GA1000 Series User Manual v1
GA1000 Series User Manual v1
GA1000 Series User Manual v1
Copyright
2012 NANJING GLARUN-ATTEN TECHNOLOGY CO. LTD All Rights Reserved.
Trademark Logo
is the registered trademark of NANJING GLARUN-ATTEN
TECHNOLOGY CO. LTD.
Declaration
Warning
The warning announcement points out the operation or condition that
may endanger the operators.
Notice
The notice announcement points out the operation or condition that
may cause damage to the product or other properties.
Terms on the product. The following terms may appear on the product:
Danger
Warning
Notice
Characters on the product. The following characters may appear on the product
Notice
Please read
the manual
Protective
Chassis
Measuring
ground terminal ground terminal ground terminal
Model
Bandwidth
Sampling Rate
GA1102CAL
100MHz
1GSa/s
GA1202CAL
200MHz
2GSa/s
GA1302CAL
300MHz
2GSa/s
Characteristics
The oscilloscope has a totally new ultrathin appearance design, and is small in
size and more portable
A 7-inch widescreen color TFT LCD displays clear, crisp and more stable
waveform display
Pass/Fail function
The pop-up menu display mode realizes more flexible and more natural for
users operations
USB Host: support storage of USB flash disc and upgrading of USB flash disc
system software
User's manual
Certificate of approval
USB cable
Content Summary
The manual introduces related information about the operation of this series of digital
oscilloscope and comprises the following chapters and sections:
Chapter Introduction introduces the front panel, the user interface, the function
check and the probe compensation of the oscilloscope.
Chapter System Prompt and Fault Recovery introduces the system prompts
and lists some simple faults and solutions so that the users are capable to
rectify simple faults.
Chapter Service and Support introduces the warranty and technical support of
this series of product.
Catalogue
Chapter 1 Introduction.1
1.1
1.2
Function check5
1.3
Probe6
1.3.1 Probe safety6
1.3.2 Probe attenuator setting...6
1.3.3 Probe compensation..7
2.2
Connector.11
2.3
Default setting.12
2.4
[Universal] knob.13
2.5
Vertical system14
2.5.1 Channels CH1 and CH2.14
2.5.2 Application of [POSITION] and [VOLT/DIV] knobs of the vertical system
.20
2.5.3 Implementation of MATH function20
2.5.4 Implementation of REF function..27
Chapter 6 Appendix96
Appendix A: Technical specification96
Appendix B: Default setting101
Introduction
DIGITAL STORAGE OSCILLOSCOPE
Function Keys
Print Screen
Universal Knob
Menu Controls
Auto Setup
Advanced Trigger
ControlsHorizontal Controls
Vertical Controls
Power Button
USB Port
Menu On/Off
Input Channel
for Analog Signal
External
Trigger Terminal
? ?
No.
Details
Trigger status
Armed: The oscilloscope is acquiring pre-triggering data. All triggers are ignored
in this state.
Ready: The oscilloscope has acquired all the pre-triggering data and is ready to
accept a trigger.
Trigd: The oscilloscope has caught a single trigger and acquires the data after
triggering.
Stop:
Auto:
Scan:
1.
2.
3.
4.
1.3 Probe
1.3.1 Probe safety
Check and obey the rated values of the probe parameters before using it.
A protective device surrounding the probe main body can prevent fingers from electric
shock.
Probe protective
device
Keep fingers behind the protective curve on the probe main body to prevent electric shock
when using the probe.
Do not contact the metal part on the top of the probe when the probe is connected to a voltage
source.
The signal measured by the oscilloscope uses ground as reference voltage, and the ground
terminal should be grounded correctly to prevent short circuit.
When the ATTENUATION switch is set as 1, the probe limits the bandwidth of the
oscilloscope within 0-10 MHz (different probes have different specifications).
Ensure that the switch is set to be 10when to use the full bandwidth of the
oscilloscope.
under compensated
suitable compensated
over compensated
4 If necessary, rotate the adjustable capacitor on the probe handle to realize suitable
compensated.
Function Introduction
and Operation
DIGITAL STORAGE OSCILLOSCOPE
Connector
Automatic setup
Default setup
Universal knob
Vertical system
Horizontal system
Triggering system
Display system
Measurement system
Save system
Auxiliary system
On-line help system
[SET TO 50%]: set the trigger electric level as midpoint of the signal amplitude.
[CURSORS]: display the CURSOR menu. The [UNIVERSAL] knob can be used for
regulating the position of the cursor when the CURSOR menu is displayed and
the cursor is triggered.
10
2.2 Connector
EXT TRIG: be used as an input connector of an external trigger source. Use [TRIG
MENU] to select EXT or EXT/5 trigger source, and the trigger signal source
can be used for triggering in the third channel while acquiring data in two
channels.
Probe compensation: The probe compensation signal is output and grounded so
This product is grounded by a protective ground wire of a power cord. For avoiding electric
shock, please ensure that the product is reliably grounded before connecting the input end
or output end of the product.
The ground wire of the probe is connected to the ground only. Please do not connect the
ground wire to high voltage.
