Pulsed Electron-Nuclear Double Resonance (Endor) at 140 GHZ
Pulsed Electron-Nuclear Double Resonance (Endor) at 140 GHZ
Pulsed Electron-Nuclear Double Resonance (Endor) at 140 GHZ
INTRODUCTION
1090-7807/99 $30.00
Copyright 1999 by Academic Press
All rights of reproduction in any form reserved.
ropy are typically dominated by the g-dispersion. Thus, microwave pulsed irradiation at arbitrary field positions in the EPR
line excites only molecules with particular orientations of the
g-tensor principal axis with respect to the magnetic field. This
leads to a simplification of the data, where spectral information
from only a few molecular orientations is observed. Further,
high magnetic fields also lead to much higher resolution in
ENDOR spectroscopy of low-gamma nuclei. For instance, in
contrast to the situation at X-band, hyperfine spectra of 13C,
15
N, 14N, and 2H are completely resolved from one another at
our field of 5 T.
Therefore, the combination of high frequency EPR and
ENDOR techniques is an attractive spectroscopic tool, as has
been demonstrated at 95 GHz (10, 15, 16) and with CW experiments at 250 GHz (17). One challenging aspect of pulsed
EPR/ENDOR at high frequencies is the design and implementation of efficient double resonance probes. The paucity of
stable millimeter-wave power sources and amplifiers requires
the use of high quality microwave cavities in order to obtain
short microwave pulses (t p/ 2 ! 200 ns). In contrast to standard
9 GHz EPR, high quality cavities (Q $ 3000) can be employed at high frequencies without degrading the temporal
resolution (i.e., dead time) of the pulsed spectrometer. The
cylindrical TE 011 cavity, introduced by Lebedev et al. (18) at
140 GHz, has been employed for pulsed EPR at millimeter
wavelengths because of its high filling factor and ease of
tuning. One disadvantage of this cavity is its miniature size,
which requires small samples and external mounting of the
radio frequency NMR coil for ENDOR, leading to compromises between RF penetration and the quality factor of the EPR
cavity.
In this contribution we describe the design and operation of
a new spectrometer for pulsed ENDOR at 140 GHz and demonstrate its capabilities with studies of the tyrosyl radical from
E. coli ribonucleotide reductase (RNR). The microwave bridge
described in detail by Prisner et al. (11) and further by Becerra
et al. (12) was modified in order to achieve high-power microwave pulses. A double resonance probe for efficient EPR
and NMR irradiation has been implemented together with a
232
233
234
BENNATI ET AL.
FIG. 1. Top: Side view of the 140 GHz cylindrical resonator used for
ENDOR displaying the different slots for improved RF penetration, the fixed
and movable plungers, and the sample tube. Bottom: Tuning curve of the
ENDOR resonator loaded with a 0.4 mm (ID) capillary and solid BDPA/
polystyrene sample.
C2
5
C1
r
,
RT
[1]
235
FIG. 2. (a) Schematic representation of the NMR circuit for the ENDOR
probe. C 1 and C 2 are variable capacitors controlled by a high-precision micrometer (M). The micrometer position is set by the computer through a digital
controller. (b) Diagram of the NMR transmission line for variable frequency
ranges and four-phase capability.
[2]
We arrive at B 1 5 0.08 mT, in agreement with the experimental 90 pulse length of 100 ns.
The efficiency of the RF irradiation was first tested by
measuring the Rabi oscillation of the nuclear sublevels in an
ENDOR experiment. As a standard sample we use the solid
organic radical bis-diphenylene-phenyl-allyl (BDPA) doped in
polystyrene (1%) at room temperature. The Rabi oscillation is
usually monitored on the maximum of the ENDOR line at 209
MHz using a Mims-ENDOR type of EPR preparation detection scheme and recording the electron spin echo intensity as a
function of the RF pulse length. A typical curve is displayed in
Fig. 3 for an ENDOR probe with a four-turn RF coil and an RF
power of 450 W. The RF field strength B 2 can be estimated
considering the expression for the nutation of a spin-21 nucleus
under the effect of a magnetic field B 2 5 v / g perpendicular to
the external field B 0 (23):
SPECTROMETER PERFORMANCE
ENDOR Probe
The performance of the ENDOR probe was assessed in
different ways. Initially, quality factors of the cylindrical cavity
were measured using a microwave frequency sweeper (Insight
Products Corp.) with a calibrated frequency output (20 180
^I z & 5
[3]
236
BENNATI ET AL.
