Mass Spectros
Mass Spectros
Mass Spectros
Dr Nizam M. El-Ashgar
1 م09:40 22/12/2012
Mass Spectrometry
2 م09:40 22/12/2012
Molecular Mass Spectrometry
• Provides information about:
• 1- The elemental composition of samples of
matter
• 2- The structures of inorganic, organic and
biological molecules.
• The qualitative and quantitative composition
of complex mixture
• The structure and composition of solid
surfaces
• Isotope ratios of atoms in samples.
3 م09:40 22/12/2012
Acuracy of MWt Mesurments
• For large samples such as biomolecules, molecular masses can be
measured to within an accuracy of 0.01% of the total molecular
mass of the sample.
• Example: an error within a 4 Daltons (Da) or atomic mass units
(amu) for a sample of 40,000 Da. This is sufficient to allow minor
mass changes to be detected, e.g. the substitution of one amino
acid for another, or a post-translational modification.
4 م09:40 22/12/2012
Where are mass spectrometers used?
Mass spectrometers are used in industry and academia for both
routine and research purposes. The following list is just a brief
summary of the major mass spectrometric applications:
Biotechnology: the analysis of proteins, peptides, oligonucleotides
Pharmaceutical: drug discovery, combinatorial chemistry,
pharmacokinetics, drug metabolism
Clinical:, haemoglobin analysis, drug testing
Environmental: water quality, food contamination
Geological: oil composition
5 م09:40 22/12/2012
Mass Spectra
• When the electron beam ionizes the molecule, the species that
is formed is called a radical cation, and symbolized as M+•.
• The radical cation M+• is called the molecular ion or parent ion.
• The mass of M+• represents the molecular weight of M.
• Because M is unstable, it decomposes to form fragments of
radicals and cations that have a lower molecular weight than
M+•.
• The mass spectrometer analyzes the masses of cations.
• A mass spectrum is a plot of the amount of each cation (its
relative abundance) versus its mass to charge ratio (m/z, where
m is mass, and z is charge).
• Since z is almost always +1, m/z actually measures the mass
(m) of the individual ions.
6 م09:40 22/12/2012
Mass Spectrometry
• The tallest peak in the mass spectrum is called the base peak.
• The base peak is also the M peak, although this may not always
be the case.
• Though most C atoms have an atomic mass of 12, 1.1% have a
mass of 13. Thus, 13CH4 is responsible for the peak at m/z = 17.
This is called the M + 1 peak.
7 م09:40 22/12/2012
• Since the molecular ion is unstable, it fragments into
other cations and radical cations containing one, two,
three, or four fewer hydrogen atoms than methane itself.
• Thus, the peaks at m/z 15, 14, 13 and 12 are due to these
lower molecular weight fragments.
8 م09:40 22/12/2012
Molecular Mass Spectra (hexane)
9 م09:40 22/12/2012
Explanation
• The collisions between energetic electrons and analyte
molecules (enough E) lead to excitation.
• Relaxation leads to fragmentation to lower masses ions
• Attraction of positive ions through the slit of a mass
spectrometer and sorted according to m/z ratios and
appear in the Mass spectrum.
• Mass spectrum is a plot of relative intensity versus m/z
• Base peak has the value of 100 (arbitrarily)
• The remaining computed as % of the base-peak height
• Modern MS programmed base-peak and normalize the
remaining peaks relative to that peak
10 م09:40 22/12/2012
11 09:40 22/12/2012م
Mass Spectrometer
All Instruments Have:
1. Sample Inlet
2. Ion Source
3. Mass Analyzer
4. Detector
5. Data System
http://www.asms.org
12 م09:40 22/12/2012
Ion sources
Starting point: Formation of gaseous analyte Ions.
Methods of ion formation: Two major categories:
1- Gas-phase sources
-The sample is first vaporized and then ionized.
-Restricted to thermally stable compounds of BP < 500 0C.
-Limited to Compounds of MWt’s <103dalton .
2- Desorption sources
The sample in a solid or liquid state is converted directly
into gaseous ions (not require volatilization of analyte
molecules)
Applicable to nonvolatile and thermally unstable samples
Applicable to analytes having of 105 dalton or larger.