11
Default setup
12
Universal knob
13
Volt/div knob
Description
Not only pass through the AC component of the input
DC
Coupling
AC
GND
Bandwidth
RUN
limit
STOP
14
Volt/div
Coarse tuning
Fine tuning
Probe
1X
10 X
100 X
vertical reading.
1000 X
Next
Table 2-4
Page1/2
Option
Inverse
setup
Description
Run
Stop
Digital filter
Back
Table 2-5
in table 2-5).
Page2/2
Option
Digital filter
Setup
Description
Run
Stop
Filter type
Filter Upper
limit
knob
Filter lower
15
limit
knob
Back
If the channel adopts a DC coupling mode, you can quickly measure the DC component of
the signal by observing the difference between the waveform and the signal ground.
If the channel adopts an AC coupling mode, the DC component in the signal is filtered. By
this mode, the AC component of the signal is displayed at a higher sensitivity.
If the channel adopts a GND coupling mode, cut off the input signal. Inside the channel, the
channel input is connected with a zero volt reference electric level.
AC identification
Figure 2-6
Set AC
DC identification
Figure 2-7
Set DC
16
GND identification
Press [CH1]Bandwidth limitOFF, and set the bandwidth limit as OFF state.
The amplitude of the high-frequency component contained in the measured
signal is unlimited, as shown in figure 2-10.
17
18
Probe coefficient
19
20
Description
+,-,
,FFT
source 2
Regulate the vertical gear of the MATH waveform by
the [UNIVERSAL] knob.
Regulate the vertical offset of the MATH waveform by
the [UNIVERSAL] knob.
Table 2-7
Operation
CH1-CH2
CH2-CH1
CH1CH2
CH1CH2
CH2CH1
FFT
Description
Add the waveform of signal source 1 with the
waveform of signal source 2
Subtract the waveform of signal source 2 from the
waveform of signal source 1
Subtract the waveform of signal source 1 from the
waveform of signal source 2
Multiply the waveform of signal source 1 with the
waveform of signal source 2
Divide the waveform of signal source 1 by the
waveform of signal source 2
Divide the waveform of signal source 2 by the
waveform of signal source 1
Fast Fourier transform operation
21
Waveform operation
Result identification
Analyze vibration.
Table 2-8
FFT Option
Signal source
Setup
CH1
CH2
Description
Select CH1or CH2 as signal source
Hanning
Window
Hamming
Rectang
Blackman
1
Horizontal
amplification
10
22
Next
Table 2-9
Page 1/2
FFT Option
Setup
Description
Vertical scale
Vrms
dBVrms
Back
Page 2/2
FFT Window
As the oscilloscope is used for FFT transformation on waveform recording with finite
length, the FFT algorithm is based on periodic signal. When the waveform period
number within the finite length is an integer, YT waveform has the same amplitude at
the start point and the end point, and no interruption is caused to the waveform. But
when the period number is not an integer, the waveform has different amplitude at the
start point and the end point, so that high-frequency transient interruption will be
caused in the joint. In a frequency domain, this effect is named as leakage. Therefore,
to avoid appearance of leakage, the original waveform is multiplied with a window
function, and forcibly make the values at the start point and the end point be 0.
Determine to use what kind of window according to the item and source signal
characteristics to be measured.
Table 2-10
Window
Rectang
Worst amplitude
23
resolution
equal
Be equivalent to no use of
window
there-before
and
there-after;
Better frequency
resolution and worse
Hanning
amplitude resolution in
noise.
comparison with
rectangular window
Frequency of window
Hamming
Hanning
Blackman
subharmonic
Access the signal to CH1 or CH2, and press [AUTO] to display a YT waveform.
According to Nyquist law, revolve [s/div] to make the sampling rate at least two
times the frequency of the input signal.
Press [MATH].
24
Regulate spectrum amplitude gear and offset at page 2 of the FFT function menu
so as to better observe FFT operation result.
Frequency interval
Window function
25
of spectrum amplitude
of spectrum frequency
26
A signal with DC component or offset will cause error or offset of an FFT waveform
component. An AC coupling mode can be selected to reduce DC components.
DBVrms vertical scale can be used for displaying the FFT waveform within a large dynamic
range, and this scale displays the vertical amplitude at a logarithm mode.
Nyquist Function: Reconstruction of the waveform needs to adopt a sampling rate that is
double of the input frequency.
Setup
Description
CH1
CH2
REF A
REF B
waveform
Save the signal source waveform in a selected
SAVE
REF A/REF B
reference position
RUN
STOP
Press [REF] to display the reference waveform menu, as shown in figure 2-22:
27
Operation procedure:
1. Press [REF] to display the menu REF WAV.
2. Select the Signal source as CH1or CH2.
3. Revolve the vertical [POSITION] and knob [Volt/div] to regulate the vertical position
and the gear of the reference waveform.
4. Select REF A or REF B as storage location of the reference waveform.
5. Press Save to save the current screen waveform as a waveform reference.
6. Select REF A ON or REF B ON to recall the reference waveform, as shown in figure
2-23.
