1207 C Manual 101
1207 C Manual 101
1207 C Manual 101
Temperature/Humidity Chamber
Operation and Service Manual
TestEquity Inc.
2450 Turquoise Circle
Thousand Oaks, CA 91320
Support:
Corporate:
800-732-3457
805-498-9933
http://www.testequity.com
Rev. 1.01
Table of Contents
Chapter 1 Safety Instructions________________________________________________ 1-1
Introduction____________________________________________________________________ 1-1
Installation Safety Notices ________________________________________________________ 1-1
Operation Safety Notices _________________________________________________________ 1-2
2-1
2-1
2-2
2-3
Overview ____________________________________________________________________________
230 V / 60 Hz Input Configuration ________________________________________________________
208 V / 60 Hz Input Configuration ________________________________________________________
Single Phase Configuration ______________________________________________________________
Three Phase Configuration_______________________________________________________________
2-3
2-5
2-5
2-5
2-5
2-7
2-7
2-7
2-8
3-2
3-2
3-2
3-2
3-2
3-2
3-3
3-3
3-11
3-12
3-12
3-12
3-12
3-13
Table of Contents
Loading the Chamber ___________________________________________________________ 3-14
Performance Considerations ____________________________________________________________ 3-14
Avoiding Moisture (non-humidity mode) __________________________________________________ 3-15
Internal Test Fixtures __________________________________________________________________ 3-15
6-1
6-1
6-1
6-1
6-1
6-1
6-2
6-2
6-2
6-2
6-3
6-4
6-5
6-6
6-7
6-8
Chapter 1 Safety
The minimum clearance you should allow for proper ventilation around the
chamber must be at least 12" from both the left and right side, and 24" from
the rear.
! CAUTION:
! CAUTION:
The Input Voltage label on the back of the chamber indicates the input voltage
configuration as shipped from the factory. If the input voltage configuration is
changed, this label must be replaced to reflect the new configuration.
Replacement labels are available from TestEquity at no charge.
! CAUTION:
This chamber must be properly configured for either 208 V or 230 V nominal
input. 208 V and 230 V are NOT the same. Do NOT guess! Do NOT assume
you have 220 V. You must verify the exact type of electrical service you
have. If there is any doubt, you must consult with a qualified electrician who
is familiar with industrial plant wiring. In addition, the input line voltage
should be measured while the chamber is operating in the COOL mode to
ensure that the expected nominal voltage of either 208 V 5/+10% or 230 V
10% is present. Operation below 198 V or greater than 253 V requires
internal transformers, which can be supplied for a nominal charge.
2 CAUTION:
Page 1-1
Chapter 1 Safety
2 CAUTION:
This chamber has a crankcase heater to protect the high-stage compressor. The
chamber must be connected to the power source AND the Main Disconnect
Switch must be ON for 3 hours prior to operating the chamber. Although it
may be safe to use the chamber immediately, this procedure ensures the
longest possible life for the high-stage compressor if the chamber has been
removed from the power source for more than 24 hours.
! CAUTION:
! CAUTION:
! CAUTION:
! CAUTION:
Do NOT select Full Defaults in the Series F4 Factory Menu. Full Defaults
are NOT the TestEquity configuration parameters for this chamber. If you
select Full Defaults, you must reconfigure all System and Operation
Parameters as documented in the TestEquity manual, NOT the Series F4
Users Manual.
! CAUTION:
Page 1-2
Chapter 1 Safety
! CAUTION:
Always verify that the Limit Controllers high and low limits are set to
temperatures that are appropriate for your test sample.
! WARNING: Do NOT put items in the chamber that could burn or explode at high
temperatures. This chamber uses open wire heating elements which generate
surface temperatures over 1000F. This is NOT an explosion-proof chamber.
! WARNING: Do NOT put items in the chamber which can emit corrosive vapors or
substances.
! WARNING: This chamber is NOT a curing oven. There are NO provisions for venting
fumes.
! WARNING: The chamber door must remain closed while the chamber is operating. If you
need to open the door while the chamber is operating, wear safety goggles to
prevent the high velocity airflow from blowing particles or objects into your
eyes.
! WARNING: This chamber operates at extreme temperatures. Avoid contact with air,
objects, and surfaces that are hot or cold to prevent severe burns or frostbite.
Protective gloves are recommended.
! CAUTION:
! CAUTION:
To prevent damage to your test sample and the chambers compressors, do not
exceed the live load rating of the chamber.
Page 1-3
Chapter 2 Installation
Chapter 2 Installation
Uncrating
Inspect the shipping container for any signs of visible damage. Notify the carrier and TestEquity
immediately if there are signs of shipping damage.
The pallet is designed with ramps so the chamber can be rolled off without the need for a forklift
or pallet jack.
1.
2.
3.
4.
Cut the two metal bands that hold the packaging to the pallet.
Remove the top cover.
Remove the plastic fasteners that hold the outer box together and remove the outer box.
Locate the retainer in the front of the pallet. Remove the two screws, then remove the
retainer. Keep in mind that there is also Velcro that holds the retainer in place. Place the
retainer in the front of the pallet on the floor.
5. Locate the two ramps under the chamber and remove them.
6. Attach the ramps to the front of the pallet using the Velcro straps. The retainer will be under
the ramps and act as a support.
7. Carefully roll the chamber off the pallet, onto the ramps. This should be done with at least
two people.
Page 2-1
Chapter 2 Installation
Installation Location
The chamber will produce a significant amount of heat during normal operation. Locate the
chamber in a room with adequate ventilation to prevent excessive heat build-up.
The chamber must be on a solid and level floor. Allow enough space around the chamber to
permit serviceability and the removal of the service access panels, which are located on each side
and the rear.
! CAUTION:
The minimum clearance you should allow for proper ventilation around the
chamber must be at least 12" from both the left and right side, and 24" from
the rear.
! CAUTION:
Page 2-2
Chapter 2 Installation
The Input Voltage label on the back of the chamber indicates the input voltage
configuration as shipped from the factory. If the input voltage configuration is
changed, this label must be replaced to reflect the new configuration.
Replacement labels are available from TestEquity at no charge.
! CAUTION:
This chamber must be properly configured for either 208 V or 230 V nominal
input. 208 V and 230 V are NOT the same. Do NOT guess! Do NOT assume
you have 220 V. You must verify the exact type of electrical service you
have. If there is any doubt, you must consult with a qualified electrician who
is familiar with industrial plant wiring. In addition, the input line voltage
should be measured while the chamber is operating in the COOL mode to
ensure that the expected nominal voltage of either 208 V 5/+10% or 230 V
10% is present. Operation below 198 V or greater than 253 V requires
internal transformers, which can be supplied for a nominal charge.
Page 2-3
Page 2-4
208V
Jump 25 to 26
203
H2 H3
2
44
230V, wire 44 on H3
208V, wire 44 on H2
F1 F2
1 Phase
Jump
F2 to F10
F8 to F9
F8 F9 F10
3 Phase
Jump
F2 to F8
INPUT
L1 L2 L3
204
(1 Ph on L1, L2 only)
Ground
INPUT
ACCESS
HOLE
Chapter 2 Installation
Chapter 2 Installation
NOTE: Refer to Figure 2-1 on the previous page for the location of the input configuration and
connection terminals on the electrical sub panel that are described below.
Page 2-5
Chapter 2 Installation
The Main Disconnect Switch on the front panel removes primary power to the entire chamber.
All branch circuits on the load side of the Main Disconnect Switch are individually fused.
However, your local electrical code may require a separate disconnect switch within sight of the
chamber.
Page 2-6
Chapter 2 Installation
Page 2-7
Chapter 2 Installation
Replace when
color change
reaches this line.
Page 2-8
Chapter 3 Operation
Chapter 3 Operation
Introduction
The Front Panel Switches control power to the chamber. The Front Panel Lights provide
indication of heat and cool functions.
The Limit Controller is a protection device. It turns the chamber OFF if the workspace
temperature exceeds either a high temperature or low temperature limit set point.
The Temperature/Humidity Controller controls the temperature of the chamber. It can function
as either a single set point controller or as a programmable profile controller. The
Temperature/Humidity Controller automatically turns the refrigeration system on or off based on
the demand for cooling.
! CAUTION:
This chamber has a crankcase heater to protect the high-stage compressor. The
chamber must be connected to the power source AND the Main Disconnect
Switch must be ON for 3 hours prior to operating the chamber. Although it
may be safe to use the chamber immediately, this procedure ensures the
longest possible life for the high-stage compressor if the chamber has been
removed from the power source for more than 24 hours.
Page 3-1
Chapter 3 Operation
This chamber has a crankcase heater to protect the high-stage compressor. The
chamber must be connected to the power source AND the Main Disconnect
Switch must be ON for 3 hours prior to operating the chamber. Although it
may be safe to use the chamber immediately, this procedure ensures the
longest possible life for the high-stage compressor if the chamber has been
removed from the power source for more than 24 hours.
Master Switch
The Master Switch enables all chamber functions. The Master Switch illuminates when it is ON.
Light Switch
The Light Switch controls the workspace light. The Light Switch illuminates when it is ON.
Silence Switch
The Limit Controller will sound an Audible Alarm in the event of an over-temperature or undertemperature condition. The Silence Switch lets you temporarily turn off the Audible Alarm.
Heat Light
The Heat Light will illuminate when the Temperature/Humidity Controller turns on the heater to
maintain the workspace temperature. The Heat Light will cycle on/off as the workspace
temperature approaches and reaches the temperature set point. This light corresponds to the 1A
indicator on the Temperature/Humidity Controller.
Page 3-2
Chapter 3 Operation
Cool Light
The Cool Light will illuminate when the Temperature/Humidity Controller turns on the cooling
valve to maintain the workspace temperature. The Cool Light will cycle on/off as the workspace
temperature approaches and reaches the temperature set point. This light corresponds to the 1B
indicator on the Temperature/Humidity Controller.
Humidity Light
The Humidity Light will illuminate when the Temperature/Humidity Controller turns on the
humidifier heater to maintain the workspace humidity. The Humidity Light will cycle on/off as
the workspace humidity approaches and reaches the humidity set point. This light corresponds to
the 2A indicator on the Temperature/Humidity Controller.
Note that the Humidity Enable Switch must be ON, the chamber workspace must be between
+7C and +90C, and the humidifier water level must be correct for the Humidity Light to
correspond to the 2A indicator on the Temperature/Humidity Controller.
The 2B indicator on the Temperature/Humidity Controller lights with dehumidification is
required to maintain the workspace humidity. There is no corresponding light on the front panel
for this function.
Page 3-3
Chapter 3 Operation
Temperature/Humidity Controller
Overview
This section provides a brief overview on how to operate the Series F4 Controller in the static set
point mode. For more detailed instructions, see the Series F4 Users Manual.
The Series F4 Controller can function as either a single set point controller (static mode) or as a
programmable profile controller. A four-line LCD display facilitates setup and programming,
and presents informative messages about status, error, and alarm conditions. Digital outputs,
profiles, and alarms can be named for easy reference. An Information Key gives you quick
information about the pages, menus, parameters and values, as well as error and alarm conditions
if they occur. The user-interface is organized into five pages of menus. The Main Page
provides access to the other four pages Operations, Profiles, Setup, and Factory.
! CAUTION:
! CAUTION:
! CAUTION:
Security Features
The Series F4 Controller has several levels of security to prevent unauthorized users from
changing critical configuration parameters. Only the Set Point and Profile menus have Full
Access. TestEquity has configured all other menus to Read Only, and have protected them
with a password.
TestEquity does not recommend that these security levels be changed for most applications.
However, there will be times when Full Access is necessary. For example, you may need to
gain access to Setup Page in order to change from C to F display, or to change the time or date.
You must call TestEquity at 877-512-3457 or 805-480-3697 to obtain the password.
Page 3-4
Chapter 3 Operation
Lower Display
Displays information
about the setup,
operation and
programming of the
controller.
Alarm Output
Indicator Lights
Lit during an alarm
state. Alarm2 is lit
when temperature
is outside the range
for humidity mode.
Cursor
Indicates selected
parameter or present
value in memory.
Moves via the four
navigation keys.
Communications
Indicator Light
Lit (pulsates) when
the controller sends
or receives valid
data.
Profile
Indicator Light
Lit when a ramping
profile runs.
When blinking, the
profile is on hold.
When not lit, the
controller operates as
a static set point
controller.
Profile Key
Summons a menu that
allows you to start, hold,
resume or terminate a profile.
Output
Indicator Lights
1A = Heat
1B = Cool
2A = Humidity
2B = Dehumidify
Scroll Bar
Appears when the
Up or Down Keys
can reveal more
information in the
Lower Display.
Information Key
Provides information
in the Lower Display
about the cursorselected parameter.
Another press toggles
the display back to
the parameter.
Page 3-5
Chapter 3 Operation
Page 3-6
4. Press the
Page.
key once to enter the new humidity set point. You are now back to the Main
Chapter 3 Operation
4. Press the key once to select the desired EventOutput. Then, press the or key to
select On or Off.
5. Press the key once to return to the Choose Event Output prompt.
6. Press the key once to return back to the Main Page.
Page 3-7
Chapter 3 Operation
Profile Programming
The Series F4 Controller can be programmed to store up to 256 steps into as many as 10 profiles.
You do not need a computer to enter a profile it can be easily done through the controllers
front panel keys.
For detailed information about how to program a profile, see Page 4.1 of the Series F4 Users
Manual.
For detailed information about how to run a profile, see Page 3.2 of the Series F4 Users
Manual.
! CAUTION:
The sample program in the Series F4 Users Manual is not applicable to this
chamber.
! CAUTION:
Programming Hints:
The first step in a program should be an initialization step of 1-second.
The next to last step establishes a condition to end on. For example, you may want to end the
program by holding at +23C, so this step would be to go to +23C.
TestEquity recommends having the end step type to be Hold or Idle.
TestEquity does NOT recommend using an end step type of All Off or Control Off. This
does not turn off the chamber fan. The chamber temperature can reach +55C just from heat
generated by the fan, and even higher if your test sample is energized.
Page 3-8
Chapter 3 Operation
1
>
>
>
>
>
>
H:M:S
00:00:01
00:30:00
00:00:01
00:30:00
00:00:01
Ch1 Ch2
SP
SP
-40.0
>
>
85.0
>
>
23.0
>
PID Set
Ch1
Ch2
1
6
1
6
1
6
1
6
1
6
-40C to 85C, uncontrolled humidity, no ramps, 30 minute soak time, 4 cycles. Ends with +23C set point.
Wait for
>
>
>
>
>
>
Description
Goes to -40C as quickly as possible.
Will hold for 30 minutes. Time will not start until chamber reaches -37C (within the 3 Guaranteed Soak Band*).
Goes to +85C as quickly as possible.
Will hold for 30 minutes. Time will not start until chamber reaches +82C (within the 3 Guaranteed Soak Band*).
Jumps back to step 1. Repeats this 3-times, for a total of 4-cycles.
Goes to +23C as quickly as possible.
End of program. Controller returns to a +23C manual set point.
Step
Step
Date
Number
Type
Day, Time
1
Ramp Time
2
Soak
3
Ramp Time
4
Soak
5
Jump
6
Ramp Time
7
End
Step
1
2
3
4
5
6
7
1
On
On
On
Off
Off
On
>
Off
7
>
>
>
>
>
>
>
>
Time
H:M:S
00:00:01
00:13:30
01:00:00
00:25:00
01:00:00
00:25:00
00:13:30
Ch1
SP
23.0
50.0
0.0
50.0
23.0
Ch2
SP
>
>
>
>
>
>
>
PID Set
Ch1
Ch2
1
6
1
6
1
6
1
6
1
6
1
6
1
6
Yes
No
No
No
Yes
No
Name
1
No
No
No
No
No
No
No
No
No
No
No
No
Name
2
No
No
End
Repeats Type
Idle
End
Repeats Type
99
Idle
Idle SP
Ch1
23.0
Idle SP
Ch1
23.0
+50C to 0C, 2C/minute ramp, uncontrolled humidity, 1-hour soak times, 100 cycles total, test sample turned ON/OFF depending on step, end with a +23C set point.
Wait for
>
>
>
>
>
>
>
>
Description
Establishes an initialization step. Goes to +23C as quickly as possible. Test sample is turned ON via Event 1.
Goes to +50C at a controlled rate of 2C per minute (50-23=27/2=13.5 or 13 minutes and 30 seconds). Test sample is turned ON via Event 1.
Will hold for 1 hour. Test sample remains ON via Event 1.
Goes to 0C at a controlled rate of 2C per minute (50-0=50/2=25 or 25 minutes). Test sample is turned OFF via Event 1.
Will hold for 1 hour. Test sample remains OFF via Event 1.
Goes to 50C at a controlled rate of 2C per minute. Test sample is turned OFF via Event 1.
Jumps back to step 3. Repeats this 99-times, for a total of 100-cycles.
After all cycles are completed, establishes the condition to end on. Goes to +23C at a controlled rate of 2C per minute. Test sample is turned OFF via Event 1.
End of program. Controller returns to a +23C manual set point. Test sample remains OFF via Event 1.
Step
Step
Date
Number
Type
Day, Time
1
Ramp Time
2
Ramp Time
3
Soak
4
Ramp Time
5
Soak
6
Ramp Time
7
Jump
8
Ramp Time
9
End
Step
1
2
3
4
5
6
7
8
9
Notes
*
Guaranteed Soak Band is set at the factory for 3.0C. Can be changed in the System Menu.
>
Means no entry or selection is required. Just scroll through this prompt to the next prompt.
Means prompt does not show for this step type.
Page 3-9
Chapter 3 Operation
RS-232C Interface
The Temperature/Humidity Controller has an RS-232C interface. A DB-9 connector is located
on the front left side. It is wired to accommodate a null-modem cable.
To communicate with the controller from a PC, you need to run software that uses the Modbus
RTU protocol. industrial data communications. This protocol works on the principle of packet
exchanges. The packet contains the address of the controller to receive the information, a
command field that says what is to be done with the information and several fields of data.
Reading from these registers retrieves all information in the controller. These registers are listed
Chapter Seven of the Series F4 Users Manual.
RS-232C and Modbus programming resources can be downloaded from
http://chamber.testequity.com/rs232.html .
The actual chamber temperature reading is Modbus register 100 (Input 1 Value).
The actual chamber humidity reading is Modbus register 104 (Input 2 Value).
The static temperature setpoint is Modbus register 300 (Set Point 1).
The static humidity setpoint is Modbus register 319 (Set Point 2).
The temperature setpoint during a profile is Modbus register 4122 (Set Point 1, Current
Profile Status).
The humidity setpoint during a profile is Modbus register 4123 (Set Point 2, Current Profile
Status).
The decimal points are implied. For example, 1005 is actually 100.5 and -230 is -23.0.
Page 3-10
Chapter 3 Operation
Limit Controller
Overview
The Series 97 Limit Controller is a protection device. It turns the chamber OFF if the workspace
temperature exceeds either a high temperature or low temperature limit. You can set these limits
to correspond to the maximum and minimum temperature that your test sample can safely
withstand. This provides protection against someone setting the Temperature/Humidity
Controller to a temperature that is unsafe for the test sample. It also provides protection in the
unlikely event of a chamber system component failure. The Limit Controller has its own
temperature sensor (thermocouple) and functions completely independent of the
Temperature/Humidity Controller.
This section provides a brief overview on how to operate the Limit Controller. For more detailed
instructions, see the Series 97 Users Manual.
! CAUTION:
! CAUTION:
Always verify that the Limit Controllers high and low limits are set to
temperatures that are appropriate for your test sample.
! CAUTION:
Page 3-11
Chapter 3 Operation
Lower Display
Indicates factory programmed message
during operation, the value for the
parameter in the lower display, or the user
programmed message.
Advance Key
Advances the lower display through
the parameters. To reverse direction,
press and hold the Advance Key while
repeatedly pressing the UP Key.
Up Key
Changes the upper display to a higher
value, or up through a list of values.
Moves from menu to menu in a page.
Down Key
Changes the upper display to a lower
value, or down through a list of values.
Moves from menu to menu in a page.
Reset Key
Returns to the Home Page
(process/actual display).
Resets the limit.
Resets an input sensor error.
Page 3-12
Chapter 3 Operation
Page 3-13
Chapter 3 Operation
! CAUTION:
To prevent damage to your test sample and the chambers compressors, do not
exceed the live load rating of the chamber.
Live Load Capacity for Model 1207C (non-humidity mode)
Temp +23C
0C
40C 55C 65C
Watts 1000 W 800 W 500 W 400 W 300 W
Performance Considerations
The performance of all chambers are significantly effected by the characteristics of your test
sample. Factors include size, weight, material, shape, and power dissipation if energized.
The test sample should be placed in the chamber in a manner that allows for air circulation. The
air plenum is located on the back wall of the chamber, where air is sucked in from the bottom
and exits from the top. You should not place the test sample directly on the chamber floor. It
should be placed on the shelf. Multiple test samples should be distributed throughout the
chamber to ensure even airflow and minimize temperature gradients. If necessary, additional
shelves should be used to evenly distribute the load. Verify that the temperature gradients are
within acceptable limits, by measuring the chamber temperature at strategic points using a
multipoint thermocouple meter or datalogger.
Page 3-14
Chapter 3 Operation
You may find that the temperature throughout the chamber is even, but always different from
what the Temperature/Humidity Controller indicates. The correct way to adjust what the
Temperature/Humidity Controller displays compared to what is measured at some point other
than the controllers sensor is with the Calibration Offset parameter. See page 6.2 of the
Series 97 Users Manual for details.
Page 3-15
Chapter 3 Operation
Humidity Operation
Humidity Range
This chamber is capable of controlling humidity from 10% RH to 95% RH over the temperature
range of +10C to +85C. The lowest possible humidity varies depending on the temperature.
For example, +45C is the lowest temperature that 10% RH can be achieved. Below is a chart
that shows the achievable range of humidity as compared to temperature. Achieving low
humidity levels require you to start with a clean, dry chamber.
100
90
80
70
Achievable Range of
Temperature/Humidity
Conditions
60
50
40
30
20
10
0
0
32
10
50
20
68
30
86
40
104
50
122
60
140
70
158
80
176
90
194
C
F
TEMPERATURE
Page 3-16
Page 4-1
Page 4-2
Im not going to use the chamber for a while. Is there anything I should do to prepare it for
storage?
Perform ALL the steps in the Preventive Maintenance Schedule before placing the chamber into
storage. This will ensure that the chamber will be ready to operate when it is taken out of storage.
If the chamber has a problem and is still under warranty, these problems should be resolved
before being placed into storage, since the warranty period starts from the date of shipment. The
chamber should be stored in a conditioned environment. Do not store it outside or where it will
be subjected to dirt or excessive moisture.
I havent used the chamber for a while. Is there anything I should do to prepare it for
operation?
Perform ALL the steps in the Preventive Maintenance Schedule before placing the chamber back
into service. This will ensure that nothing has been damaged and that a leak has not developed.
This chamber has a crankcase heater to protect the high-stage compressor. The chamber must be
connected to the power source AND the Main Disconnect Switch must be ON for 3 hours prior
to operating the chamber. Although it may be safe to use the chamber immediately, this
procedure ensures the longest possible life for the high-stage compressor if the chamber has been
removed from the power source for more than 24 hours.
Can the person who services our air conditioning also service the chamber?
Probably not. Most air conditioning mechanics are not familiar with low-temperature cascade
refrigeration systems. While this chamber is relatively easy to maintain and repair, most air
conditioning mechanics do not have the necessary refrigerants and may not be familiar with the
microprocessor-based controls. This chamber should only be serviced by a qualified mechanic
that is familiar with low-temperature cascade refrigeration systems. Call TestEquity to
recommend one in your area or to check if the one you would like to use is qualified.
