Sample Autoclave Validation Report 1-04-2013
Sample Autoclave Validation Report 1-04-2013
Sample Autoclave Validation Report 1-04-2013
Digitally signed
by Tom Holm
Date:
2013.02.06
11:12:03 -06'00'
Instruments used in this validation have been certified by or have been referred to instruments certified
by the National Institute of Standards and Technology (NIST).
Equipment
Owner:
Sample Lab
Address: 1877 125th Street
Yourtown, MN 55099
Tel:
(651) 555-1234
Eq. Type: Autoclave
Mfg:
Market Forge
Model:
STM-E
Serial#: 1097186
Location: Laboratory at above address
Val. Date: 1/04/2013
Validation Performed By
Tom Holm
BME Lab and Science
79 County Road B E
Little Canada, MN 55117
(651) 646-5339
[email protected]
Purpose.
The purpose of this validation test is to perform an annual maintenance test of the
autoclaves temperature and pressure as required by MN Rules Chapter 4740.2110 subp.
11 and the National Environmental Laboratory Accreditation Conference (NELAC now
referred to as TNI).
Test Equipment.
The test equipment consisted of 6 MPIII DataTrace NIST certified and traceable
collectors manufactured by Mesa Labs, Inc. Technical specifications for the collectors
used in this validation are included in Attachment 1.
Procedure.
To perform the stated tests, 6 data logger collectors were placed in the autoclave as
illustrated in the Attachment 2, Collector Mapping Diagram. Of the 6 probes, 2
measured pressure and temperature, while 4 measured temperature only. The autoclave
was then programmed for a 60 minute cycle time at 250 0F (122 0C). The collectors
were programmed to record readings every 60 seconds throughout the cycle, which
began at approximately 12:15 PM and ended at approximately 1:20 PM. The probes
were also used to verify the autoclave timer. All of the recorded readings from the
collectors were then downloaded to a computer and the results tabulated and graphed.
Page 1 of 2
Recommendations.
Perform regular cleaning and annual testing. This test follows closely to the results of
the test from one year ago.
Data.
Attachments and Data from this validation include the following:
Attachment 1:
Attachment 2:
Attachment 3:
Attachment 4:
Attachment 5:
Attachment 6:
Page 2 of 2
Collector Type
Serial #
Mesa Labs DataTrace Micropack III Pressure
Data Logger
M3P15108,
M3P15120,
M3P15157
Temperature
Range
Temperature
Accuracy
Pressure
Range
Pressure
Accuracy
Mesa Labs DataTrace Micropack III
Temperature Data Logger
M3T22042,
M3T22621,
M3T22629
Temperature
Range
Temperature
Accuracy
Page 1 of 1
BME Labstore
Attachment 2
Sample Customer
Manufacturer: Market Forge
Model: STM-E
Serial #: 1097186
T1
T2
Top Shelf
Bottom Shelf
Legend
TP1
TP3
T1
T2
M3T12154
M3T21982
T4
T5
TP3
TP1
M3T21990
M3T21991
M3P15157
M3P15108
T5
T4
Door
BME Labstore
Sample Customer
Manufacturer: Market Forge
Model: STM-E
Serial #: 1097186
Attachment 3
Autoclave Temperature Validation Test
Validation Date: Jan. 04, 2013
300.0
Hold Time
250.0
Come-up
Cool Down/
Exhaust
Collector
Temperature (0F)
200.0
M3T21991
M3T21990
M3T12154
M3T21982
M3P15108
M3P15157
150.0
50.0
0.0
0:00
0:10
0:20
0:30
0:40
0:50
1:00
1:10
1:20
1:30
Validation Time
BME Labstore
Sample Customer
Manufacturer: Market Forge
Model: STM-E
Serial #: 1097186
Attachment 4
Autoclave Pressure Validation Test
Validationt Date: Jan. 04, 2013
35.0
Hold Time
30.0
Cool
Down/
Come-up
Pressure (PSIA)
25.0
Collector
20.0
M3P15108
M3P15157
15.0
Hold Time Pressures
10.0
Minimum:
Median:
Maximum:
Std. Dev.:
5.0
0.0
0:00
0:10
0:20
0:30
26.2 PSIA
28.9 PSIA
30.3 PSIA
0.80
0:40
0:50
1:00
1:10
1:20
1:30
Validation Time
BME Labstore
Sample Customer
Manufacturer: Market Forge
Model: STM-E
Serial #: 1097186
Attachment 5
Autoclave F0 Value
Validation Date: Jan. 04, 2013
50.0
45.0
Cool Down
Hold Time F0 Values
40.0
Collector
Minimum: 1.0
Median:
24.5
Maximum: 46.8
35.0
M3T21991
M3T21990
M3T12154
M3T21982
M3P15108
M3P15157
F0 Value
30.0
Hold Time
25.0
20.0
15.0
F0 = 8 @ 2 min prior to Hold Time due to first
aborted start
10.0
5.0
Come-up
0.0
0:00
0:10
0:20
0:30
0:40
0:50
1:00
1:10
1:20
1:30
Validation Time
BME Labstore
Attachment 6
Validation Data Jan. 04, 2013
Sample Customer
Manufacturer: Market Forge
Model: STM-E
Serial #: 1097186
Legend
Start
Hold Time - Start/Stop
F0 =8
M3T21991
Sample Time
Reading#
1/0/00 0:00
1
1/0/00 0:01
2
1/0/00 0:02
3
1/0/00 0:03
4
1/0/00 0:04
5
1/0/00 0:05
6
1/0/00 0:06
7
1/0/00 0:07
8
1/0/00 0:08
9
1/0/00 0:09
10
1/0/00 0:10
11
1/0/00 0:11
12
1/0/00 0:12
13
1/0/00 0:13
14
1/0/00 0:14
15
1/0/00 0:15
16
1/0/00 0:16
17
1/0/00 0:17
18
1/0/00 0:18
19
1/0/00 0:19
20
1/0/00 0:20
21
1/0/00 0:21
22
1/0/00 0:22
23
1/0/00 0:23
24
1/0/00 0:24
25
1/0/00 0:25
26
1/0/00 0:26
27
1/0/00 0:27
28
1/0/00 0:28
29
1/0/00 0:29
30
1/0/00 0:30
31
1/0/00 0:31
32
1/0/00 0:32
33
Temp F
67.9
66.5
66.3
66.6
71.3
91.3
146.5
187.6
209.5
234.0
240.2
243.7
245.4
246.5
247.5
248.2
248.8
249.3
249.7
249.9
249.7
249.6
249.4
249.2
249.1
249.0
249.0
249.4
249.9
249.9
249.6
249.3
249.0
M3T21990
Fo
Temp F
0.0
64.9
0.0
64.0
0.0
64.4
0.0
65.1
0.0
70.8
0.0
97.5
0.0
136.7
0.0
177.4
0.0
207.4
0.1
233.6
0.4
239.9
0.9
243.3
1.4
245.0
2.1
246.1
2.8
247.3
3.6
248.0
4.4
248.7
5.3
249.2
6.3
249.6
7.3
249.7
8.3
249.6
9.2
249.4
10.1
249.2
11.0
249.0
11.9
248.9
12.8
248.8
13.7
249.0
14.6
249.4
15.6
249.7
16.6
249.7
17.5
249.4
18.5
249.1
19.3
248.9
M3T12154
Fo
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.1
0.4
0.8
1.3
2.0
2.7
3.4
4.3
5.2
6.1
7.1
8.1
9.0
9.9
10.8
11.6
12.5
13.4
14.3
15.3
16.2
17.1
18.0
18.9
Temp F
66.0
65.4
65.6
66.3
74.5
101.3
153.2
187.7
212.8
235.6
240.4
243.9
245.7
246.6
247.5
248.1
248.7
249.2
249.7
249.8
249.7
249.5
249.4
249.2
249.0
248.9
249.0
249.4
249.8
249.9
249.6
249.3
249.0
M3T21982
Fo
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.2