13 م09:40 22/12/2012
Basic Type Name Ionizing agent
Electron Impact (E I) Energetic electrons
Chemical Ionization (CI) Reagent gaseous ions
14
Thermospray ionization (TI) High temperature
م09:40 22/12/2012
Hard and Soft Sources (another classification)
• Hard Sources:
• Impart sufficient E to analyte molecules (become of high excited E
state).
• Relaxation involves rupture of bonds producing fragment ions with
m/z < than that of molecular ion.
• Provides useful information about kinds of functional groups and
structure information.
• Soft Sources:
• Causes little fragmentation.
• Mainly the spectrum consists of the molecular ion peak and few
peaks
• Supplies accurate information about MWt of analyte.
Both are useful for analysis
15 م09:40 22/12/2012
16 09:40 22/12/2012م
The Electron-Impact Sources (EI)
M + e- M.+ + 2e-
where, M = analyte molecule,
M.+ = molecular ion.
Relaxation then usually takes place by extensive fragmentation, giving a
large number of positive ions of various masses that are less than that of the
molecular ion. These lower mass ions are called daughter ions.
17 م09:40 22/12/2012
Gaseous sample bombarded with beam of energetic electrons.
Electrons produced at heated W or Rh wire and accelerated to energy of
about 70 eV. Typically one in every million molecules undergoes ionization.
18 م09:40 22/12/2012
Electron Impact (EI)
19 م09:40 22/12/2012
• The positive ions produced are attracted through the slit
in the first accelerating plate by a small potential diff (5
V).
• With magnetic sector instrument high potentials (103 to
104) are applied to the accelerator plates)
KE = qV = zeV
• KE of an ion is independent of its mass and depends
only upon its charge and accelerating potential
• Velocity of an ion depends on its mass
KE = 1/2m2 or = (KE/m)1/2
20 م09:40 22/12/2012
Electron Impact
Advantages
• Well-Established
• Fragmentation Libraries
• Insoluble Samples
• Interface to GC
• Non-Polar Samples
• They are convenient and produce high ion currents.
• Extensive fragmentation can lead to unambiguous identification of analytes
• Disadvantages
• Parent Identification
• Need Volatile Sample
• Need Thermal Stability
• No Interface to LC
• Low Mass Compounds (<1000 amu)
• Solids Probe Requires Skilled Operator
• The need to volatilize the sample limits this method since it excludes
analysis of thermally unstable compounds.
• Excessive fragmentation can lead to the disappearance of the molecular ion
peak therefore preventing the molecular mass of the analyte to be
determined.
21 م09:40 22/12/2012
Electron-Impact Spectra
22 م09:40 22/12/2012
EI-MS of Methylene Chloride (top) and 1-Pentanol (bottom)
23 م09:40 22/12/2012
• Types of MS Peaks:
• Peaks occur at masses that are greater than that of the molecular ion are
attributable to ions containing isotopes.
• Examples:
12C1H 35Cl : m = 84
2 2
13C1H 35Cl : m = 85
2 2
12C1H 35Cl37Cl: m = 86
2
13C1H 35Cl37Cl: m = 87
2
12C1H 37Cl : m = 88
2 2
The size of various peaks depends upon the relative natural abundance of the
isotopes.
Note: F, P, I, Na occur only as single isotopes.
Isotope peaks some times provide useful means for determining the formula for
compound.
25 م09:40 22/12/2012
26 09:40 22/12/2012م
Collision product peaks
• Ion/molecule collisions, can produce peaks at higher
mass numbers than that of the molecular ion.
• At ordinary sample pressures, however the only
important reaction of this type: transferring of H atom to
the ion to give a protonated molecular ion, enhanced (M
+1 )+ peak results (second order reaction).
• The amount of product depends strongly upon the
reactant concentration.
• The height of an (M+1)+ peak due to this reaction
increases with increase in sample pressure than do the
heights of other peaks (detected).
27 م09:40 22/12/2012
Advantages of EI-MS
Convenient to use
Produce high ion currents (gives good sensitivities)
Often unambiguous identification of compounds possible
(extensive fragmentations)
Disadvantages of EI-MS
Extensive fragmentation may not leave parent ion
(Molecular peak ion).
Sample must be volatilized (thermal degradation) of some
analytes before ionization to be occur.
