1207 C Manual 101

Download as pdf or txt
Download as pdf or txt
You are on page 1of 59

Model 1207C

Temperature/Humidity Chamber
Operation and Service Manual

TestEquity Inc.
2450 Turquoise Circle
Thousand Oaks, CA 91320
Support:

877-512-3457 Toll Free


805-480-3697

Corporate:

800-732-3457
805-498-9933

http://www.testequity.com

Copyright 2001 TestEquity Inc.

Rev. 1.01

Table of Contents
Chapter 1 Safety Instructions________________________________________________ 1-1
Introduction____________________________________________________________________ 1-1
Installation Safety Notices ________________________________________________________ 1-1
Operation Safety Notices _________________________________________________________ 1-2

Chapter 2 Installation ______________________________________________________ 2-1


Uncrating ______________________________________________________________________
Preparation For Use _____________________________________________________________
Installation Location_____________________________________________________________
Input Power Configuration _______________________________________________________

2-1
2-1
2-2
2-3

Overview ____________________________________________________________________________
230 V / 60 Hz Input Configuration ________________________________________________________
208 V / 60 Hz Input Configuration ________________________________________________________
Single Phase Configuration ______________________________________________________________
Three Phase Configuration_______________________________________________________________

2-3
2-5
2-5
2-5
2-5

Connection to the Power Source ___________________________________________________ 2-6


Input Wiring Preparation ________________________________________________________________ 2-6
Single Phase Connection ________________________________________________________________ 2-6
Three Phase Connection _________________________________________________________________ 2-6

Humidity Water Connection ______________________________________________________ 2-7


Plumbed Demineralized Water Installation __________________________________________________
Water Recirculation System Installation ____________________________________________________
Water Recirculation System Filter Cartridge Replacement ______________________________________
Water Recirculation System Assembly _____________________________________________________

2-7
2-7
2-7
2-8

Chapter 3 Operation _______________________________________________________ 3-1


Introduction____________________________________________________________________ 3-1
Summary of Chamber Operation __________________________________________________ 3-1
Front Panel Switches and Lights ___________________________________________________ 3-2
Main Disconnect Switch ________________________________________________________________
Master Switch ________________________________________________________________________
Humidity Enable Switch ________________________________________________________________
Light Switch __________________________________________________________________________
Silence Switch ________________________________________________________________________
Heat Light ___________________________________________________________________________
Cool Light ___________________________________________________________________________
Humidity Light________________________________________________________________________

3-2
3-2
3-2
3-2
3-2
3-2
3-3
3-3

Temperature/Humidity Controller _________________________________________________ 3-4


Overview ____________________________________________________________________________ 3-4
Security Features ______________________________________________________________________ 3-4
Temperature/Humidity Controller Keys and Displays __________________________________________ 3-5
Static Set Point Control _________________________________________________________________ 3-6
Digital Event Outputs___________________________________________________________________ 3-7
Digital Output Connections ______________________________________________________________ 3-7
Event Board Option ____________________________________________________________________ 3-7
Profile Programming ___________________________________________________________________ 3-8
Temperature Profile Examples____________________________________________________________ 3-9
RS-232C Interface ____________________________________________________________________ 3-10
GPIB Interface (optional)_______________________________________________________________ 3-10
Common Modbus Registers_____________________________________________________________ 3-10

Limit Controller _______________________________________________________________ 3-11


Overview ___________________________________________________________________________
Limit Controller Keys and Displays _______________________________________________________
How to Set the High and Low Temperature Safety Limits _____________________________________
Resetting a Limit Condition _____________________________________________________________
Silencing the Audible Alarm ____________________________________________________________
Protecting an Energized Test Sample______________________________________________________

3-11
3-12
3-12
3-12
3-12
3-13

Table of Contents
Loading the Chamber ___________________________________________________________ 3-14
Performance Considerations ____________________________________________________________ 3-14
Avoiding Moisture (non-humidity mode) __________________________________________________ 3-15
Internal Test Fixtures __________________________________________________________________ 3-15

Humidity Operation ____________________________________________________________ 3-16


Humidity Range ______________________________________________________________________ 3-16
Humidity Mode Considerations __________________________________________________________ 3-16

Chapter 4 Frequently Asked Questions________________________________________ 4-1


Chapter 5 Specifications ____________________________________________________ 5-1
Model 1207C Chamber Specifications ______________________________________________ 5-1
F4 Controller Specifications _______________________________________________________ 5-2
Humidity Sensor Specifications ____________________________________________________ 5-3

Chapter 6 Maintenance _____________________________________________________ 6-1


Preventive Maintenance Schedule __________________________________________________ 6-1
Daily or As Needed ____________________________________________________________________
Every 3 Months _______________________________________________________________________
Every 6 Months _______________________________________________________________________
Every 12 Months ______________________________________________________________________
How to clean the chamber interior and exterior _______________________________________________
How to listen for abnormal noise or vibration ________________________________________________
How to inspect the filter cartridge _________________________________________________________
How to inspect the door seal _____________________________________________________________
How to inspect the refrigeration machinery compartment _______________________________________
How to inspect the electrical compartment __________________________________________________
How to clean the condenser ______________________________________________________________
How to verify the chamber temperature performance __________________________________________
How to drain the humidity system _________________________________________________________
How to verify the temperature calibration ___________________________________________________
How to verify the humidity calibration _____________________________________________________
Humidity Calibration Procedure __________________________________________________________

6-1
6-1
6-1
6-1
6-1
6-1
6-2
6-2
6-2
6-2
6-3
6-4
6-5
6-6
6-7
6-8

Refrigeration System Charging Instructions _________________________________________ 6-9


R-404A High-Stage Charge ______________________________________________________________ 6-9
R-508B Low-Stage Charge ______________________________________________________________ 6-9

Series F4 Controller Setup Parameters_____________________________________________ 6-10


Series F4 System Parameters ____________________________________________________________ 6-11
Series F4 Operations Parameters _________________________________________________________ 6-12
Series F4 Set Lockout Parameters ________________________________________________________ 6-12

Series 97 Limit Controller Setup Parameters________________________________________ 6-13

Chapter 7 - Warranty________________________________________________________ 7-1


Chapter 8 Drawings________________________________________________________ 8-1

Chapter 1 Safety

Chapter 1 Safety Instructions


Introduction
Follow all CAUTION notices to prevent damage to the chamber or your test sample. Failure to
follow all CAUTION notices may void your warranty. CAUTION may also indicate a
potentially hazardous situation which, if not avoided, may result in minor or moderate personal
injury.
WARNING indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
The safety alert symbol ! precedes a general CAUTION or WARNING statement.
The electrical hazard symbol 2 precedes an electric shock hazard CAUTION or WARNING
statement.

Installation Safety Notices


! CAUTION:

The minimum clearance you should allow for proper ventilation around the
chamber must be at least 12" from both the left and right side, and 24" from
the rear.

! CAUTION:

This chamber is designed for operation in a conditioned laboratory


environment. Operation above 30C (85F) or below 16C (60F) ambient
room temperature is NOT recommended.

! CAUTION:

The Input Voltage label on the back of the chamber indicates the input voltage
configuration as shipped from the factory. If the input voltage configuration is
changed, this label must be replaced to reflect the new configuration.
Replacement labels are available from TestEquity at no charge.

! CAUTION:

This chamber must be properly configured for either 208 V or 230 V nominal
input. 208 V and 230 V are NOT the same. Do NOT guess! Do NOT assume
you have 220 V. You must verify the exact type of electrical service you
have. If there is any doubt, you must consult with a qualified electrician who
is familiar with industrial plant wiring. In addition, the input line voltage
should be measured while the chamber is operating in the COOL mode to
ensure that the expected nominal voltage of either 208 V 5/+10% or 230 V
10% is present. Operation below 198 V or greater than 253 V requires
internal transformers, which can be supplied for a nominal charge.

2 CAUTION:

This chamber should be connected to the AC power source by a qualified


electrician who is familiar with industrial plant wiring.

TestEquity 1207C Temperature/Humidity Chamber

Page 1-1

Chapter 1 Safety
2 CAUTION:

The humidity system must be supplied with demineralized water having a


specific resistance of 50,000 to 200,000 ohms/centimeter. The use of
unfiltered tap water or triple-distilled water will damage the humidity system
and chamber interior, and will void the warranty.

Operation Safety Notices


! CAUTION:

This chamber has a crankcase heater to protect the high-stage compressor. The
chamber must be connected to the power source AND the Main Disconnect
Switch must be ON for 3 hours prior to operating the chamber. Although it
may be safe to use the chamber immediately, this procedure ensures the
longest possible life for the high-stage compressor if the chamber has been
removed from the power source for more than 24 hours.

! CAUTION:

The Series F4 Users Manual is a general manual and is written by the


manufacturer, Watlow, for a wide variety of applications and configurations.
Not all features or functions are applicable. Only the capabilities of a model
F4DH-CCCC-01, as described on page A.7 of the Series F4 Users Manual
are applicable. Cascade Control as described on page 3.6 of the Series F4
Users Manual is not applicable in this configuration. The Retransmit
function is available as an option.

! CAUTION:

The Series F4 Alarm2 is configured for internal protection of the humidity


system. Do NOT change this configuration under any circumstances. The
Series F4 Alarm1 is not connected and had no protective function in this
chamber. The independent Limit Controller functions as the main system and
product protection device.

! CAUTION:

The Series F4 Controller has been properly configured by TestEquity to match


the chambers system requirements and to perform optimally over a wide
range of operating conditions. Improper modifications to these setup values
can result in erratic performance and unreliable operation. Setup examples in
the Series F4 Users Manual are not applicable to this chamber. Do not
attempt to modify the setup values, unless you thoroughly understand what
you are doing. If there is any doubt, please call TestEquity before proceeding.

! CAUTION:

Do NOT select Full Defaults in the Series F4 Factory Menu. Full Defaults
are NOT the TestEquity configuration parameters for this chamber. If you
select Full Defaults, you must reconfigure all System and Operation
Parameters as documented in the TestEquity manual, NOT the Series F4
Users Manual.

! CAUTION:

The Series 97 Users Manual is a general manual and is written by the


manufacturer, Watlow, for a wide variety of applications and configurations.
Not all features or functions are applicable. Only the capabilities of a model
97A0-DAAA-00-RG, as described on page A.20 of the Series 97 Users
Manual are applicable. Please note that only the Series 97 Limit functions
are applicable.

Page 1-2

TestEquity 1207C Temperature/Humidity Chamber

Chapter 1 Safety
! CAUTION:

Always verify that the Limit Controllers high and low limits are set to
temperatures that are appropriate for your test sample.

! WARNING: Do NOT put items in the chamber that could burn or explode at high
temperatures. This chamber uses open wire heating elements which generate
surface temperatures over 1000F. This is NOT an explosion-proof chamber.
! WARNING: Do NOT put items in the chamber which can emit corrosive vapors or
substances.
! WARNING: This chamber is NOT a curing oven. There are NO provisions for venting
fumes.
! WARNING: The chamber door must remain closed while the chamber is operating. If you
need to open the door while the chamber is operating, wear safety goggles to
prevent the high velocity airflow from blowing particles or objects into your
eyes.
! WARNING: This chamber operates at extreme temperatures. Avoid contact with air,
objects, and surfaces that are hot or cold to prevent severe burns or frostbite.
Protective gloves are recommended.
! CAUTION:

If your test sample is energized, it may be capable of raising the workspace


temperature beyond safe limits. This could occur if your test sample exceeds
the live load rating of the chamber or if the chambers refrigeration system
fails. This chamber has a set of contacts that can be used to remove power to
your test sample if the Limit Controllers temperature limits are exceeded.

! CAUTION:

To prevent damage to your test sample and the chambers compressors, do not
exceed the live load rating of the chamber.