28
s/div
Figure 2-25
Table 2-12
Option
Main time base
Description
Horizontal time scale system of the oscilloscope
Define one window region by two cursors, and use the
Window setup
horizontal
[POSITION] and the knob [s/div] to regulate the window
29
region.
Expand the window setup region to the whole screen to
Window expansion
30
31
32
[FORCE]
No matter whether the oscilloscope detects the trigger, the key can be pressed to
perform force trigger once so as to finish acquisition of the current waveform. The key
is mainly applied to Normal and Single in the trigger modes.
Pre-trigger/delay trigger
The key is used for triggering the sampled data before/after the event.
The trigger position is generally set in the horizontal center of the screen, so that the
pre-trigger and delay information can be observed. The horizontal [POSITION] knob
can be further revolved to regulate horizontal displacement of the waveform so as to
check more pre-trigger information or delay trigger information. For instance, if burrs
generated by the circuit are caught, the causes of generation of the burrs may be found
out by observing and analyzing the pre-trigger data.
Attention: pre-trigger and delay trigger are invalid at a slow scan state.
2.7.2 Types
This series of oscilloscopes provide five trigger types: Edge, Pulse, Video, Slope and
Alternation.
1. Edge trigger: the most basic as well as the most common trigger type,
as shown in figure 2-29.
33
Setup
Type
Edge
Description
Edge trigger happens when a trigger input signal
CH1
CH2
EXT
Signal
source
EXT/5
AC Line
Slope
Trigger
modes
Rise edge
Fall edge
Auto
Normal
Single
Setup
Setup
Description
34
DC
AC
Coupling
High-frequenc
y rejection
150kHz.
Low-frequency
Reject
rejection
DC
components,
and
attenuate
Trigger
hold-off
time
Reset the trigger hold-off time to be the minimal
Reset
value 100ns
Back to the homepage of the trigger menu
Back
Operation steps:
Set the types
1) Press [TRIG MENU] to display the TRIGGER menu.
2) Press Type to select Edge trigger.
Set the signal sources
3) Press Signal source to select CH1, CH2, EXT, EXT/5 or AC Line according
to the signal input.
Set the slopes
4) Press Slope to select Rise edge, Fall edge or Rise and fall edge.
Set the trigger modes
5) Press Trigger mode to select Auto, Normal or Single.
Auto:
Normal: Refresh the waveform when the waveform satisfies the trigger
condition, otherwise, do not refresh the waveform and wait for
occurrence of the next trigger event.
Single:
Acquire the waveform once when the waveform satisfies the trigger
condition, and then stop.
35
Option
Type
Setup
Description
Trigger the pulse satisfying the
Pulse
CH1
Signal source
CH2
EXT
2.7.1.
EXT/5
Positive pulse width >
Positive pulse width <
Condition
value
Pulse width
setup
20.0ns ~ 10.0s
width
36
Table 2-16
Page 1/2
function menu
Option
Setup
Type
Pulse
Trigger mode
Description
Trigger the pulse satisfying the condition when
selecting pulse.
Auto
Normal
Single
width trigger.
Setup
Back
Operation description:
Set the type:
1) Press [TRIG MENU] to display TRIGGER menu.
2) Press Type to select Pulse.
Setup of the pulse trigger is similar with that of the edge trigger.
Set the condition:
3) Press the Condition option key to select
37
trigger mode and the trigger coupling like the setup of the edge trigger.
Video trigger: Perform field or row video trigger on a standard video signal.
Table 2-17
Option
Setup
Description
Trigger an NTSC or PAL standard video
Type
Video
CH2
EXT
EXT/5
Polarity
Standard
Inverse phase
All rows
Synchronization
Specified row
Odd field
Even field
Next
Page 1/2
function menu.
Table 2-18
Figure 2-32
38
Option
Setup
Description
Trigger an NTSC or PAL standard video signal
Type
Video
Standard
NTSC
PAL
row counting.
Auto
Trigger
Normal
mode
Single
Skip to the trigger setup menu (See table
Setup
2-14).
Back
Page 2/2
menu.
Operation description
As show in figures 2-31, 2-32:
Set the types:
1) Press [TRIG MENU] key to display TRIGGER menu.
2) Press Type to select Video.
Setup of the signal source of the video trigger is like that of the edge trigger.
Set the polarity:
3) Press Polarity to select
or
Slope trigger
Table 2-19
Option
Setup
Type
Slope
Description
CH1
Signal
CH2
source
EXT
EXT/5
Positive slope>
Positive slope<
Conditions
Positive slope=
Negative slope>
Negative slope<
Negative slope=
slope time.
Set time
Page 1/2
Option
Setup
Type
Slope
Description
Upper boundary
Vertical
Lower boundary
window
boundaries
40
Auto
Trigger mode
Normal
Single
Skip to the trigger setup menu (See table
Setup
Back
2-14).