Only applicable to samples with molecular weights < 103 amu
28 م09:40 22/12/2012
Chemical ionization sources and spectra
Second most common process for ms
• In modern MS, EI ionization and chemical ionization are interchangeable.
Gaseous atoms are ionized by collision with ions produced by electron
bombardment of an excess of a reagent gas.
Positive ions usually used.
negative ion used with analyte contain very electronegative atoms.
Experimentally:
Modification of the electron beam ionization area of EI by adding
vacuum pump capacity and reducing width of the slit to the mass
analyzer so P reduced to 1 torr (Ionization area) and 10-5 torr in the
analyzer.
Gaseous regent introduced in the ionization region reagent:sample =
103-104. So electron beam reacts exclusively with reagent
molecules.
29 م09:40 22/12/2012
Common Reagents
Methane
• Reacts with high E electrons to give several ions (CH4+ and CH3+ (90 %),
CH2+).
• These ions react rapidly with additional methane molecules:
• CH4+ + CH4 CH5+ + CH3
• CH3+ + CH4 C2H5+ + H2
• Reactive collisions between sample molecules MH and CH5+ or C2H5+ take
place: (proton and hydride transfer).
• CH5+ + MH MH2+ + CH4 (proton transfer) gives (M +1)+ ion peak
• C2H5+ + MH MH2+ + C2H4 (proton transfer) gives (M +1)+ ion peak
• C2H5+ + MH M+ + C2H6 (hydride transfer) gives (M -1)+ ion peak
• C2H5+ + M M- C2H5+ gives (M + 29)+ ion peak
• Other reagents: propane, isobutane and ammonia that give different spectra
with a given analyte.
30 م09:40 22/12/2012
EI hard
source
CI Soft
source
31 م09:40 22/12/2012
Field ionization Sources
32 م09:40 22/12/2012
FI Source
33 م09:40 22/12/2012
MS of Glutamic Acid
Electron Impact
Field Ionization
Field Desorption
34 م09:40 22/12/2012
Desorption sources
• Simple spectra consist of only the molecular ion or the protonated molecular
ions.
• Exact mechanism is not understood.
35 م09:40 22/12/2012
Field Desorption Sources
36 م09:40 22/12/2012
Matrix-Assisted Laser Desorption/Ionization (MALDI)
37 م09:40 22/12/2012
38 09:40 22/12/2012م
MALDI: Matrix Assisted Laser Desorption Ionization
+/- 20 kV Grid (0 V)
39 م09:40 22/12/2012
Requirement for MALDI:
40 م09:40 22/12/2012
6. MALDI •
matrix has strong A at laser λ •
500 fmol of beta-
low mass, so can sublime •
D-galactosidase
can measure Mr of proteins and biomolecules •
mixed with nicotinic
acid on a silver
surface using Nd-
YAG 266 nm
excitation
RMM 117130
Fragments are
cluster ions [nM+H]+
and multiply-
charged ions
[nM+zH]z+
41 م09:40 22/12/2012
Electrospray ionization
44 م09:40 22/12/2012
Fast Atom Bombardment FAB
• FAB is a major role in the production of ions for ms studies of polar
high MWt. species.
• Samples in a condensed state often in a glycerol solution matrix are
ionized by bombardment with energetic (several KeV) xenon or
argon atoms.
• Both +ve and –ve analyte ions are sputtered from the surface of the
sample in a desorption process.
• Rapid sample heating by this process reduces sample
fragmentation.
• A beam of fast atoms is obtained by passing accelerated Ar or Xe
ions from an ion source or gun through a chamber containing Ar or
Xe atoms at P of about 10-5 torr.
• The high velocity ions undergo exchange reactions with the atoms
without loss of translational energy and a beam of energetic atoms
is formed.
• The lower E ions from the exchange are removed by an electrostatic
deflector.
• FAB of organic or biochemical compounds usually produces
significant amounts of molecular ions (and ion fragments) even for
high MWt and thermally unstable samples (over 10,000 da) and
45
detailed structure information obtained for samples of (3000 da).
م09:40 22/12/2012
FAB – sputtering/desorption technique.
Xe/Ar atoms
Atoms/ions
penetrate
several layers,
distribute k.e.
