Agilent 3458A Multimeter: Data Sheet
Agilent 3458A Multimeter: Data Sheet
Agilent 3458A Multimeter: Data Sheet
Data Sheet
Shattering
performance
barriers
of speed
and
accuracy
Performance Highlights
dc volts
5 ranges: 0.1 V to 1000 V
8.5 to 4.5 digit resolution
Up to 100,000 readings/sec
(4.5 digits)
Maximum sensitivity: 10 nV
0.6 ppm 24 hour accuracy
8 ppm (4 ppm optional)/year
voltage reference stability
Ohms
9 ranges: 10 to 1 G
Two-wire and four-wire ohms with
offset compensation
Up to 50,000 readings/sec
(5.5 digits)
Maximum sensitivity: 10
2.2 ppm 24 hour accuracy
ac volts
6 ranges: 10 mV to 1000 V
1 Hz to 10 MHz bandwidth
Up to 50 readings/sec with all
readings to specified accuracy
Choice of sampling or analog true
rms techniques
100 ppm best accuracy
dc current
8 ranges: 100 nA to 1 A
Up to 1,350 readings/sec (5.5 digits)
Maximum sensitivity: 1 pA
14 ppm 24 hour accuracy
ac current
5 ranges: 100 A to 1 A
10 Hz to 100 kHz bandwidth
Up to 50 readings/sec
500 ppm 24 hour accuracy
Maximum speeds
100,000 readings/sec at 4.5 digits
(16 bits)
50,000 readings/sec at 5.5 digits
6,000 readings/sec at 6.5 digits
60 readings/sec at 7.5 digits
6 readings/sec at 8.5 digits
Measurement set-up speed
100,000 readings/sec over GPIB or
with internal memory
110 autoranges/sec
340 function or range changes/sec
Post-processed math from internal
memory
Numeric/user keys
Numeric entry for constants and
measurement parameters
Shifted keys (f0 through f9) access
up to ten user-defined setups
Volts/ohms/ratio terminals
Gold-plated tellurium copper for
minimum thermal emf
2-wire or 4-wire ohms measurements
dc/dc or ac/dc ratio inputs
External output:
Programmable TTL output pulse with
5 modes for flexible system interface
Defaults to a voltmeter complete pulse
Rear input terminals
for convenient
system use
Finally!
A system multimeter with BOTH
high speed and high accuracy.
Contents
Test System Throughput / 6
Calibration Lab Precision / 7
High Resolution Digitizing /8
Technical Specifications / 9
Section 1: DC Voltage / 10
Section 2: Resistance / 11
Section 3: DC Current / 13
Section 4: AC Voltage / 14
Section 5: AC Current / 19
Section 6: Frequency/Period / 20
Section 7: Digitizing / 21
Section 8: System Specifications / 23
Section 9: Ratio / 24
Section 10: Math Functions / 24
Section 11: General Specifications / 25
Section 12: Ordering Information / 26
Faster testing
Up to 100,000 readings/sec
Internal test setups > 340/sec
Programmable integration times from
500 ns to 1 sec
Greater test yield
More accuracy for tighter test margins
Up to 8.5 digits resolution
Longer up-time
Two-source (10 V,10 k) calibration,
including ac
Self-adjusting, self-verifying autocalibration for all functions and ranges,
including ac
Easy calibration
The 3458A gives you low cost of
ownership with a simple, two-source
electronic calibration. With its superior
linearity, the 3458A is fully calibrated,
including ac, from a precision 10 Vdc
source and a precision 10 k resistor.
All ranges and functions are automatically calibrated using precise internal
ratio transfer measurements relative
to these external standards. In addition, the 3458As internal voltage
standard and resistance standard are
calibrated. Now you can perform a
self-verifying, self- or auto-calibration
relative to the 3458As low drift internal standards at any time with the
ACAL command. So, if your dmms
environment changes, auto-calibration
optimizes your measurement accuracy.
Calibration security
Unlike other dmms, the 3458A goes
to great lengths to assure calibration
security. First, a password security
code locks calibration values and
the self-calibration function. Next, you
can easily store and recall a secured
message for noting items, such as
calibration date and due date. Plus,
the 3458A automatically increments
a calibration counter each time you
unlock the dmm another safeguard against calibration tampering.
If you have a unique situation or
desire ultimate security, use the internal dmm hardwired switch to force
removal of the instrument covers to
perform calibration.
7
Section 1: DC Voltage
10
DC Voltage continued
Section 2: Resistance
11
Resistance continued
12
Section 3: DC Current
13
Section 4: AC Voltage
General Information
14
AC Voltage continued
15
AC Voltage continued
Analog Mode
16
AC Voltage continued
17
AC Voltage continued
18
Section 5: AC Current
19
AC Current continued
Section 6: Frequency/Period
20
Section 7: Digitizing
(DCV Function)
21
Digitizing continued
22
23
Section 9: Ratio
24
Operating environment
0 C to 55 C
Operating humidity range
Up to 95% RH at 40 C
Physical characteristics
88.9 mm H x 425.5 mm W x 502.9 mm D
Net weight: 12 kg (26.5 lbs)
Shipping weight: 14.8 kg (32.5 lbs)
IEEE-4888 interface
Complies with the following:
IEEE-488.1 Interface Standard
IEEE-728 Codes/Formats Standard
HPML (Multimeter Language)
Storage temperature
40 C to +75 C
Warm-up time
4 hours to published specifications
Power requirements
100/120 V, 220/240 V 10%
48-66 Hz, 360-420 Hz automatically
sensed
< 30 W, < 80 VA (peak)
Fused: 1.5 @ 115 V or 0.5 A @ 230 V
25
Accessories
10833A
10833B
10833C
10833D
GPIB cable (1 m)
GPIB cable (2 m)
GPIB cable (4 m)
GPIB cable (0.5 m)
11059A
11060A
11062A
34137A
34308A
34330A
E2308A
26
www.agilent.com
www.agilent.com/find/multimeters
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
www.axiestandard.org
AdvancedTCA Extensions for
Instrumentation and Test (AXIe) is
an open standard that extends the
AdvancedTCA for general purpose
and semiconductor test. Agilent
is a founding member of the AXIe
consortium.
www.lxistandard.org
LAN eXtensions for Instruments
puts the power of Ethernet and
the Web inside your test systems.
Agilent is a founding member of the
LXI consortium.
www.pxisa.org
PCI eXtensions for Instrumentation
(PXI) modular instrumentation
delivers a rugged, PC-based highperformance measurement and
automation system.
www.agilent.com/quality
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries