 are connected to either side of the injection transformer. CH1 measures
the disturbance signal that is applied to the feedback divider and CH2 measures the signal that
appears at the output of the converter. By dividing the voltage at CH2 by the voltage at CH1 we get
the transfer function from the feedback input to the output of the power supply. This transfer function
we call the loop gain .
In this case we use 10:1 probes to pick up the signals but any standard oscilloscope probe can be
used for this measurement.
Attention: If hazardous voltages are present, make sure that suitable probes are used to protect
operator and device from any dangerous voltage !
Furthermore, it is very important to avoid mechanical stress at soldering pads to prevent damage to
the test object. The following figures show how we have realized the modification on the demo board
and how the probes and the injection transformer are connected to the circuit.
Figure 5: The probes and the injection transformer connected to the circuit
Note: When analyzing the open loop gain for stability as it is done in text-books, positive
feedback occurs at -180 phase. Therefore the phase margin is measured by determining
the phase difference to -180.
In this measurement we measure the open loop gain in a closed loop system. The
phase margin must therefore be measured relatively to the 0 line!
This is somehow confusing but gets clearer if you imagine a signal that is injected at the
feedback input and appears at the output without any phase shift. Such a signal that
passed the loop with 0 phase will again be injected at the feedback and sum up with the
previous one. This is exactly the point where positive feedback and therefore instability
will occur in a negative feedback system.
Open the device configuration window and apply the following settings:
The external reference is switched on and both inputs are set to high impedance.
4.1 Calibration
We first check if a calibration is necessary for this measurement. Calibration would be necessary if
the two probes used to connect the Bode 100 to the circuit have a different frequency response. This
would introduce phase shift and attenuation errors in the measurements.
In order to check if our probes have the same frequency response, we connect both probes to the
same side of the injection resistor as shown in the picture below.
The measurement shows a flat line at 0dB and 0 which indicates that both probes have the same
frequency response and calibration is not necessary in this case.
60 150
40 100
20 50
TR1/dB
TR2/
0 0
-20 -50
-40 -100
-60 -150
Note: If this measurement shows a gain or phase curve deviating from and calibration is
necessary. Phase and amplitude difference can be compensated by performing a THRU
calibration in the Bode Analyzer Suite. Details on how to perform a THRU calibration can
be found in the Bode 100 User Manual.
Figure 9: Screenshot of the Bode Analyzer Suite showing the loop gain curve
The red line shows the gain magnitude and the blue curve the gain phase. Above the result
does not show much noise whereas in the lower frequencies the curve is very noisy. The reason is
the very small injection level and the high gain of . In order to reduce the noise in the low
frequency range we use the shaped level feature of the Bode 100.
We set the reference level to and increase the output level from to from -
to by entering a delta level of .
40 100
20 50
TR1/dB
TR2/
0 0
-20 -50
Figure 10: loop gain measurement ( Input voltage and load current)
By using the cursors we can read the Gain Margin and Phase Margin of the system. The
measurement indicates a Phase Margin of and a Gain Margin of .
40 100
20 50
TR1/dB
TR2/
0 0
-20 -50
-40 -100
-60 -150
Figure 11: big signal effects (nonlinearities) due to excessive injection signal
Such erroneous measurements can be avoided by reducing the injection signal level. The shaped
level feature provides the possibility to reduce the output level exactly at the frequencies where it is
necessary.
40 100
20 50
TR1/dB
TR2/
0 0
-20 -50
Figure 12: loop gain measurement ( Input voltage and load current)
The phase margin did decrease to whereas the gain margin did increase to .
60 150
100
40
50
TR1/dB
TR2/
20
0
0
-50
-20
-100
-40 -150
-60 -200
102 103 104 105
f/Hz
100mA : Mag(Gain) 500mA : Mag(Gain)
1000mA : Mag(Gain) 100mA : Phase(Gain)
500mA : Phase(Gain) 1000mA : Phase(Gain)
OMICRON Lab products stand for high quality offered at the best
price/value ratio on the market. The products' reliability and ease of use
guarantee trouble-free operation. Close customer relationship and more
than 25 years in-house experience enable the development of
innovative products close to the field.
[email protected] www.omicron-lab.com
Smart Measurement Solutions