46 No fragmentn م09:40 22/12/2012
up to 9600 amu
FAB
Sample monolayer
sputtered in few s.
Nonvolatile
glycerol allows
surface layer to be
renewed.
Ar G (Gn+H)+
(M+H)+
sample M (M-H)-
(M+G+H)+
47 م09:40 22/12/2012
Mass Spectrometer
• Several Types. Only two will be described:
• Quadrupole spectrometer and the time-of-flight spectrometer.
• General description of instrument components:
48 م09:40 22/12/2012
Inlet System:
To introduce a very small amount of sample (mol or less) into the mass
spectrometer that converted to gaseous ions. ( a mean for volatilizing solid
or liquid samples is presents).
Ion sources:
Convert the components of a sample into ions.
(generally the inlet system and the ion source are combined into a single
component). The output is a stream of +ve or –ve ions are accelerated into
the mass analyzer.
Mass analyzer
analogous to grating in an optical spectrometer.
Dispersion is based upon the mass/charge ratios of the analyte ions rather than
upon the wavelength of photons. (Different categories of MS according to
mass analyzer).
Detectors:
Convert the beam of ions into an electrical signal that can then be
processed, stored in the memory of a computer and displayed or
recorded in a variety ways.
Vacuum System: To create low pressure (10-4 to 10-8 torr) in all the
instrument components except the signal processor and readout. To
prevent interaction of components with atmosphere so destroyed.
49 م09:40 22/12/2012
Sample Inlet System
50 م09:40 22/12/2012
• For liquids a small quantity of sample is introduced into a reservoir usually
with a micrometer syringe.
• In either case vacuum system is used to achieve sample P 10-4 – 10-5 torr.
• For samples with boiling points > 1500C T must be maintained at elevated T
by oven and heating tapes of maximum T of about 3500C.
• Sample is now in the gas phase is leaked into the ionization area of the
spectrometer via a metal or glass diaphragm containing one or more
pinholes.
51 م09:40 22/12/2012
MS – Direct Sample Introduction
• Solids and non volatile liquids can be introduced into the ionization region
by means of a sample holder probe which is inserted through a vacuum
lock.
The lock system:
Designed to limit V of air pumped after probe insertion.
Probes also used for limited quantity samples (few nanograms)(less wasted
than batch system).
Probe: Sample is held on the surface of a glass or Al capillary tube, a fine wire
or a small cup.
The probe is positioned within a few mm of the Ionization source and the slit
leading to spectrometer.
Vacuum used to maintain thermally unstable compounds for spectrum
measurements before major decomposition occurs. And to elevate
nonvolatile conc. in the ionization area (carbohydrates, steroids, metal-
organic species and low molecular weight polymeric substances).
Partial pressure attained is at least of 10-8 torr before onset decomposition.
53 م09:40 22/12/2012
• Chromatographic and capillary electrophoretic inlet System.
• MS are often coupled with GC and HPLC or capillary
electrophoresis columns.
• Separation and determination of the components of complex
mixture is obtained.
• Specialized inlet systems is needed.
54 م09:40 22/12/2012
Mass Analyzers
• Several devices are available for separating ions with different m/z ratios.
Mass analyzer should be:
- Capable of distinguishing between minute mass differences.
- Allows passage of sufficient number of ions to yield measurable currents.
Resolution of mass spectrometer: R
R = m/Dm
Where, m : mass of the first peak (or mean mass of the two peaks)
Dm: mass difference between two adjacent peaks
Two peaks are considered to be separated if height of valley between them
10% of their height.
R of 4000 would resolve peaks occurring at m/z values of 400.0 and 400.1 (or
40.00 and 40.01)
55 م09:40 22/12/2012
• R needed in MS depends upon its application.
Example 1: same nominal mass ions: C2H4+, CH2N+, N2+ and CO+
(All ions of nominal mass 28 Da).
Exact masses: 28.0313, 28.0187, 28.0061 and 27.9949 Da respectively.
These requires an instrument with a resolution of several thousands.
Example 2: Low MWt ions with a unit mass difference or more: NH3+ (m = 17)
and CH4+ (m = 16) R instrument of 50 or less is sufficient.
56 م09:40 22/12/2012
Magnetic Sector Analyzers
57 م09:40 22/12/2012
58 09:40 22/12/2012م
• Operation:
• Ions formed by Electron impact are accelerated through slit B into the metal
analyzer tube internal P 10-7 torr.
