sciFLEX QuickStart
sciFLEX QuickStart
sciFLEX QuickStart
PRE-POWER UP CHECKLIST
Wash bottle filled....................................................................................... CHECK
Waste bottle empty .................................................................................. CHECK
System liquid bottle filled to appropriate level.................................... CHECK
• Use filtered (0,22µm) freshly degased millipore water
• The system liquid surface and the nozzle tip (when in front of the camera station)
!
must be on the same level.
POWERING UP
Multi-outlet power strip switch ........................................................ SWITCH ON
"Clunk"-sound of the spotter ................................................................... CHECK
• After ~20 seconds of powering the axes (S5/S11 only), a distinct sound is audible.
• In the SNAP WINDOW click: Image > show center to obtain the centered green cross.
• For better nozzle visibility activate the strobe LED by setting a pulse of 1µs and
select Continuous Dispensing and click START (image will go bright).