Aisys User Manual Part 2 Ver 3.0
Aisys User Manual Part 2 Ver 3.0
Aisys User Manual Part 2 Ver 3.0
CAUTION U.S. Federal law restricts this device to sale by or on the order of a
licensed medical practitioner. Outside the U.S.A., check local
laws for any restriction that may apply.
Datex-Ohmeda products have unit serial numbers with coded logic which
indicates a product group code, the year of manufacture, and a sequential
unit number for identification. The serial number can be in one of two formats.
M1076286 i
Aisys Carestation
ii M1076286
Table of Contents
4 Parts
M1076286 iii
Aisys Carestation
Index
iv M1076286
1 Setup and Connections
M1076286 1-1
Aisys Carestation
WARNING Always make sure that the pipeline supply hoses and the
breathing circuit components are not toxic and will not:
1-2 M1076286
1 Setup and Connections
Canister setup
6 1
5
2
AB.74p042
AB.74p043
1. Canister support pin
2. Canister handle
3. Disposable Multi Absorber canister
4. Absorbent
5. Expiratory water reservoir
6. Canister release latch
7. Reusable Multi Absorber canister
Figure 1-1 Canister
M1076286 1-3
Aisys Carestation
1-4 M1076286
1 Setup and Connections
When to change the A gradual color change of the absorbent in the canister indicates absorption
absorbent of carbon dioxide. The color change of the absorbent is only a rough indicator.
Use carbon dioxide monitoring to determine when to change the canister.
Discard the absorbent when it has changed color. If left standing for several
hours, absorbent may regain its original color giving a misleading indication of
activity.
Read the canister instructions completely before using the product.
Removing a canister 1. Hold the canister by the handle and push on the release latch to unlock
the canister.
AB.74p058
2. Remove the canister by tilting it downward and off the two support pins.
Removing an EZchange 1. Hold the canister by the handle and push the canister cradle release
Canister latch to unlock the canister cradle.
AB.75p088
M1076286 1-5
Aisys Carestation
Reusable Multi 1. Turn the canister upside down and, using your thumbs, turn the cover
Absorber canister locking ring counterclockwise to unlock it.
filling
AB.4p044
2. Push up to release the seal.
3. Lift off the cover to remove it.
AB.74p046
4. Remove and properly discard the foam filters, the absorbent, and any
water in the reservoir.
AB.74p047
1-6 M1076286
1 Setup and Connections
6. Place a new filter in the bottom of the canister, pour absorbent into the
canister and place a new filter over the absorbent before closing and
locking the cover. Wipe off any absorbent dust.
7. Align the cover slots with the canister locking tabs and press the cover
down into place. Turn the cover locking ring clockwise to lock the cover in
place. Ensure cover is properly sealed to prevent leaks and spillage.
Alignment of the arrows helps to indicate correct assembly.
WARNING The filters must be in place to help prevent dust and particles from
entering the breathing circuit.
8. When replacing the canister, make sure it is resting on both support pins
before latching it into place.
M1076286 1-7
Aisys Carestation
Electrical connections
Port The port for the battery backup of the monitor is located above the isolated
outlets.
AB.75p087
Outlets Labels show outlet voltage ratings and circuit breaker amp ratings. These are
isolated outlets. Regularly test the leakage current.
AB.75p.009
Mains inlet Arrow shows the mains power inlet and cord.
AB.75p.008
1-8 M1076286
1 Setup and Connections
Serial port The system has an RS-232C electrical interface. The RS-232C connector
allows serial input/output of commands and data. The 15-pin connector is
located on the back of the display unit.
The 15-pin female D connector - Data Communications Equipment
configuration (DCE):
Pin 1 - Monitor On/Standby
Pin 5 - Signal ground
Pin 6 - Receive data
Pin 9 - Monitor On/Standby Return
Pin 13 - Transmit data
AB.75p090
M1076286 1-9
Aisys Carestation
Pneumatic connections
CAUTION Use only medical grade gas supplies. Other types of gas supplies
may contain water, oil, or other contaminants which could affect
the operation of the pneumatic system.
The gas supplies provide gas to these optional devices through internal
connections:
venturi suction regulator (optional).
auxiliary O2 flowmeter (optional).
Pipeline Inlets
AB.75p.010
Scavenging The scavenging assembly is located below the bellows on the breathing
system. Adapters may be necessary to interface to the scavenging connector.
See Operation and Tutorial in Part 1 of the Users Reference manual for
more scavenging information.
AB.91p045
1-10 M1076286
1 Setup and Connections
Sample gas return port Connect the Datex-Ohmeda sample gas exhaust tube to the gas return port.
Exhaust gas will be directed to the scavenging system.
AB.75p.109
Pneumatic power outlet
AB.75p.007
M1076286 1-11
Aisys Carestation
Suction regulator Venturi regulators use the system air or O2 supply. Vacuum regulators must be
(optional) connected to an external vacuum supply.
3 2
1
AB.75p.007
1. Overflow safety trap
2. External vacuum (non-venturi) connection
3. Splash guard
4. Collection bottle connection
Auxiliary O2 flowmeter
(optional)
2
AB.75o011
1-12 M1076286
1 Setup and Connections
CAUTION Do not leave gas cylinder valves open if the pipeline supply is in
use. Cylinder supplies could be depleted, leaving an insufficient
reserve supply in case of pipeline failure.
M1076286 1-13
Aisys Carestation
WARNING The top of the machine has a weight limit of 45 kg (100 lb).
WARNING Fully tighten the straps. If straps are not fully tightened,
equipment can fall off the top of the machine.
Installation notes
When the system is installed, the Datex-Ohmeda representative will check the
following items and change them if necessary.
Altitude.
1-14 M1076286
2 Cleaning and Sterilization
M1076286 2-1
Aisys Carestation
Read the material safety data sheet for each cleaning agent.
2-2 M1076286
2 Cleaning and Sterilization
2 9
1 10
8
11
AB.82.045
AB.82.048
AB.82.008
1. Bellows assembly
2. APL valve ramp
3. APL valve diaphragm
4. Breathing circuit module (O2 cell not autoclavable)
5. Absorber canister (reusable only)
6. Flow sensor cover*
7. Flow sensor module (plastic flow sensors not autoclavable)
8. Exhalation valve assembly
9. Condenser module
10. Condenser
11. EZchange Canister module
Figure 2-1 Autoclavable assemblies
M1076286 2-3
Aisys Carestation
Special requirements To clean the circuit O2 cell, wipe it with a damp cloth.
If the flow sensors are plastic, refer to the How to clean and disinfect the
flow sensors procedure. Do not get the electrical connectors wet.
Autoclave metal flow sensors at 134C.
Disassemble the bellows assembly before washing. If not, it will take a
very long time to dry. Hang the bellows upside down (extended) to dry. If
not, the convolutions can stick together.
Parts marked 134C are autoclavable or washable by hand or machine
(mild detergent pH <10.5). Rinse and dry completely. All parts except the
O2 cell and disposable flow sensors can be washed.
WARNING Do not use talc, zinc stearate, calcium carbonate, corn starch or
equivalent materials to prevent tackiness. These materials can go
into the patients lungs and airways and cause irritation or injury.
CAUTION Do not put the circuit O2 cell or flow sensor connector in liquid.
w Do not insert any objects into the flow sensor to clean the interior
surfaces. Damage to the flow sensor can occur. Use a damp cloth
to clean external surfaces if needed.
2-4 M1076286
2 Cleaning and Sterilization
w Do not use high pressure gas or brushes to clean the flow sensors.
w Only use cleaning solvents that are approved for use with
polycarbonates (e.g., Do not use CIDEX Plus).
Note Autoclavable metal flow sensors are available and can be sterilized at 134C.
CIDEX disinfection Both Datex-Ohmeda and the manufacturer of CIDEX (Johnson & Johnson)
have tested this procedure.
CIDEX must be 14 day mixture, with activator vial.
One liter of this solution cleans four flow sensors.
1. Pull the latch to unlock the flow sensor module from the breathing
system.
M1076286 2-5
Aisys Carestation
1 3 2
4. Submerge the flow sensor and tubes in activated CIDEX solution. Keep
the connector dry.
5. Keep the flow sensors and tubes in the solution for the required period.
6. Submerge the flow sensor and tubes in distilled water. Again, do not get
the connector wet.
2-6 M1076286
2 Cleaning and Sterilization
9. COMPLETELY dry the flow sensor and the tubes before using the sensor.
Use a dry syringe or connect vacuum or pressure to remove all liquid from
the sensor (sensor, tubes, and connector).
CAUTION Dry for more than one minute with these precautions:
M1076286 2-7
Aisys Carestation
3 1
2. If bag arm option is present, remove the bag port elbow from the bag arm
support. Push down on the release latch and slide the bag port elbow out
of the holder.
2. Remove the canister by tilting it downward and off the two support pins.
3. Push the release button (1) and gently pull the latch handle (2) to release
the breathing system.
1
2
2-8 M1076286
2 Cleaning and Sterilization
4. Grasp the rear handle to support the breathing system. Slide the
breathing system away from the workstation pulling only with the latch
handle.
2. Pull the latch to unlock the flow sensor module from the breathing
system.
M1076286 2-9
Aisys Carestation
4. Remove the O2 cell cable from the cell. Unscrew the O2 cell
counterclockwise and remove it. Remove the O2 cell cable from the
breathing system by pressing on the connector button while pulling the
connector out.
2-10 M1076286
2 Cleaning and Sterilization
1 3 2
M1076286 2-11
Aisys Carestation
8. On the breathing circuit module, remove the check valve circuit lens (1)
by squeezing the latches (2) together and pulling up on the lens. Lift out
the check valve assemblies (3).
3 2 3
2-12 M1076286
2 Cleaning and Sterilization
9. Press the latch (1) to unlock the ramp (2). Rotate the ramp and remove
the tabs from the slots (3) to remove the ramp.
2
1
M1076286 2-13
Aisys Carestation
11. Turn over the bellows base assembly, grasp the bellows manifold with
three fingers in the openings at the points shown and pull straight away to
remove it.
12. With the breathing system removed, the exhalation valve assembly can
be removed for cleaning if desired. Loosen the two thumbscrews
indicated and lift the assembly off.
2-14 M1076286
2 Cleaning and Sterilization
3. Push the latch toward the center and remove the rim.
M1076286 2-15
Aisys Carestation
5. Push the latch toward the center and remove the locking tabs.
2-16 M1076286
2 Cleaning and Sterilization
2. Push the latch toward the center and attach the locking tabs.
4. Push the latch toward the center and install the rim. A double-click
should be heard when the rim is installed.
M1076286 2-17
Aisys Carestation
5. Attach the bottom edge of the bellows to the rim. Verify only the bottom
ring of the bellows is fitted over the rim.
6. Lower the housing and turn it clockwise to lock. Verify the housing is
secure.
2-18 M1076286
2 Cleaning and Sterilization
WARNING Objects in the breathing system can stop gas flow to the patient.
This can cause injury or death:
Do not use a test plug that is small enough to fall into the
breathing system.
Make sure that there are no test plugs or other objects caught
in the breathing system.
These tests make sure that all components are correctly assembled. These are
not an alternative to a complete system checkout. If the bellows assembly
operates correctly, complete the assembly of the breathing system. If there is
a problem, disassemble the bellows assembly. Verify proper assembly and
look for and replace damaged parts.
1. Hold the bellows assembly vertical and use the appropriate test plugs to
seal the ports shown.
M1076286 2-19
Aisys Carestation
2. Invert the bellows assembly. The bellows must not fall within one minute.
If it does:
The ports are not tightly sealed.
The bellows is incorrectly installed.
The seal inside the bellows is not correctly installed (with its groove
pointed up).
Parts are damaged.
3. Remove the plugs from the ports. Permit the bellows to fully extend.
4. Use the appropriate test plug to seal the port shown.
5. Hold the bellows assembly upright. The bellows must not fall past the
guide line within one minute. If it does:
The port is not tightly sealed.
The bellows or the pressure-relief valve is not correctly installed.
Parts are damaged.
6. If all tests pass, complete the assembly of the breathing system.
2-20 M1076286
2 Cleaning and Sterilization
2. Turn over the bellows base assembly. Replace the manifold. Be sure to
insert it correctly into the ports as shown. Then, press on the center of the
manifold to snap it into place on the bellows base assembly.
M1076286 2-21
Aisys Carestation
4. Insert the ramp tabs (1) into the slots (2). Rotate the ramp until it
locks at (3).
1
2
2-22 M1076286
2 Cleaning and Sterilization
5. On the breathing circuit module, insert the check valve assemblies (1).
Push the check valve circuit lens (2) down onto the latches (3) to lock
the lens.
1 3 1
6. Insert the breathing circuit module into the bellows assembly aligned as
shown.
7. Rotate the breathing circuit module clockwise at the point shown by the
dotted line to attach it to the bellows assembly.
M1076286 2-23
Aisys Carestation
8. Make sure the o-ring (1) is on the O2 cell. Replace the cell by screwing it in
clockwise. Connect the O2 cell cable (2).