Can/Should I put a filter in front of the condenser air inlet?
No, TestEquity does not recommend this. Just follow the maintenance procedures and clean the
condenser fins periodically.
How often should I charge the refrigeration system?
This chamber uses a closed-loop refrigeration system. Just like your refrigerator at home, it does
not need periodic charging. If the charge is low, this means that there is a leak. Leaks should be
repaired before recharging.
What kind of Freon does the chamber use?
The word Freon is a DuPont registered trade name for their CFC-based refrigerants and is
incorrectly used as a generic term for refrigerants. TestEquity chambers do not use CFC-based
refrigerants. The high-stage system uses R-404A, which is also known as DuPont Suva HP62.
The low-stage system uses R-508B, which is also known as DuPont Suva 95.
Page 4-3
Chapter 5 Specifications
Chapter 5 Specifications
Model 1207C Chamber Specifications
Temperature Range
Control Tolerance
Uniformity
75C to +175C
0.5C (0.2C Typical) (Measured at the control sensor after stabilization)
0.5C (Variations throughout the chamber after stabilization)
Humidity Range
Control Tolerance
Live Load Capacity @
+23C
0C
1000 W
800 W
(Live load capacity for non-humidity mode)
40C
500 W
55C
400 W
65C
300 W
+23C
18 min
End Temp
0C
40C
25 min
44 min
55C
54 min
65C
62 min
75C
ultimate
+23C
----1.5 min
6 min
8 min
10 min
End Temp
+50C
+85C
1.5 min
7 min
3.5 min
13 min
11 min
17 min
13 min
19 min
14 min
21 min
+125C
14 min
20 min
24 min
26 min
28 min
+150C
20 min
23 min
30 min
32 min
34 min
+175C
25 min
31 min
35 min
37 min
39 min
*Note: Transition times are measured after a 1-hour soak at the start temperature with an empty chamber, in nonhumidity mode.
Input Power Requirements
230 V 10%, 60 Hz
Phase
1 PH
3 PH
208 V -5/+10%, 60 Hz 1 PH
3 PH
Current
Draw
25 A
19 A
Minimum
Service
30 A
25 A
28 A
21 A
35 A
30 A
Input may be configured for single or three phase in the field by changing jumpers. Three phase load is semibalanced. Operation below 198 V requires internal boost transformers, which can be supplied for a nominal charge.
Call for other voltages or 50 Hz operation.
Workspace Dimensions
24" W x 21" H x 24" D (7 cubic feet)
Outside Dimensions
33.5" W x 68" H x 48" D (nominal)
Door latch adds 3" to width on right side (may be removed to permit move-in through a 36" doorway). Circulator
motor housing adds 6" to height.
Min. Installed Clearance 12" from the left and right side, 24" from the rear
Access Ports
4" Port on left and right side (two total), Supplied with foam plugs
Weight
850 pounds
NOTE: Performance is typical and based on operation at 23C (73F) ambient and nominal input voltage. Designed
for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased
cooling performance. Additional ports and shelves will also affect performance. Operation above 30C (85F) or
below 16C (60F) ambient is not recommended.
TestEquity 1207C Temperature/Humidity Chamber
Page 5-1
Chapter 5 Specifications
F4 Controller Specifications
Specifications as configured for the TestEquity 1207C Chamber
Accuracy & Sensor Conformity* 0.65C, 1 LSD with type T thermocouple input
Stability
Digital Inputs
Alarm Outputs
Digital Outputs
Communications
Displays
Data Retention
*Note: Total system accuracy in the chamber includes thermocouple wire accuracy.
Thermocouple wire accuracy is 1C or 0.75% of reading, whichever is greater. Therefore, total
system accuracy over the chambers operating range is typically 1.65C, 1 LSD.
Page 5-2
Chapter 5 Specifications
0 to 100% RH
-5C to +160C
-75C to +200C
Humidity Limits
100% RH at +85C
90% RH at +125C
25% RH at +160C
Accuracy
Repeatability
0.5% RH
Calibration Device
ER-15 Fixture
EA10, 10% RH Standard Salt Solution
EA35, 35% RH Standard Salt Solution
EA80, 80% RH Standard Salt Solution
Calibration devices available from:
Rotronic Instrument Corp.
160 E. Main Street
Huntington, NY 11743
631-427-3898
800-628-7101
www.rotronic-usa.com
Page 5-3
Chapter 6 Maintenance
Chapter 6 Maintenance
! WARNING: Maintenance must be performed by properly trained personnel only.
Every 3 Months
Every 6 Months
Every 12 Months
Clean surfaces with a damp cloth, mild detergent, or stainless-steel cleaner. Avoid cleaners
that are abrasive or leave a residue. Do NOT use steel wool.
If you clean the interior with something other than water, you may want to operate the
chamber at high temperature (approximately +125C) after cleaning. This helps to bake out
any residue. Remove the port plugs to permit the residual vapors to escape.