Page 2/2
Operation description:
Carry out the following steps to select Slope trigger type:
1. Input a signal to channel 1 or channel 2.
2. Press [AUTO].
3. Press [TRIG MENU] to skip to TRIGGER menu.
4. Press Type to select Slope.
5. Press Signal source to select CH1 or CH2.
6. Press Condition to select slope condition.
7. Press Time setup, and revolve the [UNIVERSAL] knob to regulate the slope time.
8. Press Next Page1/2 to skip to Page 2 of the slope trigger menu.
9. Press Vertical window to select window boundary.
10. Revolve the trigger electric level [LEVEL] knob until the waveform is stably
triggered.
Alternating trigger
When alternating trigger is selected, the trigger signal comes from two channels. The
mode is used for simultaneously observing two unrelated signals. Two different trigger
types (edge, pulse, video and slope) can be selected for the signals from the two
channels. During setup, the trigger types of the two channels and the trigger electric
level information are respectively displayed in the right lower corner of the screen. The
setups of the four trigger types are seen in 2.7.2.
41
2.7.3 Coupling
The Coupling option is used to determine which part of signal passes through the
trigger circuit. It is conductive to stably display the waveform. Press [TRIG MENU] to
use trigger coupling. Select the Coupling option in the Setup menu after selecting
the trigger type, including DC coupling, AC coupling, HF rejection and LF rejection. The
functions of all the options are specifically shown in table 2-14.
42
Figure 2-36
43
Measure
Display
Default setup
Auxiliary
Help
Option
Setup
Description
Sampling
Acquisition
Peak value
mode
detection
Average value
Average time
( 4, 16, 32,
64 , 128 , 256 )
Interpolation
Sine
type
Linear
Real-time
Set
sampling
sampling
Sampling mode
Sampling rate
the
sampling
mode
as
real-time
Sampling: The oscilloscope samples the signal at uniform time intervals to form a
waveform.
Advantage: Signals can be accurately shown at the mode in many
cases.
44
Peak value detection: The oscilloscope finds out the maximal value and the
minimal value of the input signal in each sampling interval and uses these values
to display the waveform.
Advantage: spike pulses that may be missed can be acquired and displayed and
signal confusion can be avoided at the mode.
Shortcoming: Loud noise is displayed at the mode.
Average value: The oscilloscope acquires a plurality of waveforms and displays the
final waveforms after averaging the waveforms.
Advantage: Random or unrelated noises in the displayed signal can be reduced at
the mode. The signal shown in figure 2-40 has loud noises, while the signal shown
in figure 2-41 adopts the average mode, so the noises are greatly reduced. What
calls for attention is that: the higher the average time is, the better the waveform
quality is, but the slower the refreshing speeds of the waveform is.
45
Real-time sampling: The real-time sampling mode realizes suffusion of the storage
space during each sampling. The real-time sampling rate is at
most 1GSa/s.
[RUN/STOP]: Press the key to start continuous data acquisition or stop acquisition.
[SINGLE]: Press the key after the oscilloscope detects the trigger and stop after wave
acquisition is finished once.
When [RUN/STOP] or [SINGLE] is pressed to start acquisition, the oscilloscope
executes the following steps:
1 Acquire enough data to fill the waveform part before the waveform is triggered,
which is also named pre-trigger.
2 Continue to acquire data when waiting for trigger.
3 Detect the trigger conditions.
4 Continue to acquire data before the storage space is full.
5 Display the lately acquired waveform.
Time base: Use the [s/div] knob to regulate the time base. The time base is a
quantization unit of the time, namely the time represented by a large
horizontal grid (this series of oscilloscope has 14 horizontal grids in total) of
the oscilloscope. The time base is closely related with the sampling rate. The
larger the time base is, and the smaller the sampling rate is.
Fake wave phenomenon: Fake wave phenomenon will appear if the oscilloscope does
not sample the signal quickly enough and does not make accurate
waveform recording, as shown in figure 2-42. To eliminate this type of
phenomenon, switch to a quick time base gear or adopt the peak value
detection mode.
Figure
46
2-42
Figure 2-44
47
Linear interpolation
Type
Description
Vector
Point
OFF
1s
Persist
2s
5s
Infinite
Set the waveform brightness by the [UNIVERSAL]
Intensity
knob.
Grid
brightness
Next
Page 1/2
48
Table 2-30
Option
Setup
Description
YT
Format
XY
Screen
Normal
Inverse
Grid
10s
20s
Infinite
Back
Page 2/2
Operation description:
Set the waveform display types:
Press [DISPLAY] to skip to the DISPLAY menu, and press Type to select Vector or
Point.
Set continue:
Press Continue to select OFF, 1s, 2s, 5s or Infinite. By utilizing the option,
some special waveforms can be observed, as shown in figure 2-47.
Figure 2-47 Utilize the continue option to observe amplitude modulated wave
49
or
50
Figure 2-48
Y-T mode
Figure 2-49
X-Y mode
The oscilloscope can acquire the waveform according to a normal YT mode at any
sampling rate and can check the corresponding waveform at XY mode.
The [Volt/div] and the vertical [POSITION] of channel 1 control the setup of the
horizontal scale and position.
The [Volt/div] and the vertical [POSITION] of channel 2 control the setup of the
vertical scale and position.
Revolve the [s/div] knob to regulate the sampling rate so as to observe the
waveform better.