4 2 3
2-24 M1076286
2 Cleaning and Sterilization
11. Push the latch closed to lock the flow sensor module onto the
breathing system.
M1076286 2-25
Aisys Carestation
2-26 M1076286
2 Cleaning and Sterilization
3. Holding the rear handle and the latch handle as shown, slide the
breathing system onto the guide pins.
4. Use the grip under the latch handle to push the breathing system in fully
until it latches firmly.
5. Install the absorber canister and bag hose.
6. Complete the tests in the Preoperative Tests section of the Users
Reference manual, Part 1.
M1076286 2-27
Aisys Carestation
2. Slide the side panel out by removing its tabs from their slots.
3. Loosen the thumbscrew and remove the reservoir.
CAUTION Do not autoclave the reservoir. Damage to the reservoir will occur.
5. Replace the filter as necessary. (See Remove the AGSS receiver filter.)
6. Do these steps in the opposite order to replace the receiver, reservoir,
and the side panel.
7. Complete the tests in the Preoperative Tests section of the Users
Reference manual, Part 1.
2-28 M1076286
2 Cleaning and Sterilization
CAUTION The AGSS filter is not autoclavable. Damage to the filter will occur.
If the AGSS receiver has a filter, remove the filter before
autoclaving the AGSS receiver.
3. Do these steps in the opposite order to replace the filter and gasket after
autoclaving the receiver and gasket. Be sure the gasket is firmly pressed
into place at all points.
4. Complete the tests in the Preoperative Tests section of the Users
Reference manual, Part 1.
M1076286 2-29
Aisys Carestation
Absorber canister
The absorber canister is available in two versions: Disposable Multi Absorber
and Reusable Multi Absorber. Only the Reusable Multi Absorber canister may
be cleaned. Refer to Removing a canister in section 1, Setup and
Connections.
Mechanical cleaning in 1. Place the reusable canister (without filters) and the lid in the washer or
washer or washer- washer-disinfector and clean them using the appropriate procedures.
High level disinfection 1. Always clean the canister before high level disinfection.
2. The canister can be steam autoclaved. Maximum recommended
temperature is 134C (273F).
3. Ensure the canister is dry before replacing the filters and refilling with
absorbent. Refer to Removing a canister in section 1, Setup and
Connections.
2-30 M1076286
2 Cleaning and Sterilization
4 1
AB.75p80
2
M1076286 2-31
Aisys Carestation
To remove the EZchange Canister system and the condenser without removing
the breathing system:
1. Remove the canister.
2. Push the release latch and pull down on the unit to remove it from the
breathing system.
3. For systems with the EZchange Canister only, pull the release latches out
to remove the cap.
2-32 M1076286
2 Cleaning and Sterilization
5. Remove the condenser reservoir by pulling the lip of the gasket away from
the reservoir around the edges.
6. When reinstalling the reservoir, ensure the curved end of the reservoir
lines up with the curved end of the seal. Push the reservoir into the
gasket.
7. Ensure the lip of the gasket fully covers all edges of the reservoir for a
leak-tight fit.
M1076286 2-33
Aisys Carestation
2-34 M1076286
3 User Maintenance
Do not use lubricants that contain oil or grease. They may burn
or explode in high O2 concentrations.
M1076286 3-1
Aisys Carestation
Repair policy
Do not use malfunctioning equipment. Make all necessary repairs or have the
equipment serviced by a Datex-Ohmeda trained service representative. After
repair, test the equipment to ensure that it is functioning properly, in
accordance with the manufacturers published specifications.
To ensure full reliability, have all repairs and service done by a Datex-Ohmeda
trained service representative. If this cannot be done, replacement and
maintenance of those parts listed in this manual may be undertaken by a
competent, trained individual having experience in the repair of devices of this
nature.
User maintenance
Minimum Frequency Maintenance
Daily Clean the external surfaces.
Perform 21% O2 calibration.
Check the condenser reservoir. Drain if needed.
Monthly Perform a Backlight test.
Perform 100% O2 calibration.
During cleaning and Inspect the parts for damage. Replace or repair as
setup necessary.
3-2 M1076286
3 User Maintenance
M1076286 3-3
Aisys Carestation
Note It may take a new O2 cell 90 minutes to stabilize. If the O2 cell calibration fails
after a new O2 cell had been installed, wait 90 minutes and repeat the
calibration.
1. Pull the latch to unlock the flow sensor module from the breathing
system.
4. Make sure the o-ring is on the cell. Install the replacement O 2 cell.
Reconnect the O2 cell cable.
5. Replace the flow sensor module on the system and push the latch closed
to secure the module.
3-4 M1076286
3 User Maintenance
100% O2 calibration 1. Ensure patient Y-piece is not plugged or there is no patient tubing
connected to the system.
2. (ACGO option only.) Set the ACGO switch to Circle.
3. Set the Bag/Vent switch to Vent.
4. Select 100% O2 and push the ComWheel to start the calibration.
5. System will flow O2 to calibrate.
M1076286 3-5
Aisys Carestation
Backlight test
1. Push the Main Menu key.
2. Select Calibration.
3. Select Backlight Test.
4. Select Start Test.
5. The display will show the test running on light 1 and then on light 2. If the
display goes completely blank or flickers during the test, one of the lights
has failed. Contact a Datex-Ohmeda trained service representative to
replace the backlights.
3-6 M1076286
4 Parts
This section lists user-replaceable parts only. For other components, refer to
the Technical Reference manual.