Clean the silicone door gaskets with a damp cloth or mild detergent.
Clean the exterior painted surfaces with a damp cloth or mild detergent. If you are using a
detergent, test a small inconspicuous area to make sure it does not damage the finish.
Page 6-1
Chapter 6 Maintenance
Repair minor tears with a high quality RTV silicone such as GE RTV167.
Check the integrity of the door seal by closing the door on a sheet of paper. With the door
closed, slowly pull the paper. You should feel the resistance getting lighter as the paper goes
past the inner gasket. Repeat this all around the door at several places.
If the seal is not tight, adjust the door latch. The stainless-steel catch (on the bracket that is
mounted to the chamber) has slotted holes to permit adjustment.
If the seal is still loose on the hinge side, adjust the door hinge. The hinges have slotted holes
(on the door side) to permit adjustment.
Page 6-2
Chapter 6 Maintenance
NOTE: You may need to clean the condenser more frequently if the chamber is in a dusty
environment. You may be able to clean the condenser less frequently if the chamber is in a very
clean environment.
Page 6-3
Chapter 6 Maintenance
Chapter 6 Maintenance
Wait at least 2 hours after running a humidity test before draining the system.
Turn all front panel switches and the Disconnect Switch OFF.
Remove the left side panel (left side if facing the front of the chamber.
Open the Drain Valve by turning its handle 90 degrees counter-clockwise.
After the system is drained, close the Drain Valve by turning its handle 90 degrees
clockwise.
Drain Valve
(closed position shown)
Page 6-5
Chapter 6 Maintenance
Page 6-6
Chapter 6 Maintenance
Page 6-7
Chapter 6 Maintenance
Chapter 6 Maintenance
Page 6-9
Chapter 6 Maintenance
! CAUTION:
! CAUTION:
Do NOT select Full Defaults in the Series F4 Factory Menu. Full Defaults
are NOT the TestEquity configuration parameters for this chamber. If you
select Full Defaults, you must reconfigure all System and Operation
Parameters as documented in the TestEquity manual, NOT the Series F4
Users Manual.
TestEquity has configured the Temperature/Humidity Controller with the following parameters.
Page 6-10
Chapter 6 Maintenance
3.0 C
(9.0 if F)
3.0 %
US
C
Yes
90%
90%
0%
0%
No
No
10 Sec
Cont
T/C
T
0.0
-75.0 C
(-103.0 if F)
175.0 C
(347.0 if F)
0.0
1.0 sec
Self-Clear
Control Output 1A
Function
Process Type
Cycle Time
Hi Power Limit
Low Power Limit
Process
4-20ma
% rh
0
0.0 %
100.0 %
7.0 %
97.0 %
0.0 %
1.0 sec
Self-Clear
Analog Input 3
Choose Sensor
Off
Retransmit Output 1
Retransmit Source
Analog Range
Low Scale
Scale Offset
Input 1
0-5V
-75.0C
(-103.0 if F)
175.0C
(347.0 if F)
0.0
Retransmit Output 2
Retransmit Source
Analog Range
Low Scale
High Scale
Scale Offset
Input 2
0-5V
0
100
0
High Scale
Control Output 1B
Function
Process Type
Cycle Time
Hi Power Limit
Low Power Limit
Cool
Fixed Time
10.0 sec
100 %
0%
Control Output 2A
Function
Process Type
Cycle Time
Hi Power Limit
Low Power Limit
Heat
Fixed Time
2.0 sec
100 %
0%
Control Output 2B
Function
Process Type
Cycle Time
Hi Power Limit
Low Power Limit
Cool
Fixed Time
10.0 sec
100 %
0%
Analog Input 2
Choose Sensor
Type
Choose Units
Choose Decimal
Scale Low
Scale High
SP Low Limit
SP High Limit
Cal Offset
Filter Time
Error Latch
Alarm Output 1
Name
Type
Alarm Output 2
Name
Type
Source
Latches
Silencing
Hysterisis
Sides
Logic
Show Message
Heat
Fixed Time
2.0 sec
100 %
0%
DIGIT IN (1-4)
Off
No
Off
No
Process
Input 1
Self-Clears
No
2.0 C
(3.6 if F)
Both
Open
No
Compressor
-2%
2%
60 sec
10 sec
Communications Menu
Main Page\Go to Setup\System\
Communications
Baud Rate
Address
19200
(9600 w/GPIB)
1
Input 2
Current File
Current Step
Set Point 1
Set Point 2
Step Type
Target SP1
Target SP2
Wait For Status
Jump Count
Time
Digital Outs
% Power 1
% Power 2
Time
Date
Page 6-11
Chapter 6 Maintenance
5.0 C (9.0 if F)
0.20 min
0.00 min
0.0 C
0.0
5.0 C (9.0 if F)
0.20 min
0.00 min
0.0
0.0
10 %
0.10 min
0.00 min
0.0 %
0.0
10.0 %
0.10 min
0.00 min
0.0 %
0.0
7.0 C (44.6 if F)
90.0 C (194.0 if F)
Set Point
Oper. Autotune PID
Oper. Edit PID
Oper. Alarm SP
Profile
Setup
Factory
Page 6-12
Full Access
Password
Password
Password
Full Access
Password
Password
Chapter 6 Maintenance
TestEquity has configured the Limit Controller with the following parameters.