Cursor
Trigger control
51
Setup
Description
Cursor mode
Manual
Voltage
Types
parameters.
52
Time
Manually
use
the
cursor
to
measure
time
parameters.
CH1
CH2
Signal sources
MATH
REF A
REF B
Cur A
Cur B
The manual cursor measurement mode is used for measuring the coordinate values
and increments of one pair of horizontal or vertical cursors. Ensure the signal source to
be set rightly when using the cursors, as shown in figure 2-50.
Voltage cursor:
Time cursor:
The time cursor appears on the display screen as a vertical line, and
it can be used for measuring horizontal parameters.
Cursor movement: Select the cursors first and use the [UNIVERSAL] knob to move
cursor A and cursor B, wherein the values of the cursors will
appear on the right upper corner of the screen during
movement.
The operation steps are as follows:
1. Press [CURSORS] to skip to the CURSOR menu.
2. Select Cursor mode as manual.
3. Press Type to select Voltage or Time.
4. Press Signal source to select CH1, CH2, MATH, REF A or REF B according to the
signal input channel.
5. Select Cur A and revolve the [UNIVERSAL] knob to regulate the position of
53
cursor A.
6. Select Cur B and revolve the [UNIVERSAL] knob to regulate the position of
cursor B.
7. Display the measured values on the left upper corner of the screen.
Figure
Option
Cursor mode
Cursor A
Setup
Description
Tracking
CH1
CH2
No cursor
Cursor B
Cur A
Cur B
At cursor tracking measurement mode, the cross cursors are displayed on the
measured waveform, the cursors are automatically located on the waveform by moving
the horizontal position between the cursors, and simultaneously the horizontal and
vertical coordinates of the current located point, and the horizontal and vertical
increments between the two cursors are displayed. The horizontal coordinate is
displayed as a time value, and the vertical coordinate is displayed as a voltage value,
as shown in figure 2-53.
54
2.10.3 Measurement
As shown in figure 2-52, [MEASURE] is a function key for measurement.
Measure
measurement
includes
three major
55
items:
measurement and delay measurement; and there are 32 measurement minor items in
total. At most five items can be displayed once. Press [MEASURE] to skip to the
measurement menu to display the measurement result at first, as shown in figure 2-53,
and press any option key to change the measurement type, as shown in figure 2-54.
Description
Voltage
measurement
Time
measurement
Delay
measurement
All measurement
Back
Table 2-36
Option
Setup
Signal
CH1, CH2
Description
Select a signal source for a
source
types
voltage test.
Maximal value, minimal value,
peak-to-peak value, amplitude,
top value, bottom value,
periodic average value,
average value,
periodic mean square root,
mean square root,
ROVERShoot, FOVERShoot,
RPREShoot, FPREShoot
56
voltage
parameters.
measurement
Setup
Signal
CH1,CH2
Description
Select a signal source for time
source
measurement.
Rise time, fall time,
frequency, period,
pulse width,
parameters.
Back
Table 2-38
Option
Setup
Description
Signal
CH1
source
CH2
measurement.
FFR,FFF,LRR,
LRF,LFR,LFF
measurement parameters.
Types
57
Back
menu 1.
Table 2-39
Option
Signal source
Setup
CH1
CH2
Run
Voltage
measurement
Stop
Run
Time
measurement
Stop
Run
Delay
measurement
Stop
Back
Description
Select the input signal channel.
Run all measurement on the voltage type
parameters.
Stop all measurement on the voltage type
parameters.
Run all measurement on the time type
parameters.
Stop all measurement on the time type
parameters.
Run all measurement on the delay type
parameters.
Stop all measurement on the delay type
parameters.
Back to the measurement function menu 1.
Description
Maximal value
Minimal value
Peak-to-peak value
Top value
Bottom value
58
Amplitude
Average value
Periodic mean
square root
Mean square root
ROVERShoot
FOVERShoot
RPREShoot
FPREShoot
Rise time
Fall time
Pulse width
Phase
59
FRF
FFR
FFF
LRR
LRF
LFR
LFF
The corresponding icons and parameter values will be displayed in the menu
corresponding to the third option key, as shown in figure 2-55.
60
6. When get back to the measurement result region in the homepage, the selected
parameters and the corresponding values will be displayed in the first option.
Other option measurement types can be changed by the same method, and the
homepage can display five parameters every time.
The voltage parameters can be measured by using the all measurement function by the
following steps:
1. Press [MEASURE] to skip to the MEASURE menu.
2. Press any option key in the homepage to skip to the measurement function menu 1.
3. Select All measurement.
4. Press Signal source to select the signal input channel.
5. Press Voltage test to select Run. At the moment, all the voltage parameter
values will be simultaneously displayed on the screen. (As shown in figure 2-56)
61
Description
Setup save
Waveform save
Image save
CSV
Factory setups
62
Figure 2-57
Save type
63
64
Press Back to the waveform display interface if the current USB flash disc has no
setup files.
The save/recall waveform acquisition step is the same as the save/recall setup.
The oscilloscope enters STOP state after the saved waveform is recalled and
displayed.