M1076286 4-1
Aisys Carestation
AB.82.019
2
4-2 M1076286
4 Parts
AB.82.021
5
M1076286 4-3
Aisys Carestation
Bellows
8 3
4
7 5
4-4 M1076286
4 Parts
Absorber canister
3
1
AB.82.017
Item Description Stock number
1 Multi absorber, reusable (includes 40 pack of foam) 1407-7004-000
(does not include absorbent)
2 Cover assembly, CO2 canister 1009-8240-000
3 Foam, CO2 canister (pack of 40) 1407-3201-000
4 O-ring 1407-3204-000
5 Canister, CO2, with handle 1407-3200-000
Multi absorber, disposable, white to violet, 8003138
(pack of 6)
Multi absorber, disposable, pink to white, (pack of 6) 8003963
M1076286 4-5
Aisys Carestation
AB.82.035
Description Stock number
Exhalation valve assembly 1407-7005-000
4-6 M1076286
4 Parts
AGSS
Description Stock number
Common
Cap 3.18 barb silicone 1406-3524-000
Connector, inlet 30 mm male to 19 mm male M1003947
Connector, inlet 30 mm male to 30 mm male M1003134
O-ring for connector, 21.95 ID 1406-3558-000
O-ring for receiver, 22 ID 1407-3104-000
O-ring for thumbscrews, 4.47 ID 1407-3923-000
Reservoir scavenger 1407-3903-000
Seal, down tube scavenger 1407-3904-000
Seal, receiver scavenger 1407-3901-000
Thumbscrew M6 X 28.5 1406-3305-000
Thumbscrew, M6 X 43 1406-3304-000
Valve, unidirectional (complete assembly) 1406-8219-000
Passive AGSS
Adapter, outlet 30 mm female to 19 mm male (pack of 5) 1500-3376-000
Exhaust hose 8004461
Plug assembly 30 mm ISO 1407-3909-000
Screw, shoulder 4 dia X 4 L M3 X 0.5 sst 1407-3915-000
Active AGSS, adjustable flow
Bag with 30 mm male connector 8004460
Plug assembly 30 mm ISO 1407-3909-000
Active AGSS, high flow
Filter, 225 micrometer nylon screen AGSS 1406-3521-000
Seal, filter scavenger 1407-3902-000
Active AGSS, low flow
Filter, 225 micrometer nylon screen AGSS 1406-3521-000
Seal, filter scavenger 1407-3902-000
M1076286 4-7
Aisys Carestation
3 1
AB.82.044
AB.82.057
Item Description Stock number
1 EZchange Canister module, includes valve and cap 1407-7021-000
2 Valve 1407-3126-000
3 Cap 1407-3130-000
4 Condenser 1407-7024-000
EZchange Canister module with condenser 1407-7027-000
4-8 M1076286
4 Parts
Condenser
AB.82.045
Item Description Stock number
Condenser assembly (includes module and condenser) 1407- 7026-000
1 Condenser module 1407-7025-000
2 Condenser 1407-7024-000
M1076286 4-9
Aisys Carestation
4-10 M1076286
5 Specifications and Theory of Operation
Note All specifications are nominal and subject to change without notice.
M1076286 5-1
5-2
75
Aisys Carestation
23
72
22
NO
24 51
NC 53
21 25 71
20 70
26 52
73 74
1 8
9 55
86 53
18 44
P 11 28 76
10
3
P 17
NC
29 56
5 79 77
12 78
4
30 54
6 P 46
P 31 80
13 15 57
4 NO 53
32 68
P 14 33
66 45
81 58 60 59
3
62
P 5
T P
P 37 13
4
NO 54
6 7 61
P
34 38 69
P
16 67
3 82 39 41
NO
P 5 35 40 NC
66
4
7 64
P P P P
P 83
47 48 49 50
36 42
84 84
65
43 63
85
AB.75.036
M1076286
5 Specifications and Theory of Operation
M1076286 5-3
Aisys Carestation
Gas supplies Pressurized gas supplies enter the system through a pipeline or cylinder
connection. All connections have indexed fittings, filters, and check valves.
A regulator decreases the cylinder pressures to the appropriate system
pressure. A pressure relief valve helps protect the system from high pressures.
To help prevent problems with the gas supplies:
Install yoke plugs on all empty cylinder connections.
When a pipeline supply is connected, keep the cylinder valve closed.
Disconnect the pipeline supplies when the system is not in use.
WARNING Do not leave gas cylinder valves open if the pipeline supply is in
use. Cylinder supplies could be depleted, leaving an insufficient
reserve supply in case of pipeline failure.
Air and N2O Pipeline or regulated cylinder pressure supplies air directly to the air channel
of the gas mixer and to the ventilator if air is configured as drive gas. If the
pressure is too low, an alarm appears on the display.
Pipeline or regulated cylinder pressure supplies N2O directly to the N2O
channel of the gas mixer. If the pressure is too low, an alarm appears on the
display.
Mixed gas Either Air or N2O can be selected as the balance gas. The balance gas flow is
controlled by the mixer. When Alternate O2 control is in use, only O2 is flowing
and balance gas is disabled.
The mixed gas goes from the mixer outlet through the vaporizer to the fresh gas
outlet and into the breathing system. A pressure-relief valve sets the
maximum outlet pressure.
5-4 M1076286
5 Specifications and Theory of Operation
EZchange Canister When activated, this mode permits continued ventilation and rebreathing of
exhaled gases without any gas passing through the absorbent.
Condenser The condenser removes water in the system that is produced from the reaction
of CO2 gas with the absorbent during low flow anesthesia (fresh gas flows less
than 1.5 l/min). The condenser is connected between the outlet of the
absorber canister and the inlet of the circuit module. Moisture in the gas is
condensed into water droplets, which run into the condenser's reservoir.
Pneumatic specifications
Gas supplies
Pipeline gases O2, Air, N2O
Primary regulator Pin indexed: The primary regulator is set to pressure less
output pressure than 345 kPa (50 psi).
DIN-477: The primary regulator is set to pressure less
than 414 kPa (60 psi)
M1076286 5-5
Aisys Carestation
Pneumatic power outlet DISS power outlet is shown as dot curve. ISO power outlet is shown as box
curve. Inlet pressure constant at 241.5 kPa.
AB.75.057
2
1. O2 flow in l/min
2. Outlet pressure in kPa
Figure 5-2 Pneumatic power outlet graph
ACGO port relief A relief valve in the vaporizer limits fresh gas pressure at the ACGO port to
55 kPa (8 psi) at 45 l/min.
Non-circle circuit relief A relief valve limits fresh gas pressure at the inspiratory port to 27 kPa (4 psi)
at 55 l/min when non-circle circuit is selected.
5-6 M1076286
5 Specifications and Theory of Operation
Electrical power
Power cord
Length 5 meters
Voltage rating 100 to 240 V ac
Current carrying capacity 10 A for 220-240 V ac
15 A for 100-120 V ac
Type Three conductor power supply cord (medical grade
where required).
WARNING Unplug the system power cord to run the system on the battery
power if the integrity of the protective earth conductor is in doubt.
Battery information The system is not a portable unit; sealed lead acid batteries supply backup
power in the event of a power failure.
Capacity to operate for 90 minutes under typical operating conditions;
30 minutes under heavy workload.
System functions to specifications through the transition to battery
power.
Only Datex-Ohmeda trained service representatives are to replace the battery.
Batteries must be disposed of in accordance with applicable regulatory
requirements in effect at the time and place of disposal.