0.5
0.9
1.5
2.1
2.9
3.7
4.5
5.4
6.4
7.3
8.3
9.2
10.2
11.1
11.9
12.8
13.7
14.6
15.6
16.6
17.5
18.5
19.3
Temp F
65.4
64.9
65.1
65.9
71.6
95.4
146.7
184.8
209.9
233.9
240.2
243.8
245.4
246.5
247.4
248.1
248.8
249.3
249.7
249.8
249.7
249.5
249.3
249.2
249.0
249.0
249.1
249.5
249.8
249.8
249.5
249.2
249.0
Page 5 of 7
M3P15108
Fo
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.1
0.4
0.9
1.4
2.1
2.8
3.6
4.4
5.3
6.3
7.3
8.2
9.2
10.1
11.0
11.9
12.7
13.6
14.6
15.6
16.5
17.5
18.4
19.2
Temp F
62.0
62.4
62.8
63.1
63.7
66.7
80.4
143.9
187.7
218.6
235.9
241.3
243.9
245.8
247.5
248.5
249.0
249.2
249.3
249.3
249.3
249.3
249.3
249.3
249.4
249.5
249.6
249.6
249.5
249.3
249.2
249.1
249.1
Fo
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.2
0.5
1.0
1.6
2.3
3.1
4.0
4.9
5.8
6.7
7.6
8.5
9.5
10.4
11.3
12.2
13.2
14.1
15.1
16.0
16.9
17.8
18.7
M3P15157
PSIA
15.3
15.2
15.2
15.2
15.2
15.2
15.6
17.8
23.5
30.6
30.5
30.5
30.2
30.0
29.9
29.8
29.8
30.1
30.3
30.3
30.2
30.1
29.9
29.8
29.7
29.6
29.7
29.9
30.2
30.4
30.0
29.9
29.7
Temp F
58.9
59.9
60.5
61.0
61.9
65.8
78.5
127.4
179.8
216.7
235.9
241.2
243.8
245.6
247.3
248.3
248.9
249.2
249.2
249.3
249.2
249.2
249.2
249.3
249.3
249.5
249.6
249.6
249.5
249.3
249.1
249.1
249.1
Fo
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.2
0.5
1.0
1.5
2.2
3.0
3.9
4.8
5.7
6.6
7.5
8.4
9.3
10.2
11.2
12.1
13.1
14.0
14.9
15.9
16.7
17.6
18.5
PSIA
15.3
15.2
15.2
15.2
15.2
15.2
15.4
17.8
23.5
30.6
30.3
30.5
30.2
30.0
29.9
29.8
30.0
30.1
30.3
30.3
30.2
30.1
29.9
29.8
29.7
29.6
29.7
29.9
30.2
30.2
30.0
29.9
29.7
BME Labstore
Attachment 6
Validation Data Jan. 04, 2013
1/0/00 0:33
1/0/00 0:34
1/0/00 0:35
1/0/00 0:36
1/0/00 0:37
1/0/00 0:38
1/0/00 0:39
1/0/00 0:40
1/0/00 0:41
1/0/00 0:42
1/0/00 0:43
1/0/00 0:44
1/0/00 0:45
1/0/00 0:46
1/0/00 0:47
1/0/00 0:48
1/0/00 0:49
1/0/00 0:50
1/0/00 0:51
1/0/00 0:52
1/0/00 0:53
1/0/00 0:54
1/0/00 0:55
1/0/00 0:56
1/0/00 0:57
1/0/00 0:58
1/0/00 0:59
1/0/00 1:00
1/0/00 1:01
1/0/00 1:02
1/0/00 1:03
1/0/00 1:04
1/0/00 1:05
1/0/00 1:06
1/0/00 1:07
1/0/00 1:08
1/0/00 1:09
1/0/00 1:10
1/0/00 1:11
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
248.8
248.7
248.9
249.3
249.6
249.8
249.5
249.3
249.0
248.8
248.6
248.5
248.5
248.7
249.1
249.4
249.5
249.5
249.4
249.3
249.0
248.8
248.7
248.5
248.4
248.3
248.3
248.2
248.2
248.2
248.2
245.2
211.1
210.9
182.2
108.8
112.3
91.8
20.2
21.0
21.9
22.8
23.8
24.7
25.7
26.6
27.5
28.3
29.2
30.0
30.8
31.7
32.6
33.5
34.4
35.4
36.3
37.2
38.1
38.9
39.8
40.6
41.4
42.2
43.0
43.8
44.6
45.4
46.2
46.8
46.8
46.8
46.8
46.8
46.8
46.8
248.7
248.6
248.8
249.2
249.5
249.6
249.3
249.1
248.8
248.6
248.4
248.4
248.4
248.7
249.0
249.3
249.4
249.4
249.2
249.0
248.9
248.7
248.5
248.3
248.2
248.2
248.1
248.1
248.1
248.1
248.1
245.2
211.1
210.8
199.7
94.3
98.3
92.7
19.7
20.6
21.4
22.3
23.3
24.2
25.1
26.0
26.9
27.7
28.5
29.4
30.2
31.0
31.9
32.8
33.7