E5
N
,
~S/N! z n z DH 2pp
[4]
E5
N
e 22 t /T 2,
~S/N!~2m I 1 1!DH 1/ 2
[5]
ENDOR Performance
Two standard pulsed ENDOR experiments, Davies- and
Mims-ENDOR, have been employed in order to demonstrate
the performance of the spectrometer. As typical experimental
parameters we use p/2 microwave pulses of 100 ns and RF p
pulses of 10 ms. Baseline correction is performed by subtracting two consecutive sequences and toggling the RF pulse.
Spectra from 1% BDPA in polystyrene at room temperature
were obtained from a single scan of the RF frequency with an
integration averaging of 100 shots/point.
237
FIG. 4. 1H ENDOR spectra of BDPA in polystyrene (1%) at room temperature, n 0 5 211.7 MHz. Mims-ENDOR (top), Davies-ENDOR (middle),
Davies-ENDOR simulation (bottom). Experimental parameters: MW p pulse: 200 ns, t (Mims) 5 200 ns, t (Davies) 5 400 ns, RF p pulse: 10 ms, RF power:
400 W, 100 averages/point, single scan spectra. Simulation parameters used are: g x 5 2.00263, g y 5 2.00260, g z 5 2.00257. The following hyperfine
constants are in megahertz and represent the absolute values: 1 A x 5 1.0, 1 A y 5 1.0, 1 A z 5 1.26, 2 A x 5 7.7, 2 A y 5 5.3, 2 A z 5 2.0. Inset: Structure of BDPA
radical.
238
BENNATI ET AL.
effects arising from preparation and detection pulses. Roomtemperature spectra are displayed in Fig. 4.
Mims-ENDOR, which is a stimulated echo sequence with a RF
pulse at a nuclear transition inserted between the second and third
mw-pulses (p/2tp/2RFp/2), is known to preferentially detect small hyperfine couplings. This is demonstrated by the experimental comparison of Mims- with Davies-ENDOR spectra in
Figs. 4a and 4b. Additionally, a strong 1H matrix peak is observed
at the Larmor frequency. However, the Mims-ENDOR intensity
suffers from a periodic dependence on the pulse spacing t, given
by I (ENDOR) ; 1 2 cos(2pAt), where A is the hyperfine constant
and t is the time between the two p/2 excitation pulses. The
spectrum represented in Fig. 4a was recorded with t 5 200 ns,
which leads to clearly visible intensity minima at Dn > 62.5
MHz. Davies-ENDOR (pRFp/2tp) (2) is the more attractive technique because of limited lineshape distortion, essential for
accurate simulations. Spectra are affected by an intensity hole
centered at Dn 5 0, which reflects the excitation band width of the
preparation pulse. Note that the doublet at 60.7 MHz is much
weaker than the second one at 62.5 MHz, although they are
determined by the same number of protons. We find that the
intensity of the center increases for longer and weaker preparation
pulses, but the sensitivity also degrades. The overall signal-tonoise is comparable to that of Mims-ENDOR, and therefore
Davies-ENDOR was typically used. A more detailed discussion
of the lineshape is given in the next section.
n iI~6! 5 A i ~6!
O m ~ A ~1! 2 A ~2!!
n
i
I
for EPR
[6]
i51
for ENDOR
I ENDOR~D v I! 5 I N(DvI)
[7]
[8]
where g and A are the angular-dependent g-factor and hyperfine constant, respectively, the summation runs over the i
1`
dD v Sg~D v S!
2`
[9]
g b eB 0
nS 5
2
h
v *21s
,
Dv 2I 1 v *21s
[10]
with v*1s ' 0.7 v 1s, where the phenomenological scaling factor
of 0.7 includes relaxation by spin diffusion effects. Finally, RF
power broadening leads to ENDOR linewidths on the order of
1/t RF, where t RF is the RF pulse length. This can be derived
directly from Eq. [3]. For a 10 ms pulse the ENDOR linewidth
is therefore ;0.1 MHz. For faster computation, the ENDOR
lineshape is approximated by a Lorentzian line and the lineshape convolution is performed in the time domain.