M1076286 5-7
Aisys Carestation
6
9 10
1 2 3 4 5 76 8
68 70 78
82 77
81 69 71 79
83
67 72
66 73
65 7 74
64 75
30
28 12 11
31 13
29 27
14
33
34 32 54
80
35
40
36
41
37 47
45
38 42 48
49
39 50
43 44
46 52 53 51
57
18
25 59 58
16
62
26
19
55
15
60
56
63
17
AB.75.101
20 24 23 22 21
61
5-8 M1076286
5 Specifications and Theory of Operation
M1076286 5-9
Aisys Carestation
Flow specifications
Alternate O2
Flow range 500 ml/min to 10 l/min
Indicator Flow tube
Indicator accuracy 5% full scale
Fresh gas
Flow range 0 and 200 ml/min to 15 l/min
Total flow accuracy 10% or 40 ml/min of setting (larger of)
O2 flow accuracy 5% or 20 ml/min of setting (larger of)
Balance gas flow accuracy 5% or 20 ml/min of setting (larger of) Air/N2O
O2 concentration range 21%, 25% to 100%
(25% to 35% adjustment in 1% increments, 35% to
100% adjustment in 5% increments)
O2 concentration accuracy 5 V/V for flows < 1 l/min
2.5% setting for flows >1 l/min
Electronic mixer response 500 ms (10% to 90% flow step)
time
Compensation Temperature and atmospheric pressure compensated
to standard conditions of 20C and 101.3 kPa
(14.7 psi)
Hypoxic guard Electronic
Leakage from one gas inlet to another gas inlet is < 10 ml/hour.
5-10 M1076286
5 Specifications and Theory of Operation
M1076286 5-11
Aisys Carestation
5-12 M1076286
5 Specifications and Theory of Operation
Gas scavenging
All scavenging
Passive scavenging
Active scavenging
Disposal system type Outlet connector * Flow range Pressure
Adjustable flow DISS EVAC Up to 30 l/min 305 mmHg
(12 inHg)
Minimum vacuum
High flow, low vacuum BS6834 50 to 80 l/min 1.6 kPa
(12 mmHg)
Low flow, high vacuum DISS EVAC 36 4 l/min 305 mmHg
(12 inHg)
Minimum vacuum
Low flow, low vacuum 12.7 mm hose barb 40 to 60 l/min n/a (venturi)
Low flow, low vacuum 25 mm barb 40 to 60 l/min n/a (venturi)
Low flow, low vacuum 30 mm ISO taper 40 to 60 l/min n/a (venturi)
*Other market-specific connectors may be available.
Particle filter at the outlet has a pore size of 225 microns. All flow data uses a new filter.
Physical specifications
All specifications are approximate values and can change without notice.
M1076286 5-13
Aisys Carestation
Environmental requirements
Gas specifications
Airway humidity 0 to 100% condensing
Sampling delay 2.5 seconds typical with a 3 m sampling line
Total system response time 2.9 seconds typical with a 3 m sampling line,
including sampling delay and rise time
Warm-up time 2 minutes for operation with CO2, O2, and N2O
5 minutes for operation of anesthetic agents
3O minutes for full specifications
5-14 M1076286
5 Specifications and Theory of Operation
Typical performance
CO2 Measurement range 0 to 15 vol% (0 to 15 kPa, O to 113 mmHg).
Measurement rise time <400 ms typical.
Accuracy (0.2 vol% +2% of reading).
Gas cross effects <2 vol% (O2, N2O, anesthetic agents).
O2 Measurement range 0 to 100 vol%.
Measurement rise time <400 ms typical.
Accuracy (1 vol% +2% of reading).
Gas cross effects <1 vol% anesthetic agents, <2 vol% N2O.
N2O Measurement range 0 to 100 vol%.
Measurement rise time <450 ms typical.
Accuracy (2 vol% +2% of reading).
Gas cross effects <2 vol% anesthetic agents.
Anesthetic agents Measurement range Hal, Enf, Iso 0 to 6 vol%.
Measurement range Sev 0 to 8 vol%.
Measurement range Des 0 to 20 vol%.
Measurement rise time <400 ms typical.
Accuracy (0.15 vol% +5% of reading).
Gas cross effects <0.15 vol% N2O.
M1076286 5-15
Aisys Carestation
Ventilator theory
The ventilator pneumatics are at the rear of the breathing system. A precision
valve controls gas flow to the patient. During inspiration, this gas flow closes
the exhalation valve and pushes the bellows down. During expiration, a small
flow pressurizes the exhalation diaphragm to supply PEEP pressure.
Volume and pressure measurements come from flow sensors in the flow
sensor module. Two tubes from each sensor connect to a transducer that
measures the pressure change across the sensor, which changes with the
flow. A third transducer measures airway pressures at the inspiratory flow
sensor.
The ventilator uses the data from the flow sensors for volume-related
numerics and alarms. The ventilator also uses the flow sensors to adjust its
output for changes in fresh gas flow, small leaks, and gas compression
upstream of the breathing circuit. There is adjustment for compression in the
patient circuit.
For better precision a small quantity of gas bleeds through a resistor to help
keep the pressure on the exhalation valve constant. At high airway pressures,
this can cause a slight hiss during inspiration.
5-16 M1076286
5 Specifications and Theory of Operation
3
TI TE
AB.91.038
4
1. Paw
2. Pmax
3. PEEP
4. Time
Figure 5-3 Volume control diagram
Volume control supplies a set tidal volume. The ventilator calculates a flow
based on the set tidal volume and the length of the inspiratory time (T I) to
deliver that tidal volume. It then adjusts that output by measuring delivered
volumes at the flow sensors. Since the ventilator adjusts output, it can
compensate for breathing system compliance, fresh gas flow, and moderate
breathing system leaks.
A typical volume-controlled pressure waveform increases throughout the
entire inspiratory period, and rapidly decreases at the start of expiration. An
optional inspiratory pause is available to improve gas distribution.
Volume control mode settings:
TV
RR
I:E
Tpause
PEEP
Pmax
M1076286 5-17
Aisys Carestation
AB.91.039
5
1. Paw
2. Pmax
3. Pinsp
4. PEEP
5. Time
Figure 5-4 Pressure control diagram
5-18 M1076286
5 Specifications and Theory of Operation
SIMV/PSV mode
3
1
4
2
5 TI
AB.91.040
6
1. Mandatory SIMV breath
2. Spontaneous pressure supported breath
3. Paw
4. Psupport
5. PEEP
6. Time
Figure 5-5 SIMV/PSV diagram
Synchronized Intermittent Mandatory Ventilation (SIMV) is a mode in which
periodic volume breaths are delivered to the patient at preset intervals (time-
triggered). Between the machine delivered breaths, the patient can breathe
spontaneously at the rate, tidal volume and timing that the patient desires.