TestEquity 1207C Temperature/Humidity Chamber

Page 1-3

Chapter 2 Installation

Chapter 2 Installation
Uncrating
Inspect the shipping container for any signs of visible damage. Notify the carrier and TestEquity
immediately if there are signs of shipping damage.
The pallet is designed with ramps so the chamber can be rolled off without the need for a forklift
or pallet jack.
1.
2.
3.
4.

Cut the two metal bands that hold the packaging to the pallet.
Remove the top cover.
Remove the plastic fasteners that hold the outer box together and remove the outer box.
Locate the retainer in the front of the pallet. Remove the two screws, then remove the
retainer. Keep in mind that there is also Velcro that holds the retainer in place. Place the
retainer in the front of the pallet on the floor.
5. Locate the two ramps under the chamber and remove them.
6. Attach the ramps to the front of the pallet using the Velcro straps. The retainer will be under
the ramps and act as a support.
7. Carefully roll the chamber off the pallet, onto the ramps. This should be done with at least
two people.

Preparation For Use


1.
2.
3.
4.
5.
6.

Inspect the chamber for signs of shipping damage.


Read this entire manual.
Select a suitable location to install the chamber.
Verify the input voltage and phase configuration.
Connect to the power source.
Connect to humidity water source and drain OR connect to the recirculation system and fill
reservoir with water.
7. Perform following the procedures as described in the Preventive Maintenance section:
a. Inspect the electrical compartment.
b. Inspect the refrigeration machinery compartment.
c. Verify the chamber performance.
The following pages discuss these steps in detail.

TestEquity 1207C Temperature/Humidity Chamber

Page 2-1

Chapter 2 Installation

Installation Location
The chamber will produce a significant amount of heat during normal operation. Locate the
chamber in a room with adequate ventilation to prevent excessive heat build-up.
The chamber must be on a solid and level floor. Allow enough space around the chamber to
permit serviceability and the removal of the service access panels, which are located on each side
and the rear.
! CAUTION:

The minimum clearance you should allow for proper ventilation around the
chamber must be at least 12" from both the left and right side, and 24" from
the rear.

! CAUTION:

This chamber is designed for operation in a conditioned laboratory


environment. Operation above 30C (85F) or below 16C (60F) ambient
room temperature is NOT recommended.

Page 2-2

TestEquity 1207C Temperature/Humidity Chamber

Chapter 2 Installation

Input Power Configuration


Overview
This chamber is designed to be easily configured for operation from a Single Phase or Three
Phase power source, and either 208 V / 60 Hz or 230 V / 60 Hz. Other input voltages and 50 Hz
operation are available as special options, and are not covered in these instructions.
Your chamber was configured prior to shipment for the particular voltage that was specified at
time of order. These instructions should be used to verify the input voltage configuration prior to
installation, or to change the input voltage from one configuration to another.
! CAUTION:

The Input Voltage label on the back of the chamber indicates the input voltage
configuration as shipped from the factory. If the input voltage configuration is
changed, this label must be replaced to reflect the new configuration.
Replacement labels are available from TestEquity at no charge.

! CAUTION:

This chamber must be properly configured for either 208 V or 230 V nominal
input. 208 V and 230 V are NOT the same. Do NOT guess! Do NOT assume
you have 220 V. You must verify the exact type of electrical service you
have. If there is any doubt, you must consult with a qualified electrician who
is familiar with industrial plant wiring. In addition, the input line voltage
should be measured while the chamber is operating in the COOL mode to
ensure that the expected nominal voltage of either 208 V 5/+10% or 230 V
10% is present. Operation below 198 V or greater than 253 V requires
internal transformers, which can be supplied for a nominal charge.

TestEquity 1207C Temperature/Humidity Chamber

Page 2-3

Page 2-4
208V
Jump 25 to 26

203

H2 H3
2

44

230V, wire 44 on H3
208V, wire 44 on H2

F1 F2

1 Phase
Jump
F2 to F10
F8 to F9

F8 F9 F10

3 Phase
Jump
F2 to F8

INPUT
L1 L2 L3

204

(1 Ph on L1, L2 only)

Ground

INPUT
ACCESS
HOLE

Chapter 2 Installation

Figure 2-1 Location of Input Configuration and Connection Terminals

TestEquity 1207C Temperature/Humidity Chamber

Chapter 2 Installation
NOTE: Refer to Figure 2-1 on the previous page for the location of the input configuration and
connection terminals on the electrical sub panel that are described below.

230 V / 60 Hz Input Configuration


1. Turn the Main Disconnect switch to the OFF position.
2. Remove the lower door retaining screws located on the right side. Open the lower door.
3. Locate the Control Transformer, item 203. Wire number 44 must be connected to terminal
H3 on the Control Transformer.
4. Locate the terminal strip on the lower left side. There must be NO jumper between terminals
25 and 26.

208 V / 60 Hz Input Configuration


1. Turn the Main Disconnect switch to the OFF position.
2. Remove the lower door retaining screws located on the right side. Open the lower door.
3. Locate the Control Transformer, item 203. Wire number 44 must be connected to terminal
H2 on the Control Transformer.
4. Locate the terminal strip on the lower left side. There must be a jumper between terminals 25
and 26.

Single Phase Configuration


1. Turn the Main Disconnect Switch to the OFF position.
2. Remove the lower door retaining screws located on the right side. Open the lower door.
3. Locate the row of fuses on the upper right side. There must be a jumper across the top of the
fuseholder terminals F2 and F10, and another jumper across fuseholder terminals F8 and F9
(2 jumpers total).

Three Phase Configuration


1. Turn the Main Disconnect Switch to the OFF position.
2. Remove the lower door retaining screws located on the right side. Open the lower door.
3. Locate the row of fuses on the upper right side. There must be a jumper across the top of the
fuseholder terminals F2 and F8.

TestEquity 1207C Temperature/Humidity Chamber

Page 2-5

Chapter 2 Installation

Connection to the Power Source


2 CAUTION:

This chamber should be connected to the AC power source by a qualified


electrician who is familiar with industrial plant wiring.

The Main Disconnect Switch on the front panel removes primary power to the entire chamber.
All branch circuits on the load side of the Main Disconnect Switch are individually fused.
However, your local electrical code may require a separate disconnect switch within sight of the
chamber.

Input Wiring Preparation


1. Ensure that the chamber is properly configured for Single Phase or Three Phase input as
described on the previous page.
2. Remove the chambers right side panel.
3. Mount the input wire through the hole on the rear of the chamber, using an appropriate
bushing. An additional 3-feet of wire will need to pass through the Input Access Hole in
order to reach to the terminals on the electrical subpanel. The Input Access Hole is located in
the corner of the partition which separates the compressor compartment from the electrical
compartment.

Single Phase Connection


1. Locate the Main Disconnect Switch, item 204. Remove the terminal cover from the top of
this switch.
2. Connect the Hot input wires to terminals L1 and L2 on the Main Disconnect Switch. (the
switch is marked as terminals 1 and 3 respecively).
3. Put the terminal cover back on.
4. Connect the Ground wire (NOT A NEUTRAL) to the Ground terminal.
5. Put the chamber side panel back on and close the lower door.

Three Phase Connection


1. Locate the Main Disconnect Switch, item 204. Remove the terminal cover from the top of
this switch.
2. Connect the Hot input wires to terminals L1, L2 and L2 on the Main Disconnect Switch
(the switch is marked as terminals 1, 3 and 5 respecively).
3. Put the terminal cover back on.
4. Connect the Ground wire (NOT A NEUTRAL) to the Ground terminal.
5. Put the chamber side panel back on and close the lower door.

Page 2-6

TestEquity 1207C Temperature/Humidity Chamber

Chapter 2 Installation

Humidity Water Connection


2 CAUTION:

The humidity system must be supplied with demineralized water having a


specific resistance of 50,000 to 200,000 ohms/centimeter. The use of
unfiltered tap water or triple-distilled water will damage the humidity system
and chamber interior, and will void the warranty.

Plumbed Demineralized Water Installation


Follow these instructions if your chamber does not have a factory installed recirculation system,
and you are connecting to a plumbed source of demineralized water.
1. Attach a supply of demineralized water to the Water Supply fitting on the chamber.
2. The chamber drain water is not under pressure, and is fed by gravity. Therefore, it must
empty into an open floor drain. Alternatively, the chamber drain can empty into a condensate
pump. You can purchase a condensate pump from suppliers such as Grainger
(www.grainger.com).

Water Recirculation System Installation


The optional water recirculation system provides a source of demineralized water for the
humidity system, using ordinary tap or pre-filtered water. Condensate from the chamber is
returned back to the water tank. The pump operates only when needed to fill the chambers
humidifier to the correct level. All water connections should be finger-tight only do not use a
wrench. Detailed assembly pictures are shown on the next page.
1. Attach filter cartridge. Be sure to place the rubber washers between the filter ends and the
fittings. Install the cartridge with the label upside down.
2. Attach the pump outlet hose to the bottom fitting of the filter.
3. Place the pump in the water tank.
4. Attach the chamber Water Supply hose to the top fitting of the filter.
5. Attach the chamber Drain hose and position it so it goes into the water tank.
6. Fill the water tank with clean water.
7. Place the cover on the water tank.
8. Plug the pump electrical line into receptacle on the chamber.

Water Recirculation System Filter Cartridge Replacement


The demineralizer filter cartridge must be replaced when it is used up. The cartridge color
changes gradually from the bottom up to a yellow-amber through continued use. When all but
2-inches have completely changed color, it must be replaced. Replacement cartridges are
available from TestEquity (part number TE-0011 for a carton of 6 cartridges).
The demineralizer filter cartridge will last longer if pre-filtered water is used. Rather than filling
the recirculation system reservoir with tap water, you can use filtered drinking water (not spring
water) or distilled water (supermarket quality, not triple distilled lab quality).

TestEquity 1207C Temperature/Humidity Chamber

Page 2-7

Chapter 2 Installation

Water Recirculation System Assembly

Replace when
color change
reaches this line.

Install with label


upside down.
Be sure to place the rubber washers
between the filter ends and the fittings.
Finger-tighten all connections.

Pump placement in water tank.

Water Supply and Drain


connections to chamber.
Finger-tighten all connections.

Page 2-8

Completed installation. Plug into receptacle, turn clockwise to lock.


Installation behind chamber is not recommended because it may block
the air vents.

TestEquity 1207C Temperature/Humidity Chamber

Chapter 3 Operation

Chapter 3 Operation
Introduction
The Front Panel Switches control power to the chamber. The Front Panel Lights provide
indication of heat and cool functions.
The Limit Controller is a protection device. It turns the chamber OFF if the workspace
temperature exceeds either a high temperature or low temperature limit set point.
The Temperature/Humidity Controller controls the temperature of the chamber. It can function
as either a single set point controller or as a programmable profile controller. The
Temperature/Humidity Controller automatically turns the refrigeration system on or off based on
the demand for cooling.
! CAUTION:

This chamber has a crankcase heater to protect the high-stage compressor. The
chamber must be connected to the power source AND the Main Disconnect
Switch must be ON for 3 hours prior to operating the chamber. Although it
may be safe to use the chamber immediately, this procedure ensures the
longest possible life for the high-stage compressor if the chamber has been
removed from the power source for more than 24 hours.

Summary of Chamber Operation


1.
2.
3.
4.
5.
6.

Turn the Main Disconnect Switch ON.


Enter the appropriate high and low temperature safety limits on the Limit Controller.
Enter the desired set points (or program) on the Temperature/Humidity Controller.
Load your test sample in the chamber.
If you are running a controlled humidity condition, turn the Humidity Enable Switch ON.
Turn the Master Switch ON.