Image save
As shown in figure 2-61, waveform images can be saved in the USB flash disc but
cannot be recalled by the oscilloscope. The images are in standard BMP format and
can be opened by related software in computer.
The images are saved in the root directory of the USB flash disc, and the file name is
defaulted as ASxxxx.BMP, such as GAS0001. BMP.
Table 2-41 Image save function menu:
Option
Setup
Description
Type
Image save
Print button
Save image
65
CSV save
Table 2-42 CSV save function menu:
Option
Setup
Description
Type
CSV
Screen
Data length
Parameter save
Run
Stop
Save
As shown in figure 2-62, the following steps can be carried out to save the CSV file in
the USB flash disc:
1. Press [SAVE/RECALL] to skip to the SAVE/RECALL menu.
2. Press Type to select CSV.
3. Plug the USB flash disc, and prompt USB save equipment is connected
successfully after the connection is normal.
4. Press Data length to select Screen or Internal memory.
5. Press Parameter save to select Stop or Run.
6. Press the Save option to save the CSV.
The data is saved in the root directory of the USB flash disc, and the file name is
defaulted as GASxxxx.CSV, such as GAS0001.CSV.
66
67
Setup
Description
System state
Sound
Frequency meter
Language
Next
Run
Stop
Simplified Chinese
Simplified Chinese
English
English
Page 1/4
function menu
Table 2-44
Option
Setup
Description
Carry out self correction operation for
Self correction
Self test
function menu
channel correction.
Screen test
Keyboard test
Lightening test
68
Computer
interface
Next
Table 2-45
Page 2/4
Option
Setup
Description
Upgrade the software by the USB flash
Upgrade
firmware
minutes.
Press the key to skip to the Pass/Fail
Pass/Fail
menu.
Waveform
recording
recording menu.
Interface setup
Next
Page 3/4
69
Setup
Description
1min
2min
5min
10min
Screen saving
15min
20min
1hour
2hour
5hour
Stop
Next
Page 4/4
70
2.12.4
Self test
Press [UTILITY], and press Self test in page 2 of the UTILITY menu. The self test
includes screen test, keyboard test and lightening test.
Operation steps:
1. Screen test:
Select Screen test to skip to a screen test interface, as shown in figure 2-70. At the
moment, the screen displays a prompt message of Press SINGLE key to continue,
Press RUN/STOP key to exit, namely start the test when prompting to press [SINGLE].
Different colors are displayed on the screen. Observe whether the screen has serious
color cast or other display errors.
71
A prompt message of Press RUN/STOP key three times to exit is displayed at the bottom
of the screen to remind the user to exit the test method.
continue, Press RUN/STOP key to exit. After continuously pressing the [SINGLE] key,
the corresponding region on the screen will be in green when the key is lightened. RUN
and STOP share one key, so the key is in green at RUN state and in red at STOP state.
2.12.6 Pass/fail
Pass/fail is used for judging whether the input signal is in a built rule range and
73
outputting the past or failed waveform so as to detect the change condition of the
signal.
Table 2-47 Page 1 of the pass/fail function menu:
Option
Test allowed
Signal source
selection
Operation
Setup
Description
Run
Stop
CH1
CH2
Run pass/fail
Stop pass/fail
Open waveform
information.
Display
information
pass/fail
time
display
Next
Output
Output stop
Setup
Description
Pass
Fail
Run
Stop
74
Rule setup
Back
Back
Page 2/2
Setup
Description
Horizontal
regulation
Vertical
regulation
Build rules
setups above.
Save position
Next
Inside
Outside
Page 1/2
Description
Save
Recall
Back
Back
Page 2/2
75
to start.
As shown in figure 2-77, after the rules are built completely, PASS starts counting if
the signal of CH2 satisfies the rules during start, and Pass would stop counting
while Fail would start counting when the signal surpasses the set template (the
amplitude of the signal in the figure is smaller).
Pass/Fail output
The pass/fail function can be used for outputting a negative pulse train through a
Pass/Fail BNC interface on a rear panel of the oscilloscope.
76
End frame
Setup
Description
Run
Stop
Record
Replay
CH1
CH2
recording.
Time interval
Operation
As shown in figure 2-78, the operation steps of waveform recording are as follows:
1. Press [UTILITY] to skip to the [UTILITY] menu.
2. Press Next to skip to page 3 of the auxiliary menu.
3. Press Waveform recording to skip to the RECORD menu.
77
Option
Setup
Description
Mode
Replay
Operation
Replay mode
Time interval
Next
frame.
Page 1/2
replay menu
replay menu
Setup
Description
Start frame
Current frame
End frame
Back
Back
Page 2/2
During replay, the current frame number is displayed on the screen; after replay is
stopped, the [UNIVERSAL] knob is used to observe all frames of waveforms between
the start frame and the end frame.
The current recorded waveforms can be replayed by carrying out the following steps:
1. Press [UTILITY] to skip to the [UTILITY] menu.
2. Select Mode as Replay.
3. Select Replay mode as
or
l.
4. Press Time interval to set the time interval of the replayed waveform frame to
frame.