At the specified time interval, the ventilator will wait for the next inspiratory
effort from the patient. The sensitivity of this effort is adjusted using the flow
trigger level. When the ventilator senses the beginning of inspiration it
synchronously delivers a volume breath using the set tidal volume, and
inspiratory time that is set on the ventilator. If the patient fails to make an
inspiratory effort during the trigger window time interval, the ventilator will
deliver a machine breath to the patient. The ventilator will always deliver the
specific number of breaths per minute that the clinician has set.
In SIMV, the spontaneous breaths can be pressure supported to assist the
patient in overcoming the resistance of the patient circuit and the artificial
airway. When the Psupport level is set, the ventilator will deliver the pressure
support level to the patient during inspiration. PEEP can also be used in
combination with this mode.
Spontaneous breaths that occur during this mode are indicated by a color
change in the waveform.
SIMV/PSV mode settings:
TV
RR
Tinsp
Tpause
Psupport
PEEP
Pmax
Trig. Window
Flow Trigger
End of Breath
Rise Rate
M1076286 5-19
Aisys Carestation
PSVPro mode
1
AB.91.041
TI
3
1. Paw
2. PEEP
3. Time
Figure 5-6 PSVPro diagram
5-20 M1076286
5 Specifications and Theory of Operation
1
4
2
5
AB.91.042
TI
6
1. Mandatory pressure control breath
2. Spontaneous pressure supported breath
3. Paw
4. Psupport
5. PEEP
6. Time
Figure 5-7 Backup mode SIMV-PC+PSV
When the ventilator switches to the backup mode, the alarm text Backup
Mode Active will be displayed and will remain in the low priority message site
until PSVPro is reinstated or until another mode is selected. To reactivate the
PSVPro mode the user must go into the Ventilation Mode menu and select
PSVPro. Upon selecting PSVPro the ventilator will immediately begin providing
pressure supported breaths to the patient using the established settings.
Spontaneous breaths that occur during this mode are indicated by a color
change in the waveform.
PSVPro mode settings:
Psupport
PEEP
Trig. Window
Flow Trigger
End of Breath
Pmax
Backup Time
Pinsp
RR
Tinsp
Rise Rate
M1076286 5-21
Aisys Carestation
5-22 M1076286
5 Specifications and Theory of Operation
PCV-VG mode In PCV-VG, a tidal volume is set and the ventilator delivers that volume using a
decelerating flow and a constant pressure. The ventilator will adjust the
inspiratory pressure needed to deliver the set tidal volume breath-by-breath
so that the lowest pressure is used. The pressure range that the ventilator will
use is between the PEEP + 2 cmH2O level on the low end and 5 cmH2O below
Pmax on the high end. The inspiratory pressure change between breaths is a
maximum of 3 cmH2O.
This mode will deliver breaths with the efficiency of pressure controlled
ventilation, yet still compensate for changes in the patients lung
characteristics. PCV-VG begins by first delivering a volume breath at the set
tidal volume. The patients compliance is determined from this volume breath
and the inspiratory pressure level is then established for the next PCV-VG
breath.
PCV-VG mode settings:
TV
RR
I:E
PEEP
Pmax
Rise Rate
2 3
1
5
7
6
AB.98.034
1. Paw waveform
2. Tinsp
3. Texp
4. Variable pressure to deliver desired TV
5. PEEP
6. Flow waveform
7. TV
Figure 5-8 PCV-VG waveforms
M1076286 5-23
Aisys Carestation
Pneumatics
Gas source Anesthesia system
Gas composition Medical air or O2
Nominal supply pressure 350 kPa (50 psi)
Pressure range at inlet 240 to 700 kPa (35 to 101 psi)
Peak gas flow 120 l/min at 240 kPa (35 psi), 0.75 seconds
Continuous gas flow 80 l/min at 240 kPa (35 psi)
Flow valve range 1 to 120 l/min at 240 kPa (35 psi)
Fresh gas
compensation Flow compensation range 200 ml/min to 15 l/min
Gas composition O2, N2O, Air, anesthetic agents
Pressure
Patient airway pressure range -20 to +120 cmH2O, +/-1 cmH2O resolution
High pressure alarm set range 12 to 100 cmH2O, 1 cm increment
Sustained pressure alarm range 6 to 30 cmH2O, 1 cm increment
Display range -20 to 120 cmH2O
5-24 M1076286
5 Specifications and Theory of Operation
Volume
Tidal volume display range 0 to 9999 ml, 1 ml resolution
Setting range 20 to 1500 ml
Minute volume 0.0 to 99.9 liters, 0.1 liter resolution
Breath rate (breaths per minute) 4 to 100 bpm (non-spontaneous)
2 to 60 bpm (spontaneous)
1 bpm resolution
Volume sensor type Variable flow orifice
Oxygen
Display range 0 to 100% O2
Display resolution 1% increments
Sensor type Galvanic fuel cell
Measurement range 0 to 100% O2
Measurement accuracy Better than 3% of full scale
Cell response time 35 seconds
Note: Response time of cell and adapter is
measured using the test method described in
ISO 7767 (1997).
Low O2 alarm range 18% to 99%
High O2 alarm setting 19% to 100% or Off
Note: Low O2 limit may not be set above the high
O2 limit, nor may the high O2 limit be set below
the low O2 limit.
Expected cell life Four months of shelf life (23C room air) and one
year of normal operation.
M1076286 5-25
Aisys Carestation
Volume delivery accuracy > 210 ml tidal volume - accuracy better than 7%
< 210 ml but > 60 ml tidal volume -
accuracy better than 15 ml
< 60 ml tidal volume -
accuracy better than 10 ml
Volume monitoring accuracy > 210 ml tidal volume - accuracy better than 9%
< 210 ml but > 60 ml tidal volume -
accuracy better than 18 ml
< 60 ml tidal volume -
accuracy better than 10 ml
Inspiratory pressure delivery greater of 10% or 3 cmH2O
accuracy
PEEP delivery accuracy 1.5 cmH2O
Pressure monitoring accuracy greater of 5% or 2 cmH2O
Note: Gas composition errors may be in addition to the above normalized accuracy.
When adding errors, positive errors can have the effect of nulling out negative
errors.
Note: Use of anesthetic agent could affect the errors by approximately -0.95%/%
volume agent in normal mode.
% O2
20 30 40 50 60 70 80 90 100
0
10 > +10%
20
30
> 5%
% N2O 40
50
60
> -10%
70
AB.74.027
80
5-26 M1076286
5 Specifications and Theory of Operation
M1076286 5-27
Aisys Carestation
Aladin2 cassettes
Filling
Filling system Easy-Fil: Adapter filler system for Halothane, Isoflurane,
Enflurane, and Sevoflurane.
Quick-Fil: Filler system compatible to Abbotts system for
Sevoflurane.
Saf-T-Fil: Filler system compatible to the Datex-Ohmeda
Saf-T-Fil bottle for Desflurane.
Filling speed > 2 ml/s
Overfilling protection Overfilling prevention systems built into the cassettes.
Cassette
Empty weight 2.8 kg
Height 7 cm
Depth 24 cm
Width 14 cm
5-28 M1076286
5 Specifications and Theory of Operation
Aladin cassettes
Filling
Filling system Adapter filling: Rectangular keyed adapter filler system
for Halothane, Isoflurane, Enflurane, and Sevoflurane.
Quick-Fil: Filler system compatible to Abbotts system for
Sevoflurane.
Cylindrical keyed adapter: Filler system compatible to the
Datex-Ohmeda Saf-T-Fil bottle for Desflurane.
Filling speed > 2 ml/s
Overfilling protection Overfilling prevention systems built into the cassettes.
Liquid capacity
Maximum 250 ml
Normal fill when indicator shows empty 150 ml
(residual volume 100 ml)
Cassette
Empty weight 2 kg: Enflurane, Isoflurane, Sevoflurane with keyed
filler.
2.5 kg: Halothane with keypad filler, Sevoflurane with
Quik Fil.
3 kg: Desflurane.
Height 7 cm
Depth 23 cm
Width 14 cm
16 cm with keyed filler
M1076286 5-29
Aisys Carestation
Guidance and The system is suitable for use in the specified electromagnetic environment.
manufacturer's The customer and/or the user of the system should assure that it is used in an
electromagnetic environment as described below.
declaration -
electromagnetic
emissions
Emissions test Compliance Electromagnetic environment guidance
RF emissions Group 1 The system uses RF energy only for its internal function. Therefore, its RF emissions are
CISPR 11 very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions Class B The system is suitable for use in all establishments, including domestic
CISPR 11 establishments and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic emissions Class A
IEC 61000-3-2 Class A
Voltage fluctuations/ Complies
flicker emissions
IEC 61000-3-3
5-30 M1076286
5 Specifications and Theory of Operation
Guidance and The system is suitable for use in the specified electromagnetic environment.
manufacturer's The customer and/or the user of the system should assure that it is used in an
electromagnetic environment as described below.
declaration -
electromagnetic
immunity
Power immunity
Immunity test IEC 60601-1-2 test level Compliance level Electromagnetic environment guidance
M1076286 5-31
Aisys Carestation
Radiated immunity
5-32 M1076286
5 Specifications and Theory of Operation
Recommended The system is intended for use in the electromagnetic environment in which
separation distances radiated RF disturbances are controlled. The customer or the user of the
system can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications
equipment (transmitters) and the system as recommended below, according
to the maximum power of the communications equipment.
Note 1: At 80 MHz to 800 MHz the separation distance for the higher frequency range applies.
Note 2: The ISM (Industrial, Scientific and Medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553
MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
Note 3: An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the ISM
frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz to decrease the likelihood that
mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas.
Note 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects, and people.
M1076286 5-33
Aisys Carestation
Electrical safety
The system provides connections for items such as printers, visual displays
and hospital information networks. When these items (non-medical
equipment) are combined with the system, these precautions must be
followed:
Dont place items not approved to IEC 60601-1 closer than 1.5 m to the
patient.
All items (medical electrical equipment or non-medical electrical
equipment) connected to the system by a signal input/signal output
cable must be supplied from an AC power source which uses a separating
transformer (in accordance with IEC 60989) or be provided with an
additional protective earth conductor.
If a portable multiple socket outlet assembly is used as an AC power
source, it must comply with IEC 60601-1. The assembly must not be
placed on the floor. Using more than one portable multiple socket outlet
assembly is not recommended. Using an extension cord is not
recommended.
Do not connect non-medical electrical equipment directly to the AC outlet at
the wall instead of an AC power source which uses a separating transformer.
Doing so may increase enclosure leakage current above levels allowed by IEC
60601-1 in normal conditions and under single-fault conditions. This may
cause an unsafe electrical shock to the patient or operator.
After connecting anything to these outlets, conduct a complete system
leakage current test (according to IEC 60601-1).
IEC-60601-1 Classification
The system is classified as follows.
Class I equipment.
Type B equipment.
Type BF equipment.
Ordinary equipment.
Not for use with flammable anesthetics.
Continuous operation.
5-34 M1076286
6 Super User Mode
M1076286 6-1
Aisys Carestation
Install/Service menu
Important Changes made in the super user mode will affect the system configuration. All
changes made are permanent and preserved until changed again.
Several settings can be changed in the super user mode. These changes
should only be made by the person responsible for the configuration of the
system.
The Installation and Service menus are password protected.
The system must be turned off in order to exit the Install/Service menu. Go to
Main Menu - Screen Setup - Install Service. Enter the password16, 4, 34.
Install/Service - Page 1
Trends Setup
Colors and Units
Cumulative Gas Usage
Show Alarm Limits Yes Yes or No
External Gas Monitor Yes Yes or No
Save Default Case
Factory Defaults
Parameter Settings
Install/Service - Page 2
Exit
Install/Service - Page 2
Fresh Gas Controls O2% O2%, Flow, User
VCV Cardiac Bypass No Yes or No
Installation
Service
Install/Service - Page 1
Exit
6-2 M1076286
6 Super User Mode
Using super user mode Set the Default Trend and configure the graphic trend pages through the
Trends Setup menu.
Set the colors and the units of parameters through the Colors and Units
menu.
View the gas used and reset the gas usage in the Cumulative Gas Usage
menu. Select Reset Usage to set all cumulative agent totals to 0.0, all
cumulate gas usage to 0, and to set the displayed date to the current date in
the inactive right column.
Set Show Alarm Limits to Yes display the alarm limits next to the measured
values on the normal screen.
Set the External Gas Monitor to Yes only if the system uses a stand-alone
monitor for O2, agent, and CO2. Selecting Yes will disable the O2 limit alarms
and the O2 monitoring not connected alarm when there is no O 2 cell or
airway module detected in the system. The AA, CO2 monitoring not
connected alarm will also be disabled.
Select Save Default Case to save the air, N2O, circuit, ventilator settings, and
alarms settings from the last case as defaults.
Factory defaults are the settings installed by the manufacturer. These cannot
be changed. To return the system to the factory defaults, select Factory
Defaults.
Set the volume conditions and the CO2 humidity compensation through the
Parameters Setup menu. Set the volume calculation conditions to be based
on ATPD (ambient temperature and pressure, dry humidity conditions) or
BTPS (body temperature, ambient pressure, saturated humidity condition).
Set the humidity compensation type in CO2 partial pressure values to wet or
dry.