TestEquity 1207C Temperature/Humidity Chamber

Page 3-1

Chapter 3 Operation

Front Panel Switches and Lights


Main Disconnect Switch
The Main Disconnect Switch controls power to the entire chamber and provides a mechanical
safety interlock to the lower door.
In the ON position (clockwise) primary power is connected. The Temperature/Humidity
Controller, Limit Controller, and (optional) Chart Recorder are always functional when the Main
Disconnect Switch is ON, regardless of the Master Switch position. The Main Disconnect Switch
should be left ON if the chamber is usually used on a daily basis.
In the OFF position (counter clockwise) primary power is disconnected. The Main Disconnect
Switch can be left OFF if the chamber is not used on a daily basis. It can also be locked in the
OFF position with a padlock.
! CAUTION:

This chamber has a crankcase heater to protect the high-stage compressor. The
chamber must be connected to the power source AND the Main Disconnect
Switch must be ON for 3 hours prior to operating the chamber. Although it
may be safe to use the chamber immediately, this procedure ensures the
longest possible life for the high-stage compressor if the chamber has been
removed from the power source for more than 24 hours.

Master Switch
The Master Switch enables all chamber functions. The Master Switch illuminates when it is ON.

Humidity Enable Switch


The Humidity Enable Switch activates the controlled humidity mode. In this mode, the cooling
capacity is reduced. The Humidity Enable Switch illuminates when it is ON (if the chamber
temperature is between +7C and +90C).
If the chamber temperature is less that +7C or more than +90C, the Humidity system is
disabled. This switch should ONLY be ON if you are performing a controlled humidity test
within the humidity range of the chamber.

Light Switch
The Light Switch controls the workspace light. The Light Switch illuminates when it is ON.

Silence Switch
The Limit Controller will sound an Audible Alarm in the event of an over-temperature or undertemperature condition. The Silence Switch lets you temporarily turn off the Audible Alarm.

Heat Light
The Heat Light will illuminate when the Temperature/Humidity Controller turns on the heater to
maintain the workspace temperature. The Heat Light will cycle on/off as the workspace
temperature approaches and reaches the temperature set point. This light corresponds to the 1A
indicator on the Temperature/Humidity Controller.
Page 3-2

TestEquity 1207C Temperature/Humidity Chamber

Chapter 3 Operation

Cool Light
The Cool Light will illuminate when the Temperature/Humidity Controller turns on the cooling
valve to maintain the workspace temperature. The Cool Light will cycle on/off as the workspace
temperature approaches and reaches the temperature set point. This light corresponds to the 1B
indicator on the Temperature/Humidity Controller.

Humidity Light
The Humidity Light will illuminate when the Temperature/Humidity Controller turns on the
humidifier heater to maintain the workspace humidity. The Humidity Light will cycle on/off as
the workspace humidity approaches and reaches the humidity set point. This light corresponds to
the 2A indicator on the Temperature/Humidity Controller.
Note that the Humidity Enable Switch must be ON, the chamber workspace must be between
+7C and +90C, and the humidifier water level must be correct for the Humidity Light to
correspond to the 2A indicator on the Temperature/Humidity Controller.
The 2B indicator on the Temperature/Humidity Controller lights with dehumidification is
required to maintain the workspace humidity. There is no corresponding light on the front panel
for this function.

TestEquity 1207C Temperature/Humidity Chamber

Page 3-3

Chapter 3 Operation

Temperature/Humidity Controller
Overview
This section provides a brief overview on how to operate the Series F4 Controller in the static set
point mode. For more detailed instructions, see the Series F4 Users Manual.
The Series F4 Controller can function as either a single set point controller (static mode) or as a
programmable profile controller. A four-line LCD display facilitates setup and programming,
and presents informative messages about status, error, and alarm conditions. Digital outputs,
profiles, and alarms can be named for easy reference. An Information Key gives you quick
information about the pages, menus, parameters and values, as well as error and alarm conditions
if they occur. The user-interface is organized into five pages of menus. The Main Page
provides access to the other four pages Operations, Profiles, Setup, and Factory.
! CAUTION:

The Series F4 Users Manual is a general manual and is written by the


manufacturer, Watlow, for a wide variety of applications and configurations.
Not all features or functions are applicable. Only the capabilities of a model
F4DH-CCCC-01, as described on page A.7 of the Series F4 Users Manual
are applicable. Cascade Control as described on page 3.6 of the Series F4
Users Manual is not applicable in this configuration. The Retransmit
function is available as an option.

! CAUTION:

The Series F4 Alarm2 is configured for internal protection of the humidity


system. Do NOT change this configuration under any circumstances. The
Series F4 Alarm1 is not connected and had no protective function in this
chamber. The independent Limit Controller functions as the main system and
product protection device.

! CAUTION:

The Series F4 Controller has been properly configured by TestEquity to match


the chambers system requirements and to perform optimally over a wide
range of operating conditions. Improper modifications to these setup values
can result in erratic performance and unreliable operation. Setup examples in
the Series F4 Users Manual are not applicable to this chamber. Do not
attempt to modify the setup values, unless you thoroughly understand what
you are doing. If there is any doubt, please call TestEquity before proceeding.

Security Features
The Series F4 Controller has several levels of security to prevent unauthorized users from
changing critical configuration parameters. Only the Set Point and Profile menus have Full
Access. TestEquity has configured all other menus to Read Only, and have protected them
with a password.
TestEquity does not recommend that these security levels be changed for most applications.
However, there will be times when Full Access is necessary. For example, you may need to
gain access to Setup Page in order to change from C to F display, or to change the time or date.
You must call TestEquity at 877-512-3457 or 805-480-3697 to obtain the password.

Page 3-4

TestEquity 1207C Temperature/Humidity Chamber

Chapter 3 Operation

Temperature/Humidity Controller Keys and Displays


Upper Display
Displays actual chamber temperature
during operation. Displays error
information if errors occur.

Lower Display
Displays information
about the setup,
operation and
programming of the
controller.

Alarm Output
Indicator Lights
Lit during an alarm
state. Alarm2 is lit
when temperature
is outside the range
for humidity mode.

Cursor
Indicates selected
parameter or present
value in memory.
Moves via the four
navigation keys.

Communications
Indicator Light
Lit (pulsates) when
the controller sends
or receives valid
data.

Profile
Indicator Light
Lit when a ramping
profile runs.
When blinking, the
profile is on hold.
When not lit, the
controller operates as
a static set point
controller.
Profile Key
Summons a menu that
allows you to start, hold,
resume or terminate a profile.

Output
Indicator Lights
1A = Heat
1B = Cool
2A = Humidity
2B = Dehumidify

Scroll Bar
Appears when the
Up or Down Keys
can reveal more
information in the
Lower Display.

Information Key
Provides information
in the Lower Display
about the cursorselected parameter.
Another press toggles
the display back to
the parameter.

Left and Right Keys


Move right to select the
choice to the right of the
cursor and proceed
to the next screen.
Move left to exit.

Up and Down Keys


Move the cursor position in the
Lower Display through the
software in the direction of the
key arrow. Increase or decrease
a value, or change a letter in a
user-nameable field, such as
events and profile names.

Figure 4.1 Temperature/Humidity Controller Keys and Displays

The actual chamber workspace temperature is indicated on the Upper Display.


The actual chamber workspace humidity is displayed on the Lower Display at the Input2
prompt.
The temperature set point is entered at the SP1 prompt in the Lower Display.
The humidity set point is entered at the SP2 prompt in the Lower Display.

TestEquity 1207C Temperature/Humidity Chamber

Page 3-5

Chapter 3 Operation

Static Set Point Control


The Temperature/Humidity Controller is in Static Mode when it is not controlling a Profile.
When in a Static Mode, the Profile Indicator Light is off.
The Static Set Point prompt is accessed from the Main Page on the Lower Display.

To enter a Static Temperature Set Point:


1. Press the or key to position the Cursor next to the SP1 prompt. You may already be at
this prompt.
2. Press the key once. You will see Static Set Point1 in the lower display with the current
set point indicated below.
3. Press the or key to change the temperature set point.
4. Press the key once to enter the new temperature set point. You are now back to the Main
Page.

To enter a Static Humidity Set Point:


1. Press the or key to position the Cursor next to the SP2 prompt. You may already be at
this prompt.
2. Press the key once. You will see Static Set Point2 in the lower display with the current
set point indicated below.
3. Press the or key to change the humidity set point.

Page 3-6

4. Press the
Page.

key once to enter the new humidity set point. You are now back to the Main

TestEquity 1207C Temperature/Humidity Chamber

Chapter 3 Operation

Digital Event Outputs


The Temperature/Humidity Controller has digital outputs which are configured as Event Outputs
to turn remote devices on and off. There are seven Event Outputs which are available for
customer use. Output number eight is configured to control the refrigeration compressors and is
not available for customer use. The Event Output control is accessed from the Main Page.

To control the Event Outputs:


1. Press the or key to position the Cursor next to the DigitalOut prompt.
2. Press the key once. You will see Choose Event Output prompt in the lower display.
3. Press the or key to position the Cursor next to the EventOutput prompt which you
want to change. You will be able to select from EventOutput1 through EventOutput7.

4. Press the key once to select the desired EventOutput. Then, press the or key to
select On or Off.
5. Press the key once to return to the Choose Event Output prompt.
6. Press the key once to return back to the Main Page.

Digital Output Connections


See page 12.9 of the Series F4 Users Manual for details on how to connect to the Digital
Outputs.

Event Board Option


Your chamber may have been ordered with the Event Board option. The event board contains
solid state relays to control AC operated devices.
The solid state relays are rated for 24 to 240 VAC, 3.0 Amps. However, TestEquity recommends
derating the maximum current to 2.5 Amps. The solid state relay mounting board is UL
recognized/CSA certified for 120 VAC max. with the board-mounted fuses; 250 VAC max. with
a #22 solid copper jumper wire instead of the fuses.
Connecting to the Event Board
1. Turn the Main Disconnect switch to the OFF position.
2. Remove the lower door retaining screws located on the right side. Open the lower door.
3. The Event Board is located on the top of the lower door. Connect to the event outputs as
follows:
Event Output
Event 1
Event 2
Event 3
Event 4
Event 5
Event 6
Event 7

Event Board Terminals


1, 2
3, 4
5, 6
7, 8
9, 10
11, 12
13, 14

Event Board Terminals

4. The lead wires can be brought out through


the access hole on the left side of the
chamber (next to the interface connector).
TestEquity 1207C Temperature/Humidity Chamber

Page 3-7

Chapter 3 Operation

Profile Programming
The Series F4 Controller can be programmed to store up to 256 steps into as many as 10 profiles.
You do not need a computer to enter a profile it can be easily done through the controllers
front panel keys.
For detailed information about how to program a profile, see Page 4.1 of the Series F4 Users
Manual.
For detailed information about how to run a profile, see Page 3.2 of the Series F4 Users
Manual.
! CAUTION:

The sample program in the Series F4 Users Manual is not applicable to this
chamber.

! CAUTION:

The Series F4 Controller has been properly configured by TestEquity to match


the chambers system requirements and to perform optimally over a wide
range of operating conditions. Improper modifications to these setup values
can result in erratic performance and unreliable operation. Setup examples in
the Series F4 Users Manual are not applicable to this chamber. Do not
attempt to modify the setup values, unless you thoroughly understand what
you are doing. If there is any doubt, please call TestEquity before proceeding.

Programming Hints:
The first step in a program should be an initialization step of 1-second.
The next to last step establishes a condition to end on. For example, you may want to end the
program by holding at +23C, so this step would be to go to +23C.
TestEquity recommends having the end step type to be Hold or Idle.
TestEquity does NOT recommend using an end step type of All Off or Control Off. This
does not turn off the chamber fan. The chamber temperature can reach +55C just from heat
generated by the fan, and even higher if your test sample is energized.

Page 3-8

TestEquity 1207C Temperature/Humidity Chamber

Chapter 3 Operation

Temperature Profile Examples

The following are examples of two typical temperature-only profiles.