5. Press Next Page1/2 to skip to page 2 of the replay function menu.
6. Press Start frame, and revolve the [UNIVERSAL] knob to regulate the start frame
number of the replayed waveforms.
7. Press End frame, and use the [UNIVERSAL] knob to regulate the end frame number
of the replayed waveforms.
8. Press the Back Page 2/ 2 option key to get back to page 1 of the waveform replay
function menu.
9. Press the Operation option
79
80
Application examples
DIGITAL STORAGE OSCILLOSCOPE
and
these
simplified
examples
Simple measurement
Cursor measurement
Screen capture
Set the attenuation coefficient as 10X in the probe option by [CH1], and set the
switch on the probe as 10.
The corresponding icons and measurement values will be displayed in the third option.
Measure peak-to-peak value of the signal
81
The corresponding icons and measured values will be displayed in the third option.
Display of the measurement results on the screen is changed with the change of the
measured signal.
Please regulate [Volt/div] or [s/div] if the reading of the value is displayed as ****.
82
of spike pulse
of spike pulse
83
5. Press the CurAoption key, and revolve the [UNIVERSAL] knob to place the cursor A
at the highest wave peak of the spike pulse.
6. Press the CurBoption key, and revolve the [UNIVERSAL] knob to place the cursor B
at the lowest point of the spike pulse.
As shown in figure 3-2the following measurement results will be displayed on the
display screen:
Voltage at cursor A
Voltage at cursor B
84
85
86
The phase difference angle theta is equal to +/- arcsin (A/B) or theta is equal to +/arcsin (C/D) according to the formula sin theta=A/B or C/D, wherein theta represents
the phase difference angle between the channels, and A, B, C and D are defined in the
figure above.
87
The two signals have the same amplitude if the frequency or phase difference of the
two measured signals is integer multiples of pi/4. The Lissajous pattern of the special
signals is shown in figure 3-4, wherein m:n is a frequency ratio, K represents a phase
difference coefficient, and theta=K x pi/4. Figure 2-49 is a measured Lissajous pattern
when the two signals have the same frequency and amplitude and the phase difference
is pi/2.
88
Two probes should be compensated at first, and the differences of the probe
compensation may cause errors of the measurement results.
89
90
Trigger electric level reaches the limit: It prompts that the current trigger electric
level has already reached the limit and cannot be regulated any more.
Horizontal position reaches the limit: It prompts that the horizontal movement
position has already reached the limit and cannot be regulated any more.
Voltage gear reaches the limit: It prompts that the vertical voltage gear has
already reached the lowest gear 2mV/div or the highest gear 5mV/div and cannot
be regulated any more.
Voltage displacement reaches the limit: The system will pop up this prompt
message if channel offset is regulated to be too high or too low by revolving the
knob [VERTICAL POSITION].
Time base gear reaches the limit: The system will prompt this message if the
horizontal time base is still regulated after being regulated to the fastest or
slowest gear.
Hold-off time reaches the limit: The system will prompt this message when the
hold-off time is regulated to a smallest or largest value by revolving the knob
[UNIVERSAL].
Function unavailable: The system cannot set some functions at some special
modes.
No signal: This message is prompted when the signal does not satisfy the
automatic setup condition during automatic setup.
Setup reaches the limit: Regulation cannot be continued if the pulse width
reaches the minimal value 20.0ns or the maximal value 10.0s when the knob
[UNIVERSAL] is used to regulate the pulse width.
Blank storage unit: This message appears when the option key [CALLOUT] is
pressed if no waveform or setup is stored at the storage position in waveform
storage or setup storage.
USB storage device is connected successfully: This message will appear when
the USB flash disc is plugged in a USB Host interface of the oscilloscope.
USB storage device is disconnected: This message will appear when the USB
flash disc is unplugged.
91
Data is stored successfully: It means that the setup data, waveform data and
image data are already stored in the oscilloscope or the USB flash disc.
Data is read successfully: It means that the setup data or the waveform data is
successfully called out from the oscilloscope or the USB flash disc.
USB storage device is unconnected: This message will be prompted if the USB
flash disc is not plugged for storage when the option store to is set as USB
flash disc or the option PRINT BUTTON is set as stored image in the menu
SAVE/RECALL.
Waveform recording is finished: The system will prompt this message when
waveform recording is ended.
92
(2)
(3)
(4)
2. Treat the fault according to the following steps if the waveform of the signal does
not appear in the picture after the signal is acquired:
(1)
Check whether the probe is normally connected to the signal connecting wire.
(2)
Check whether the signal connecting wire is normally connected to the BNC.
(3)
(4)
(5)
3. Check whether the channel attenuation coefficient accords with the probe
attenuation ratio actually used if the measured voltage amplitude is 10 times larger
or 10 times smaller than the actual value.
4. If waveform exists but not stable:
(1)
Check whether the signal source option of the trigger panel accords with the
signal channel actually used.
(2)
Check the trigger type: general signals apply EDGE TRIGGER mode, video
signals apply VIDEO TRIGGER mode, and the waveform is only stably
displayed when suitable trigger mode is applied.
(3)
93
afterglow time.