An example of the calculated Davies-ENDOR spectrum of
BDPA for a 200 ns 180 preparation pulse is displayed in Fig.
4 together with the experimental spectrum. The calculation is
performed for 16 nuclear spins coupled to the electron spin.
Although there are only two inequivalent sets of protons, the
full spin space must be considered for computation of the EPR
line and for proper orientation selection. The weak intensity of
the doublet at 60.7 MHz compared to the doublet at 62.5
MHz is a direct consequence of the preparation pulse. The
matrix peak is simulated by a simple Gaussian line and is
weighted in order to obtain the observed intensity. The experimental spectrum was recorded in the center of the EPR line,
which results in excitation of a wide range of molecular orientations. This is reflected in the broad, almost featureless
powder pattern for the larger hyperfine coupling. The calcu-
239
A major application of high-field ENDOR consists of performing hyperfine measurements at different field positions in
the EPR line corresponding to excitation of highly selected
molecular orientations. The g-anisotropy dispersion scales
with the external magnetic field, and therefore the 140 GHz
ENDOR spectra reveal field-dependent powder patterns that
provide a large number of constraints for the determination of
the hyperfine couplings. We demonstrate the advantages of 140
240
BENNATI ET AL.
Protons
Ax
(MHz)
Ay
(MHz)
Az
(MHz)
f
(deg)
2 or 6
2 or 6
b2
b2
3 or 5
b1
15.0
14.9
12.1
22.1
226.4
58.0
17.6
17.7
25.0
6.2
28.1
55.4
12.1
11.9
24.0
21.0
219.3
52.4
44
25
226
230
c
(deg)
u10u
28
Ref.
28
This work
28
This work
This work b and 26
This work b and 26
241
FIG. 9. Simulated ENDOR spectra of the tyrosyl radical at the low magnetic field edge ( g x ) for different angles f of the 2,6-proton bond (which defines
the direction of the A y principal axis) to the direction of the g y principal axis. Simulation curves (dotted lines) are displayed for f 5 10, 20, 25, 30, 35,
40. The angle f is related to the Euler angle a by a 5 90 2 f.
242
BENNATI ET AL.
SUMMARY
ACKNOWLEDGMENTS
Thanks are accorded to Professor D.J. Singel for conversations pointing out
the advantages of high frequency ENDOR. We are grateful to K. Kreischer for
the loan of the frequency sweeper used in the cavity quality measurements and
to Jan Henrik Ardenkjaer-Larsen (NYCOMED Innovation AB) for the gift of
the trityl radical. We would like to acknowledge P. Hofer for helpful discussions and comments on this manuscript. M. Bennati thanks the Deutsche
Forschungsgemeinschaft for a postdoctoral fellowship. C.T. Farrar was supported by a National Institutes of Health postdoctoral fellowship (GM-18790).
This research was supported by the National Institutes of Health (grants
GM-38352 and RR-00995).
REFERENCES
1. G. Feher, Phys. Rev. 103, 834 (1956).
2. E. R. Davies, Phys. Lett. 47A, 12 (1974).
3. W. B. Mims, Proc. Roy. Soc. A283, 452 457 (1965).
4. A. Grupp and M. Mehring, in Modern Pulsed and CW Electron
Spin Resonance (L. Kevan and M. Bowman, Eds.), p. 195, Wiley,
New York (1990).
5. C. Gemperle and A. Schweiger, Chem. Rev. 91, 1481 (1991).
6. H. Thomann and M. Bernardo, Isr. J. Chem. 32, 323337 (1992).
7. W. Bryan Lynch, K. A. Earle, and J. H. Freed, Rev. Sci. Instrum. 59,
13451351 (1988).
8. R. T. Weber, J. A. J. M. Disselhorst, L. J. Prevo, J. Schmidt, and
W. T. Wenckebach, J. Magn. Reson. 81, 129 144 (1989).
9. A. Y. Bresgunov, A. A. Dubinskii, V. N. Krymov, Y. G. Petrov, O. G.
Poluektov, and Y. S. Lebedev, Appl. Magn. Reson. 2, 715728
(1991).
10. O. Burghaus, M. Rohrer, T. Gotzinger, M. Plato, and K. Mobius,
Meas. Sci. Technol. 3, 765774 (1992).
19.
20.
21.
22.
23.
24.
243