1
>
>
>
>
>
>

Choose Digit Out


2 3 4 5 6
> > > > >
> > > > >
> > > > >
> > > > >
> > > > >
> > > > >

7
>
>
>
>
>
>

H:M:S
00:00:01
00:30:00
00:00:01
00:30:00

00:00:01

Ch1 Ch2
SP
SP
-40.0
>

>
85.0
>

>

23.0
>

PID Set
Ch1
Ch2
1
6
1
6
1
6
1
6

1
6

-40C to 85C, uncontrolled humidity, no ramps, 30 minute soak time, 4 cycles. Ends with +23C set point.

Wait for
>
>
>
>
>
>

Description
Goes to -40C as quickly as possible.
Will hold for 30 minutes. Time will not start until chamber reaches -37C (within the 3 Guaranteed Soak Band*).
Goes to +85C as quickly as possible.
Will hold for 30 minutes. Time will not start until chamber reaches +82C (within the 3 Guaranteed Soak Band*).
Jumps back to step 1. Repeats this 3-times, for a total of 4-cycles.
Goes to +23C as quickly as possible.
End of program. Controller returns to a +23C manual set point.

Step
Step
Date
Number
Type
Day, Time
1
Ramp Time

2
Soak

3
Ramp Time

4
Soak

5
Jump

6
Ramp Time

7
End

Step
1
2
3
4
5
6
7

1
On
On
On
Off
Off
On
>
Off

Choose Digit Out


2 3 4 5 6
> > > > >
> > > > >
> > > > >
> > > > >
> > > > >
> > > > >
> > > > >
> > > > >

7
>
>
>
>
>
>
>
>

Time
H:M:S
00:00:01
00:13:30
01:00:00
00:25:00
01:00:00
00:25:00

00:13:30

Ch1
SP
23.0
50.0

0.0

50.0

23.0

Ch2
SP
>
>
>
>
>
>

>

PID Set
Ch1
Ch2
1
6
1
6
1
6
1
6
1
6
1
6

1
6

Guar. Soak Jump to


Ch1
Ch2 Profile Step
No
No

Yes
No

No
No

Yes
No

Name
1
No
No

Guar. Soak Jump to


Ch1
Ch2 Profile Step
No
No

No
No

No
No

No
No

No
No

No
No

Name
2
No
No

End
Repeats Type

Idle

End
Repeats Type

99

Idle

Idle SP
Ch1

23.0

Idle SP
Ch1

23.0

+50C to 0C, 2C/minute ramp, uncontrolled humidity, 1-hour soak times, 100 cycles total, test sample turned ON/OFF depending on step, end with a +23C set point.

Wait for
>
>
>
>
>
>
>
>

Description
Establishes an initialization step. Goes to +23C as quickly as possible. Test sample is turned ON via Event 1.
Goes to +50C at a controlled rate of 2C per minute (50-23=27/2=13.5 or 13 minutes and 30 seconds). Test sample is turned ON via Event 1.
Will hold for 1 hour. Test sample remains ON via Event 1.
Goes to 0C at a controlled rate of 2C per minute (50-0=50/2=25 or 25 minutes). Test sample is turned OFF via Event 1.
Will hold for 1 hour. Test sample remains OFF via Event 1.
Goes to 50C at a controlled rate of 2C per minute. Test sample is turned OFF via Event 1.
Jumps back to step 3. Repeats this 99-times, for a total of 100-cycles.
After all cycles are completed, establishes the condition to end on. Goes to +23C at a controlled rate of 2C per minute. Test sample is turned OFF via Event 1.
End of program. Controller returns to a +23C manual set point. Test sample remains OFF via Event 1.

Step
Step
Date
Number
Type
Day, Time
1
Ramp Time

2
Ramp Time

3
Soak

4
Ramp Time

5
Soak

6
Ramp Time

7
Jump

8
Ramp Time

9
End

Step
1
2
3
4
5
6
7
8
9

Notes
*
Guaranteed Soak Band is set at the factory for 3.0C. Can be changed in the System Menu.
>
Means no entry or selection is required. Just scroll through this prompt to the next prompt.
Means prompt does not show for this step type.

Page 3-9

TestEquity 1207C Temperature/Humidity Chamber

Chapter 3 Operation

RS-232C Interface
The Temperature/Humidity Controller has an RS-232C interface. A DB-9 connector is located
on the front left side. It is wired to accommodate a null-modem cable.
To communicate with the controller from a PC, you need to run software that uses the Modbus
RTU protocol. industrial data communications. This protocol works on the principle of packet
exchanges. The packet contains the address of the controller to receive the information, a
command field that says what is to be done with the information and several fields of data.
Reading from these registers retrieves all information in the controller. These registers are listed
Chapter Seven of the Series F4 Users Manual.
RS-232C and Modbus programming resources can be downloaded from
http://chamber.testequity.com/rs232.html .

GPIB Interface (optional)


The optional GPIB interface consists of an internal board which converts the controllers serial
interface to GPIB. GPIB programming resources can be downloaded from
http://chamber.testequity.com/gpib.html .

Common Modbus Registers

The actual chamber temperature reading is Modbus register 100 (Input 1 Value).
The actual chamber humidity reading is Modbus register 104 (Input 2 Value).
The static temperature setpoint is Modbus register 300 (Set Point 1).
The static humidity setpoint is Modbus register 319 (Set Point 2).
The temperature setpoint during a profile is Modbus register 4122 (Set Point 1, Current
Profile Status).
The humidity setpoint during a profile is Modbus register 4123 (Set Point 2, Current Profile
Status).
The decimal points are implied. For example, 1005 is actually 100.5 and -230 is -23.0.

Page 3-10

TestEquity 1207C Temperature/Humidity Chamber

Chapter 3 Operation

Limit Controller
Overview
The Series 97 Limit Controller is a protection device. It turns the chamber OFF if the workspace
temperature exceeds either a high temperature or low temperature limit. You can set these limits
to correspond to the maximum and minimum temperature that your test sample can safely
withstand. This provides protection against someone setting the Temperature/Humidity
Controller to a temperature that is unsafe for the test sample. It also provides protection in the
unlikely event of a chamber system component failure. The Limit Controller has its own
temperature sensor (thermocouple) and functions completely independent of the
Temperature/Humidity Controller.
This section provides a brief overview on how to operate the Limit Controller. For more detailed
instructions, see the Series 97 Users Manual.
! CAUTION:

The Series 97 Users Manual is a general manual and is written by the


manufacturer, Watlow, for a wide variety of applications and configurations.
Not all features or functions are applicable. Only the capabilities of a model
97A0-DAAA-00-RG, as described on page A.20 of the Series 97 Users
Manual are applicable. Please note that only the Series 97 Limit functions
are applicable.

! CAUTION:

Always verify that the Limit Controllers high and low limits are set to
temperatures that are appropriate for your test sample.

! CAUTION:

If your test sample is energized, it may be capable of raising the workspace


temperature beyond safe limits. This could occur if your test sample exceeds
the live load rating of the chamber or if the chambers refrigeration system
fails. This chamber has a set of contacts that can be used to remove power to
your test sample if the Limit Controllers temperature limits are exceeded.

TestEquity 1207C Temperature/Humidity Chamber

Page 3-11

Chapter 3 Operation

Limit Controller Keys and Displays


Active Output Indicator Lights
Lit when the corresponding output trips.
Only output 1 is applicable for this configuration.
Upper Display
Indicates actual process values during
operation, the value for the parameter
in the lower display, or the user
programmed message.

Lower Display
Indicates factory programmed message
during operation, the value for the
parameter in the lower display, or the user
programmed message.
Advance Key
Advances the lower display through
the parameters. To reverse direction,
press and hold the Advance Key while
repeatedly pressing the UP Key.

Up Key
Changes the upper display to a higher
value, or up through a list of values.
Moves from menu to menu in a page.

Down Key
Changes the upper display to a lower
value, or down through a list of values.
Moves from menu to menu in a page.

Reset Key
Returns to the Home Page
(process/actual display).
Resets the limit.
Resets an input sensor error.

Figure 4.1 Limit Controller Keys and Displays

How to Set the High and Low Temperature Safety Limits


1. Press the and Keys together for three seconds, until the [OPER] prompt appears in the
Lower Display. The Upper Display should say [LIM].
2. Press the Key once to get the [L~Lo] prompt. This is the Low Limit Set Point prompt.
3. Press the or Key to enter the desired Low Limit Set Point. Make sure it is at least a few
degrees lower than your lowest temperature set point to prevent nuisance tripping.
4. Press the Key once to get the [L~Hi] prompt. This is the High Limit Set Point prompt.
5. Press the or Key to enter the desired High Limit Set Point. Make sure it is at least a few
degrees higher than your highest temperature set point to prevent nuisance tripping.
6. Press the RESET Key to save these values and return to the Home Page.
If you press the Key after the [L`Hi] prompt, you will get the [CAL1] prompt, which is the
Calibration Offset. This is an adjustment to eliminate a difference between the indicated value
and the actual process value.

Resetting a Limit Condition


If either the High or Low Limit is exceeded, the Limit Controller will indicate either [L~Hi] or
[L~Lo] respectively in the lower display. It will also shut down all chamber functions and
sound the Audible Alarm. The Limit Controller cannot be reset until the temperature returns to
within the limit set points. Then, you must press the RESET Key to resume normal operation.

Silencing the Audible Alarm


The SILENCE switch on the chamber front panel lets you temporarily turn off the Audible
Alarm, even though the High or Low Limit condition may still exist.

Page 3-12

TestEquity 1207C Temperature/Humidity Chamber

Chapter 3 Operation

Protecting an Energized Test Sample


If your test sample is energized, it may be capable of raising the workspace temperature beyond
safe limits. This could occur if your test sample exceeds the live load rating of the chamber or if
the chambers refrigeration system fails.
This chamber has a set of safety contacts that can be used to remove power to your test sample if
the Limit Controllers temperature limits are exceeded.
The safety contacts are rated as follows:
Resistive: 10 A, 250 VAC or 10 A, 28 VDC
Inductive: 7 A, 250 VAC
To access the safety contacts:
1. Turn the Main Disconnect Switch to the OFF position.
2. Remove the lower door retaining screws located on the right side. Open the lower door.
3. Locate the terminal strip on the lower left side. Connections to the safety contacts are at
terminals 8 and 9.

TestEquity 1207C Temperature/Humidity Chamber

Page 3-13

Chapter 3 Operation

Loading the Chamber


! WARNING: Do NOT put items in the chamber that could burn or explode at high
temperatures. This chamber uses open wire heating elements which generate
surface temperatures over 1000F. This is NOT an explosion-proof chamber.
! WARNING: Do NOT put items in the chamber which can emit corrosive vapors or
substances.
! WARNING: This chamber is NOT a curing oven. There are NO provisions for venting
fumes.
! WARNING: The chamber door must remain closed while the chamber is operating. If you
need to open the door while the chamber is operating, wear safety goggles to
prevent the high velocity airflow from blowing particles or objects into your
eyes.
! WARNING: This chamber operates at extreme temperatures. Avoid contact with air,
objects, and surfaces that are hot or cold to prevent severe burns or frostbite.
Protective gloves are recommended.
! CAUTION:

If your test sample is energized, it may be capable of raising the workspace


temperature beyond safe limits. This could occur if your test sample exceeds
the live load rating of the chamber or if the chambers refrigeration system
fails. This chamber has a set of contacts that can be used to remove power to
your test sample if the Limit Controllers temperature limits are exceeded.