• Ions of different mass can be scanned across the exit slit by varying the
field strength of the magnet or the accelerating potential between slits
A and B.
• Ions passing through the exit slit fall on a collector electrode, ion current
resulted that amplified and recorded.
• Translation or KE of an ion of mass m bearing z exiting slit B is given by:
KE = ZeV = ½ mv2
Where V is voltage between A and B
V is velocity of the ion after acceleration
e is the electronic charge = 1.60x10-19 C
Note: all ions having the same number of charges are assumed to have the
same KE after accelerating regardless of their mass. (approximately true).
All ions leaving the slit have approximately same KE, the heavier ions must
travel at lower velocities.
The path in sector by ions of a given m/z represents a balance between two
forces acting upon them.
59 م09:40 22/12/2012
The magnetic force FM:
FM =BzeV
Where B is MF strength
The balancing centripetal force:
Fc = mv2/r
Where r is the radius of curvature of the magnetic sector.
For ion to traverse the circular path to the collector,
FM and FC must be equal
FM =BzeV = Fc = mv2/r
v = Bzer/m
Substituting the previous equation in ZeV = ½ mv2
m/z = B2r2e/2V
Mass Spectra:
Varying one of three variables (B, V or r) while holding the other two constants.
Modern MS ions are sorted by holding V and r constant while varying the
current in the magnet and thus B.
In sector MS (using photographic recording) B and V are constants, r is the
variable.
60 م09:40 22/12/2012
• What accelerating potential will be required to direct a singly charged water molecule
through the exit slit of a magnetic mass spectrometer if B = 0.240 T and r of curvature
of the ion through the magnetic field is 12.7 cm?
SI units:
ez = 1.60x10-19 Cx 1 r = 0.127 m
18.02 g H 2O / mol kg
m x103
6.02 x1023 H 2O / mol g
m = 2.99x10-26kg
B = 0.240 T = 0.240 W/m2
m/z = B2r2e/2V or V = B2r2ez/2m
= [0.240 W/m2]2[0.127m]2[1.60x10-19C]
2x2.99x10-26kg
= 2.49 x103 W2C/m2kg or volts
(In SI base units, the dimensions of the weber are kg.m2/S2.A . In derived units, they
are volt-seconds V.s , or joules per amp J/A.
The weber is a large unit, equal to 1 T m2 = 108 maxwells.
•
61 م09:40 22/12/2012
Double focusing spectrometers
62 م09:40 22/12/2012
Double Focusing Spectrometers:
Correction for both the directional and E distribution of ions.
Both directional and E aberrations of a population of ions are simultaneously
minimized by use of carefully selected combinations of electrostatic and
magnetic fields.
63 م09:40 22/12/2012
• The ion beam is first passed through an electrostatic analyzer (ESA).
• Consists of two smooth curved metallic plates across which a dc potential is
applied which limiting the KE of the ions reaching the magnetic sector to a
closely defined range.
• Ions with E greater than average strike the upper side of ESA slit and lost to
the ground.
• Ions with E less than average strike the lower side of the ESA slit and are
thus removed.
• Directional focusing occurs along the focal plane (d).
• Energy focusing takes place along the plane e.
• Only ions of one m/z are double focused at the intersection of d and e for
any given V and B.
• The collector slit is located at this locus of double focus.
64 م09:40 22/12/2012
Quadrupole Mass Spectrometer
• Comparing with magnetic sector instrument it is:
• Less expensive and more rugged.
• More compact
• Found in commercial bench top ms
• Low scan times (<100 ms) which is useful for
chromatography.
• Most common mass analyzers in use today.
• Quadrupole mass analyzer is responsible for filtering the samples
ions.
• Consists of four parallel metal rods.
• Each rod pair is connected together electrically.
• A radio frequency voltage is applied between one pair of rods then
the other.