It is normal.
7. If waveform display is ladder-like:
(1)
(2)
The horizontal time base gear may be too low, and the display can be improved
by increasing the horizontal time base to increase the horizontal resolution.
(3)
The display type may be VECTOR, and the linked line between the sampling
points may cause ladder-like display of the waveform. This problem can be
solved by setup the display type as POINT display mode.
94
This
chapter
mainly
introduces
the
related
95
Appendix
DIGITAL STORAGE OSCILLOSCOPE
The oscilloscope should run for at least 30min within a specified operation
temperature range.
All the specifications are ensured to satisfy the requirement except that marked with
TYPICAL sign.
Technical specifications
Input
Input coupling
Input impedance
Maximum input
voltage
Probe attenuation
Real-time sampling
Single channel 2GSa/s; dual channel 1GSa/s
Sampling rate
(Only to GA1302CAL/GA1202CAL)
Single channel 1GSa/s, dual channel 500MSa/s (only to
GA1102CAL)
Storage depth
Acquire mode
Average time
96
Vertical system
Vertical sensitivity
Channel voltage offset
range
Vertical resolution
8bit
Channel amount
2
300MHz(only to GA1302CAL)
Bandwidth
200MHz(only to GA1202CAL)
100MHz(only to GA1102CAL)
AC plus accuracy
DC measurement
accuracy
Rise time
Vertical coupling
AC,DC,GND
Arithmetical operation
+, -, , , FFT
FFT
Bandwidth limit
Horizontal system
2ns/div ~ 50s/div, step 32 gears according to sequence
1-2-5 (only to GA1102CAL/GA1202CAL)
Time base
Horizontal
displacement range
Display mode
100div
Y-T mode, X-Y mode
97
3 Degrees
Point display, vector display
Trigger system
Trigger type
Trigger mode
Trigger coupling
CH1, CH2:10div
EXT: 1.5V
EXT/5: 7.5V
CH1, CH2: 1div
Trigger sensitivity
EXT: 0.15V
Hold-off range
100ns ~10s
Edge trigger
EXT/5: 0.75V
Video trigger
Slope trigger
Alternating trigger
Measurement system
Automatic
measurement (32
98
kinds)
Cursor measurement
Save/recall
6 bits
Alternating-current coupling, from 10Hz to the maximal
bandwidth
All sources capable of being triggered in pulse trigger or
edge trigger type
General specifications
Display system
Display type
Display color
64k color
99
500:1
2
300 Cd/m
14 8 grids
Afterglow
Menu display
Screen saver
Interpolation mode
Display language
Power supply
Power voltage
AC
power
frequency range
Consumed power
45Hz to 440Hz
50VA Max
Environment
Temperature
Operating: 10 C to +40 C
Non operating: -20 C to +60 C
Cooling
Humidity
90% below 40 C
Height
Mechanical specifications
Size
Weight
About 2.3kg
100
CH1 CH2
Default setup
Coupling
Direct current
Bandwidth limit
OFF
Volt/div
Coarse tunning
Probe
Inverse phase
OFF
Volt/div
200mV
Digital filter
OFF
Operation
CH1+CH2
FFT operation
MATH
Signal source
CH1
Window
Hanning
Horizontal
1
amplification
HORIZONTAL
Vertical scale
dBVrms
Window
Position
0.0s
S/div
100s
OFF
Signal
CH1
Voltage
2 G
Time
2G
Signal source
CH1
CURSOR
MEASURE
101
Type
Peak-to-peak value
Average value
Amplitude
Period
Rise time
ACQUIRE
sampling
Average time
Interpolation type
Sine
type
vector
Go on
OFF
Waveform brightness
100%
Grid brightness
100%
Format
YT
Menu display
Infinite
Type
Save setup
Save to
Equipment
Set
No.1
Signal source
CH1
REFA/REFB
REFA
REFA
OFF
REFB
OFF
Source
OFF
Frequency meter
ON
Computer
DISPLAY
SAVE/RECALL
REF
UTILITY
102
Test passed
OFF
Waveform recording
OFF
RS-232
Type
Edge
Signal source
CH1
Slope
Rise
Trigger mode
Auto
Coupling
Direct current
Electric level
0.00mV
Type
Pulse
Signal source
CH1
TRIGGER (EDGE)
TRIGGER
Condition
(PULSE)
20.00ns
Trigger mode
Auto
Coupling
Direct current
Type
Video
Signal source
CH1
TRIGGER
Polarity
(VIDEO)
Synchro
All rows
Standard
NTSC
Trigger mode
Auto
TRIGGER
Type
Slope
(SLOPE)
Signal source
CH1
Condition
103
Time setup
20.00ns
Trigger mode
Auto
Type
Alternation
Signal source
CH1
Trigger mode
Edge
Coupling
Direct current
Slope
Rise edge
TRIGGER
(ALTERNATION)
104
To avoid damage of the surface of the instrument of probe, please do not use any corrosive
reagent or chemical cleaning reagent.
Before use, after the power on, please confirm that the instrument is completely dry so as to
prevent short circuit or even personal injury caused by moisture.
105