! CAUTION:

To prevent damage to your test sample and the chambers compressors, do not
exceed the live load rating of the chamber.
Live Load Capacity for Model 1207C (non-humidity mode)
Temp +23C
0C
40C 55C 65C
Watts 1000 W 800 W 500 W 400 W 300 W

Performance Considerations
The performance of all chambers are significantly effected by the characteristics of your test
sample. Factors include size, weight, material, shape, and power dissipation if energized.
The test sample should be placed in the chamber in a manner that allows for air circulation. The
air plenum is located on the back wall of the chamber, where air is sucked in from the bottom
and exits from the top. You should not place the test sample directly on the chamber floor. It
should be placed on the shelf. Multiple test samples should be distributed throughout the
chamber to ensure even airflow and minimize temperature gradients. If necessary, additional
shelves should be used to evenly distribute the load. Verify that the temperature gradients are
within acceptable limits, by measuring the chamber temperature at strategic points using a
multipoint thermocouple meter or datalogger.

Page 3-14

TestEquity 1207C Temperature/Humidity Chamber

Chapter 3 Operation
You may find that the temperature throughout the chamber is even, but always different from
what the Temperature/Humidity Controller indicates. The correct way to adjust what the
Temperature/Humidity Controller displays compared to what is measured at some point other
than the controllers sensor is with the Calibration Offset parameter. See page 6.2 of the
Series 97 Users Manual for details.

Avoiding Moisture (non-humidity mode)


Any time the ambient air is subjected to temperatures below the dewpoint, moisture will
condense out of the air. The effect is ice or frost during low temperature operation. When the
chamber is heated above 0C, the ice or frost will turn into water.
To avoid moisture condensation, make sure the port plugs are inserted at all times and properly
seal around any wires that go through the port. Also, avoid opening the chamber door while the
chamber is operating at temperatures below room ambient. When a low temperature test is
completed, warm the chamber to at least room ambient before opening the chamber door and
before removing your test sample.
You CANNOT use the humidity system to control moisture at low temperatures. The humidity
system is disabled below 7C.

Internal Test Fixtures


Some applications require internal fixtures to support test samples and provide a convenient
method of connecting wires and sensors. Fixtures must be designed to minimize their impact on
chamber functionality and performance.
Fixtures should be designed for easy removal to permit maintenance and cleaning of the
chamber. The chamber liner should never be drilled or screwed into. This will compromise the
integrity of the liner and permit moisture migration due to condensation into the insulation,
which will eventually impact performance and lead to premature rusting of the outer cabinet.
Fixtures should be constructed of stainless steel. This also applies to all screws and fasteners. All
welds should be passivated. To prevent rust and corrosion, never use iron or mild steel even if it
is painted or plated. Aluminum may be used. However, since the specific heat of aluminum is
double that of steel, it represents a greater load and will have more impact on the chamber
performance.
Make sure that all connectors, wiring, pc boards, and auxiliary components can withstand the
temperature extremes that they will be subjected to. In some cases, these components may not be
able to last after repeated tests and should be considered expendable.

TestEquity 1207C Temperature/Humidity Chamber

Page 3-15

Chapter 3 Operation

Humidity Operation
Humidity Range
This chamber is capable of controlling humidity from 10% RH to 95% RH over the temperature
range of +10C to +85C. The lowest possible humidity varies depending on the temperature.
For example, +45C is the lowest temperature that 10% RH can be achieved. Below is a chart
that shows the achievable range of humidity as compared to temperature. Achieving low
humidity levels require you to start with a clean, dry chamber.
100

RELATIVE HUMIDITY (% RH)

90
80
70
Achievable Range of
Temperature/Humidity
Conditions

60
50
40
30
20
10
0
0
32

10
50

20
68

30
86

40
104

50
122

60
140

70
158

80
176

90
194

C
F

TEMPERATURE

Humidity Mode Considerations


This chamber is in controlled humidity mode when the Humidity Enable switch is ON. Also, the
chamber workspace must be between +7C and +90C. If the chamber is outside this temperature
range, the Temperature/Humidity Controllers Alarm2 light will turn ON and the chamber will
operate as though the Humidity Enable switch was OFF.
It takes several minutes for the humidifier to fill to the correct level after the water supply is
initially connected to an empty system. The humidifier function is disabled until the water level
is correct.
If you are running a high humidity condition, it can take approximately 15-20 minutes until the
humidifier heats up from an initial cold start. During most of that time, it might appear that
nothing is happening because there is little increase in the humidity reading (the Input2
prompt) on the Temperature/Humidity Controller. Be patient - as long as the Humidity Light is
ON, then the humidity will eventually begin to rise after this initial heat-up period.

Page 3-16

TestEquity 1207C Temperature/Humidity Chamber

Chapter 4 Frequently Asked Questions

Chapter 4 Frequently Asked Questions


The input voltage label says 230 (or 208) VAC. I thought I had 220 (or 240) VAC. Is that
ok?
220 V is a misnomerthere is no such standard as nominal 220 V in the United States. You
must verify the exact type of electrical service you have. If there is any doubt, you must consult
with a qualified electrician who is familiar with industrial plant wiring. In addition, the input line
voltage should be measured while the chamber is operating in the COOL mode to ensure that the
expected nominal voltage of either 208 V 5/+10% or 230 V 10% is present. Also, make sure
the chamber is properly configured for either 208 V or 230 V nominal input as described in
Chapter 2 - Input Power Configuration. If you have a 208 V line that measures under 198 V, the
chamber will require boost transformers. If you have a 240 V line that measures over 252 V (a
240 V line that is 10% high could measure up to 264 V), the chamber will require bucking
transformers. These transformers are available from TestEquity (part number TE-0702F) for a
reasonable cost and are easy to install in the field.
Why doesnt the chamber come with a power cord and plug?
Most local electrical codes require permanent wiring for this type of equipment. If used as a
portable device, a flexible wire with a plug may be acceptable, but local codes may limit the
length to 6 feet. TestEquity recommends that the appropriate method for your installation be
determined by a qualified electrician who is familiar with industrial plant wiring.
I need to send the chamber outside North America. Will it work with their power?
Outside North America, most countries have 50 Hz. Standard voltages in Europe are typically
220 V to 240 V, while Japan is 200 V. Operation at 220 V to 240 V / 50 Hz requires bucking
transformers to lower the voltage at the compressors to their 50 Hz rating of 200 V. These
transformers are available from TestEquity for a reasonable cost and is easy to install in the field.
Note that the cooling performance will be reduced by 17% at 50 Hz. Please call TestEquity for
details on reconfiguring for 50 Hz operation.
Why does my chamber heat or cool slower than the published specifications?
Performance is significantly effected by the characteristics of your test sample. Factors include
size, weight, material, shape, and power dissipation if energized. The test sample should be
placed in the chamber in a manner that allows for air circulation. You should not place the test
sample directly on the chamber floor. It should be placed on the shelf. Multiple test samples
should be distributed throughout the chamber to ensure even airflow and minimize temperature
gradients. If necessary, additional shelves should be used to evenly distribute the load. You can
determine if the chamber is operating properly by following the procedure in How to verify the
chamber performance.
How can I modify the chamber to cool faster?
Unfortunately, there is little you can do to improve upon the designed-in performance.
TestEquity does NOT recommend using CO2 or LN2 in this chamber to achieve faster cooling
due to reliability and safety considerations, so it is NOT an available option. Modifying the
chamber to add CO2 or LN2 will void the warranty.

TestEquity 1207C Temperature/Humidity Chamber

Page 4-1

Chapter 4 Frequently Asked Questions


Why is there water/ice/snow in the chamber?
Any time the ambient air is subjected to temperatures below the dewpoint, moisture will
condense out of the air. The effect is ice or frost during low temperature operation. When the
chamber is heated above 0C, the ice or frost will turn into water. To avoid moisture
condensation, make sure the port plugs are inserted at all times. Also, avoid opening the chamber
door while the chamber is operating at temperatures below room ambient. When a low
temperature test is completed, warm the chamber to at least room ambient before opening the
chamber door and before removing your test sample.
My test specification requires convection heat only. Can I turn the circulator motor off?
NO! This will damage the heating and refrigeration systems and void the warranty. You need a
gravity convection oven for that kind of test.
How accurate is the chamber?
Thats a loaded question! There is no chamber accuracy specification as such. The answer
requires an understanding of several performance parameters.
Control Tolerance The Temperature/Humidity Controller uses a thermocouple control sensor,
which is located in the discharge airflow. Control tolerance is a measure of how much the
temperature varies after stabilization at the control sensor. It is a measure of the relative
variations, NOT the absolute accuracy of the readout. The control tolerance specification for this
chamber is 0.5C, or a total of 1C. For example, the temperature set point may be 25.0C.
The actual temperature varies between 25.4C and 24.5C. This corresponds to 0.4C and
+0.5C or a total of 0.9C of RELATIVE variations. These specifications are for an empty
chamber. The addition of a test sample may effect the control variations. In some instances, the
test sample will reduce these variations.
Uniformity Also known as Gradients. This is a measure of variations in temperature at
different locations throughout the chamber interior, at the same time, after stabilization. The
uniformity specification for this chamber is 0.5C or a total of 1C, when measured at least 2"
away from the chamber interior walls. These specifications are for an empty chamber. The
addition of a test sample may effect the temperature uniformity. For example, an energized test
sample will produce a higher temperature near the sample.
Controller Accuracy This is the ability of the Temperature/Humidity Controller to accurately
display a temperature measurement when compared to a standard. The controller display
accuracy is 0.65C, 1 LSD. However, the total measurement accuracy in the chamber includes
the thermocouple sensor wire accuracy. Thermocouple wire accuracy is 1C or 0.75% of
reading, whichever is greater. Therefore, total system accuracy over the chambers typical
operating range is typically 1.65C, 1 LSD. This is not a measurement of chamber
performance.
Can I tilt the chamber to move it?
You should be able to tilt the chamber 45 degrees to move it. After tilting it and moving it into
place, perform the steps as outlined in How to inspect the refrigeration machinery
compartment and How to check the refrigerant charge before placing the chamber back into
service to make sure that no damage has occurred.

Page 4-2

TestEquity 1207C Temperature/Humidity Chamber

Chapter 4 Frequently Asked Questions

Im not going to use the chamber for a while. Is there anything I should do to prepare it for
storage?
Perform ALL the steps in the Preventive Maintenance Schedule before placing the chamber into
storage. This will ensure that the chamber will be ready to operate when it is taken out of storage.
If the chamber has a problem and is still under warranty, these problems should be resolved
before being placed into storage, since the warranty period starts from the date of shipment. The
chamber should be stored in a conditioned environment. Do not store it outside or where it will
be subjected to dirt or excessive moisture.
I havent used the chamber for a while. Is there anything I should do to prepare it for
operation?
Perform ALL the steps in the Preventive Maintenance Schedule before placing the chamber back
into service. This will ensure that nothing has been damaged and that a leak has not developed.
This chamber has a crankcase heater to protect the high-stage compressor. The chamber must be
connected to the power source AND the Main Disconnect Switch must be ON for 3 hours prior
to operating the chamber. Although it may be safe to use the chamber immediately, this
procedure ensures the longest possible life for the high-stage compressor if the chamber has been
removed from the power source for more than 24 hours.
Can the person who services our air conditioning also service the chamber?
Probably not. Most air conditioning mechanics are not familiar with low-temperature cascade
refrigeration systems. While this chamber is relatively easy to maintain and repair, most air
conditioning mechanics do not have the necessary refrigerants and may not be familiar with the
microprocessor-based controls. This chamber should only be serviced by a qualified mechanic
that is familiar with low-temperature cascade refrigeration systems. Call TestEquity to
recommend one in your area or to check if the one you would like to use is qualified.
Can/Should I put a filter in front of the condenser air inlet?
No, TestEquity does not recommend this. Just follow the maintenance procedures and clean the
condenser fins periodically.
How often should I charge the refrigeration system?
This chamber uses a closed-loop refrigeration system. Just like your refrigerator at home, it does
not need periodic charging. If the charge is low, this means that there is a leak. Leaks should be
repaired before recharging.
What kind of Freon does the chamber use?
The word Freon is a DuPont registered trade name for their CFC-based refrigerants and is
incorrectly used as a generic term for refrigerants. TestEquity chambers do not use CFC-based
refrigerants. The high-stage system uses R-404A, which is also known as DuPont Suva HP62.
The low-stage system uses R-508B, which is also known as DuPont Suva 95.