65 م09:40 22/12/2012
• Ions travel down the
quadrupole in between the rods
• Electric field separates ions
• Ions are subjected to complex
forces
• Only Ions of a particular m/z
reaches the detector
Advantages
• Inexpensive
• Easily Interfaced to Many
Ionization Methods
Disadvantages
• Low Resolution (< 4000)
• Low Accuracy (>100 ppm)
• MS/MS requires multiple
analyzers
• Low Mass Range (< 4000)
• Slow Scanning
66 م09:40 22/12/2012
Time-of-Flight MS (TOF)
• The +ve ions are produced periodically by bombardment of the
sample with brief pulses of electrons, secondary ions or laser
generated photons.
• The ions are then accelerated into a field-free drift tube by an
electric field pulse of 103 – 104 V
• Separation of ions on the basis of mass occurs during the
transition of the ions to the detector located at the end of the tube.
• All ion have same KE but their velocities are inversely
proportional to their masses
• Lighter particles arrived earlier.
• Typical flight times are 1 to 30 s.
67 م09:40 22/12/2012
Time-of-Flight MS (TOF)
68 م09:40 22/12/2012
69 09:40 22/12/2012م
Advantages
• Simplicity and ruggedness.
• Ease of accessibility of the ion source.
• Extremely High Mass Range (>1 MDa)
• Fast Scanning
Disadvantages
• Low Resolution (4000)
• Low sensitivity
• Low Accuracy (>200ppm)
• MS/MS not possible
70 م09:40 22/12/2012
Ion Trap Analyzers
• Gaseous anions or cations can be formed and confined for extended
periods by electric and/or magnetic fields.
• Several types, A simpler type of ion trap that used for GC/MS
• Now used to obtain mass spectra of a variety of analytes.
71 م09:40 22/12/2012
72 09:40 22/12/2012م
• It consists of a central doughnut-shaped ring electrode and a pair of endcap
electrodes.
• A variable RF voltage is applied to the ring electrode while the two end-cap
electrodes are grounded.
• Ions of appropriate m/z value circulate in a stable orbit within the cavity
surrounded the ring.
• By increasing RF voltage the orbit of heavier ions become stabilized , while
lighter become destabilized causing them to collide with the wall of the ring
electrode.
• By RF voltage scanning the trapped ions destabilized and leave the ring
electrode cavity via openings in the lower end cap then emitted to a
transducer.
• Rugged compact and less costly than sector or quadrupole instruments.
• Capable of resolving ions that differ in mass by unit in the mass range of
500-1000 Da
73 م09:40 22/12/2012
Operation of an Ion Trap MS
74 م09:40 22/12/2012
Fourier Transform (FT) instruments
• FTMS provide:
• improved signal to noise ratios
• Greater speed.
• Higher sensitivity.
• Higher resolution.
The heart of FTMS is an ion trap within which ions can circulate in a
well-defined orbits for extended periods.
75 م09:40 22/12/2012
The Ion Cyclotron Resonance (ICR)
• Gaseous ion drifts into or formed in a strong MF.
• The motion becomes circular in a plane that is to the direction of
the field.
• The angular frequency of this motion is called cyclotron frequency
c.
c= v/r = zeB/m
• In fixed field the c depends only upon the inverse of the m/z value.
• Increases the velocity of an ion will be accompanied by increase of
rotation of the ion.
• This circulated trapped ion in the MF is capable of absorbing E from
an ac EF.
• So the EF frequency matches the c.
• The absorbed E increases v of the ion and r of travel without
disturbing c
76 م09:40 22/12/2012
MS – Fourier Transform Analyzer
Ion Cyclotron Resonance
Magnetic Field
77 م09:40 22/12/2012
• Inner solid line represents the original path to the ion.
• Dashed line shows spiral path when switch is moved briefly to
position 1.
• Outer solid line is new circular path when switch again opened.
• For ensemble ions of the same m/z ratio between the two plates;
when ac signal is applied the cyclotron resonance frequency sets
all the particles into coherent motion that is in phase with the field.
• Ions of different m/z ratios (different c) are unaffected by the ac
field.
Measurement of the ICR signal:
Coherent circular motion of resonant ions creates image current
observed current after termination of the frequency sweep signal
(from position 1 to 2).
The current decreases exponentially with time.
It is a capacitor current induced by circular movement of a packet of
ions with the same m/z ratios.
78 م09:40 22/12/2012
Example:
+ve ions approaches the upper plate electrons attracted from circuit
common to this plate causing a momentary current.
This current reversed at the other plate as ions reach the other plate.