M1076286 6-3
Aisys Carestation
Menus
The following is a list of the menu selections available.
Not all menu items are available depending on the system configuration.
Inactive menu items are grayed out and are not selectable.
Menu selections shown are the factory default values. The options are listed to
the right of the menu shown.
Trends Setup
Default Trend Num Num, Graph, Set
Graphical Trends
Previous Menu
Graphical Trends
Page 1
Page 2
Page 3
Page 4
Page 5
Previous Menu
Page 1
Field 1 Pres * Off, Pres, MVexp, CO2, O2, Bal, AA1, AA2,
Field 2 TVexp MAC, N2O, TVexp, Spont, Pmean, rr+co2,
Field 3 CO2 Compl, VO2
Previous Menu
*The Field 1, Field 2, and Field 3 selections are the same for all fields.
6-4 M1076286
6 Super User Mode
Page 2
Field 1 O2 * Off, Pres, MVexp, CO2, O2, Bal, AA1, AA2,
Field 2 N2O MAC, N2O, TVexp, Spont, Pmean, rr+co2,
Field 3 AA1 Compl, VO2
Previous Menu
Page 3
Field 1 AA2 * Off, Pres, MVexp, CO2, O2, Bal, AA1, AA2,
Field 2 N2O MAC, N2O, TVexp, Spont, Pmean, rr+co2,
Field 3 MAC Compl, VO2
Previous Menu
Page 4
Field 1 Bal * Off, Pres, MVexp, CO2, O2, Bal, AA1, AA2,
Field 2 Mac MAC, N2O, TVexp, Spont, Pmean, rr+co2,
Field 3 MVexp Compl, VO2
Previous Menu
Page 5
Field 1 rr+co2 * Off, Pres, MVexp, CO2, O2, Bal, AA1, AA2,
Field 2 Compl MAC, N2O, TVexp, Spont, Pmean, rr+co2,
Field 3 Off Compl, VO2
Previous Menu
*The Field 1, Field 2, and Field 3 selections are the same for all fields.
M1076286 6-5
Aisys Carestation
Colors
Paw Yellow Yellow, White, Green, Red or Blue
Flow Green Yellow, White, Green, Red or Blue
Resp White Yellow, White, Green, Red or Blue
CO2 White Yellow, White, Green, Red or Blue
Previous Menu
Parameter Settings
TV Based on ATPD ATPD or BTPS
CO2 Numbers Dry Dry or Wet
Previous Menu
6-6 M1076286
Warranty
This Product is sold by Datex-Ohmeda under the warranties set forth in the
following paragraphs. Such warranties are extended only with respect to the
purchase of this Product directly from Datex-Ohmeda or Datex-Ohmedas
Authorized Dealers as new merchandise and are extended to the Buyer
thereof, other than for the purpose of resale.
For a period of twelve (12) months from the date of original delivery to Buyer or
to Buyers order, but in no event for a period of more than two years from the
date of original delivery by Datex-Ohmeda to a Datex-Ohmeda Authorized
Dealer, this Product, other than its expendable parts, is warranted against
functional defects in materials and workmanship and to conform to the
description of the Product contained in this Users Reference manual and
accompanying labels and/or inserts, provided that the same is properly
operated under the conditions of normal use, that regular periodic
maintenance and service is performed and that replacements and repairs are
made in accordance with the instructions provided. This same warranty is
made for a period of thirty (30) days with respect to expendable parts. The
foregoing warranties shall not apply if the Product has been repaired other
than by Datex-Ohmeda or in accordance with written instructions provided by
Datex-Ohmeda, or altered by anyone other than Datex-Ohmeda, or if the
Product has been subject to abuse, misuse, negligence, or accident.
Datex-Ohmedas sole and exclusive obligation and Buyers sole and exclusive
remedy under the above warranties is limited to repairing or replacing, free of
charge, at Datex-Ohmedas option, a Product, which is telephonically
reported to the nearest Datex-Ohmeda Customer Service Center and which, if
so advised by Datex-Ohmeda, is thereafter returned with a statement of the
observed deficiency, not later than seven (7) days after the expiration date of
the applicable warranty, to the Datex-Ohmeda Customer Service and
Distribution Center during normal business hours, transportation charges
prepaid, and which, upon Datex-Ohmedas examination, is found not to
conform with above warranties. Datex-Ohmeda shall not be otherwise liable
for any damages including but not limited to incidental damages,
consequential damages, or special damages.
There are no express or implied warranties which extend beyond the
warranties hereinabove set forth. Datex-Ohmeda makes no warranty of
merchantability or fitness for a particular purpose with respect to the product
or parts thereof.
Index
A E
Absorber canister 2-30, 4-5 Electromagnetic compatibility 5-30
cleaning 2-30 EZchange Canister
disinfection 2-30 removal 1-5
filling 1-6
removing 1-5 F
setup 1-3
AGSS Flow sensors 2-3, 4-2
remove receiver 2-27 cleaning and disinfection 2-5
remove receiver filter 2-29 prevent water buildup 3-6
B I
Backlight test 3-6 Install/Service
Bellows assembly menu 6-2
assemble 2-17 password 6-2
disassemble 2-15
test 2-19 M
Breathing system Mains inlet 1-8
assembly 2-21 Maintenance schedule 3-2
cleanable parts 2-3
disassembly 2-9 O
installing 2-26
removal 2-8 O2 cell
remove bag hose 2-8 calibration 3-5
specifications 5-11 replacement 3-4
C P
Calibration Pneumatic
100% O2 3-5 connections 1-10
21% O2 3-5 specifications 5-5
backlight test 3-6 system pneumatic circuits 5-2
O2 cell 3-5
Circuit R
non-circle relief 5-6
Repair policy 3-2
Cleaning
cautions 2-2
Condenser cleaning 2-31
S
Connections Scavenging 1-10
electrical 1-8 specifications 5-13
pneumatic 1-10 Serial port 1-9
Cylinder
installation 1-13
M1076286 I-1
Aisys Carestation
Specifications
airway module 5-14
breathing system 5-11
flow 5-10
gas 5-14
physical 5-13
pneumatic 5-5
ventilation operating 5-24
Suction regulator 1-12
Super User mode 6-1
V
Ventilation modes 5-17
pressure control 5-18
PSVPro 5-20
SIMV/PSV 5-19
SIMV-PC 5-22
volume control 5-17
Ventilator
accuracy data 5-26
modes 5-17
operating specifications 5-24
theory 5-16
W
Water buildup
prevention 3-6
I-2 M1076286
Aisys
Users Reference Manual
English
M1076286
05 06 18 05 02
Printed in USA
Datex-Ohmeda, Inc.
All rights reserved