TestEquity 1207C Temperature/Humidity Chamber

Page 4-3

Chapter 5 Specifications

Chapter 5 Specifications
Model 1207C Chamber Specifications
Temperature Range
Control Tolerance
Uniformity

75C to +175C
0.5C (0.2C Typical) (Measured at the control sensor after stabilization)
0.5C (Variations throughout the chamber after stabilization)

Humidity Range

10%RH to 95%RH, in the temperature range of 10C to 85C.


Low humidity range limited by a 6C dewpoint.
2%RH (Measured at the control sensor after stabilization)

Control Tolerance
Live Load Capacity @

+23C
0C
1000 W
800 W
(Live load capacity for non-humidity mode)

40C
500 W

55C
400 W

65C
300 W

+23C
18 min

End Temp
0C
40C
25 min
44 min

55C
54 min

65C
62 min

75C
ultimate

+23C
----1.5 min
6 min
8 min
10 min

End Temp
+50C
+85C
1.5 min
7 min
3.5 min
13 min
11 min
17 min
13 min
19 min
14 min
21 min

+125C
14 min
20 min
24 min
26 min
28 min

+150C
20 min
23 min
30 min
32 min
34 min

+175C
25 min
31 min
35 min
37 min
39 min

Cool Down Transition Time*


From +85C to

Heat Up Transition Time*


Start Temp to
+23C
0C
40C
55C
65C

*Note: Transition times are measured after a 1-hour soak at the start temperature with an empty chamber, in nonhumidity mode.
Input Power Requirements
230 V 10%, 60 Hz

Phase
1 PH
3 PH

208 V -5/+10%, 60 Hz 1 PH
3 PH

Current
Draw
25 A
19 A

Minimum
Service
30 A
25 A

28 A
21 A

35 A
30 A

Input may be configured for single or three phase in the field by changing jumpers. Three phase load is semibalanced. Operation below 198 V requires internal boost transformers, which can be supplied for a nominal charge.
Call for other voltages or 50 Hz operation.
Workspace Dimensions
24" W x 21" H x 24" D (7 cubic feet)
Outside Dimensions
33.5" W x 68" H x 48" D (nominal)
Door latch adds 3" to width on right side (may be removed to permit move-in through a 36" doorway). Circulator
motor housing adds 6" to height.
Min. Installed Clearance 12" from the left and right side, 24" from the rear
Access Ports
4" Port on left and right side (two total), Supplied with foam plugs
Weight
850 pounds
NOTE: Performance is typical and based on operation at 23C (73F) ambient and nominal input voltage. Designed
for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased
cooling performance. Additional ports and shelves will also affect performance. Operation above 30C (85F) or
below 16C (60F) ambient is not recommended.
TestEquity 1207C Temperature/Humidity Chamber
Page 5-1

Chapter 5 Specifications

F4 Controller Specifications
Specifications as configured for the TestEquity 1207C Chamber
Accuracy & Sensor Conformity* 0.65C, 1 LSD with type T thermocouple input
Stability

0.1C/C rise in ambient

Digital Inputs

(Four) Contact closure or dc voltage, 10 k impedance

Retransmit Outputs (Optional)

(Two) User-selectable ranges:


0 to 10 VDC, 0 to 5 VDC, 1 to 5 VDC
0 to 20 mA, 4 to 20 mA

Alarm Outputs

(Two) Electromechanical relay;


Form C, 2 A @ 20 VDC or 240 VAC max.
(Alarm2 is used for system functions and not available for
customer use.)

Digital Outputs

(7 available for customer use) Open collector output


OFF: 42 VDC @ 10 A max.
ON: 0.2 VDC @ 50 mA sink max.
Internal supply: 5 VDC @ 80 mA

Communications

EIA-232 and EIA-485 serial communications with


Modbus RTU protocol

Safety & Agency Approvals

UL/c-UL 916-listed, File #E185611


CE to EN61010
NEMA 4X and IP65
CE EMC to EN50082-2
CE EMC to EN55011

Displays

Process: 5, seven-segment red LED.


Interface Display: 4-line high-definition green LCD;
selectable C or F

Data Retention

Retention upon power failure via nonvolatile memory


(seven years for battery-backed RAM)

*Note: Total system accuracy in the chamber includes thermocouple wire accuracy.
Thermocouple wire accuracy is 1C or 0.75% of reading, whichever is greater. Therefore, total
system accuracy over the chambers operating range is typically 1.65C, 1 LSD.

Page 5-2

TestEquity 1207C Temperature/Humidity Chamber

Chapter 5 Specifications

Humidity Sensor Specifications


Type

Rotronic Model H290 Capacitive Humidity Sensor

Humidity Measuring Range

0 to 100% RH

Sensor Operating Range

-5C to +160C

Survival Temperature Range

-75C to +200C

Humidity Limits

100% RH at +85C
90% RH at +125C
25% RH at +160C

Accuracy

2% RH (at 20C to 25C)

Repeatability

0.5% RH

Humidity Sensor Stability

Better than 1% RH over a year

Calibration Adjustment Points

10%, 35%, 80% RH

Calibration Device

ER-15 Fixture
EA10, 10% RH Standard Salt Solution
EA35, 35% RH Standard Salt Solution
EA80, 80% RH Standard Salt Solution
Calibration devices available from:
Rotronic Instrument Corp.
160 E. Main Street
Huntington, NY 11743
631-427-3898
800-628-7101
www.rotronic-usa.com

TestEquity 1207C Temperature/Humidity Chamber

Page 5-3

Chapter 6 Maintenance

Chapter 6 Maintenance
! WARNING: Maintenance must be performed by properly trained personnel only.

Preventive Maintenance Schedule


Daily or As Needed

Clean chamber interior and exterior.


Listen for abnormal noise or vibration.
Inspect the demineralizer cartridge (if provided).

Every 3 Months

Inspect the door seal.


Inspect the refrigeration machinery compartment.
Verify the chamber performance.

Every 6 Months

Inspect the electrical compartment.


Clean the condenser.
Drain the humidity system.

Every 12 Months

Verify the calibration.

How to clean the chamber interior and exterior

Wipe or vacuum out all debris.

Clean surfaces with a damp cloth, mild detergent, or stainless-steel cleaner. Avoid cleaners
that are abrasive or leave a residue. Do NOT use steel wool.

If you clean the interior with something other than water, you may want to operate the
chamber at high temperature (approximately +125C) after cleaning. This helps to bake out
any residue. Remove the port plugs to permit the residual vapors to escape.

Clean the silicone door gaskets with a damp cloth or mild detergent.

Clean the exterior painted surfaces with a damp cloth or mild detergent. If you are using a
detergent, test a small inconspicuous area to make sure it does not damage the finish.

How to listen for abnormal noise or vibration


You should become familiar with normal operating noises. Being able to recognize changes from
normal operating noises can be a valuable way to identify problems and prevent further damage.
Examples of noises to be aware of include:
Circulator motor and fan noises (with compressors off).

Compressor start-up and running noises, sequential starting of compressors.

Condenser and desuperheater fan noises.

Relay and valve cycling noises when cool light is cycling.

TestEquity 1207C Temperature/Humidity Chamber

Page 6-1

Chapter 6 Maintenance

How to inspect the filter cartridge


The filter cartridge color changes gradually to a yellow-amber through continued use. When all
but 2-inches have completely changed color, it must be replaced. Replacement cartridges are
available from TestEquity (part number TE-0011 for 6 cartridges). Follow the instructions in
Chapter 2, Water Recirculation System Assembly.
The humidity system should be drained whenever the demineralizer filter cartridge is replaced.

How to inspect the door seal


The door has two silicone gaskets to minimize thermal losses and moisture migration.
Inspect the gaskets for dirt and tears.

Repair minor tears with a high quality RTV silicone such as GE RTV167.

Check the integrity of the door seal by closing the door on a sheet of paper. With the door
closed, slowly pull the paper. You should feel the resistance getting lighter as the paper goes
past the inner gasket. Repeat this all around the door at several places.

If the seal is not tight, adjust the door latch. The stainless-steel catch (on the bracket that is
mounted to the chamber) has slotted holes to permit adjustment.

If the seal is still loose on the hinge side, adjust the door hinge. The hinges have slotted holes
(on the door side) to permit adjustment.

How to inspect the refrigeration machinery compartment


! WARNING: Wear safety goggles when inspecting the machinery compartment to protect
against a refrigerant line which could break.
1.
2.
3.
4.
5.
6.
7.

Turn the Main Disconnect Switch OFF.


Remove the side and rear panels.
Inspect for signs of water leaks.
Inspect for signs of refrigeration tubing abrasion.
Inspect for oil around refrigeration valves, fittings and joints. This may be a sign of leaks.
Inspect for loose hardware and tighten as required.
Inspect for signs of insect or rodent infestation. Yes, it does happen!

How to inspect the electrical compartment


1.
2.
3.
4.

Disconnect the chamber from the power source.


Turn the Main Disconnect Switch to the OFF position.
Remove the lower door retaining screw located on the right side. Open the lower door.
Check for loose components, loose wires, burned insulation near terminals, and burned or
excessively pitted contacts on contactors.

Page 6-2

TestEquity 1207C Temperature/Humidity Chamber

Chapter 6 Maintenance

How to clean the condenser


1.
2.
3.
4.

Disconnect the chamber from the power source.


Turn the Main Disconnect Switch to the OFF position.
Remove the lower door retaining screw located on the right side. Open the lower door.
Clean the condenser and desuperheater fins with a vacuum cleaner.

NOTE: You may need to clean the condenser more frequently if the chamber is in a dusty
environment. You may be able to clean the condenser less frequently if the chamber is in a very
clean environment.

TestEquity 1207C Temperature/Humidity Chamber

Page 6-3

Chapter 6 Maintenance

How to verify the chamber temperature performance


These tests verify the performance of the heating, refrigeration, electrical controls,
Temperature/Humidity Controller, and air circulation systems. The chamber should meet all
published performance specifications if all of these tests are successfully passed.
These tests assume that the Temperature/Humidity Controllers setup and tuning values have not
been changed from the values as shipped from TestEquity. Also, the Limit Controller high limit
must be set to over +85C (+88C would be fine), and the low limit set to 75C.
If the chamber fails any of these tests, it should be removed from service to prevent further
damage until the cause of the problem is determined and resolved.
1. The chamber interior should be empty and at ambient temperature, approximately +23C.
2. The Humidity Enable Switch must be OFF.
3. Set the Temperature/Humidity Controller Set Point (SP1) to +85C and turn the Master
Switch ON.
4. The Heat Light should be ON continuously and the Cool Light should be OFF.
5. The chamber should heat up to about +80C and begin controlling (Heat Light cycles
ON/OFF) within 10 minutes.
6. The chamber temperature should slowly increase and stabilize to +85C. It should NOT
overshoot beyond +85C by more than a few tenths of a degree, and the compressors should
NOT need to turn ON in order to maintain +85C.
7. After stabilization, the chamber temperature should vary no more than 0.5C, or a total of
1C.
8. Let the chamber stay at +85C for one hour.
9. After two hours at +85C, set the Temperature/Humidity Controller Set Point to 65C.
10. The high-stage compressor should turn ON within a few seconds. Then, the low-stage
compressor should turn ON within approximately 30 seconds, and the Cool Light should be
ON continuously and the Heat Light should be OFF.
11. The chamber should cool down to about 60C and begin controlling (Cool Light cycles
ON/OFF) within approximately 1 hour.
12. The chamber temperature should slowly decrease and stabilize to 65C. It should NOT
undershoot beyond 65C by more than a few tenths of a degree, and the compressors should
NOT need to turn OFF in order to maintain 65C.
13. After stabilization, the chamber temperature should vary no more than 0.5C, or a total of
1C.
14. Set the Temperature/Humidity Controller Set Point to 73C.
15. The chamber should approach 73C and control (Cool Light cycles ON/OFF) within
approximately 5 minutes.
16. Set the Temperature/Humidity Controller Set Point to +23C. The chamber should begin to
heat up. The compressors should turn off within approximately 1 minute.
17. This concludes the temperature performance verification tests.
18. Let the chamber heat up to +23C before turning the Master Switch OFF.
Page 6-4

TestEquity 1207C Temperature/Humidity Chamber

Chapter 6 Maintenance

How to drain the humidity system


1.
2.
3.
4.
5.