An ac current produced depends on number of ions in the packet
Frequency of ac current is characteristic of m/z value of the ions in the
packet.
This current measures conc. of ions.
The induced image current decays with time (few tenth of s to several
s) by losing energy with collisions between ions (ions reach the
thermal equilibrium) (time domain signal).
79 م09:40 22/12/2012
Diagram of the cell used in pulsed ICR & in FT-MS
81 م09:40 22/12/2012
Fourier Transform M.S.
Mass analysis performed by detecting cyclotron frequencies of
ions (depend on m/z) in uniform B, in time domain. Then F.T. to
frequency domain (mass spectrum).
Advantages:
No slits or ion optic lenses to adjust.
All ions detected simultaneously for a single ionizing pulse.
Can produce very high mass resolution using slow scan.
e.g. m/m ~ 220000 for m/z = 84 at B = 1.2T
Detection sensitivity independent of m.
Resolution α1/m (α B).
82 م09:40 22/12/2012
FT-ms sequence of operations
1. Sample (plus R for CI) admitted to chamber.
2. Short e pulse causes EI/CI of sample.
3. Ions trapped by strong magnetic field, inducing ion
cyclotron motion.
4. Then RF sweep (1 ms) coherently excites all m/z to larger
r.
5. Then decay of cyclotron motion induces image currents in
receiver circuit in cell walls. Observed for ms to s.
6. Quench pulse expels ions.
7. Decay signal in 5 amplified, phase-detected, filtered,
digitized, stored. FT from time to frequency domain to get
mass spectrum.
83 م09:40 22/12/2012
apply V (rf) for t s
Radius expands
under V(rf) to
r = V(rf)t/(2dB)
84 م09:40 22/12/2012
what is B/z range?
High vac. So
not FAB
wide-range
Ion image current Rapid scan ICR ms
generation
86 م09:40 22/12/2012
Identification of pure compounds
Several kinds of data obtained from the spectrum of a pure compound.
1- MWt of the compound.
2- Molecular formula of the compound.
3- presence or absence of various functional groups (from fragmentation).
Molecular weights from mass spectra:
By identification of the ion peak or (M+1)+ or the (M-1)+ peak.
Caution: when electron-impact source is used the molecular ion peak may be
absent or small relative to impurity peaks.
Molecular formula from exact molecular weights:
The molecular ion peak can provide the exact mass.
Requires high resolution instrument which capable of detecting mass
differences of a few thousands of mass unit.
Example:
Molecular ions m/z of : Purine (m =120.044), benzamidine C7H8N2 (m
=120.069), ethyltoluene C9H12 (m = 120.096) and acetophenon C8H8O
(m=120.058).
87 م09:40 22/12/2012
If the measured mass of the molecular ion peak is 120.070 (0.005) the all but
C7H8N2 are excluded as possible formula. (precision 40 ppm) which
achieved by high resolution double focusing instruments.
Molecular formulas from isotopic ratios
Data from a low resolution instruments can yield useful information about the
formula of a compound if the molecular ion peak is sufficiently intense that
its height and the heights of the (m+1)+ and (M+2)+ isotope can be
determined accurately.
Example:
Calculate the ratios of the (m+1)+ to M+ peak heights for the following two
compounds: dinitrobenzene, C6H4N2O4 (m = 168) and olefin C12H24 (m
=168).
13C isotope = 1.08%
88 م09:40 22/12/2012
MS – Isotopes Abundance
89 م09:40 22/12/2012
13C 6x1.08 = 6.48%
2H 4x0.015 = 0.060%
15N 2x0.37 = 0.74%
17O 24x0.04 = 0.16%
So (M+1)+/M+ = 7.44%
For C12H24
Thus measuring the heights of (M+1)+ and M+ the two compounds can be
discriminated (same MWt.)
90 م09:40 22/12/2012
• Determination of Molecular Formulas and Molecular Weights
The Molecular Ion and Isotopic Peaks
• The presence of heavier isotopes one or two mass units above the
common isotope yields small peaks at M+.+1 and M+.+2
• The intensity of the M+.+1 and M+.+2 peaks relative to the M peak
can be used to confirm a molecular formula
• Example: In the spectrum of methane one expects an M+.+1 peak of
1.17% based on a 1.11% natural abundance of 13C and a 0.016%
natural abundance of 2H
91 م09:40 22/12/2012
Structural information from fragmentation
It is seldom possible or (desirable) to account for all the peaks in the spectrum.