Home Page
Setup Page
Operations Page
[LIM] Limit Menu
[OPEr] Operations Page
[L~Lo] Low Limit Set Point
[L~hi] High Limit Set Point
[CAL1] Calibration Offset
[~-75]
[~175]
[`~`0]
-[SAFE]
Factory Page
[~~tc]
[~~~t]
[~-75]
[~175]
[`~`0]
[`~`0]
[`Pr1]
-[L~St]
---
[``C]
[`LAt]
Page 6-13
Chapter 7 Warranty
Chapter 7 - Warranty
TestEquity Inc. Limited Warranty
TestEquity Inc. (TestEquity) warrants Environmental Chambers (Equipment) manufactured by TestEquity and
supplied under this contract to be free from defects in materials and workmanship under normal use and proper
maintenance.
TestEquity will repair or replace any defective part for a period of THREE YEARS from the date of invoice.
TestEquity reserves the right to require any defective part be returned, freight prepaid, to TestEquitys factory or to
inspect any defective part at the Purchasers site. TestEquity shall have sole discretion to determine whether any part
is defective and whether any defective part will be repaired or replaced. This limited warranty shall extend to any
standard chamber accessory and component part that is normally sold by TestEquity. Non-standard accessories and
component parts specified by the Purchaser shall be warranted only to the extent of the original manufacturers
warranty, if any exists.
If the repair or replacement is performed in the FIRST YEAR from the date of invoice, TestEquity will also pay for
the labor associated with the repair at the Purchasers site, subject to TestEquitys prior approval. During the
SECOND and THIRD YEAR of the warranty period, Purchaser will be responsible for the installation and cost of
installation of replacement or repaired parts. Purchaser shall notify TestEquity in writing of any alleged defect
within 10 days after its discovery within the warranty period. TestEquity reserves the right to satisfy the labor
portion of this limited warranty either through its own service personnel or an authorized agent. In order to provide
expeditious service, TestEquity reserves the right to satisfy its limited warranty obligation by sending replacement
parts to be installed by the Purchaser if they can be installed easily without special tools or training. TestEquity
reserves the right to satisfy this limited warranty by requiring the Purchaser to return the Equipment to TestEquity
when such return is feasible.
The following parts are excluded from this limited warranty and are sold as-is or are considered expendable: interior
light bulb, viewing window, paint and cosmetic surface finishes and treatments, and port plugs.
This limited warranty shall extend in full to Equipment installed within continental United States and Canada. For
all other locations, Purchaser is responsible for all labor costs for repairs or parts installation, and for all shipping
costs associated with providing replacement parts.
This limited warranty does not cover: (1) Defects or damages arising as the result of shipment by common carriers
or private transportation, unless TestEquity undertakes shipment and transportation of the Equipment to Purchasers
site or contractually assumes the risk of damage to the Equipment in shipment; (2) Defects or damages arising out
of, or as the result, of mishandling, modification, or improper start up, installation or maintenance of the Equipment
(including start up, installation or maintenance not in accordance with TestEquitys written procedures); (3) Defects
or damages resulting from, or arising out of, abuse, misuse, neglect, intentional damage, accident, fire, flood,
earthquake, or any other act of God.
This warranty as to Equipment is LIMITED to repair or replacement of parts or Equipment in the determination of
TestEquity Inc. THE FORGOING LIMITED WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES
INCLUDING THE IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE AND
MERCHANTABILITY. TestEquity Inc. DISCLAIMS ANY LIABILITY FOR ANY DAMAGES RESULTING
FROM DELAY OR LOSS OF USE IN SERVICE OR REPAIR, OR FOR INCIDENTAL OR CONSEQUENTIAL
DAMAGES ARISING OUT OF OR IN CONNECTION WITH THE USE OR PERFORMANCE OF THE
EQUIPMENT, EXCEPT AS STATED IN THIS PARAGRAPH.
This limited warranty cannot be modified in any way except in writing by both TestEquity and Purchaser.
Invalidation of any one or more of the provisions of this limited warranty shall in no way affect any of the other
provisions hereof, which remain in full force and effect.
This limited warranty shall be extended only to the first Purchaser of this Equipment and is not transferable.
Page 7-1
Chapter 8 Drawings
Chapter 8 Drawings
Page 8-1
214A
214C
Terminal Strip
214B
231
232
215
213A
206F
213B
F11
203
206E
206A
202A
206B
202B
206C
206D
202C
250
202D
F1 F2 F3 F4 F5 F6 F7 F8 F9 F10
204