Wait at least 2 hours after running a humidity test before draining the system.
Turn all front panel switches and the Disconnect Switch OFF.
Remove the left side panel (left side if facing the front of the chamber.
Open the Drain Valve by turning its handle 90 degrees counter-clockwise.
After the system is drained, close the Drain Valve by turning its handle 90 degrees
clockwise.

Drain Valve
(closed position shown)

TestEquity 1207C Temperature/Humidity Chamber

Page 6-5

Chapter 6 Maintenance

How to verify the temperature calibration


TestEquity recommends verifying the calibration before attempting to actually perform a
calibration. The state-of-the-art instrumentation used in TestEquity chambers is of the highest
quality and seldom goes out of calibration. If you try to calibrate the instrumentation before
determining that calibration is necessary, you may make it worse if done incorrectly.
Variations in temperature throughout the chamber interior is NOT a measurement of accuracy.
These variations, called gradients, are a function of the physical design of the chamber and its
airflow, the characteristics of the test sample, and how it is oriented in the chamber. You cannot
calibrate to improve gradients. The common practice of measuring multiple points in the
chamber and adjusting the Temperature/Humidity Controllers calibration to correct for these
errors is incorrect! The correct way to adjust what the Temperature/Humidity Controller
displays compared to what is measured at some point other than the controllers sensor, is with
the Calibration Offset parameter. See page 6.2 of the Series 97 Users Manual for details.
Calibration verification should be performed with the Calibration Offset set to 0.0 (zero).
Total system accuracy in the chamber includes the controller plus the thermocouple wire
accuracy. Total system accuracy over the chambers operating range is typically 1.65C, 1
LSD.
The easiest way to verify the instrumentation accuracy is with an independent calibrated
temperature sensor and display. Place the sensor near the chambers sensors, which are located
towards the right side of the conditioner fan grille. If the readings agree within the specified
limits, then no calibration adjustments are necessary.
If calibration of the Temperature/Humidity Controller is necessary, refer to page 9.2 of the
Series F4 Users Manual and follow the instructions for Thermocouple Input Procedure. If
calibration of the limit controller is necessary, refer to page A.11 of the Series 97 Users
Manual and follow the instructions for Thermocouple Input Procedure.

Page 6-6

TestEquity 1207C Temperature/Humidity Chamber

Chapter 6 Maintenance

How to verify the humidity calibration


As long as the probe is operated in an environment free of chemical contaminants and within the
operating limits, the long term stability of the humidity sensor is typically better than 1% RH per
year. For maximum accuracy, calibration of the unit may be verified every 6 to 12 months.
Applications where the unit is exposed to significant levels of chemical contaminants may
require more frequent verifications
When calibrating a relative humidity instrument, temperature stability is the single most
important requirement. Calibration should be done at room temperature (60 to 80F or 15 to
26 C) and temperature should be stable to 0.25C or better during the period of time required
for each calibration point
Calibration Device
The ER-15 calibration device is a small airtight container that slips over the probe of the
Humidity Sensor Probe and seals around the humidity sensor. During calibration, a known
reference humidity is produced inside the calibration device by means of a humidity standard
(usually an aqueous salt solution).
Humidity Standard Salt Solutions
The Rotronic certified humidity standards permit calibration by non-skilled personnel. These
standards are available in boxes of 5 glass ampoules of the same value, which can be stored
indefinitely.
Required Device and Standards
The following are required to calibrate the humidity sensor:
ER-15 Calibration Device
EA10, 10% RH Standard Salt Solution
EA35, 35% RH Standard Salt Solution
EA80, 80% RH Standard Salt Solution
Available from:
Rotronic Instrument Corp.
160 E. Main Street
Huntington, NY 11743
631-427-3898
800-628-7101
www.rotronic-usa.com

TestEquity 1207C Temperature/Humidity Chamber

Page 6-7

Chapter 6 Maintenance

Humidity Calibration Procedure


1.
2.
3.
4.

Turn the Main Disconnect Switch to the OFF position.


Remove the lower door retaining screw located on the right side. Open the lower door.
Unplug relay 206A. This will prevent the chamber from accidentally running.
Remove the air plenum cover from the inside of the chamber workspace. The Humidity
Sensor Probe is located in the upper left side.
5. Locate the Humidity Sensor Conditioning Board on the upper left
corner of the Electrical Subpanel. You will need to adjust the
potentiometers on this board.
H80
H35
6. Turn the Disconnect Switch ON.
HM
WARNING: The exposed electrical wiring will be energized. Do not
touch any exposed electrical terminals.
7. Slip an ER15 calibration device over the Humidity Sensor Probe and
make sure it seals tightly on the probe. The receptacle of the
calibration device should be under the probe tip. Remove the
Adjustment Potentiometers
receptacle from the calibration device.
8. Set the H80 and HM potentiometers in mid position.
9. Place one fiber disc n the receptacle of the calibration device. The purpose of this disc is to
prevent accidental spilling of the solution inside the calibration device or on the Humidity
Sensor Probe.
10. Tap the top of one ampoule of 35% RH solution so that all liquid drops to the bottom of the
ampoule. Snap off top and empty contents on fiber disc. Since the ampoule is made of glass,
use proper caution (gloves, safety glasses) when snapping off the top.
11. Put the receptacle back on the calibration device and make sure that the solution does not
come in contact with the sensor: The solution inside the calibration device should never be
on top of the sensor.
12. Read the humidity at the Input2 prompt on the Temperature/Humidity Controller
13. Allow at least 60 minutes to insure that the calibration device, the solution and the sensor are
in a state of equilibrium. This is verified by monitoring the humidity reading.
14. At equilibrium (stable reading), adjust H35 potentiometer until the humidity reads 35%.
15. Remove the receptacle from the calibration device. Throw away the wet disc. Thoroughly
wash and dry the receptacle, removing all traces of the humidity standard. Use 80% RH as
the second calibration value as this provides the best overall accuracy over the full range of
measurement.
16. Repeat the procedure used for the 35% RH adjustment with an 80% RH standard. Allow at
least 60 minutes for equilibrium.
17. At equilibrium, adjust the H80 potentiometer until the humidity reads 80%.
18. Remove the receptacle from the calibration device and clean thoroughly. Use 10% RH as the
last calibration value.
19. Repeat the procedure with a 10% RH standard. Allow at least 60 minutes for equilibrium.
20. At equilibrium, adjust the HM potentiometer until the humidity reads 10%.
21. Remove the receptacle from the calibration device and clean thoroughly.
Page 6-8

TestEquity 1207C Temperature/Humidity Chamber

Chapter 6 Maintenance

Refrigeration System Charging Instructions


! WARNING: Repair of the refrigeration system must be performed only by a properly
trained mechanic who is experienced in repairing cascade refrigeration
systems. Do NOT substitute any component. Do NOT substitute refrigerants.
Improper repairs will void the warranty.
These instructions are intended as guidelines for repairing TestEquity chambers. Details such as
how to attach a gauge manifold are not covered. These are NOT do-it-yourself instructions!

R-404A High-Stage Charge


TestEquity does NOT recommend charging the system by relying on a clear sight glass only.
Although a clear sight glass generally means the system is fully charged, it can be misleading.
For example, if the system is charged on a cool day or with an empty chamber, it could be
undercharged for hot days or with a heavy load. The proper charging procedure is as follows:
1. Repair any leaks before recharging.
2. Attach a vacuum pump and manifold gauge to the suction and discharge ports.
3. Evacuate the system to at least 100 microns. DO NOT GUESS! You must use a micron
gauge.
4. Use a charging scale to weigh in 4 pounds of R-404A.
5. Verify the cooling performance as outlined in How to verify the chamber performance.
NOTE: If the Temperature/Humidity Controller is cycling (Cool Light cycles on/off), the
sightglass may appear 1/2 to 2/3 full. This is normal.

R-508B Low-Stage Charge


1. Repair any leaks before recharging.
2. Attach a vacuum pump and manifold gauge to the suction, discharge, and expansion tank
ports. Attaching to the expansion tank is very important because it is otherwise very difficult
to evacuate the tank through the pressure regulator or capillary tube that is connects it to the
system.
3. Evacuate the system to at least 100 microns. Do NOT guess! You must use a micron gauge.
4. Do NOT put any additives in the system. Pentane is NOT necessary or desirable.
5. Charge the system until the standby pressure is 125 PSIG. Allow time for the charge to
equalize as read on the suction and discharge gauges. This is 10 PSIG higher than the target
amount of 115 PSIG. This corresponds to about 12 ounces. See NOTE below.
6. Verify the cooling performance as outlined in How to verify the chamber performance.
NOTE: If the low-stage has been evacuated and recharged, the standby pressure should be
rechecked after 24 hours to make sure it is 115 PSIG. This is because the refrigerant mixes with
the POE oil in the compressor, causing a lower standby pressure than was initially observed. Do
not mistake this initial loss of pressure with a leak. After verifying that there is no leak, you may
need to top-off the charge if the pressure is too low.
TestEquity 1207C Temperature/Humidity Chamber

Page 6-9

Chapter 6 Maintenance

Series F4 Controller Setup Parameters


! CAUTION:

The Series F4 Users Manual is a general manual and is written by the


manufacturer, Watlow, for a wide variety of applications and configurations.
Not all features or functions are applicable. Only the capabilities of a model
F4DH-CCCC-01, as described on page A.7 of the Series F4 Users Manual
are applicable. Cascade Control as described on page 3.6 of the Series F4
Users Manual is not applicable in this configuration. The Retransmit
function is available as an option.

! CAUTION:

The Series F4 Controller has been properly configured by TestEquity to match


the chambers system requirements and to perform optimally over a wide
range of operating conditions. Improper modifications to these setup values
can result in erratic performance and unreliable operation. Setup examples in
the Series F4 Users Manual are not applicable to this chamber. Do not
attempt to modify the setup values, unless you thoroughly understand what
you are doing. If there is any doubt, please call TestEquity before proceeding.

! CAUTION:

Do NOT select Full Defaults in the Series F4 Factory Menu. Full Defaults
are NOT the TestEquity configuration parameters for this chamber. If you
select Full Defaults, you must reconfigure all System and Operation
Parameters as documented in the TestEquity manual, NOT the Series F4
Users Manual.

TestEquity has configured the Temperature/Humidity Controller with the following parameters.