34.7
35.6
36.5
37.3
38.2
39.0
39.8
40.6
41.4
42.2
43.0
43.7
44.5
45.3
45.9
45.9
45.9
45.9
45.9
45.9
45.9
248.8
248.7
248.9
249.3
249.6
249.7
249.5
249.2
249.0
248.7
248.5
248.5
248.5
248.7
249.1
249.3
249.5
249.5
249.4
249.2
249.0
248.8
248.6
248.5
248.4
248.3
248.2
248.2
248.1
248.1
248.2
245.2
211.1
210.9
199.4
120.5
123.7
104.5
108.6
Sample Customer
Manufacturer: Market Forge
Model: STM-E
Serial
20.2 #: 1097186
248.8
20.1
21.0
21.9
22.8
23.8
24.7
25.7
26.6
27.5
28.3
29.1
29.9
30.8
31.6
32.5
33.4
34.4
35.3
36.2
37.1
38.0
38.9
39.7
40.5
41.3
42.1
42.9
43.7
44.5
45.3
46.1
46.6
46.6
46.6
46.6
46.6
46.6
46.6
46.6
Page 6 of 7
248.8
249.0
249.4
249.6
249.7
249.4
249.2
248.9
248.7
248.6
248.5
248.6
248.8
249.2
249.4
249.5
249.5
249.3
249.1
248.9
248.7
248.6
248.5
248.4
248.3
248.2
248.2
248.2
248.2
248.3
245.4
211.4
211.0
204.1
113.0
121.7
103.8
100.9
21.0
21.8
22.8
23.7
24.7
25.6
26.5
27.4
28.2
29.1
29.9
30.7
31.6
32.5
33.4
34.3
35.3
36.2
37.1
38.0
38.8
39.6
40.5
41.3
42.1
42.9
43.7
44.5
45.3
46.1
46.6
46.6
46.6
46.6
46.6
46.6
46.6
46.6
249.2
249.3
249.4
249.4
249.3
249.1
249.0
248.9
248.8
248.9
249.0
249.1
249.2
249.2
249.2
249.1
249.0
248.9
248.8
248.7
248.6
248.6
248.6
248.6
248.6
248.7
248.7
248.8
248.8
248.8
248.9
247.1
224.2
217.2
215.2
205.4
192.1
181.3
172.1
19.6
20.5
21.4
22.3
23.3
24.2
25.0
25.9
26.8
27.6
28.5
29.4
30.3
31.2
32.1
33.0
33.9
34.7
35.6
36.5
37.3
38.1
39.0
39.8
40.6
41.5
42.3
43.2
44.0
44.9
45.8
46.5
46.5
46.5
46.5
46.5
46.5
46.5
46.5
29.6
29.5
29.6
29.8
30.0
30.1
30.0
29.8
29.7
29.5
29.4
29.4
29.6
29.5
29.7
29.9
29.7
30.0
30.0
29.9
29.7
29.6
29.5
29.4
29.4
29.3
29.3
29.2
29.2
29.2
29.2
27.5
13.7
14.1
14.2
13.9
14.3
14.5
14.5
249.2
249.3
249.4
249.4
249.3
249.1
249.0
248.9
248.8
248.9
248.9
249.1
249.2
249.2
249.2
249.1
249.0
248.9
248.8
248.7
248.6
248.6
248.6
248.6
248.6
248.7
248.7
248.8
248.8
248.8
248.9
247.1
225.3
218.3
215.5
201.2
184.7
173.1
164.4
19.4
20.3
21.3
22.2
23.1
24.0
24.9
25.8
26.6
27.5
28.3
29.2
30.1
31.0
31.9
32.8
33.7
34.6
35.4
36.3
37.1
38.0
38.8
39.6
40.5
41.3
42.2
43.0
43.9
44.7
45.6
46.3
46.3
46.4
46.4
46.4
46.4
46.4
46.4
29.6
29.5
29.6
29.8
30.0
30.1
30.0
29.8
29.7
29.5
29.4
29.4
29.4
29.5
29.7
29.9
30.0
30.0
30.0
29.9
29.7
29.6
29.5
29.4
29.4
29.3
29.3
29.2
29.2
29.2
29.2
27.5
13.7
14.1
14.2
13.9
14.3
14.5
14.5
BME Labstore
Attachment 6
Validation Data Jan. 04, 2013
Min
Max
Median
Std. Dev.
Min. All
Median All
Max. All
Std. Dev. A
245.2
249.9
249.0
0.95
Temp.
243.8
249.0
249.9
0.9
1.4
46.8
24.7
13.78
F0
1.0
24.5
46.8
245.0
249.7
248.9
0.97
1.3
45.9
24.2
13.53
245.2
249.9
249.0
0.92
Sample Customer
Manufacturer: Market Forge
Model: STM-E
Serial #: 1097186
1.5
46.6
24.7
13.72
PSIA
27.5
29.8
30.4
0.4
Page 7 of 7
245.4
249.8
249.0
0.93
1.4
46.6
24.7
13.74
243.9
249.6
249.1
0.93
1.0
46.5
24.2
13.75
27.5
30.4
29.7
0.44
243.8
249.6
249.1
0.96
1.0
46.3
24.0
13.72
27.5
30.3
29.8