Instead characteristic patterns of fragmentation are sought.
92 م09:40 22/12/2012
• High-Resolution Mass Spectrometry
• Low-resolution mass spectrometers measure m/z
values to the nearest whole number
• High-resolution mass spectrometers measure m/z
values to three or four decimal places
• The high accuracy of the molecular weight
calculation allows accurate determination of the
molecular formula of a fragment
• Example
– One can accurately pick the molecular formula of a
fragment with a nominal molecular weight of 32 using
high-resolution MS
93 م09:40 22/12/2012
The exact mass of certain nuclides is shown –
below
94 م09:40 22/12/2012
• Fragmentation
• In EI mass spectrometry the molecular ion is highly energetic and
can break apart (fragment)
• Fragmentation pathways are predictable and can be used to
determine the structure of a molecule
• The processes that cause fragmentation are unimolecular
• The relative ion abundance is extremely important in predicting
structures of fragments
– Fragmentation by Cleavage at a Single Bond
• Cleavage of a radical cation occurs to give a radical and a cation
but only the cation is observable by MS
• In general the fragmentation proceeds to give mainly the most
stable carbocation
– In the spectrum of propane the peak at 29 is the base peak
(most abundant) 100% and the peak at 15 is 5.6%
95 م09:40 22/12/2012
– Fragmentation Equations
• The M+. Ion is formed by loss of one of its most loosely held
electrons
– If nonbonding electron pairs or pi electrons are present, an
electron from one of these locations is usually lost by electron
impact to form M+.
– Loosely held nonbonding electrons on nitrogen and oxygen,
and p electrons in double bonds are common locations for an
electron to be lost (i.e., where the remaining unshared electron
in M+. resides)
• In molecules with only C-C and C-H bonds, the location of the lone
electron cannot be predicted and the formula is written to reflect this
using brackets
96 م09:40 22/12/2012
Example: The spectrum of hexane •
97 م09:40 22/12/2012
• Example: spectrum of neopentane
– Fragmentation of neopentane shows the propensity of cleavage
to occur at a branch point leading to a relatively stable
carbocation
– The formation of the 3o carbocation is so favored that almost no
molecular ion is detected
98 م09:40 22/12/2012
– Fragmentation by Cleavage of 2 Bonds
• The products are a new radical cation and a neutral molecule
• Alcohols usually show an M+.-18 peak from loss of water
99 م09:40 22/12/2012
Compounds identification from comparison spectra
1- Determination of MWt. and isotopic distribution and fragmentation patterns.
2- Narrowing the possible structures.
3- Comparing the unknown spectrum with the available reference compounds.
Procedure assumptions:
1-mass fragmentation patterns are unique.
2- Controlling experimental conditions to produce reproducible spectra.
The EI ionization is the method of choice for spectral comparison.
Limitations:
Heights of mass spectral peaks dependent upon:
E of electron beam.
Location of the sample with respect to the beam.
Sample pressure and temperature.
General geometry of the mass spectrometer.
Generally it is desirable to confirm the identity of a compound by comparison
of its spectrum to the spectrum of an authentic (standard) compound
obtained with the same instrument under identical conditions.
Source: Skoog, Holler, and Nieman, Principles of Instrumental Analysis, 5th edition, Saunders College Publishing.
107 م09:40 22/12/2012
Applications of Tandem
• Analysis of complex mixtures either organic or biological.
• Faster than GC/MS separation complete in milliseconds.
• No dilution with mobile phase is needed (interferences occurs) as that of
GC/MS and LC/MS.
• So tandem MS is more sensitive than GC/MS or LC/MS (smaller noises).
• Used for quant. And qualit. determination of the components of a wide
variety of complex materials either in nature or industry.
Examples:
• identification and determination of drug metabolites, insect pheromones,
alkaloids in plants, trace contaminants in air , alkaloids in plants, polymer
sequences, petrochemicals, polychlorinated biphenyls, prostaglandins,
diesel exhausts and odors in air.
• It will find wider applications in future.
Disadvantage:
Greater cost.