Page 6-10

TestEquity 1207C Temperature/Humidity Chamber

Chapter 6 Maintenance

Series F4 System Parameters


System Menu

Control Output Menu

Retransmit Output Menu

Main Page\Go to Setup\System

Main Page\Go to Setup\System\


Control Output (1A-2B)

(if retransmit option is ordered)


Main Page\Go to Setup\System\
Retransmit Output (1-2)

Guaranteed Soak Band 1


Guaranteed Soak Band 2
PID Units
F or C
Show F or C
Ch1 Autotune SP
Ch2 Autotune SP
Input 1 Fail
Input 2 Fail
Open Loop Ch1
Open Loop Ch2
Power Out Time
Power-Out Action

3.0 C
(9.0 if F)
3.0 %
US
C
Yes
90%
90%
0%
0%
No
No
10 Sec
Cont

Analog Input Menu


Main Page\Go to Setup\System\
Analog Input (1-3)
Analog Input 1
Choose Sensor
Type
Choose Decimal
SP Low Limit
SP High Limit
Cal Offset
Filter Time
Error Latch

T/C
T
0.0
-75.0 C
(-103.0 if F)
175.0 C
(347.0 if F)
0.0
1.0 sec
Self-Clear

Control Output 1A
Function
Process Type
Cycle Time
Hi Power Limit
Low Power Limit

Process
4-20ma
% rh
0
0.0 %
100.0 %
7.0 %
97.0 %
0.0 %
1.0 sec
Self-Clear

Analog Input 3
Choose Sensor

Off

Digital Input Menu

Retransmit Output 1
Retransmit Source
Analog Range
Low Scale

Scale Offset

Input 1
0-5V
-75.0C
(-103.0 if F)
175.0C
(347.0 if F)
0.0

Retransmit Output 2
Retransmit Source
Analog Range
Low Scale
High Scale
Scale Offset

Input 2
0-5V
0
100
0

High Scale
Control Output 1B
Function
Process Type
Cycle Time
Hi Power Limit
Low Power Limit

Cool
Fixed Time
10.0 sec
100 %
0%

Control Output 2A
Function
Process Type
Cycle Time
Hi Power Limit
Low Power Limit

Heat
Fixed Time
2.0 sec
100 %
0%

Control Output 2B
Function
Process Type
Cycle Time
Hi Power Limit
Low Power Limit

Cool
Fixed Time
10.0 sec
100 %
0%

Alarm Output Menu


Main Page\Go to Setup\System\
Alarm Output (1, 2)

Analog Input 2
Choose Sensor
Type
Choose Units
Choose Decimal
Scale Low
Scale High
SP Low Limit
SP High Limit
Cal Offset
Filter Time
Error Latch

Alarm Output 1
Name
Type
Alarm Output 2
Name
Type
Source
Latches
Silencing
Hysterisis
Sides
Logic
Show Message

Main Page\Go to Setup\System\


Digital Input (1-4)
Name
Function

Heat
Fixed Time
2.0 sec
100 %
0%

DIGIT IN (1-4)
Off

TestEquity 1207C Temperature/Humidity Chamber

No
Off

No
Process
Input 1
Self-Clears
No
2.0 C
(3.6 if F)
Both
Open
No

Digital Output Menu


Main Page\Go to Setup\System\
Digital Output (1-8)
Digital Output 1 through 7
Name
No
Function
Event Output
Digital Output 8
Function
Comp. On % Pwr.
Comp. Off % Pwr.
Comp. Off Delay
Comp. On Delay

Compressor
-2%
2%
60 sec
10 sec

Communications Menu
Main Page\Go to Setup\System\
Communications
Baud Rate
Address

19200
(9600 w/GPIB)
1

Custom Main Page Menu


Main Page\Go to Setup\System\
Custom Main Page
P1
P2
P3
P4
P5
P6
P7
P8
P9
P10
P11
Remaining
P12
P13
P14
P15
P16

Input 2
Current File
Current Step
Set Point 1
Set Point 2
Step Type
Target SP1
Target SP2
Wait For Status
Jump Count
Time
Digital Outs
% Power 1
% Power 2
Time
Date

Page 6-11

Chapter 6 Maintenance

Series F4 Operations Parameters


PID Set Channel 1 Menu
Main Page\Go to Operations\Edit PID\PID Set Channel 1
PID Set 1-5
Prop. Band A
Reset A
Rate A
Dead Band A
Hysteresis
Prop. Band B
Reset B
Rate B
Dead Band B
Hysteresis

5.0 C (9.0 if F)
0.20 min
0.00 min
0.0 C
0.0
5.0 C (9.0 if F)
0.20 min
0.00 min
0.0
0.0

PID Set Channel 2 Menu


Main Page\Go to Operations\Edit PID\PID Set Channel 2
PID Set 6-10
Prop. Band A
Reset A
Rate A
Dead Band A
Hysteresis
Prop. Band B
Reset B
Rate B
Dead Band B
Hysteresis

10 %
0.10 min
0.00 min
0.0 %
0.0
10.0 %
0.10 min
0.00 min
0.0 %
0.0

Alarm Setpoints Menu


Main Page\Go to Operations\Alarm Setpoints
Alarm2
Alarm2 Low SP
Alarm2 High SP

7.0 C (44.6 if F)
90.0 C (194.0 if F)

Series F4 Set Lockout Parameters


The Series F4 Controller has several levels of security to prevent unauthorized users from
changing critical configuration parameters. Only the Set Point and Profile menus have Full
Access. TestEquity has configured all other menus to Read Only, and have protected them
with a password.
TestEquity does not recommend that these security levels be changed for most applications.
However, there will be times when Full Access is necessary. For example, you may need to
gain access to Setup Page in order to change from C to F display, or to change the time or date.
You must call TestEquity at 877-512-3457 or 805-480-3697 to obtain the password.
Set Lockout Menu
Main Page\Go to Factory\Set Lockout

Set Point
Oper. Autotune PID
Oper. Edit PID
Oper. Alarm SP
Profile
Setup
Factory

Page 6-12

Full Access
Password
Password
Password
Full Access
Password
Password

TestEquity 1207C Temperature/Humidity Chamber

Chapter 6 Maintenance

Series 97 Limit Controller Setup Parameters


! CAUTION:

The Series 97 Users Manual is a general manual and is written by the


manufacturer, Watlow, for a wide variety of applications and configurations.
Not all features or functions are applicable. Only the capabilities of a model
97A0-DAAA-00-RG, as described on page A.20 of the Series 97 Users
Manual are applicable. Please note that only the Series 97 Limit functions
are applicable.

TestEquity has configured the Limit Controller with the following parameters.

Home Page

Setup Page

[``97] Process Value

[InP1] Input 1 Menu

[`LOC] Lockout Menu

[Safe] Limit Status

[`SEt] Setup Page


[SEn1] Sensor Type 1
[In`1] Input 1
[rL`1] Range Low 1
[rh`1] Range High 1
[dEC1] Decimal 1
[Ftr1] Input Software Filter 1

[FctY] Factory Page

Operations Page
[LIM] Limit Menu
[OPEr] Operations Page
[L~Lo] Low Limit Set Point
[L~hi] High Limit Set Point
[CAL1] Calibration Offset

[~-75]
[~175]
[`~`0]

[Mon] Monitor Menu


[OPEr] Operations Page
[Pr1`] Process 1
[L`St] Limit Status

-[SAFE]

Factory Page
[~~tc]
[~~~t]
[~-75]
[~175]
[`~`0]
[`~`0]

[OPEr] Operations Page Mode [chng]


[`SEt] Setup Page Lock
[chng]
[`CAL] Calibration Menu Lock [chng]

[Out1] Ouput 1 Menu


[`SEt] Setup Page
[LSid] Set Limit Active Sides [both]
[LhyS] Limit Hysteresis
[`~`1]

[diSP] Display Menu


[`SEt] Setup Page
[UdSP] Upper Display
[UP`L] Upper Display
User Limit Message
[LdSP] Lower Display
[Lo`S] Lower Display
User Safe Message
[Lo`L] Lower Display
User Limit Message

[`Pr1]
-[L~St]
---

[GLbL] Global Menu


[`SEt] Setup Page
[C-F`] C or F
[`Err] Input Error Latching

TestEquity 1207C Temperature/Humidity Chamber

[``C]
[`LAt]

Page 6-13

Chapter 7 Warranty

Chapter 7 - Warranty
TestEquity Inc. Limited Warranty
TestEquity Inc. (TestEquity) warrants Environmental Chambers (Equipment) manufactured by TestEquity and
supplied under this contract to be free from defects in materials and workmanship under normal use and proper
maintenance.
TestEquity will repair or replace any defective part for a period of THREE YEARS from the date of invoice.
TestEquity reserves the right to require any defective part be returned, freight prepaid, to TestEquitys factory or to
inspect any defective part at the Purchasers site. TestEquity shall have sole discretion to determine whether any part
is defective and whether any defective part will be repaired or replaced. This limited warranty shall extend to any
standard chamber accessory and component part that is normally sold by TestEquity. Non-standard accessories and
component parts specified by the Purchaser shall be warranted only to the extent of the original manufacturers
warranty, if any exists.
If the repair or replacement is performed in the FIRST YEAR from the date of invoice, TestEquity will also pay for
the labor associated with the repair at the Purchasers site, subject to TestEquitys prior approval. During the
SECOND and THIRD YEAR of the warranty period, Purchaser will be responsible for the installation and cost of
installation of replacement or repaired parts. Purchaser shall notify TestEquity in writing of any alleged defect
within 10 days after its discovery within the warranty period. TestEquity reserves the right to satisfy the labor
portion of this limited warranty either through its own service personnel or an authorized agent. In order to provide
expeditious service, TestEquity reserves the right to satisfy its limited warranty obligation by sending replacement
parts to be installed by the Purchaser if they can be installed easily without special tools or training. TestEquity
reserves the right to satisfy this limited warranty by requiring the Purchaser to return the Equipment to TestEquity
when such return is feasible.
The following parts are excluded from this limited warranty and are sold as-is or are considered expendable: interior
light bulb, viewing window, paint and cosmetic surface finishes and treatments, and port plugs.
This limited warranty shall extend in full to Equipment installed within continental United States and Canada. For
all other locations, Purchaser is responsible for all labor costs for repairs or parts installation, and for all shipping
costs associated with providing replacement parts.
This limited warranty does not cover: (1) Defects or damages arising as the result of shipment by common carriers
or private transportation, unless TestEquity undertakes shipment and transportation of the Equipment to Purchasers
site or contractually assumes the risk of damage to the Equipment in shipment; (2) Defects or damages arising out
of, or as the result, of mishandling, modification, or improper start up, installation or maintenance of the Equipment
(including start up, installation or maintenance not in accordance with TestEquitys written procedures); (3) Defects
or damages resulting from, or arising out of, abuse, misuse, neglect, intentional damage, accident, fire, flood,
earthquake, or any other act of God.
This warranty as to Equipment is LIMITED to repair or replacement of parts or Equipment in the determination of
TestEquity Inc. THE FORGOING LIMITED WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES
INCLUDING THE IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE AND
MERCHANTABILITY. TestEquity Inc. DISCLAIMS ANY LIABILITY FOR ANY DAMAGES RESULTING
FROM DELAY OR LOSS OF USE IN SERVICE OR REPAIR, OR FOR INCIDENTAL OR CONSEQUENTIAL
DAMAGES ARISING OUT OF OR IN CONNECTION WITH THE USE OR PERFORMANCE OF THE
EQUIPMENT, EXCEPT AS STATED IN THIS PARAGRAPH.
This limited warranty cannot be modified in any way except in writing by both TestEquity and Purchaser.
Invalidation of any one or more of the provisions of this limited warranty shall in no way affect any of the other
provisions hereof, which remain in full force and effect.
This limited warranty shall be extended only to the first Purchaser of this Equipment and is not transferable.

TestEquity 1207C Temperature/Humidity Chamber

Page 7-1

Chapter 8 Drawings

Chapter 8 Drawings

TestEquity 1207C Temperature/Humidity Chamber

Page 8-1

214A

214C

Terminal Strip

214B

231

232

215
213A

206F

213B

F11

203

206E

206A

202A

206B

202B

206C

206D

202C

250

202D

F1 F2 F3 F4 F5 F6 F7 F8 F9 F10

204

Electrical Subpanel Component Location

You might also like